Lightmed LightLas 532 User manual

LightLas532 –Operators Manual Rev. No 01 Page 1of 115
Operator’s Manual
LightLas 532
Green Laser Photocoagulator With LCD
Control Panel

LightLas 532 –Operator's Manual Rev. No 01 Page 2of 115
Operator’s Manual
for the
LightLas 532
Green Laser Photocoagulator With LCD
Control Panel
Directive 93/42/EEC
as amended by 2007/47/EC
Doc. No. : DC1001
Rev. No. : 01

LightLas 532 –Operator's Manual Rev. No 01 Page 3of 115
Operator’s Manual for the LightLas 532 Green Laser
Photocoagulator
Clinicians or Doctors should ensure that they are adequately knowledge of the
operation procedures prior to using the LightLas 532 Laser system.
This Operators Manual should be studied and understood before proceeding to
operate the equipment on patients.
CAUTIONS - Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous radiation exposure.
CAUTIONS - Any modification to the Ophthalmic Laser will result in the necessity for
it to be reclassified
CAUTIONS - U.S. law restricts this device to sell by or on the order of a physician
This Operators Manual contains confidential and proprietary information of the
Manufacturer.
Manufactured by: LightMed Corporation, No.1-1, Ln. 1, Pao-An St. Sec. 3,Shulin Dist.,
New Taipei City 23861, Taiwan
USA Address: 1030 Calle Cordillera, Suite 101, San Clemente, CA 92673
Tel No.: 949-218-9555 Fax No.: 949-218-9556
Copyright ©LightMed Corporation
EU Representative: Medical Device Safety Service GmbH
Schiffgraben 41, 30175 Hannover, Germany
Document Title:
Operators Manual for the LightLas 532 Ophthalmic Laser
Document Number:
DC1001
Document Revision History:
DRAFT
March 2014
Draft prepared
01
August
Release 01 version

LightLas 532 –Operator's Manual Rev. No 01 Page 4of 115
Table of Contents
Section 1 INTRODUCTION.................................................................................. 9
Section 2 SAFETY............................................................................................. 10
2.1 Product Classifications ............................................................... 10
2.2 Warnings and Precautions......................................................... 11
2.3 Optical Hazards ......................................................................... 12
2.3.1 Nominal Ocular Hazard Distance (NOHD)........................... 12
2.3.2 Avoid Exposure to Laser beams.......................................... 13
2.4 Electrical Hazards ...................................................................... 13
2.5 Safety Controls and Features ..................................................... 14
2.6 Product Labeling ........................................................................ 14
2.6.1 Console system.................................................................. 14
2.6.2 Integrated Slitlamp LDU.................................................... 16
2.6.3 Laser Indirect Ophthalmoscope LDU................................... 18
2.6.4 Attachment LDU ................................................................ 19
2.6.5 Safety Filter (Manually) ..................................................... 20
Section 3 PRODUCT SPECIFICATIONS............................................................... 21
3.1 LightLas 532 System Specification ............................................. 21
3.2 CSO SL980 and SL990 Slitlamp Specifications ............................. 24
Section 4 PRINCIPLES OF OPERATION .............................................................. 25
4.1 General Description ................................................................... 25
4.2 LightLas 532 Laser Controls and Displays ................................... 33
4.2.1 Laser Console.................................................................... 33
4.2.2 Slitlamp Delivery Unit ....................................................... 40
4.2.3 Laser Indirect Ophthalmoscope (LIO)................................. 44
4.2.4 Two types of Integrated CSO SL980 and SL990
Slitlamp Controls Breakdown............................................ 47
Section 5 INSTALLATION.................................................................................. 51
5.1 Introduction and Requirements ................................................. 51
5.2 Unpacking and Receiving Inspection .......................................... 52
5.3 Tools and Equipment ................................................................. 54
5.4 Setting Up the Laser System Parts.............................................. 55
5.4.1 Slitlamp Integrated LDU.................................................... 56
5.4.2 Laser Indirect Ophthalmoscope LDU (LIO).......................... 68
5.4.3 Endoprobes LDU ............................................................... 69
5.5 Pre-check / Alignment Procedures ............................................. 72
5.5.1 Pre-check Laser console operation .................................... 72
5.5.2 Pre-check / Alignment LDU ............................................... 74

