National Instruments 9792 User manual

GETTING STARTED GUIDE
NI Wireless Sensor Network Devices
This guide describes how to install, configure, and verify the
National Instruments Wireless Sensor Network (WSN) devices
and NI software.
Figure 1. NI WSN System Components
1NIWSN-32xx Nodes
2 NI 9792 WSN Real-Time Gateway
3 NI WSN-9791 Gateway
4 PC Running NI Software
11
1
2
4
3
or

NI Wireless Sensor Network Devices 2 ni.com
The NI WSN system contains four main components: a PC running
Windows and NI software, a NI WSN gateway, NI WSN-32xx
node(s), and an Ethernet network as shown in Figure 1.
What You Need To Get Started
• PC running Windows and NI software
• NI-WSN software installation disc, which contains
NI Measurement & Automation Explorer (MAX)
• NI LabVIEW
– the NI WSN-9791 Gateway requires LabVIEW 8.6.1 or
later
– the NI 9792 WSN Real-Time Gateway requires
NI LabVIEW 2009 Service Pack 1 (SP1) (32-bit) or later,
and NI LabVIEW 2009 SP1 (32-bit) Real-Time Module or
later
– the NI WSN-3226 node requires LabVIEW 2010 or later
• NI WSN gateway and appropriate power supply
• NI WSN-3202/3212 nodes and four AA alkaline batteries or a
9–30 V power supply

© National Instruments Corp. 3 NI Wireless Sensor Network Devices
• NI WSN-3226 nodes and four AA alkaline batteries or a
5–30 V power supply
• RJ45 Ethernet cable/Ethernet network
Related Documentation
For a complete list of documentation related to NI WSN, refer to
ni.com/info and enter rdwsnrd. In addition to this guide, the
following documents may be useful when configuring your
NI WSN system.
•NI WSN-9791 User Guide and Specifications or NI 9792 WSN
Real-Time Gateway User Guide and Specifications
•NI WSN-32xx User Guide and Specifications
•MAX Help for WSN Devices, available in MAX from Start»
Programs»National Instruments»NI-WSN»Configuring
WSN in MAX
• Refer to the NI-WSN book on the Contents tab of the LabVIEW
Help, which is available by selecting Help»Search the
LabVIEW Help in LabVIEW.
•NI-WSN Readme, available on the installation disc included
with your device

NI Wireless Sensor Network Devices 4 ni.com
Caution Before installing your device, read all product
documentation to ensure compliance with safety, EMC,
and environmental regulations.
Step 1. Unpack and Set Up the Devices
Remove the devices from the packaging and inspect them. Contact
NI if the devices appear damaged. Do not install damaged devices.
Caution The devices are static sensitive. Always
properly ground yourself and the equipment when
handling or connecting to the devices.
Powering on and Mounting the NI WSN Devices
Refer to the NI WSN device documentation for more information
about powering on and mounting your devices. Before setting up
NI WSN nodes, remove and save the serial number sticker located
on the back of NI WSN device for future use when configuring
your device in the software.

© National Instruments Corp. 5 NI Wireless Sensor Network Devices
Step 2. Install the NI-WSN Software on the
PC Host
You must be an Administrator to install NI software and devices on
your computer. To use the NI WSN-9791 Gateway, you must have
NI LabVIEW 8.6.1 or later installed. To use the NI 9792 WSN
Real-Time Gateway, you must have NI LabVIEW 2009 SP1
(32-bit) or later, and NI LabVIEW 2009 SP1 (32-bit) Real-Time
Module or later installed. To use the NI WSN-3226 node, you must
have NI LabVIEW 2010 or later installed. To install the NI-WSN
software, complete the following steps:
1. Close all other applications.
2. Install NI LabVIEW.
3. For the NI 9792, refer to the instructions included with your
NI LabVIEW software and install the NI LabVIEW 2009 SP1
(32-bit) Real-Time Module or later.
4. Insert the NI-WSN software installation disc into your
computer.
5. Follow the onscreen instructions to install the NI-WSN
software.

NI Wireless Sensor Network Devices 6 ni.com
Step 3. Install NI-WSN Software on the NI-9792
If using a NI 9792, complete the following steps to install NI-WSN
software on the NI 9792 WSN Real-Time Gateway:
1. Launch MAX.
2. Expand Remote Systems to confirm that MAX detects the
NI 9792. If MAX is not detecting your device, refer to step 2
of Step 4. Configure the NI WSN System.
3. Click on Software and select Add/Remove.
4. Install NI-WSN for LabVIEW software.
Step 4. Configure the NI WSN System
You must configure the NI WSN devices in MAX before using
NI LabVIEW to take measurements. For information about
configuring the WSN system, refer to Start»Programs»National
Instruments»NI-WSN»Configuring WSN in MAX. To confirm
device recognition and configure the NI WSN system, complete
the following steps:
1. Launch MAX.

