OPERATION
yRemove the protective cap.
yTurn the digital pen on with the
ON/OFF switch located on the top.
000
CD97
yImmerse into solution up to the
maximum immersion level.
12
yStir gently and wait until the display
stabilizes. CD97 compensates for the
temperature variance automatically.
yRead the value on the display.
CALIBRATION
yThe calibration procedure is very
simple and fast.
yImmerse the pen up to the maximum
level into the appropriate optional
Calibration Solution M10080B
correspondig to 800ppm.
yAllow the reading to stabilize.
yBy turning the calibration trimmer
with a small screwdriver adjust the
reading to match the solution value
of 800ppm (normally to the value at
25°C).
800
600
yCalibration is now complete.
BATTERY REPLACEMENT
When the CD97 cannot be switched
on or the display fades, pull out the
battery compartment and replace all
four 1.5V batteries, paying attention
to their polarity. Batteries should only
be replaced in a non‑hazardous area
using the battery types specified in this
instruction manual.
CD97
RECOMMENDATION
Before using this product, makesure that
it is entirely suitable for the environment
in which they are used. Operation of
these instruments in residential areas
could cause unacceptable interferences
to radio and T.V. equipment. The pins at
the end of the sensor are sensitive to
electrostatic discharges. Avoid touching
this metal band at all times. During
operation of instruments ESD wrist
straps should be worn to avoid possible
damage to the sensor by electrostatic
discharge. Any variation introduced
by the user to the supplied equipment
may degrade the instrument EMC
performance. To avoid electrical shock,
do not use the instrument when voltage
at the measurement surface exceed
24VAC or 60VDC. To avoid damages or
burns, do not perform any measurement
in microwave ovens.