LightLas 532 –Operator's Manual Rev. No 01 Page 5of 115
Section 6 CLINICAL USE.................................................................................... 87
6.1 Different types of LDUs Indication / Contraindication Use....88
6.1.1 Slitlamp Delivery Unit ........................................................ 88
6.1.2 Laser Indirect Ophthalmoscope (LIO).................................. 88
6.1.3 Endoprobe devices............................................................. 89
6.2 General Warnings...................................................................... 90
6.3 Possible side-effect or adverse reactions .................................... 92
Section 7 MAINTENANCE................................................................................. 93
7.1 Operator / User Maintenance..................................................... 93
7.2 Laser Beam Alignment Check...................................................... 94
7.3 System Output Power Checking Procedure................................. 95
Section 8 TROUBLESHOOTING .......................................................................... 97
8.1 Symptom / Warning ................................................................... 97
8.2 Warning ..................................................................................... 99
8.3 Error Codes............................................................................... 102
Section 9 European Community Issues............................................................. 106
Section 10 EMC Test Tables ............................................................................... 108

LightLas 532 –Operator's Manual Rev. No 01 Page 6of 115
LIST OF DRAWINGS / FIGURES
Figure
Description
Page
2.1
Laser Console Safety and Controls Labels
14
2.2
Laser Console Safety labels
15
2.3(a)
Integrated Slitlamp LDU with Labels (RH side)
16
2.3(b)
Integrated Slitlamp LDU with Labels (LH side)
17
2.4
LIO LDU with Labels
18
2.5
Attachment LDU with Labels
19
2.6
Microscope Doctor Safety Filter with Labels
19
2.7
Microscope Safety Filter with labels
20
4.1
System software start up screen shot
29
4.2
System software boot up ok and system is "Standby" mode
29
4.3
System software is ready to shutdown
32
4.4
Laser Console Controls and LCD Displays
33
4.5
LCD Panel Displays and Control
34
4.6
Integrated Slitlamp LDU Controls
40
4.7(a)
Attachment Slitlamp LDU Controls
41
4.7(b)
Truspot Attachment Slitlamp LDU Controls (New version)
41
4.8(a)
Front View of LIO LDU Controls
44
4.8(b)
Top View of LIO LDU Controls
44
4.9
SL980 Slitlamp Parts List and Controls
47/48
4.10
SL990 Slitlamp Parts List and Controls
49/50
5.1
Packing Carton for Integrated LDU and Slitlamp
52
5.2
Portable Carry Cases for Console and LDU’s
53
5.3
Laser System Parts
55
5.4
Slitlamp table top ass'y
57
5.5(a)
Rear view box connection
57
5.5(b)
Removal fuse socket
58
5.5(c)
Removal fuses
58
5.5(d)
Removal voltage setting block and reinsert it back upon completion
59
5.6
Slitlamp Ass’y
59
5.7
Assembled Upper and Lower Housing Screw
60
5.8
Slitlamp mounted onto Table Top ready for Integrated LDU
60
5.9
Slitlamp arm ready for Delivery housing
61
5.10
Fit Delivery unit housing to the Slitlamp
62
5.11
Lock the housing securely to the Slitlamp when it is pushed against
Chrome Stop screw
62

