Neomeris Basic pH Pocket-Tester User manual

Max-Planck-Straße 16 | D-31135 Hildesheim | Fon: +49 (0) 5121 7609-0 | Fax: +49 (0) 5121 7609-44 | www.neomeris.de
Instruction manual
Pocket – Tester


2
3
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Table of contents
Data sheet 3
Product description 4
Instructions for pH1 / Conductivity 1 6
Calibration procedure 7
Instructions for pX4 / pH5-Food / ORP5 / Cond5 / PC5 / PC6 8
Calibration procedure 9
ERRORS DESCRIPTION 12

2
3
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Data sheet
Basic pH
Pocket-
Tester
(pH 1)
Basic
Conductivity /
TDS
Pocket-Tester
(Cond. 1)
Advanced
pH / mV /
Temp. Food
Pocket-
Tester
(pH 5 food)
Advanced
pH / ORP
/ Temp.
Pocket-
Tester
(pX 4)
Advanced
ORP / Temp.
Pocket-
Tester
(ORP 5)
Advanced
Conductivity /
TDS /
Salt. / Temp.
Pocket-Tester
(Cond. 5)
Advanced
pH / mV
/ Cond. /
TDS /Salt.
/ Temp.
Pocket-
Tester
(PC 5)
Advanced
pH / mV /
ORP / Cond.
/ TDS / Salt.
/ Temp.
Pocket-
Tester
(PC6)
pH measuring range 0-14 - -2 - 16 - - -2 - 16
Resolution 0,1 / +0,2 - 0,01 / +00.2 - - 0,01 / +0,02
Points of calibration 1 - 2 - 1 - 3 - - 1 - 3
Buers 3 USA - 5 USA - - 5 USA
mV measuring range - - - 1000... + 1000 mV - - - 1000... + 1000 mV
Resolution - - 0,1 / 1 - - 0,1 / 1
mV (ORP) measuring range - - - 1000... + 1000 mV - - - 1000... + 1000 mV
Resolution - - 0,1 / 1 - - 0,1 / 1
Points of calibration - - 1 dened by the user - - 1 dened by
the user
Conductivity measuring
range -0,01μS - 199,9 - - - 0,01μS - 199,9 mS
Resolution -Automatic scale
/ +2 % fs - - - Automatic scale / +2 % FS
Points of calibration - 1 - 2 - - - 1 - 3
Temperature coecient -1,91 %/°C - - - 0,00 - 4,00%/°C
Reference temperature - 25 °C - - - 20 / 25 °C
TDS measuring range -0,01ppm - 199,
9 ppt - - - 0,01 ppm - 199,9 ppt
TDS factor - 0,40 - 1,00 - - - 0,40 - 1,00
Salinita measuring range - - - - - 0,01 mg/l - 100,0 g/l
Temperature range 0 - 60°C (nv) 0 - 60 °C
Resolution - 0,1 / +0,2°C
Measure unit °C / °F
System indication -Si -Si
Auto O After 8 minutes
Display LCD 3 color LCD backlight display
IP-protection IP 67
Power supply 2 x 1,5V AAA batteries
Data sheet

4
5
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Product description
KEYPAD
Button Function Action
Press to switch on/o the meter.
ESC Press to escape from setup menu or calibration provedure.
During measurement: Press to turn on/o the back light.
MODUS During measurement: Press to switch between
pH -> mV (pH) -> mV (ORP) -> Cond -> TDS -> SAL
During Setup: Press to scroll in the menu or increase the value of the selected parameter.
CAL During measurement: press to start the calibration of the selected parameter.
Press to conrm the calibration and setup value.
Keypad Functions for Tester pX 4 / pH 5-Food / ORP 5 / Cond 5 / PC 5 / PC 6
Pocket-Tester Advanced Pocket-Tester Basic
pX 4 / pH 5-Food / ORP 5 / Cond 5 / PC 5 / PC 6 pH 1 / Cond 1
Keypad Functions for Tester pH 1 / Cond 1
Button Function Action
Press to switch on/o the meter.
During Setup: Press to scroll in the menu or increase the value of the selected parameter.
CAL During measurement: press to start the calibration of the selected parameter.
Press to conrm the calibration and setup value.
Product description

