Myron L ULTRAPEN PTBT2 User manual

4 5
PTBT2
S
ACCURATE
S
RELIABLE
S
SIMPLE
Professional Water Analysis
myronl.com
PACKAGE CONTENTS:
PTBT2 Pocket Tester•
Pen - battery installed
Soaker Cap•
Scoop•
Pocket Clip•
Holster•
Lanyard•
Operating Instructions•
MADE IN USAMADE IN USA
ORIGINAL
EDITED FOR USE
Minimum height 5 mm
pH and Temperature Pen
for use with your mobile device
ULTRAPEN™
• Accuracyof±0.01pH
• ProprietarySensorDesign
• 1,2and3pointcalibration
• AutomaticTemperatureCompensation
• TemperatureReadout
• Waterproof
• Poweredby1NTypebattery(included)
• MeasurementandCalibrationstoragewith
time,dateandlocation.
• Exportstoredmeasurementdataviaemail.
Enabled
®
ULTRAPEN™
PTBT3
ORP & Temperature
ULTRAPEN™ PTBT4
Free Chlorine Equivalent
(FC
E
) & Temperature
ULTRAPEN™ PTBT1
Conductivity, TDS, Salinity
& Temperature
Coming soon
Also Available
Available at
Apple App Store
Vi sit our w eb site:
myronl.com/PTBT2.htm

1
INTRODUCTION
Thank you for purchasing the Myron L® ULTRAPEN™ PTBT2 pH Pen. This
instrumentisdesignedtobeextremelyaccurate,fastandsimpletouseindiverse
waterqualityapplications.Advancedfeaturesinclude1,2and3pointcalibration;
automatic temperature compensation; highly stable microprocessor-based
circuitry;waterproofhousing;andMobileDeviceApplicationthatisuserintuitive.
ThePTBT2iseasytocalibrateandeasytouse.
Downloading the Mobile Device App
1. GototheAppleAppStore.
2. Searchforthe“ULTRAPENPTBTx”iPhoneApp.
3. OpentheULTRAPENpageandtaptheInstallbutton.
4. BesuretoreadtheEndUserLicenseAgreement(EULA)andtheDataUseLiability
Agreement(DULA).
Opening the ULTRAPEN App
1. TapULTRAPEN App icononthemobiledevice’sHOMEscreen.
2. TheApp’smainMeasurementscreenwillappear.
NOTE: AppleiOSdevices:ActivateSIRIandsay,“ULTRAPEN.”
FEATURES
PTBT2
1. Push Button -turnsPTBT2on;enterscalibrationmode,activatesPTBT2preference
menus.
2. Battery Cap -providesaccesstobatteryforreplacement.
3. Pocket Clip -holdspentoshirtpocketforsecurestorage.
4. Wireless Transceiver Window -Bluetoothantennaislocatedhere.
5. LED Indicator Light -Tellstheuserwhentodip&swirlPTBT2duringmeasurementsand
calibration.IndicatesPTBT2isinConfigurationmode.
6. pH Sensor-measureshydrogenionconcentrationofsolution.
7. Soaker Cap-ContainspHSensorStorageSolutiontomaintainsensorhydration.To
remove,twistthecapwhilepullingoff.Usecautionnottospillsolution.Toreplace,fill
thecaphalf full ONLYwithsolution.Twistthecapwhilepushingon.Becareful,excess
solutionmaysquirtout.
DoNOTpushthesoakercapbeyondtheCapStopassensordamageWILLoccur.
NOTE:TheformationofKClcrystalsaroundthesoakercapisnormalanddoesnotaffectthe
sensorlife,performance,oraccuracyprovidedtheyarerinsedoffwithwaterpriortoatest.
8. Cap Stop -ShowshowfartopushthesoakercapwhenputtingitonthePTBT2.
9. Scoop-Usedtoholdsamplesolutionwhendippingisnotpossible.Touse,pushthe
scoopontothesensorwhileshiftingside-to-side.andholdthescoopdirectlyundera
verticalstream,avoidingbubbles.Toremove,pullthescoopoffwhileshiftingside-to-side.
IfthesensorisnotfullyseatedinthePTBT2,reinstallperpHSensorReplacementsection
onpage5.
Not shown: LanyardandHolster.
ULTRAPEN App Measurement Screen
1. Mobile Device Status Bar -standardStatusBarforyourmobiledevice.
2. Measurement Value Field -displaysthemeasuredvalueofthesolution.When
theULTRAPENturnsoff,displaysthemessage,“ULTRAPEN is Ofine”andvalues
grayout.
3. Units of Measure -displayspHunits.
4. ULTRAPEN Battery Level -ashesREDwhenULTRAPENbatteryis≤25%.
5. Solution Temperature -measuredtemperatureofthesolution.
6. Memory Store Button -tapheretorecordthemeasurementintheApp’sdatabase.
7. Memory Replace Button -tapheretoreplaceapreviouslystoredmeasurement
withdatafromanewmeasurement.
8. Feature Navigation Bar -thesebuttonsactivatevariousAppfeatures.
CAUTION:

1
OPERATING INSTRUCTIONS
The following are basic instructions for operating the PTBT2 ULTRAPEN &App. To
accessafullOperationsManual:
1. InsidetheApptapthePreferencesbuttonontheFeature
NavigationBar.
2. SelectHelpfromtheSettingsMenuscreen.
3. ThebrowserofyourmobiledeviceswillopenandlinktotheMyronL®Company
websitepagewherethePTBTxOperationsManualcanbeaccessed.
I. Connecting the ULTRAPEN to the Mobile Device
TopairandconnectthePTBT2toyourmobiledevice:
1. InsidetheApptaptheConnectbuttonintheFeatureNavigationBar.
2. PushandreleasethebuttononthePTBT2toturnitON.
3. WhenthePTBT2’snameappearsontheConnectScreen,selectthatline.
4. AcheckmarkwillappearnexttoPTBT2’sname.
NOTE: The default name for all PTBT2’s is, “MLC-PTBTx”.To edit the name, Tap the
Preferences button and go to ULTRAPEN Settings section. See the PTBTx
OperationsManualforfullinstructiions.
TounpairfromaPTBT2:
1. TaptheredUnpairbuttonnexttothePTBT2’sname.
2. AnUnpairInstrumentdialogueboxwillappear.
3. TaptheDisconnectbuttoninthedialoguebox.
II. Temperature Unit Selection
ThePTBT2allowsyoutoselecteitherCentigrade(°C)orFarenheit(°F)whendisplaying
temperatureontheiOSscreens.
ToselecttheTemperatureUnits:
1. InsidetheApptapthe PreferencesbuttonontheFeatureNavigationBar.
2. SelectApplication Preferencesfromthelistofsettingstopics.
3. SlidetheTemperatureUnitsswitchleftorrighttodisplay
temperaturevaluesaseither°Cor°F.
III. pH Mode Selection
ThePT2allowsyoutoselectbetweentwopHmeasurementmodes:
Hold(Default):Oncestabilized,thereadingiscapturedthendisplayed.
LIVE:Real-timereadingsaredisplayedcontinuouslyduringmeasurementuntilthe
ULTRAPENtimesout.
ToselectthepHMode:
1. InsidetheApptap the PreferencesbuttonontheFeature
NavigationBar.
2. SelectULTRAPEN Settingsfromthelistofsettingstopics
3. PushandreleasethebuttononthePTBT2toturnitON.
4 When the PTBT2 LED begins to ash, TheApp will display, “Press and hold the
button on the ULTRAPEN to enter Preferences”
5. PressandholdthePTBT2’spushbutton.
6. WhentheLEDgoestosolidONtheULTRAPENSettingsscreenwillappear.
7. SelectPen ModefromtheULTRAPENSettingsscreen.ThePenModescreenwill
appear.
8. TapthePen Modefieldonthescreen.
9. Choosethedesiredmode(HoldorLive)fromthelistthatappears.
10. Tapthe Applybutton.
IV. pH Measurement
ThefollowingtableexplainstheLEDIndicatorsignalsandtypicaldurations.
LED Indicator Signal Meaning Duration
RapidFlashing Dippeninsolution 6sec
SlowFlashing SwirlPen.Measurementinprocess
InLIVEmode,thereadingsare
dispayed.
10-30secinHoldmode
5minutesinLivemode
SolidONLight Measurementiscomplete
Finalreadingisdisplayed.
≤6sec
To measure solution at the extremes of the specified temperature range,
allowthepen to equilibrateby submerging thecellin the samplesolution of thesame
temperaturefor1minutepriortotakingameasurement.
NOTE:Ifyoucannotdipthepeninthesamplesolution,pourthesampleintoaclean
container.Ifyoudon’thaveasamplecontainerorneedtotestaverticalstreamof
solution,usethescoop.
2
CAUTION:

1
1. Rinsethepen.Ifmeasuringfromacontainer,submergethepenandswirlitaround
inFRESHsamplesolution3times.Alternatively,swirlfor30secondsunder
astreamorinabodyofwater.
2. IftheMeasurementscreenisnotdisplayed,taptheMeasurement
buttonintheFeatureNavigationBar.Thescreenwillsay,
“Paired ULTRAPEN is Ofine“.
3. Graspthepennearthepencaptoavoidsamplecontamination.
4. Removepenfromtherinsesolution.
5. PressandreleasethePTBT2’spushbuttontoturnitON.
6. WhentheLEDashesrapidly,IMMEDIATELYdipthepeninthesamplesolutionso
thatthecelliscompletelysubmerged.Ifyoudonotsubmergethecellinsolution
beforetheashingslows,allowthepentopoweroffandretakethereading.
7. TheAppscreenwilldisplay“Measuring ...”
8. WhentheLEDashesslowly,swirlthepentoremovebubbles,keepingthecell
submergedandpreferablyatleast1inch(2½cm)awayfromsides/bottomof
container.
a.InHoldmodewhentheLEDturnsonsolidthedisplaywillshowthefinalpHand
temperaturereadings.
b.InLIVEmodeallowthepentoremaininsolutionwhiletheLEDashesslowly.
ThedisplaywillshowlivepHandtemperaturereadings.Thepenwilltimeout
after5minute.Pushandreleasethepen’sbuttontopoweritOFFatany
timeduringLIVEmeasurement
9. TheULTRAPENwillturnOFFoncethemeasurementiscompleted.
V. Recording and Recalling Measurements
1. Onceameasurementisfinished,taptheMS button(MemoryStore).
2. TheAppscreenwilldisplaythemeasurementasanunsavedrecord.
3. TaptheSavebuttonintheupperrightcorneroftherecorddisplay.
4. TherecordwillbesavedandtheAppwillreturntothe
Measurementscreen.
5. TaptheMR button(MemoryRecall)intheFeatureNavigationBarto
displayalistofstoredrecords.
6. Swipeupanddowntorevealthefulllist.Tapanyrecordonthelist
toviewit’sdetails.
VI. Calibration
The Myron L® Company recommends calibrating twice a month depending on usage.
However, you should check the calibration whenever measurements are not as
expected.3-pointWetCalibrationismostaccurateandisrecommended.
NOTE:IfthemeasurementisNOTwithincalibrationlimits,“Error. Clean sensor. Check
Solution.”willdisplay.ChecktomakesureyouareusingaproperpHbuffersolution.If
thesolutioniscorrect,cleantheglassbulbofthesensorwithacottonswabsoakedin
isopropylalcohol.Restartcalibration.
NOTE:Smallbubblestrappedinthesensormaygiveafalsecalibration.
NOTE: If at any point during calibration, you do not submerge the sensor in solution
beforetheashingslows,allowthepentopoweroffandstartover.
A. Calibration Preparation
1. Formaximumaccuracy,fill2cleancontainerswitheachpHbufferyouwillbeusing.
Arrangetheminsuchawaythatyoucanclearlyrememberwhichistherinsesolution
and which is the calibration buffer. Always rinse the pH sensor between buffer
solutions.
2. EnsurethepHsensoriscleanandfreeofdebris.
NOTE:Ifyoudonothaveenoughbuffer,youcanuse1containerofeachbufferfor
calibrationand1containerofcleanwaterforallrinsing.
B. Multi-Point Calibration (2 and 3 point Calibration).
Toperformamulti-PointCalibrationyoumustalwaysstartwith7.0pHbuffer.
1. TaptheCALbuttonintheFeatureNavigationBar.
TheAppscreenwillappearandsay,“Paired ULTRAPEN is Ofine”.
2. Graspthepennearthebatterytoavoidsamplecontamination.
3. Removepenfromtherinsesolution.
4. Pressandreleasethepen’spushbutton.TheCalibrationscreenwilldisplay,“Press
and hold the button on the ULTRAPEN to enter CAL mode.”
5. PressandHoldthepen’sbutton.TheLEDwillgotosolidONandtheApp’sCalibration
screenwillappear
6. TapthegreenCALbuttonontheCalibrationscreen.
7. Whenthepen’sLEDashesrapidly,IMMEDIATELYdip
thepenincalibration7.0pHbuffersothatthecellis
completelysubmerged.TheCalibrationscreenwilldisplay.“Calibrating pH ...”
3
Continued on page 4 ...