LightLas 532 –Operator's Manual Rev. No 01 Page 7of 115
Figure
Description
Page
5.12
Align the keyways then fit the Zoom unit to the Delivery housing
63
5.13
Rotate the Zoom housing to the end stop
63
5.14(a)
Original Illumination Tower
64
5.14(b)
New modified Illumination Tower fitted
64
5.15(a)
Haag Streit Attaching Tonometer mount ready for Attachment LDU
65
5.15(b)
Zeiss 30SL Attaching the Tonometer mount and Attachment LDU
Mounting Arm
65
5.15(c1)
Lightmed SYL9000 Attaching Tonometer mount ready for Attachment
LDU
66
5.15(c2)
Lightmed SYL9000 Attaching the Tonometer mount and Attachment
LDU Mounting Arm
66
5.16(a)
Haag Streit Mounting the whole attachment LDU to Tonometer mount
& micromanipulator
67
5.16(b)
Zeiss Style 30SL Mounting the whole attachment LDU to Tonometer
mount & micromanipulator
67
5.16(c)
Lightmed SYL9000 Mounting the whole attachment LDU to Tonometer
mount & micromanipulator
67
5.17
Fiber tips cleaning
68
5.18
LIO connections in console
68
5.19
LIO LDU fiber connection
68
5.20
Endoprobe plug installation
69
5.21(a)
Standard System connection set-up
71
5.21(b)
Combo System Y-Joint connection set-up
71
5.21(c)
Sample of Delivery Fiber and Delivery Key Connection
71
5.22(a)
System boot up sequential display #1
72
5.22(b)
System boot up sequential display #2
72
5.22(c)
Initial Power on displays and outputs
72
5.22(d)
Sample of Delivery Fiber and Delivery Key Connection
72
5.23
Safety filter checks
73
5.24
Attach the Delivery fiber to the Zoom unit
73
5.25
Upper / Lower Slit Alignment Aiming Integrity and Focus Checkout
75
5.26
Realignment set screw location (X-Y axis)
76

LightLas 532 –Operator's Manual Rev. No 01 Page 8of 115
Figure
Description
Page
5.27(a)
Checking Field of View set-up
76
5.27(b)
Checking manipulator operation on the target
76
5.28(a)
New Type Version Check Slitlamp focus on Target Rod and Laser focus
of aiming beam
79
5.28(b)
Old type Version Adjust the Attachment LDU Spot Size Focus using the
knob on the mounting arm
80
5.29
Old type version Adjusting Micromanipulator Arm to verify Laser Spot is
in the Aperture Center
80
5.30
Old / New type version Adjusting sideways movement of Attachment
LDU during alignment
81
5.31(a)
Fiber and Delivery Key fitted to Console Front Panel
82
5.31(b)
LIO Intensity Control Knob
83
8.1
Sample of Error code display
102
Record Sheet
Installation Record Sheet
84/85/86
Power Meter Calibration Record Sheet
96
Appendix I
All the Accessories Listing Detailed
113/114

LightLas 532 –Operator's Manual Rev. No 01 Page 9of 115
Section 1 INTRODUCTION
This manual is intended to provide the operator with an overview of the operation
and safety requirements for the LightLas 532 Ophthalmic Laser. This manual is not
intended to provide instructions on actual treatment procedures and it is expected
that users will have undertaken training prior to using the equipment.
The Manufacturer and Distribution organization assume no liability through the use
of this Laser system.
All care has been taken in the preparation and checking of this manual however there
is no guarantee provided that all information is correct. The information provided in
this manual is subject to change without notice.
Only approved or authorized accessories may be used in the LightLas 532. The
Manufacturer and Distribution organization shall not be held liable or responsible for
damages or injury caused as a result of using non-approved accessories. This includes
all Optical Fiber systems, Laser Delivery Units, Safety Filters, Safety Glasses and Table
units.
All maintenance and service work must be carried out by authorized and trained
service agents and only those procedures outlined in the operator and service
manual are allowed. Any service work carried out by unauthorized persons will void
all warranties.
No circuit diagrams or component part lists are to be supplied for the LightLas 532. If
you require technical documentation that is not provided in this manual then please
contact the manufacturer or your local distributor in writing with your reasons for
wanting them and then a copy of the service manual may be provided.
Before using the LightLas 532 Photocoagulator Laser system the operator should
read this manual carefully and pay particular attention to the sections of Safety,
Operation and Maintenance.