4
5
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Product description
DISPLAY
The device is endowed with LCD display for Series 1; 3 colors backlight LCD for Series 4, 5 und 6.
• GREEN: Setup or measurement mode
• BLUE: Calibration mode
• RED: Error/Alarm
CALIBRATION POINTS INDICATOR
When a calibration is performed these icons indicate the points calibrated.
The instruments mentioned are supplied with a 475 mV solution.
Icon Working range pH - mode Conductivity mode
ᴏLow 4,01 pH 84 µS
ᴏMedium 7,00 pH* 1413 µS
ᴏHigh 10,01 pH 12,88 µS
*First point for pH calibration is always 7.00 pH
Note: ORP calibration on pX4, ORP5 and PC6 instruments is possible on only 1 point denable by the user.
POWER SUPPLY
These testers use 2 x 1.5V AAA alkaline batteries (already supplied).
Make sure to insert the batteries in the correct direction, following the silk-screened indications on the trans-
parent body of the Tester itself.
• Properly dispose of waste batteries according to current legislation.
• Replace all batteries together with batteries of the same type.

6
7
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Instructions
Instructions for pH1 / Conductivity 1
POWER ON
Press key once, the meter will switch On and the display will show all the segments active for 2 sec.
Then it will display the following:
All segments ON -> model name + software release -> measurement
POWER OFF
To turn o the meter press key for 3 seconds, the meter will switch o.
SETUP-MENÜ FOR PH1 / CONDUCTIVITY 1
1. With the meter turned o, press and hold the CAL key and press once.
2. The meter will switch on with all the segments active, release CAL key, the meter will
go into the Setup Menu.
3. The display will show SLT on primary display with COND ashing (only for Cond1).
4. Press to select the parameter between COND or TDS to be used for measurements
and press to conrm (Only for Cond1). Only if the TDS is selected then the display will
show TDS Fct ashing, press to change this factor and then press to conrm
(Only for Cond1).
5. The display will show rSt (RESET): nO ashing.
6. Press and select YES if a reset of the meter is required and then press to conrm.
7. At this point the meter nishes SETUP menu and switches o.
Note: To skip the changing of the value simply conrm the ashing value with key, the meter will go to the
next Parameter.
MEASUREMENT
Rinse the electrode with distilled water or sample prior to start measurement.
Fill the measuring cap with sample, switch on the meter and immerse the Tester in sample and wait for stabi-
lity, when stability icon appears on display take the reading.
During measurement make sure that pH electrode membrane is free from air bubbles, and that there
is not any air bubble around or between conductivity sensor.

6
7
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Calibration procedure
Calibration procedure
CONDUCTIVITY (COND1)
1. Power ON the meter by pressing the key.
2. Rinse the probe with distilled water.
3. Immerse the probe in the calibration solution (1413µS or 12.88 mS), wait for stability.
4. Press CAL key.
5. The meter will start calibration procedure and will recognise automatically the standard
used.
6. When calibration is stable press to conrm and complete the calibration.
7. The standard value will ash for 3 times and then the meter will go into the measure-
ment mode.
8. If the 2nd point calibration is required then rinse the probe with distilled water and im-
merse in the second standard solution (1413µS or 12.88 mS), wait for stability .
9. Repeat the points 4 to 7.
10. The calibration process is completed and the meter is ready to use.
Note: Anytime press key to abort and exit from calibration procedure.
• Reading based on theoretical cell value C=1
• Standard solution
TDS (COND 1)
When the meter is set to read TDS (see Paragraph “Setup Menu” at page 5), then the calibration is done on
TDS with 1 or 2 points.
The calibration procedure for TDS is the same as for conductivity.
PH (PH 1)
1. Power ON the meter by pressing the key.
2. Rinse the electrode with distilled water.
3. Immerse the electrode in the 1st buer solution pH7.0 and wait for stability.
4. Press CAL key.
5. The meter will start calibration procedure and will recognise automatically the standard
used.
6. When calibration is stable press to conrm and complete the calibration.
7. The standard value will ash for 3 times and then the meter will ask for next point for calibration.
If only 1-point calibration is required then press to nish and exit.
8. If the 2nd point calibration is required then rinse the electrode with distilled water and immer-
se in pH4.0 or pH10.0, wait for stability.
9. Repeat the points 4 to 7.
10. The calibration process is completed and the meter is ready to use.
Note: Anytime press key to abort and exit from calibration procedure.
Note 2: When the rst point calibration is conrmed (point 7) if the sensor is not removed from the buer
solution, the instrument may give wrong buer error.

