manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Milwaukee
  6. •
  7. Measuring Instrument
  8. •
  9. Milwaukee MW803 User manual

Milwaukee MW803 User manual

This manual suits for next models

1

Other Milwaukee Measuring Instrument manuals

Milwaukee MW11 User manual

Milwaukee

Milwaukee MW11 User manual

Milwaukee SM700 User manual

Milwaukee

Milwaukee SM700 User manual

Milwaukee 2270-20 User manual

Milwaukee

Milwaukee 2270-20 User manual

Milwaukee SMS510 User manual

Milwaukee

Milwaukee SMS510 User manual

Milwaukee MI405 User manual

Milwaukee

Milwaukee MI405 User manual

Milwaukee Mi170 User manual

Milwaukee

Milwaukee Mi170 User manual

Milwaukee MA871 User manual

Milwaukee

Milwaukee MA871 User manual

Milwaukee LDM 45 User manual

Milwaukee

Milwaukee LDM 45 User manual

Milwaukee SM106 User manual

Milwaukee

Milwaukee SM106 User manual

Milwaukee MW150 User manual

Milwaukee

Milwaukee MW150 User manual

Milwaukee 2205-40 User manual

Milwaukee

Milwaukee 2205-40 User manual

Milwaukee pH51 User manual

Milwaukee

Milwaukee pH51 User manual

Milwaukee 48-22-5112 User manual

Milwaukee

Milwaukee 48-22-5112 User manual

Milwaukee 2235-20 User manual

Milwaukee

Milwaukee 2235-20 User manual

Milwaukee 3712 User manual

Milwaukee

Milwaukee 3712 User manual

Milwaukee SM700 User manual

Milwaukee

Milwaukee SM700 User manual

Milwaukee pH54 User manual

Milwaukee

Milwaukee pH54 User manual

Milwaukee SM101 User manual

Milwaukee

Milwaukee SM101 User manual

Milwaukee M12 CPL User manual

Milwaukee

Milwaukee M12 CPL User manual

Milwaukee MW105 MAX User manual

Milwaukee

Milwaukee MW105 MAX User manual

Milwaukee MA882 User manual

Milwaukee

Milwaukee MA882 User manual

Milwaukee MW700 PRO User manual

Milwaukee

Milwaukee MW700 PRO User manual

Milwaukee 2217-40 User manual

Milwaukee

Milwaukee 2217-40 User manual

Milwaukee 48-22-9801 User manual

Milwaukee

Milwaukee 48-22-9801 User manual

Popular Measuring Instrument manuals by other brands

Suaoki D40 user manual

Suaoki

Suaoki D40 user manual

Elster BK-G1.6 operating instructions

Elster

Elster BK-G1.6 operating instructions

Endress+Hauser Deltabar FMD71 technical information

Endress+Hauser

Endress+Hauser Deltabar FMD71 technical information

Duratool D03128 manual

Duratool

Duratool D03128 manual

Vega VEGAPULS 65 Quick setup guide

Vega

Vega VEGAPULS 65 Quick setup guide

Tonghui TH2618B Operation manual

Tonghui

Tonghui TH2618B Operation manual

Tenmars TM-801 user manual

Tenmars

Tenmars TM-801 user manual

MSR 145 user manual

MSR

MSR 145 user manual

TPS WP-82 manual

TPS

TPS WP-82 manual

PCB Piezotronics 003C30 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 003C30 Installation and operating manual

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Endress+Hauser Proline Promag 50 Brief operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 50 Brief operating instructions

Siemens SITRANS F Coriolis FCT030 Function manual

Siemens

Siemens SITRANS F Coriolis FCT030 Function manual

KLINGER CMF V Series instruction manual

KLINGER

KLINGER CMF V Series instruction manual

Keysight M8290A Getting started guide

Keysight

Keysight M8290A Getting started guide

Acterna MS1400 Operation manual

Acterna

Acterna MS1400 Operation manual

R&S ZPH manual

R&S

R&S ZPH manual

Datacolor ColorReader DC10-2 user guide

Datacolor

Datacolor ColorReader DC10-2 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