LightLas 532 –Operator's Manual Rev. No 01 Page 10 of 115
Section 2 SAFETY
This Laser system has been designed and tested to function in a safe and correct
when used as indicated in this manual.
Do not use this laser before reading and understanding completely this Operators
Manual.
It is important to remember that this laser emits high levels of visible laser radiation
which can cause permanent and irreparable eye and tissue damage. Always observe
precautions for laser safety including using warning signs, safety glasses and only
operating the laser in a treatment room that provides protection to casual observers.
2.1 Product Classifications
The LightLas 532 Photocoagulator Laser is a Class IV laser product as specified in the
standard IEC60825-1 (2007) and the USA 21 CFR’s 1040.10, 1040.11.
The LightLas 532 Photocoagulator Laser is classified as Class I Type B Electromedical
equipment as specified in the IEC60601-1 standard.
The LightLas 532 Photocoagulator Laser is classified as a Class II device according to
the FDA CFR21 regulations.
The LightLas 532 Photocoagulator Laser is classified as a Class II Type B Medical
Device according to the MDD 93/42/EEC (as amended by 2007/47/EC).
The LightLas 532 has been designed to comply with the following standards:
Laser standards
IEC 60825-1 (2007)
USA 21 CFR 1040.10, 1040.11 (1997)
IEC60601-2-22 (1995)
Electrical standards
IEC 60601-1:2005
EN 60601-1:2006
EN 60601-1-2:2007
IEC60601-1-2 :2007
USA UL 2601
JIS T1001 (1992) and T1002 (1992)
Others
MDD 93/42/EEC (as amended by 2007/47/EC)
EN 60601-1-6:2010
IEC 60601-1-6 :2010

LightLas 532 –Operator's Manual Rev. No 01 Page 11 of 115
EN62366 :2008
IEC62366 :2007
EN 980:2008
ISO14971 (2012)
2.2 Warnings and Precautions
The following warnings and precautions apply to the LightLas 532 Laser System and
should be observed by all users at all time:
DO NOT look directly into the laser beam or at laser reflections since direct
and reflected laser light from the laser aperture can cause permanent eye
injury.
DO NOT operate the laser unless observers are using the correct protective
eyewear. The protective eyewear must have an optical density of OD4 or
more at 532 nanometers wavelength. This information must be present on
the eyewear.
DO NOT use objects that can readily reflect light in the vicinity of the laser
beam to avoid reflecting the beam in a hazardous manner.
DO NOT fire the Laser directly onto flammable agents or gasses as the
focused laser beam may cause ignition. There is no AP/ APG protection.
DO NOT try to service or repair the laser other than what is included in this
manual. Service should only be performed by an authorized and trained
agent of the manufacturer.
DO NOT fire the laser on a patient without first checking the operation of the
laser and verifying the optical alignment of the treatment Laser beam.
ALWAYS use the lowest power settings possible when treating a patient
with the laser and start the treatment with minimum level of power.
ALWAYS set the correct spot size and/or use the most appropriate one for
the power setting and type of procedure that is to be performed.
DO NOT put the laser into ‘TREAT’mode until ready to operate on the
patient.
DO NOT inhale any laser plume generated by the Laser during surgery.
Personnel should take an extreme measured precaution, such as wearing
surgical masks or use plume evacuation systems when a treatment is
undergoing. Caution - Laser plume may contain viable tissue particulates.
ALWAYS take particular care of the optical fibers that connect the Laser
Delivery Units to the Console to make sure they do not get damaged.

