MSI T HOLD 1000LS User manual

6056 Rev B
3/4/09
T° HOLD
OPERATORS MANUAL
Medical Solutions, Inc.
3901 Centerview Drive, Suite L, Chantilly, VA 20151
(800) 435-7600 Phone • (703) 834-0039 Fax • www.warmiv.com
®

1 6056 Rev B
TABLE OF CONTENTS
1.0 Intended Use.......................................................................................................... 3
2.0 Product Description .............................................................................................. 3
2.1 Important Labels and Directions ...................................................................... 4
2.2 Important Labels Heater Plate and Power Box................................................. 5
3.0 Specifications and Operating Information.......................................................... 6
3.1 Specifications.................................................................................................... 6
3.2 Storage and Operating Conditions.................................................................... 6
3.3 Electrical Requirements.................................................................................... 6
4.0 Operating Instructions.......................................................................................... 7
5.0 Cleaning Instructions............................................................................................ 8
6.0 Troubleshooting .................................................................................................... 8
7.0 Preventive Maintenance Plan ............................................................................... 9
8.0 Electrical Schematic............................................................................................ 10
Appendix A: Limited Warranty ………………………………………………………..11
A.1 Warranty Registration Card ……………………………………………..12
Appendix B: Product Labels List ……………………………………………………...13

2 6056 Rev B
TABLE OF FIGURES
Figure 1 - General Set-up ...............................................................................................3
Figure 2 - Product Labels (Heater Plate) ......................................................................4
Figure 2A (Front)……………………………………………………………….4
Figure 2B (Back) ……………………………………………………………….4
Figure 2C (Side Bottom & Upper Left) ……………………………………….4
Figure 3 - Product Labels (Power Box).........................................................................5
Figure 3A (Front).............................................................................................................5
Figure 3B (Back)..............................................................................................................5
Figure 4 - I.V. Mounting Pole .......................................................................................7
Figure 5 - Hook Swivel and Bag Mounting ..................................................................7
Figure 5A (Lower Hook)................................................................................................7
Figure 5B (Upper Hook) ...............................................................................................7
Figure 6 - Power Indicator Light and LED Indicators .................................................8
Figure 6A (Power Switch) .............................................................................................8
Figure 6B (LED Lights) ................................................................................................8
Figure 7 - Electrical Schematic.................................................................................... 10

3 6056 Rev B
T°
HOLD
®Operators Manual
Congratulations on your selection of the Medical Solutions, Inc. (MSI) T° HOLD®1000LS.
You have selected one of the components of our state-of-the-art Fluid Warming System,
designed to maintain normothormic fluids for your patients. The use of the T° HOLD®, like all
of the components of our proprietary System, helps reduce complications due to inadvertent
hypothermia and increases patient safety and comfort. The MSI Fluid Warming System helps
keep the patient warm from admission to discharge. We hope that both you and your patients
enjoy the benefits of this exclusive technology.
Please read the entire Operators Manual before using the T°HOLD®1000LS.
1.0 INTENDED USE
The T° HOLD®is designed to hold, warm, and maintain 1000mL I.V. fluid bags at a
temperature of 40°C (104°F).
2.0 PRODUCT DESCRIPTION
The T° HOLD®consists of two components: the Power Box and the Heater Plate. The
Power Box can be mounted on an I.V. pole or bed rail or attached to a wall. The
Heater Plate is designed to hang on an I.V. pole.
The I.V. fluid bag is hung from either hook at the top of the Heater Plate and the
elastic strap holds the I.V. fluid bag in place. Two LED Indicators on the bottom of the
Heater Plate illuminate when the Heater Plate is warming.
FIGURE 1 - GENERAL SET - UP
NOTE: Please review the I.V. fluid bag manufacturer’s recommendations on the
appropriate time and temperature for the fluid being warmed
I.V. Pole Hook
HEATER PLATE
I.V. Bag Strap
LED Indicators
Power Switch
POWER BOX

4 6056 Rev B
2.1 Important Labels and Directions
The following labels are found on the T° HOLD®.
oModel/ Serial Number.
oDo Not Open.
oElectrical Shock Symbol.
oUnited States Patents.
oMade in the USA.
oWarning: Risk of Fire. Replace Fuse as Marked: 250V 2.0 AMP.
oCaution: Electrical shock hazard. Do not remove protective panel. Refer servicing to
qualified personnel.
oCaution: Verify temperature of warmed items before use.
oCaution: Hot.
oGrounding Reliability.
oDanger: Risk of explosion if used in the presence of flammable anesthetics.
oWarming: Unit not intended for blood or blood products.
oFactory Pre-set and Locked at 40°C.
oMET and/or UL Listing.
Figure 2B Figure 2B
Figure 2C
FIGURE 2 - PRODUCTS LABELS (HEATER PLATE)
Label-
“Do Not Open”
Label-
“Caution: Hot”
Label-
“United States
Patents”
Label-
“UL and/or
MET Listing”
Label-
“Factory Pre-Set”
Label
“Electrical
Shock Symbol”
Label-
“Do Not Open”