USER MANUAL
MW803 • MW804
Pocket-size
pH/EC/TDS/Temperature
Meters
SPECIFICATIONS
Range
0.00 to 14.00 pH
 0to3999μS/cm(MW803)
 0to2000ppm(MW803)
 0.00to20.00mS/cm(MW804)
 0.00to10.00ppt(MW804)
 0.0to50.0°C/32.0to122.0°F
Resolution
0.01 pH
 1μS/cm/1ppm(MW803)
 0.01mS/cm/0.01ppt(MW804)
 0.1°C/0.1°F
Accuracy (@25°C)
0.05 pH
 2%FS(EC/TDS)
 0.5°C/±1°F
TDS Factor
 0.45to1.00(CONV)
Temperature Compensation
 Automatic,withß=0.0to2.4%/°C
Calibration
 Automatic,1pointforECand1or
 2pointsforpH
Probe Replaceable MI60P
Environment
 0to50°C;100%RHmax.
Battery Type
 4x1.5V;IECLR44,A76
Battery Life
 Approx.100hoursofuse
Auto-off
 After8min.ofnon-use
Dimensions / Weight
 200xdia38mm/100g
Makesurethegasketisinplacebeforescre-
wing back the cap. Batteries should only be
replacedinasafeareausingthebatterytype
speciedinthisinstructionmanual.
ACCESSORIES
MI60P Replaceableprobefor
 MW803&MW804
M10030B12.88mS/cmcalibration
 solution,20mlsachet,25pcs
M10031B1413μS/cmcalibrationsolution,
 20mlsachet,25pcs
M10032B1382ppmTDSCalibrationSolution,
 20mlsachet,25pcs.
M10038B6.44pptTDSCalibration,
 20mlsachet,25pcs.
MA906012.88mS/cmcalibrationsolution,
 230mlbottle
MA9061 1413μS/cmcalibrationsolution,
230mlbottle
MA9062 1382ppmTDSCalibration
 Solution,230mLbottle
M10016BCleaningsolution,20mlsachet,
 25pcs
M10000BRinsesolution,20mlsachet,
 25pcs
M10004BpH4.01buffer,20mlsachet,
25pcs.
M10007BpH7.01buffer,20mlsachet,
 25pcs.
M10010BpH10.01buffer,20mlsachet,
 25pcs.
MA9004pH4.01buffer,230mlbottle
MA9007pH7.01buffer,230mlbottle
MA9010pH10.01buffer,230mlbottle
MA9006 pH6.86buffer,230mlbottle
MA9009 pH9.18buffer,230mlbottle
MA9015Electrodestoragesolution,230ml
• ThesecondaryLCDdisplays“ESC”for1se-
condandthenthemeterreturnstothenormal
measurementmode.
• To reset to the default values and clear a
previous calibration, press the SET/HOLD
button after entering the calibration mode
andbeforetherstpointisaccepted.
• The secondary LCD displays “CLR” for 1
second,themeterresetstothedefaultcalib-
rationandtheCALtagontheLCDturnsoff.
• Ifa validbuffer value is detected, the REC
messageisdisplayedandthemetercomp-
letesthecalibrationprocedure.
• TheLCDshowstheacceptedvaluewiththe
“OK2”messageandtheinstrumentreturns
to the normal measurement mode. Other-
wise,ifnovalidbufferisdetected,themeter
displaystheWRNGmessage.
PROBE REPLACEMENT
• Removetheprotectivecapandunscrewthe
plasticringonthetopoftheprobe.
• PullouttheMI60Pprobeandreplaceitwith
anewone.
• Make sure the gaskets are in place before
screwingbackthering.
BATTERY REPLACEMENT
Whenthebatteriesbecomeweak,thebattery
symbolontheLCDwilllightuptoadvisethat
onlyafewhoursofworkingtimeisremaining.
ThemeterisalsoprovidedwithBEPS(Battery
Error Prevention System), which avoids any
erroneous readings due to low battery level
byautomaticallyswitchingthe meter off.It is
recommendedtoreplacethebatteriesimme-
diately.
To replace the batteries unscrew the battery
compartmentcapandreplaceallfour1.5Vbat-
terieswhilepayingattentiontotheirpolarity.
OPERATIONAL GUIDE
• Removetheprobecapandturnthemeteron
bypressingtheON/OFF/CALbutton.Allthe
usedsegmentsontheLCDwillbevisiblefor
1secondoraslongasthebuttonispressed.
• Immersetheprobeinthesolutiontobetes-
tedand select eitherpH, ECor TDSmode
withSET/HOLDbutton.
• Stirgentlyandwaitforthereadingtostabilize,
i.e.thestabilityindicatorontheLCDturnsoff.
The pH and EC (or TDS) values are auto-
maticallycompensatedfortemperatureand
will be displayed on the main LCD, while
thetemperatureisshownonthesecondary
LCD.
• To freeze the display,whileinmeasurement
mode, press and hold the SET/HOLD but-
ton. The “HOLD” message appears on the
secondary display and  the reading will be
frozenontheLCD.
• Press  any  button to  return  to  normal
mode.
• To turn the meter off,presstheON/OFF/CAL
button. The “OFF” message will appear on
thesecondarydisplay.Releasethebutton.
Notes:
• Before taking any measurement, make sure
themeteriscalibrated(theCALtagison).
• After use always turn the meter OFF, rinse
theprobewithwaterandstoreitwiththepro-
tectivecap.
Measuring unit
on main LCD
Main LCD
Temperature reading
Secondary LCD
Low battery indicator
Automatic temperature
compensation