LightLas 532 –Operator's Manual Rev. No 01 Page 12 of 115
Additional clinical warnings may be found in Section 6.1.4 of this Manual.
ALWAYS try to let the laser have its own or dedicated power outlet.
Additional items may be plugged in a Multiple Portable Socket Outlet, which
may be plugged into an additional outlet.
DO NOT use the Laser Console if the ambient temperature is outside the
range of 20 to 35°C. This temperature range is the rated operating
temperature limits where the Laser system can be guaranteed to operate
without any interruptions to normal use. Outside this range of temperature
it is possible that the Laser will generate an error condition where the word
“hold” is displayed and the system goes to Standby until the internal
temperature returns to within normal limits then the Laser can be used again
but the error condition may reoccur unless the rated temperature comes
within limits.
2.3 Optical Hazards
Guidance for the safe use of Lasers and Laser systems is found in the standard
IEC60825-1, the USA 21CFR 1040.10, 1040.11 and ANSI Z136.1 - 1986.
During normal operation of the LightLas 532 the operator is protected from Laser
hazards by built in optical absorption Safety Filters. All other personnel in the area
should wear protective eyewear to eliminate the risk of eye injury occurring.
The optical density (OD) of eye protection must be greater than or equal to 4 and the
wavelength 532nm range is also specified on it. It is shown in the following format:
OD4+ @ 532nm
Otherwise, the safety glasses are NOT suitable for this purpose of eye protection.
Safety Glasses are required to have the CE mark applied if used in the EU.
The LightLas 532 uses a Class II Laser Diode Aiming beam. Its wavelength range from
635 to 650 nanometers (nm) and the maximum power output is set at the factory to
be less than 1mW delivered to the patients cornea. However it is always
recommended to use the lowest aiming beam intensity during treatments.
The LightLas 532 Photocoagulator Laser has been classified as a Class 4 and its
classification specified accordingly to the above quoted standards. This classification
is also based on the Accessible Emission Limits (AEL) as calculated according to the
standards.
2.3.1 Nominal Ocular Hazard Distance (NOHD)
The Nominal Ocular Hazard Distance (NOHD) is the distance between the equipment

LightLas 532 –Operator's Manual Rev. No 01 Page 13 of 115
and a person’s eye for which the optical power, from the equipment, entering the
dilated pupil of the person will be less than or equal to the Maximum Permissible
Exposure (MPE) as specified in the standards (i.e. less than a Class 1 Laser Output).
The calculated NOHD for the LightLas 532 with different Laser Delivery Units is:
5 meters at maximum power settings for the Endoprobes
18 meters at maximum power settings and 1000m spot size for Slitlamp Delivery
Units
20 meters at maximum power settings for LIO
Therefore when the laser is in operation, all persons that are closer than these
distances to the equipment should be wearing eye protection.
Patients, where possible, should have the untreated eye covered or protected from
laser reflections.
2.3.2 Avoid Exposure to Laser beams
Reassembly or maintenance of the laser system should only be performed by
authorized and trained personnel. The external housing of the laser system should
never be removed otherwise you or standby observers could be exposed to
dangerous levels of laser radiation and potentially lethal electrical voltages.
Eye safety filters are designed to protect physician’s eyes from back scattered laser
beams must always be used. They are integrated into the Slitlamp and LIO Delivery
Units. When using the endoprobes a separate filter that attaches to the operating
microscope must be used.
For other personnel that may be exposed to reflections or backscatter they must
wear safety glasses or goggles. In any case NEVER look directly at the treatment
laser beam as severe eye injury is likely. This means avoid looking into the aperture of
any of the laser delivery units or the console.
2.4 Electrical Hazards
The Lightlas 532 Photocoagulator laser has been designed to apply the International
Standards for Medical Equipment. The laser system is designed to operate with three
terminal prongs AC voltage where the third prong pin is the earth-grounded prong.
Warning: It is not safe to operate the Lightlas 532 photocoagulator laser without an
earth-grounded receptacle. There is possible risk of electric shock.
No cover or housing need to be removed by the operator or user. Only the
authorized and trained service or agent can remove the cover or housing assembly

