
•The minimum concentration of a combustible gas in air that can
ignite is defined as the Lower Explosive Limit (LEL). A
combustible gas reading of "100", or a %CH4reading within the
range of approximately 5% to 10% indicates that an explosion
hazard exists. Extreme caution must be exercised in this
situation.
•Do not use the Orion G Multigas and Leak Detector to test for
combustible or toxic gases in the following atmospheres as this
may result in erroneous readings:
•Oxygen-deficient or oxygen-rich atmospheres
•Reducing atmospheres
•Furnace stacks
•Inert environments
•Atmospheres containing combustible airborne mists
or dusts.
•Do not use the Orion G Multigas and Leak Detector to test for
combustible gases in atmospheres containing vapors from
liquids with a high flash point (above 100oF) as this may result
in erroneously low readings.
•Allow sufficient time for unit to display accurate reading.
Response times vary based on the type of sensor being utilized
(see "Performance Specifications" section of manual).
Additionally, allow a minimum of 0.7 seconds per foot of sample
line to allow the sample to be drawn through to the sensors.
•Keep the probe tip above liquid surfaces; otherwise, liquid may
enter the system and block the sample flow, causing inaccurate
readings and/or internal damage.
•All instrument readings and information must be interpreted by
someone trained and qualified in interpreting instrument
readings in relation to the specific environment, industrial
practice and exposure limitations.
•Replace alkaline cells or recharge NiMH battery-pack in
non-hazardous area only. Use only battery chargers listed in
this manual; other chargers may damage the battery pack and
the unit. Dispose of batteries in accordance with local health
and safety regulations.
•Do not alter this instrument or make any repairs beyond those
specified in this manual. Only MSA-authorized personnel may
repair this unit; otherwise, damage may result.
Operating Manual Chapter 1, Safety Limitations and Certifications
1-3