5 6056 Rev B
2.2 Important Labels Heater Plate and Power Box
Figure 3A - Front
Figure 3B - Back
FIGURE 3 - PRODUCT LABELS (POWER BOX)
Label-
“Not Intended
for Blood or
Blood Products”
Label-
“Risk of Fire”
Label-
“Grounding
Reliability”
Label-
“Model/Serial
Number”
Label-
“Risk of
Explosion”
Label-
“Electrical Shock
Hazard”
Label-
“Verify
Temperature”
Label-
“United States
Patents”
Label-
“Electrical Shock
S
y
mbol”
Label-
“Made in the USA”
Label-
“UL and/or
MET Listing”

6 6056 Rev B
3.0 SPECIFICATIONS AND OPERATING INFORMATION
3.1 Specifications
Classification Class 1
Dimensions: Power Box 10.00˝tall x 6.50˝wide x 3.00˝deep
Heater Plate 14.25˝tall x 5.00˝wide x 1.125˝deep
Heater Plate Capacity 1000 mL Industry standard IV fluid bags.
Mounting: Heater Plate
Power Box
Hook for I.V. Pole
I.V. Pole C-Clamp, Bed Rail Bracket or Wall
Mount Bracket
Construction Powder Coat Finish Aluminum
Factory Preset Set Point Temperature 40°C (104°F)
3.2 Storage and Operating Conditions
3.3 Electrical Requirements
Storage Temperature -40 to 80°C (-40 to 176°F)
Humidity Conditions 0 to 90% up to 40°C (104°F) non-condensing.
10 to 55% at 55°C (131°F) non-condensing.
Ambient Operating Temperature -10 to 55°C (14 to 131°F)
Power Requirements 120 VAC
60 HZ
Fused 250 VAC
2.0 AMP Time Delay
Amperage Consumed 1.09 AMPS
Heater Wattage 53 watts
Thermal Cutout Temperature 41°C (106°F) Automatic reset

7 6056 Rev B
4.0 OPERATING INSTRUCTIONS
1. Remove the T° HOLD®Heater Plate from its packaging.
2. Check the Heater Plate to ensure that all of the packaging material is removed.
3. Mount the Power Box to the I.V. pole support using the supplied C-Clamp.(Ref: Fig 4)
4. Hang the Heater Plate from the I.V. pole hook. (Ref: Fig 5A and 5B).
FIGURE 4 - I.V. POLE MOUNTING 1
5. After ensuring the power switch is “off” plug the power supply cord into an
approved hospital grade outlet.
6. Place a 104°F pre-warmed bag from a Medical Solutions, Inc. fluid warmer (16N1®,
Temp 1®, Temp 2®, Temp 3®, Heat Stack®) without the sterile over-wrap on the upper
or lower hook so that the bag’s ports are accessible below the Heater Plate.
Figure 5A Figure 5B
FIGURE 5 - HOOK SWIVEL AND BAG MOUNTING 1
C
–
Clamp
Hook Swivel – Rotate Bag
For User
View Preference
Lower Hook
(Shorter Bag)
Upper Hook
(Longer Bag)
Not available
On certain
models

8 6056 Rev B
7. Pull the elastic strap across the fluid bag and secure by affixing the I.V. Bag Strap.
8. Turn the unit “on” by pressing the Power Switch. The switch and the LED indicators on
the bottom of the Heater Plate will illuminate when energized. (Ref: Figure 1).
If fluid spills inside the Heater Plate or Power Box, unplug the unit to prevent an
electrical shock hazard and refer to the Cleaning Instructions (Section 5.0).
5.0 CLEANING INSTRUCTIONS
1. Turn the main power “off” and unplug the T° HOLD®from the power source.
2. Wipe down the Heater Plate with a damp cloth using a general hospital cleaner
(Isopropyl Alcohol) as needed, following the cleaner manufacturer’s instructions for use.
3. If excessive fluid enters the Heater Plate, disconnect power to the unit and refer to
qualified Bio- Med personnel for service.
4. Contact Medical Solutions, Inc., at 800-435-7600 for further instruction.
6.0 TROUBLESHOOTING
1. If the T° HOLD®fails to warm, check to ensure that:
a) The power cord is plugged into an approved hospital grade outlet.
b) The power switch is turned on at the Power Box with the Power Indicator Light
illuminated. (Ref: Figure 6A).
c) The green LED Indicators on bottom of the Heater Plate are illuminated.
(Ref: Figure 6B).
2. If the Power Indicator Light is not illuminated please have a Bio Medical technician
verify that the two fuses on the Power Box are functional. (Ref: Figure 3A).
3. If the power is “on” and the T° HOLD®fails to warm, please contact Medical
Solutions, Inc. at 800-435-7600.
Figure 6A Figure 6B
FIGURE 6- POWER INDICATOR LIGHT AND LED INDICATORS
Power Switch
LED Lights