LightLas 532 –Operator's Manual Rev. No 01 Page 14 of 115
since there is possible of exposing laser radiation and high current or voltages.
2.5 Product Labeling
All the labels on the LightLas 532 comply with the requirements of the various
regulatory standards referred to previously.
2.5.1 Console system
A full facsimile of all the safety and control labels is shown in the figures 2.1 and 2.2.
EXPLOSIONSGEFAHR
Dieses Gerat ist nicht fur den Betrieb in
explosiongefahrdeten bereichen bestimmt
exposition dangereuse au rayonnement
Complies to the requirements of
21CFR, Chapter 1, Subchapter J
connect only to with a "Hospital
d'anesthesiques inflammables
Risques d'explosion si utilise en presence
Ne pas ouvrir L'appar eil risque de chocs electriques
Laser direct ou dif fuse reparations par service
For grounding reliability
Gefahr eines elektri schen Schlages
GEHAUSETEILE NICHT ENTFERNEN
Wartungsarbeiten nur durch qualifizierten
accessible laser radiation. Refer servicing to
Risk of explos ion if used in the presence of
Do not remove covers. Shock hazard and
qualified service personnel.
Laserstrahlung zuganglich.
flammable anesthetics.
Kundendienst.
ACHTUNG!
DANGER!
technique qualifie.
Grade receptacle".
DANGER!
FOOTSWITCH INTERLOCK
DOOR
fuse as marked above
WARNING
Risk of fire
Replace only with
AC MAINS INPUT
50/60Hz, 400VA
100-230V ~ (Fuse:T3.15AH250V)INPUT:
Danger/Caution
DIRECT OR SCATTERED RADIATION
AVOID EYE OR SKIN EXPOSURE TO
Visible and Invisible laser radiation when open
LIO
LIO
LIGHTMED
STOP
Delivery
Key
STOP
DIRECT OR SCATTERED RADIATION
AVOID EYE OR SKIN EXPOSURE TO
Diode Aiming Laser 635-650nm 1mW CW Max
Green Laser 532nm 2.0W CW Max
CLASS 4 LASER PRODUCT
LASER RADIATION DIRECT OR SCATTERED RADIATION
AVOID EYE OR SKIN EXPOSURE TO
Diode Aiming Laser 635-650nm 1mW CW Max
Green Laser 532nm 2.0W CW Max
CLASS 4 LASER PRODUCT
LASER RADIATION
Figure 2.1
Laser Console Safety and Control Labels

LightLas 532 –Operator's Manual Rev. No 01 Page 15 of 115
2.5.1 Console system continues...
Lightmed Corporation
SERIAL NO:
MODEL NO:
Lightmed Corporation
SERIAL NO:
MODEL NO:
MANUFACTURED: Manufactured by LightMed Corporation
No1-1, Lane 1, Pao-An Street, Section 3
Shulin City, Taipei 238, TAIWAN
Schiffgraben 41, 30175 Hannover, Germany
Medical Device Safety Service GmbH
(MDD 93/42/EEC as amended by 2007/47/EC)
EU Authorized Representative
EC REP
"Instructions Under Lid"
Lightmed Corporation
MODEL NO:
SERIAL NO:
Coiling
EXPLOSIONSGEFAHR
Dieses Gerat ist nicht fur den Betrieb in
explosiongefahrdeten bereichen bestimmt
exposition dangereuse au rayonnement
Complies to the requirements of
21CFR, Chapter 1, Subchapter J
connect only to with a "Hospital
d'anesthesiques inflammables
Risques d'explosion si utilise en presence
Ne pas ouvrir L'appareil risque de chocs electriques
Laser direct ou diffuse reparations par service
For grounding reliability
Gefahr eines elektrischen Schlages
GEHAUSETEILE NICHT ENTFERNEN
Wartungsarbeiten nur durch qualifizierten
accessible laser radiation. Refer servicing to
Risk of explosion if used in the presence of
Do not remove covers. Shock hazard and
qualified service personnel.
Laserstrahlung zuganglich.
flammable anesthetics.
Kundendienst.
ACHTUNG!
DANGER!
technique qualifie.
Grade receptacle".
DANGER!
FOOTSWITCH INTERLOCK
DOOR
fuse as marked above
WARNING
Risk of fire
Replace only with
AC MAINS INPUT
50/60Hz, 400VA
100-230V ~ (Fuse:T3.15AH250V)INPUT:
Danger/Caution
DIRECT OR SCATTERED RADIATION
AVOID EYE OR SKIN EXPOSURE TO
Visible and Invisible laser radiation when open
Figure 2.2
Laser Console Safety Labels