4 5
8. WhentheLEDashesslowly,swirlthepenaroundtoremovebubbles,keepingthe
cellsubmerged.TheCalibrationscreenwilldisplay.“Calibrating 7.0 pH ...”
9. WhentheLEDlightgoessolidON,theCalibrationscreenwilldisplay
“7.0 ph CAL SAVED”alongwiththebeforeandaftercalibrationvalues.
10. DONEandCONTINUEButtonswillappearatthebottomoftheCalibrationScreen.
NOTE:TaptheDONEbuttontoendthecalibrationatthispoint,or;
11. Rinse the pH sensor with the next pH buffer (it can be either 4.0 pH or 10.0 pH
buffer).
12. TaptheCONTINUEButtontocalibratethenextpoint.
13. Repeatsteps7trough10.Theinstrumentwill
automaticallyrecognizethecalibrationpointbasedonthebufferused.
14. Afterthe3rdcalibrationpointissuccessfullysavedonlythe
DONEbuttonwillappear.Tappingthisbuttonendsthe
calibration.
C. Single-Point Calibration.
1. Usingeither4,7,or10pHbuffer,performthecalibrationasdescribedinsteps1
through9ofSectionB.
2. Afterthecalibrationpointissuccessfullysaved,taptheDONEbutton
D. Factory Calibration
WhenpHbuffersarenotavailable,thePTBT2canbereturnedtofactorydefault
calibrationusingtheFACCALfunction.Thiswillerasestoredwetcalibrationconstants.
NOTE:defaultfactorycalibrationresetstheelectronicsonlyanddoesNOTtakethe
conditionofthesensorintoconsideration.
1. InsidetheApptaptheCALbuttonintheFeatureNavigationBar.TheAppscreen
willsay,“Paired ULTRAPEN is Ofine.”
2. Pressandreleasethepushbutton.Thescreenwilldisplay,“Press and hold the
button on the ULTRAPEN to enter CAL mode”.
3. PressandHoldtheULTRAPEN’sbuttontheLEDwillgotosolidONandtheApp’s
Calibrationscreenwillappear.
4. TaptheredFAC CALbutton.Awarningdialoguebox
willappear.
5. TaptheOKbuttoninthedialoguebox.TheApp’sCalibratingscreenwilldisplay,
“FAC CAL Reset”.
MAINTENANCE
I. Battery Replacement
ThePTBT2Appdisplayhas an indicatorthatdepictsthepenbattery’s charge level.
Whenthechargelevelfallsbelow25%,theindicatorwillashred.Imediatelyreplace
thebatterywithannewNtypebattery.
1. InaCLEAN DRYplaceunscrewthebatterycapinacounter-clockwisemotion.
2. Slidethebatteryandbatteryhousingoutofthepen.
3. Removethedepletedbatteryfromitshousing.
4. Insertanewbatteryintothebatteryhousingorientedwiththenegativeendtouching
thespring.
5. Alignthegroovealongthebatteryhousingwiththeguidebumpinsidethepencase
andslidethebatteryhousingbackin.
6. Screwthebatterycapbackoninaclockwisedirection.
... Continued from page 3
II. Routine Maintenance
1. ALWAYSrinsethepHsensorwithcleanwateraftereachuse.
2. ALWAYSreplacethesoakercaphalffilledwithpHSensorStorageSolutionto
preventthesensorfromdryingoutaftereachuse.
3. Ifthesensorbecomesdirty,cleanthesensorsurfacewithanisopropylsoaked
cottonswab.Thenrinsethoroughlywithcleanwater.
4. Donotdrop,throworotherwisestrikethepen.Thisvoidsthewarranty.
5. Donotstorethepeninalocationwheretheambienttemperaturesexceedits
specifiedOperating/StorageTemperaturelimits.
Do not over tighten.