Calibration indicator
Stability indicator
SETUP
Setup mode allows the selection of tempera-
ture (°C or °F), calibration buffer set, in pH
modeonly,TDSconversionfactor(CONV)and
temperature coefcient (BETA), in EC mode
only.ToentertheSetup mode,presstheON/
OFF/CALbuttonuntil“CAL”onthesecondary
LCD is replaced by “TEMP” and the current
temperatureunit(e.g.TEMP°C).Then:
In EC and pH mode:
• for °C/°F selection:
usetheSET/HOLDbutton,thenpresstheON/
OFF/CALbuttontogointhesettingsmenuand
toreturntothenormalmeasurementmode.
InECmodeonly:
• to change the TDS factor value:
aftersettingthetemperatureunit,pressON/
OFF/CALbuttononcetoshowthecurrentval-
ue(e.g.0.50CONV).Selectthedesiredvalue
byusingtheSET/HOLDbutton,thenpressON/
OFF/CALbutton twice toreturnto the normal
measurementmode.
• to change the temperature coefcient:
after setting the TDS factor, press ON/OFF/
CAL button to show the current value of the
temperaturecoefcientß(e.g.2.1BETA).Use
theSET/HOLDbuttontosetthedesiredvalue,
thenpressON/OFF/CALbuttontoreturntothe
normalmeasurementmode.
InpHmodeonly:
• to change the calibration buffer set:
after setting the temperature unit, press ON/
OFF/CALbuttononceandselectthebufferset
(“pH7.01BUFF”forpH4.01/7.01/10.01or“pH
6.86 BUFF” for NIST set 4.01/6.86/9.18) by
using the SET/HOLD button. Press ON/OFF/
CAL button to return to the normal measure-
mentmode.
EC MEASUREMENT AND CALIBRATION
PROCEDURE
Select the EC or TDS mode with the SET/
HOLDbutton.
Submerge the electrode in the solution to be
tested. The measurements should be taken
whenthestabilityindicator⌛onthetopleftof
theLCDdisappears.
• For better accuracy, frequent calibration of
the tester is recommended. Calibration is
alsonecessaryafterprobereplacement,after
testingaggressive chemicalsand where ex-
tremeaccuracyisrequired.FromnormalEC
operationmode,pressandholdtheON/OFF/
CALbuttonuntilthe“OFF”messageonthe
secondaryLCDissubstitutedby“CAL”.Re-
leasethebutton.
• Immerse the probe in the proper calibration
solution: M10031 (1413 μS/cm) for MW803
andM10030(12.88mS/cm)forMW804.
• Oncethe calibration hasbeen automatically
performed,theLCDwillshow“OK”for1sec-
ondandthemeterwillreturntonormalmea-
surementmode.
• Sincethereisaknownrelationshipbetween
ECandTDSreadings,itisnotnecessaryto
calibratethemeterinTDSifitwaspreviously
calibratedinECmode.
pH MEASUREMENT AND CALIBRATION
SelectthepHmodewiththeSET/HOLD
button.Submergetheelectrodeinthesolution
tobetested.Themeasurementsshouldbeta-
kenwhenthestabilityindicator⌛onthetopleft
oftheLCDdisappears.
Forbetteraccuracy,frequentcalibrationofthe
tester is recomanded. Calibration is also ne-
cessaryafterelectrodereplacement,aftertes-
tingaggressivechemicalsandwhereextreme
accuracyisrequired.
• From normal operation mode, press and hold
theON/OFF/CALbuttonuntiltheOFFmessage
onthesecondaryLCDissubstitutedby“CAL“.
Releasethebutton.
• The instrument enters the calibration mode
by displaying “pH 7.01 USE“ (or ‘‘pH 6.86
USE“iftheNISTbuffersetwasselected).
• For a single-point calibration, immerse the
electrodeinanybuffer,i.e.pH4.01,7.01(or
6.86),10.01(or9.18).
• The meter activates the automatic buffer
recognition.Ifnovalidbufferisdetected,the
meterkeepstheUSEindicationactivefor12
seconds,andthenreplacesitwithWRNGin-
dicating that the sample being measured is
notavalidbuffer.Otherwise,ifavalidbuffer
is detected, then its value is shown on the
main display, and REC appears on the se-
condaryLCD.
• IfthepH7.01(or6.86)wasused,pressthe
SETbuttontoexittheCalibrationmodeand
the “OK 1” message will appear on the dis-
play. The calibration point is stored and the
meterreturnstonormalmeasurementmode.
Forbetteraccuracy,itisalwaysrecommended
toperforma2-pointcalibration.
• For a 2-point calibration, immerse the elec-
trodeinpH7.01(or6.86)buffersolution.
• After the rst point has been accepted, the
meterwillthenaskforthesecondbufferand
themessage“pH4.01USE”willappear.
• Rinsetheelectrodeandimmerseitinthese-
condsolution(pH4.01,10.01or9.18).
Notes:
• When the calibration procedure is completed,
theCALtagisturnedon.
• To quit the procedure and return to the last
calibrationdata,afterenteringthecalibration
modepresstheON/OFF/CALbutton.