© National Instruments Corp. 7 NI Wireless Sensor Network Devices
2. Expand Remote Systems to confirm that MAX detects the
NI WSN-9791 or NI 9792 gateway. If plugged into an Ethernet
port on the local subnet or directly into the PC host, the device
appears in Remote Systems. If the device is plugged into a
port that is not on your subnet, try manually adding it by
right-clicking Remote Systems and selecting Create New»
WSN Gateway.
• If your DHCP server is set up to automatically register
hostnames, the hostname is registered as
NI-WSN9791<serial number> or NI-NI9792-<serial
number>. Enter the hostname and click Finish.
• If you know the device IP address, such as 192.168.0.2,
enter it and click Finish.
Note If you are still unable to detect your device, wait until
the POWER LED on the device is green and the STATUS
LED turns off. Wait one minute to allow the device to
connect to the network and then press <F5> to refresh
MAX. If the device does not appear, contact your system
administrator to confirm that the network is working and
that a firewall is not interfering with device discovery. For
more information about the ports used for device discovery
and communication, refer to the NI WSN-9791 User Guide

NI Wireless Sensor Network Devices 8 ni.com
and Specifications or the NI 9792 WSN Real-Time
Gateway User Guide and Specifications.
3. Select the NI WSN Gateway and open the Network Settings tab.
a. Confirm or change the default settings.
b. If the NI WSN gateway is connected to a DHCP server,
select Obtain an IP address automatically. For static
IP addresses, select Use the Following IP address.
c. Click Apply to save changes.
4. To add a WSN node to a gateway, complete the following:
Adding a Node to the NI WSN Gateway
1. Select the gateway you want to use:
•(NI 9792 WSN Real-Time Gateway) Expand NI-9792, select
Devices and Interfaces, and then select WSN.
•(NI WSN-9791 Gateway) Select the NI WSN-9791 and then
click the WSN Nodes tab.
2. Click Add Node.
3. In the Add Node dialog box:
a. Select the Type.

© National Instruments Corp. 9 NI Wireless Sensor Network Devices
b. Enter the NI WSN-32xx device Serial Number, which is
listed on the sticker on the underside of the device, and an
ID Number (number between 1 and 50).
4. Click Done if you are done adding nodes. If you want to add
other nodes, click Add Another.
Note Upon completion, these changes are automatically
applied.
5. After nodes are added in MAX, nodes connect automatically
when they are in range of a gateway.

NI Wireless Sensor Network Devices 10 ni.com
Note If your node does not connect automatically, refer
to ni.com/info and enter exxhqv for troubleshooting
information.
Step 5. Attach Sensors and Signal Lines
Attach sensors and signal lines to the terminal block or accessory
terminals for each device. For device pinouts, refer to the device
documentation.
Step 6. Start Using the NI WSN System
To use the NI WSN system, open a new LabVIEW project. If the
gateway is on the local subnet, complete the following steps:
1. Right-click the project name and select New»Targets and
Devices. In the Add Targets and Devices on My Project dialog
box, select Existing target or device and expand the
WSN Gateway folder. When you expand the WSN Gateway
folder, the network is scanned for any connected NI WSN
gateway devices. Wait for device detection.
2. Once detected, click the NI WSN gateway device you wish to
use and click OK. Clicking OK adds the NI WSN-9791 or
NI 9792 device to the Project Explorer window.

© National Instruments Corp. 11 NI Wireless Sensor Network Devices
3. Expand the listing for the NI WSN gateway in the
NI LabVIEW project to see the NI WSN-32xx nodes
connected to it. Expand each node to list the I/O variables.
If the gateway is not on the local subnet, complete the
following steps:
a. Right-click the project name and select New»Targets and
Devices,click Existing target or device, and click
Specify a target or device by IP address.
b. Select IP Address and enter the IP address.
c. Expand the WSN Gateway folder and select the Gateway
type and click OK.
d. To add the nodes, right click the gateway in the project and
select New»Targets and Devices. Expand the WSN Node
folder, select the node, and click OK.

NI Wireless Sensor Network Devices 12 ni.com
4. To configure the I/O variables properties (such as Range and
Sample Interval), right-click on the node and select
Properties.
5. To acquire data, create a new VI, and drag the I/O variable to
the block diagram, create an indicator, and click Run. For
more information about configuring the WSN system in
NI LabVIEW, select Help»Search the LabVIEW Help and
enter WSN.

© National Instruments Corp. 13 NI Wireless Sensor Network Devices
Programming Examples
NI LabVIEW includes example programs to help you develop an
application. To locate LabVIEW examples, go to ni.com/info
and enter wsnexamples. For additional examples, refer to
zone.ni.com.
Troubleshooting
• Go to ni.com/support/install.
• If you need to return your National Instruments hardware for
repair or device calibration, go to ni.com/info and enter
redesign to start the Return Merchandise Authorization
(RMA) process.
Worldwide Technical Support
For support information, refer to the Technical Support Information
document available at ni.com/manuals. Also visit ni.com/
support or ni.com/zone. For support available at the National
Instruments worldwide offices, visit ni.com. National Instruments
corporate headquarters is located at 11500 North Mopac
Expressway, Austin, Texas, 78759-3504.

© 2009–2011 National Instruments Corp. All rights reserved.
372781D-01 May11
LabVIEW, National Instruments, NI, ni.com, the National Instruments corporate logo, and the Eagle logo are
trademarks of National Instruments Corporation. Refer to the Trademark Information at ni.com/trademarks
for other National Instruments trademarks. Other product and company names mentioned herein are trademarks or
trade names of their respective companies. For patents covering National Instruments products/technology, refer
to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the
National Instruments Patent Notice at ni.com/patents. Refer to the Export Compliance Information at
ni.com/legal/export-compliance for the National Instruments global trade compliance policy and
how to obtain relevant HTS codes, ECCNs, and other import/export data.
Other manuals for 9792
1
Table of contents
Other National Instruments Gateway manuals

National Instruments
National Instruments NI 9792 User manual

National Instruments
National Instruments NI 9795 WSN C Series User manual

National Instruments
National Instruments WSN-9791 User manual

National Instruments
National Instruments NI 9792 WSN User manual

National Instruments
National Instruments WSN-9791 User manual

National Instruments
National Instruments NI WSN C Series User manual

National Instruments
National Instruments 9792 User manual

National Instruments
National Instruments NI 9796 User manual

National Instruments
National Instruments MON-10496 User manual