9 6056 Rev B
7.0 PREVENTIVE MAINTENANCE PLAN
The T° HOLD®temperature controller should be verified once every five (5) years).
The following items are needed to accurately measure the bag temperature for
verification:
oFluke 51 II Digital Thermometer.
o6.00˝long .093˝Diameter Type T Needle Thermocouple calibrated to the
Thermometer.
To verify the temperature controller.
1. Pre-warm to 40°C (104° F) a 1000 ml I.V. fluid bag for the size T° HOLD®being
calibrated to 40°C (104° F).
2. Remove the outer wrap from the fluid bag.
3. Pierce the bag through the needle injection port on the bag with an 18 ga needle to
make it easier to insert the Thermocouple.
4. Place the bag into the T° HOLD®and secure the bag to the Heater Plate with the
I.V. Bag Strap.
5. Insert the Thermocouple into the bag and position the probe in the center of the
bag, both front to back and from side to side.
6. Turn on the T° HOLD®. Allow the unit to operate for a minimum of one (1) hour
before verifying the temperature.
7. After the allotted time, the temperature reading on the Fluke Thermometer should
be within one (1) degree Celsius (two [2] degrees Fahrenheit) of the factory preset
temperature.
8. If the two temperature readings are within one degree (1) Celsius (two [2] degrees
Fahrenheit) of each other, the temperature controller has been verified. If they are
not, please contact Medical Solutions, Inc., at 800-435-7600.

10 6056 Rev B
8.0 ELECTRICAL SCHEMATIC
FIGURE 7- ELECTRICAL SCHEMATIC

11 6056 Rev B
A P P E N D I X A : L I M I T E D W A R R A N T Y
For one year from the date of shipment to the Buyer (the
“Warranty Period”), Medical Solutions, Inc. (“MSI”) warrants that
the Equipment provided to the Buyer shall be free from defects in
workmanship and materials. In the event of a breach of this
warranty during the Warranty Period, MSI shall, at its option,
repair or replace the Equipment. Buyer’s remedies are limited to
such repair or replacement.
MSI DISCLAIMS ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING THE WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL
MSI BE LIABLE FOR DIRECT, INDIRECT, PUNITIVE, ECONOMIC,
CONSEQUENTIAL OR OTHER DAMAGES, INCLUDING BUT NOT
LIMITED TO LOST PROFITS.
To enable MSI to administer the Warranty properly, Buyer shall
(1) register the Warranty using the form provided below within
thirty days of receipt of Equipment, and (2) promptly notify MSI of
any claim hereunder. As a condition to the Warranty, Buyer shall
use and maintain the Equipment according to the provisions of
the Operators Manual.

12 6056 Rev B
________________________________________________________________________
A.1 IMPORTANT: THIS IS YOUR T°HOLD® WARRANTY REGISTRATION
Please register your Warranty by copying this page, providing the requested information, and
sending the completed form to Medical Solutions, Inc. within thirty (30) days of purchase.
Hospital Name:
Street
Address:
Mailing Address
City: State: Zip Code:
Contact Name: Phone: Email:
Date Purchased:
Serial Number:
Department Purchased
For:
COPY THIS PAGE AND MAIL, FAX, OR EMAIL TO:
Medical Solutions, Inc.
ATTN: Warranty Supervisor
3901 Centerview Drive, Suite L
Chantilly, VA 20151
FAX NUMBER: 703-834-0039

13 6056 Rev B
A P P E N D I X B : P R O D U C T L A B E L S L I S T
Label: UL and/or MET Listing
Location: Back of Heater Plate & Power Box
Label: Factory Pre-Set
L
oc
ati
o
n
:
Fr
o
nt
o
f H
e
at
e
r Plat
e
Label: Caution: Hot
Location: Front of Heater Plate
Label: United States Patents
Location: Back of Power Box & Back of Heater Plate
Label: Electric Shock Symbol
Location: Back of Heater Plate, Back of
Power Box & Inside Power Pox
Label: Made in the USA
Location: Top of Power Box
Label: Not Intended for Blood or Blood Products
Location: Front Bottom of Power Box

14 6056 Rev B
Label: Fuse
Location: Bottom Side of Power Box
Label: Verify Temperature
Location: Front Bottom of Power Box
Label: Ground Reliability
Location: Back of Power Box
Label: Do Not Open
Location: Left Upper & Bottom Front
of Heater Plate
Label: Ground
Location: Inside Power Box &
Inside Heater Plate
Labels: Electrical Shock Hazard & Risk of
Ex
p
losion Label: Model/Serial Number
Location: Side of Power Box
Table of contents
Popular Medical Equipment manuals by other brands

Clearlight
Clearlight HALOONE user manual

Arjo
Arjo Prioma Couch Electric Instructions for use

Maquet
Maquet Betamaquet 1140 Series Maintenance and Repair instruction

BTL
BTL 1500 user manual

WillowWood
WillowWood LimbLogic Patient instructions

COMPASS HEALTH
COMPASS HEALTH TWIN STIM PLUS OTC instruction manual