4 5 PTBT2TFM 08-14
SPECIFICATIONS
MYRON L®COMPANY
2450ImpalaDrive•Carlsbad,Ca92010-7226•Phone:+1-760-438-2021
E-Mail:CustomerService–[email protected]•TechnicalSupport–[email protected]
Website:www.myronl.com
MADEINUSA ©2014 MYRON L® COMPANY
Do not over tighten.
AppRequirements: AppleiOS6oriOS7.OptimizedforiOS7.
MeasurementRange: 0.00–14.00pH;Temperature:0°-71°C/32-160°F
Accuracy(AfterWetCalibration): ±0.01pH;Temperature±0.1ºC/±0.1ºF
Repeatability: ±0.01pH;Temperature±0.1ºC/±0.1ºF
Resolution: 0.01pH;Temperature±0.1ºC/±0.1ºF
TimetoReadingStabilization: 10-30seconds
ActiveModePowerConsumption: ActiveMode137mA,SleepMode2µA
TemperatureCompensation: Automaticto25ºC
PhysicalDimensions: 17.15cmLx1.59cmDor6.75in.Lx.625in.D
Weight: 54gor1.94oz
CaseMaterial: AnodizedAircraftAluminumwithProtectiveCoating
BatteryType: Ntype,Alkaline1.5V
CalibrationStandardSolution: 4.0pH,7.0pH,10.0pH
Operating/StorageTemperature: 0-55ºC/32-131ºF
WaterResistance: IP67andNEMA6
EN61236-1:2006-AnnexA:
2008;1999/5/ECAnnexIII
Electrostaticdischargetocaseofinstrumentmay
causePTBT2tospontaneouslypoweron.Inthiscase,
thePTBT2willpoweroffafterseveralseconds
FCCID:T7VPAN17 TheBluetoothtransceiverdevicemeetstherequirementsfor
modulartransmitterapprovalasdetailedinFCCpublicNotice
DA00-1407.
Canada(IC),license:
IC:216Q-PAN17
TheBluetoothtransceiverdevicemeetstherequirementsfor
modulartransmitterapprovalasdetailedinRSS-GEN.
ORIGINAL
EDITED FOR USE
Minimum height 5 mm
QUICK REFERENCE INSTRUCTIONS
1. TurnonULTRAPENApp:TapULTRAPENicononmobiledevice’shomescreen.
2. TurnonPTBT2:Pressandreleasepushbuttononpen.
3. LEDashesrapidly:Dippeninsamplesolutionsocellistotallysubmerged.
4. LEDashesslowly:Swirlpentoremovebubbles,keepcellsubmerged.
5. LEDturnssolidONthenturnsOFF:Measurementiscompleteandreadingisdisplayed.
Feature Navigation Bar
ThisbuttonalwaysreturnstheApptotheMeasurementScreen.
Thisbuttondisplaysalistofsavedmeasurements.
ThisbuttonplacestheAppinCalibrationmode.
ThisbuttonopenstheApp’sBluetoothConnect/Disconnectscreen.
ThisbuttonopenstheApp’spreferencesandsettingsmode.
AninternetlinktofullOperationsManual(Helpbutton)isfoundhere.
ACCESSORIES
STANDARDSOLUTIONSFORCALIBRATION:TheULTRAPEN™PT2requirespH4,pH7,andpH10
buffersolutionsforwetcalibrationandpHSensorStorageSolutionforproperstorage.
— Order MODEL#s: PH4, PH7, PH10, and PHSS.
ReplacementpHSensor(withinstructions)— Order Model: RPT2.
III. pH Sensor Replacement (Follow the instructions that come with your Replacement Sensor.)
OnlyRemove/ReplacethepHsensorinaCLEANandDRYenvironment!
ToremovethepHsensor:
1. Removethesoakercap;makesurethePTBT2(includingthepHsensor)isclean
anddry.Loosenthebatterytraytoallowpressureequalization.
2. ThenfirmlygraspthepHsensorbodyandslowlypullthepHsensorout.
ToinstallanewpHsensor:
1. LineupthealignmenttabsonthepHsensorwiththealignmentslotsonthePT2
unit.
2. GentlypushthepHsensorintoposition,closebatterycap.
CAUTION:
MYRON L®WARRANTY
Exceptingsensorandbattery,theMyronL®CompanyPTBT2pHULTRAPENiswar-
rantedtobefreefrommanufacturedefectfor1year.ThepHsencerhasasix(6)month
limittedwarrenty.
Other manuals for ULTRAPEN PTBT2
2
This manual suits for next models
3
Table of contents
Popular Temperature Controllers manuals by other brands

Siemens
Siemens Landis & Staefa RAV12 Series operating instructions

La Crosse Technology
La Crosse Technology Alerts TX60U-IT quick start guide

ELEKTRA
ELEKTRA ELR-10 installation manual

SUNWARD
SUNWARD SUN-15T user manual

LK Armatur
LK Armatur SmartComfort LK 120 instruction manual

TECH
TECH Sinum MC-02 quick start guide