LightLas 532 –Operator's Manual Rev. No 01 Page 16 of 115
2.5.2 Integrated Slitlamp LDU
VISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED
RADIATION
CLASS 4 LASER PRODUCT IEC/EN 60825-1:2007
CAUTION/DANGER
VISIBLE LASER RADIATION
AVOIDEYE OR SK IN EXPOSURE
TODIRE CT OR SCATTERED
RADIATION
CLASS4 L ASER PRODUCT I EC/EN 6082 5-1:2007
Figure 2.3(a)
Integrated Slitlamp LDU with labels (RH side)

LightLas 532 –Operator's Manual Rev. No 01 Page 17 of 115
2.5.2 Integrated Slitlamp LDU continues...
AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
0434
Lightmed Corporation
OD4@532nm
Safety Filter Serial No:
Model No:
Visible laser radiation when open
DANGER / CAUTION
DIRECT OR S CATTERED RADIATION
AVOID EYE OR SKIN EXPOSURE TO
Visible laser radiation when open
DANGER / CAUTION
Figure 2.3(b)
Integrated Slitlamp LDU with labels (LH side)

LightLas 532 –Operator's Manual Rev. No 01 Page 18 of 115
2.5.3 Laser Indirect Ophthalmoscope LDU
Figure 2.4 LIO LDU with labels

LightLas 532 –Operator's Manual Rev. No 01 Page 19 of 115
2.5.4 Attachment LDU
Manufactured:
DANGER / CAUTION
Visible laser radiation when open
AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
CAUTION/DANGER
VISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED
RADIATION
CLASS 4 LASER PRODUCT IEC/EN 60825-1:2007
Manufactured:
Lightmed Corporation
Safety Filter OD4@532nm
Model No:
Serial No:
Lightmed Corporation
Serial No:
Safety Filter OD4@532nm
Model No:
Figure 2.5
Attachment LDU with labels
400300 500
500
VISIBLELASER RADIATION
AVOIDEYE OR SKIN EXPOSURE
TODIRECT OR SCATTERED
RADIATION
CLASS4 LASER PRODUCT IEC825-1(1993)
CAUTION/DANGER
CAUTION/DANGER
VISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED
RADIATION
CLASS 4 LASER PRODUCT IEC60825-1 2007
Manufactured:
Serial No:
Model No:
Lightmed Corporation
Safety Filter OD4@532nm
Safety Filter OD4@577nm
Lightmed Corporation
Manufactured:
Serial No:
Model No:
Figure 2.6
TruSpot Attachment LDU Labeling

LightLas 532 –Operator's Manual Rev. No 01 Page 20 of 115
2.5.5 Safety Filter (Manually)
Manufactured:
Manufactured:
OD4@532nm
Safety Filter
Lightmed Corporation
Serial No:
Model No: FILTER
Lightmed Corporation
Serial No:
Model No:
FIL
OD4@532nm
Safety Filter
Figure 2.7
Microscope Safety Filter with labels
Other manuals for LightLas 532
2
Table of contents
Other Lightmed Medical Equipment manuals