8
9
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Instructions
Instructions
pX 4 / pH 5-Food / ORP 5 / Cond 5 / PC 5 / PC 6
POWER ON
Press key once, the meter will switch On and performs a test of 3 colours backlight display and then show
all the segments active for 2 sec.
Then it will display the following:
SETUP MENU FOR PX4 / PH5-FOOD / ORP5 / COND5 / PC5 / PC6
1. With meter switched o, press and hold CAL key and press key once..
2. The meter will switch on with all the segments active, release CAL key, the meter will go into the Setup
menu (green backlight during setup).
3. Press key to select the parameter to be changed between:
Function Primary display Secondary display Default value
Temperature unit (°C/°F) T. u. - °C
Reference temperature for conductivity trE 20 - 25 °C 25°C
Coecient for temperature compensation 0 - 4 %/°C tCC 1,91
TDS Factor 0,40 - 1,00 Fct 0,71
Reset to factory default Nein - Ja rSt Nein
MEASUREMENT
Rinse the electrode with distilled water or sample prior to start measurement.
Fill the measuring cap with sample, switch on the meter with and press MODE key to select desired para-
meter to be measured (green backlight during measurement). Immerse the Tester in sample and wait for
stability, when stability icon appears on display take the reading.
During measurement make sure that pH electrode is free from air bubbles, and that there isn’t any air
bubble around or between conductivity sensor.
4. Press to enable the value changing of the selected parameter.
5. The value of selected parameter will start ashing.
6. Press to change the value and then press to conrm.
7. The value stops ashing.
8. Press to select other parameters or press ESC to exit the setup menu.
Note: Any time press ESC key to exit from SETUP menu.

8
9
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Calibration procedure
CONDUCTIVITY (COND5 / PC5 / PC6)
1. Power ON the meter by pressing the key..
2. Rinse the probe with distilled water.
3. Immerse the probe in the calibration solution (84µS or 1413µS or 12.88mS), wait for stability.
4. Press CAL key, (blue backlight during calibration).
5. The meter will start calibration procedure and will recognise automatically the standard
used.
6. When calbration is stable press to conrm and complete the calibration.
7. The standard value will ash for 3 times and then the meter will go into the measurement
mode.
8. If the 2nd point calibration is required then rinse the electrode with distilled water and im-
merse in the 2nd Standard, wait for stability.
9. Repeat the points 4 to 7.
10. If the 3rd point calibration is required then rinse the electrode with distilled water and im-
merse in the 3rd Standard, wait for stability.
11. Repeat the points 4 to 7.
12. The calibration process is completed and the meter is ready to use.
1. Power ON the meter by pressing the key.
2. Rinse the electrode with distilled water.
3. Immerse the electrode in the 1st buer solution pH7.00 and wait for stability.
4. Press CAL key (blue backlight during calibration).
5. The meter will start calibration procedure and will recognise automatically the standard
used.
6. When calibration is stable, press to conrm and complete the calibration.
7. The standard value will ash for 3 times and then the meter will ask for next point for calibrat-
ion. If only 1-point calibration is required then press to nish and exist.
8. If the 2nd point calibration is required then rinse the electrode with distilled water and im-
merse in pH4.01 or pH10.01, wait for stability, otherwise press ESC to nish and exit.
9. Repeat the points 4 to 7.
10. If the 3rd point calibration is required then rinse the electrode with distilled water and im-
merse in the last buer, wait for stability, otherwise press ESC to nish and exit.
11. Repeat the points 4 to 7.
12. The calibration process is completed and the meter is ready to use.
Note: Anytime press ESC key to abort and exit from calibration procedure.
Note: Anytime press ESC key to abort and exit from calibration procedure.
Note 2: In case multipoint calibration is performed it is better to start from the lower value standard rst and
then go increasing.
PH (PX4 / PH5 / PC5 / PC6)
Calibration procedure

10
11
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
ORP (PX4 / ORP5 / PC6)
POWER OFF
To turn o the meter press key for 3 seconds, the meter will switch o. The instrument cannot be switched
o during calibration.
1. Switch on the instrument by pressing the key .
2. Rinse the electrode with distilled water and pat dry gently.
3. Immerse the electrode in the available Redox solution (es: 475 mV); wait for the stability.
4. Press CAL button (LED Blu backlight).
5. The instrument enters calibration mode and recognizes the Redox solution, with a certain
degree of uncertainty.
6. At the center of the display, the value ashes, in demonstration of the possibility, by the
user, to adjust this value by +/- 75 mV compared to what is shown on the display during
calibration. Make this operation, keep pressing MODE button. Initially the value will move by
units, then by tens.
Note: The value will only move upwards. If the initial value on the display is higher than the required
one (ex: 490 mV with respect to the 475 mV buer), keep pressing MODE until the value becomes lower
than the desired (-75 mV with respect to the value shown on the display); at that point it will be possible to
reach the desired value by keeping MODE pressed.
7. Press to conrm and complete calibration.
8. Calibration is complete and the instrument automatically returns to measuring mode.
Note: Press at any time to exit calibration.
SENSOR MAINTENANCE
After long time dry storage, put and leave the probe in storage solution (or pH 4) for at least 30 minutes to
reactivate the sensor.
Conductivity 1 - Conductivity 5: • Store the sensor dry after each use, once rinsed
with distilled water.
• Never touch conductivity probe with paper or any
tools (expecially the internal part); for cleaning
purpose only rinse with distilled water. If touched
the probe may damage.
pH1 – pX4 - pH5 – pH5 Food – ORP5 – PC5 – PC6: • Rinse the probe with distilled water before each
use.
• Store the sensor into STORAGE Solution (or pH 4)
after each use, once rinsed with distilled water.
• Never store pH sensor in distilled water!!!
Calibration procedure

