Position Electrophoresis Gels/Films or Transparencies
To scan the electrophoresis gels/lms (such as 1D/2D gels, SDS-PAGE, western/northern/southern
blots, or membranes) and transparent lms, please use the Leak-Free Glass Holder together.
1. Take the Leak-Free Glass Holder out from the box, and turn the side of the holder with a sticker
facing up.
2. Place the electrophoresis gel or lm to be scanned on the glass surface of the holder.
If necessary, secure the electrophoresis lm to the glass surface using the Vinyl strips included
in the box when you position the lm on the holder.
LEAK-FREE GLASS HOLDER
Thisholder is specifically designed for
theuse of the gel and transparent film scanning.
LEAK-FREE GLASS HOLDER
This holder is specifically designed for
the use of the gel and transparent film scanning.
LEAK-FREE GLASS HOLDER
This holder is specifically designed for
the use of the gel and transparent film scanning.
Electrophoresis lm
Vinyl strip
The side of the holder
labeled with a sticker
should be facing up.
Keep Calibration strip clear
and free of any obstruction
3. Insert the holder with loaded electrophoresis
gel or lm all the way into the transparency
bay (lower scan bed) of the scanner.
NOTE:
• MakesurethecorrectsideoftheLeak-Free Glass Holder is facing up when you insert it into the scanner’s
transparency bay (lower scan bed). The correct side of the holder is labeled with a sticker on it.
• MakesurethattheCalibration strip on the holder is kept clear and free of obstruction at all times; no
material should ever be placed on this area.
• Pleasetakenotethatthehightemperaturemaycausedamagetoyourelectrophoresisgelsandtransparent
lms.Therefore,turnoffthescannerandremovethescannedgelsandlmsfromthescanneronceyou
nishthescanningjob.
• TheLeak-Free Glass Holder requires regular cleaning. To clean, use a mild glass cleaning solution, and wipe
theglassplategentlywithalint-free,lens-cleaningclothtopreventleavingberresidue.