Speed Control/ Change Direction of Rotation
Turn the control ring to change the speed and the rotation direction.
Disinfektion
The MK-dent Air Motor can be disinfected externally. Only use DGHM listed chemical disinfectant for spray and wipe disinfection.
Don‘t place the Air Motor in disinfectant baths. Results are mulfunctions.
Lubrication with Oil
Use “MK-dent Premium Service Oil” for care and lubrication. Remove the Air Motor
from the tubing, attach the enclosed MK-dent oil adapter on the oil can and spray for
1-2 seconds in the smaller one of the two large pipelines (see drawing).
Sterilization
Sterilize the MK-dent Air Motor in an autoclave up to max. 135°C (2 bar) / 275°F (29 psi).
Autoclave three times with an initial vacuum for at least 4 minutes at 134°C +/- 1
Autoclave using the gravitation method for at least 10 minutes at 134°C +/- 1
Autoclave using the gravitation method for at least 60 minutes at 121°C +/- 1
Follow the manufacturer‘s instructions for use.
The prepared instrument must be stored protected germ-free from dust in a dry, dark and cool room.
Observe the expiration date of the sterilized item. No chlorine-based products and no chemical solvents or cleaning materials
must be used. Do not use dryer (post-heat). Do not sterilize the packing box.
Caution
Stop working with the contra angle handpiece by higher and irregular running noise, signicant vibrations, overheating,
and optical damaging. Please contact in these cases an authorized by the manufacturer repair center or your dental-depot.
Warranty
MK-dent provides the end user a warranty for proper function, immaculate material and workmanship for the period of 12 months after purchasing.
Relevant for this is the invoice date.
In case of justied complaints MK-dent carries out a free repair or a possible free replacement. This will be decided by MK-dent.
Other claims of any kind, in particular damage compensation are excluded. In the event of default, gross negligence or intent, this shall only apply unless
there are compelling legal regulations.
MK-dent is not liable for defects and their consequences that have arisen whose arising may be due to natural wear, improper handling, improper
cleaning, or maintenance, non-compliance with operating or manual instructions,
calcination or corrosion, contaminated air and water supply or chemical or electrical inuences that are unusual or not permitted according to
MK-dent’s instruction for use or to other manufacturer‘s instructions.
The warranty does usually not cover lamps, light conductors made of glass bers, glassware, rubber parts and the color fastness of plastic parts.
All liability is excluded if defects or their consequences arise because of interventions or modications on the product by the end user or by a third party
not authorized by MK-dent.
Warranty claims will only be accepted if the product is submitted with a proof of purchase in the form of an invoice or a copy of the delivery note. Dealer,
purchase date, model and serial number must be clearly visible.