10
11
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Calibration procedure
REPLACEMENT OF SENSOR
SETUP MENU FUNCTIONS
FOR ALL TESTERS
Tester 4, 5 and 6 series have replacement sensors which can be replaced in case it expires of damages.
1. To replace the sensor unscrew the dial in anti-clock wise.
2. Pull out sensor from unit body.
3. Put a new sensor by correctly matching the sign of dent.
4. Be sure that all the gaskets are good and in correct position.
5. Screw the dial tightly.
Function pH 1 Conductivity pX4 / pH 5 / ORP 5 Conductivity 5 PC 5 / PC 6 Reset
COND / TDS
selection
- - - Yes Yes -
TDS factor - 0,40 - 1,00 - 0,40 - 1,00 0,40 - 1,00 0,71
°C / °F Yes, nv Yes, nv Yes Yes Yes °C
T ref for COND - 25 °C - 20 / 25 °C 20 /25 °C 25 °C
T coecient - 1,91 % - 0 - 4,00 % / °C 0 - 4 % / °C 1,91 %
Reset Yes Yes Yes Yes Yes /

12
MF
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Max-Planck-Straße 16 I D- 31135 Hildesheim I Fon: +49 (0) 5121 7609-0 I Fax: +49 (0)5121 7609-44 I www.neomeris.de
Error Description
ERRORS DESCRIPTION
Error Contents Checking
Er 1 Wrong pH buer solution or the recognition of
calibration solution out of range.
1. Check whether buer solution is correct.
2. Check whether the meter connects the
electrode well.
3. Check whether the electrode is damaged.
Er 2 Press key when measuring value is not
stable during calibration.
Press key when the icon appears.
Er 3 During calibration, the measuring value is not
stable for ≥3min.
1. Check whether there are bubbles in glass
bulb.
2. Replace with new electrode.
Er 4 Electrode zero electric potential out of range
(<-60mV or >60mV)
1. Check whether there are bubbles in glass
bulb.
2. Check whether pH buer solution is correct.
3. Replace with new pH electrode.
Er 5 Elektrodensteilheit außerhalb des Bereichs.
(<85% oder >110%)
Er 6 pH measuring range out of range
(<0.00 pH or >14.00pH) 1 Series
(<-2.00 pH or >16.00pH) 5 Series
1. Check whether the electrode is suspended.
2. Check whether the meter connects the
electrode well.
3. Check whether the electrode is damaged.
The electrical and electronic equipment marked with this symbol cannot be disposed of in
public landlls.
According to the UE Directive 2002/96/EC, the European users of electrical and electronic
equipment can return it to the dealer or manufacturer upon purchase of a new one.
The illegal disposal of electrical and electronic equipment is punished with an administrative
ne.
DISPOSAL OF ELECTRONIC DEVICES
SAFETY INSTRUCTIONS
• Read this instruction manual carefully before using your new tester.
• The membrane of pH electrode is made of glass and can be danger in case it breaks. To avoid damage
check the electrode tip after each measurements.
• Replace all batteries together with same type.
• The manufacturer of these instruments cannot be held responsible for any improper use.
• Verication of the measuring results is the responsibility of the operator and the manufacturer does not
respond to any direct or indirect damage occurred while using this instrument.
This manual suits for next models
7
Table of contents
Other Neomeris Test Equipment manuals
Popular Test Equipment manuals by other brands

AEMC
AEMC OX 7102 Quick start user guide

Nitty-Gritty
Nitty-Gritty PASSI TEST PLUS owner's manual

BC Biomedical
BC Biomedical ULT-2000 series user manual

Apera Instruments
Apera Instruments ZenTest PH60S-Z user manual

WIKA
WIKA CalibratorUnit CPU6000 Series operating instructions

Fujikura
Fujikura AFL Telecommunications NOYES TURBOTEST 500B user guide