manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. NeuLog
  6. •
  7. Accessories
  8. •
  9. NeuLog NUL-223 User manual

NeuLog NUL-223 User manual

NEULOG DROP COUNTER LOGGER SENSOR GUIDE
NeuLog drop counter logger sensor
NUL-223
The NeuLog drop counter sensor can be used for any science experiment which
utilizes fluid drip rates, particularly in the fields of Chemistry and Biology.
The sensor comes pre-calibrated so you can start experimentation right out of the box
using this guide.
The NeuLog drop counter sensor excels in the laboratory but can also be used in the
field for a number of experiments in various configurations. For example, titrations
become much more accurate as you can determine both the total number of drops
and the total volume which has passed through the drop counter sensor.
The drop counter sensor's measurement units are:
Drops: The number of times a drop of liquid passes through the drop
counter sensor.
Milliliters (mL): The SI unit of volume. The NeuLog drop counter sensor
calculates the rough volume of drops which pass through it.
Setting up the drop counter sensor:
There are a number of ways to set up the NeuLog drop counter sensor to give
accurate readings. The drop counter gate includes three threaded holes and a
threaded utility rod which can be attached to various lab equipment to offer many
different orientations.
The most important aspect while setting up the drop counter sensor is to ensure that
the drops of liquid are passing through the sensor correctly. As in the figure below,
you will want to orient the sensor so that drops pass through the infrared beam
transmitted between the ends of the drop counter sensor.
Quick start procedure:
PC or Mac Computer
Materials needed:
NUL-223 Drop Counter Sensor
USB-200 USB Module
USB to mini USB cable (included with the USB-200 module)
Your drop counter sensor needs to be connected to a USB-200 module. The USB-200
module then connects to a computer via a USB to mini-USB cable. Please note that
you cannot plug the drop counter sensor directly into the computer.
The browser based application can be downloaded for free at www.NeuLog.com as
well as a full application user guide.
Choose "Downloads" on the main menu and then choose "Software and Application".
Procedure:
1. Install the NeuLog application.
2. Connect the USB-200 module to the PC or Mac.
3. Connect the drop counter sensor to the USB-200 module (they directly plug
together). Please note that no calibration is required for this sensor.
4. Open the NeuLog application by clicking on the shortcut on the screen.
5. Once a drop counter sensor module box appears on the left side of the screen
the probe has been automatically identified and you can begin experimentation.
6. If the drop counter sensor is not automatically identified then click the “Search
for sensors” icon to find the sensor.
7. Select the “On-line experiment” icon; this will open a graph below.
8. Click on the “Module setup” button located in the drop counter sensor module
box to change the sensor settings if need be.
9. Click on the "Experiment setup" button to change the experiment settings if need
be (experiment duration for example).
10. The drop counter sensor will give a live reading in the module box to the left of
the screen while plugged in.
11. To run an experiment and collect data click the “Run experiment” icon.
12. To end data collection early, click the “Stop experiment” icon.
NEULOG DROP COUNTER LOGGER SENSOR GUIDE
Tablet, smart phone device
Materials needed:
NUL-223 Drop Counter Sensor
WIFI-201 WIFI Module
BAT-200 Battery Module
Your drop counter sensor needs to be connected to a WIFI-201 module. The WIFI-201
module will create a closed NeuLog WiFi network which will stream the NeuLog data
to a device of your choosing. Once your device is wirelessly connected to the NeuLog
network, you can run experiments and collect data through a browser of your
choosing.
Procedure:
1. Connect the drop counter sensor directly to the left side of a WIFI-201 (no wires
required).
2. Connect a BAT-200 module to the right side of the WIFI-201 module.
3. Although not required, we recommend plugging the BAT-200 to an outlet using a
USB to mini USB charger (such as a typical cell phone charger). The WIFI-201
module will run for 60-90 minutes (depending on the sensor) without being
plugged in.
4. The WIFI-201 can be powered directly using a mini to USB cord and plugging it
into your computer or a wall charger. Please note this function is only available
on WIFI-201 modules with the USB icon on the cover.
5. For further WIFI-201 instructions or the WIFI-201 quick start guide please visit:
www.NeuLog.com; Choose "Products" on the main menu,
then choose "WiFi Communication module".
6. The WIFI-201 indicator lights will flash; take no action until the LED to the far left
turns blue; this can take up to one minute.
7. Take your tablet or smart phone, go to the WiFi settings and select the NeuLog
network (NeuLog0184 for example) which matches the ID found on the back of
the WIFI-201 module (0184 for example).
8. Give your device about 20 seconds to connect to the WIFI-201.
9. Once the device is connected, go to your browser and type the website
wifi201.com into the URL bar, then wait for 30-60 seconds.
10. You will see a “Control mode” icon in the application, click on it.
11. The application will then load a new screen and begin to auto detect the
sensor(s); this can take a minute. (You can stop the search when the sensor is
found.)
12. If the application does not auto detect the sensor(s), select “Search for sensors”.
13. Once the sensor is found you will see a module box on the left side of the screen
for the drop counter sensor. In this box, real time data will be displayed.
14. Click on the “Module setup” button located on the drop counter sensor icon in
the module window to change the drop counter probe settings if need be.
15. Click on the “Experiment setup” icon to change the experiment settings if need
be (Experiment duration for example).
16. To run an experiment and collect data, click the “On-Line experiment” icon, then
click the "Run experiment" icon.
17. If you have a newer generation WIFI-201 (with the USB icon on the cover) this
can be used as a USB-200, meaning you can tether the probes directly to the
computer using the WIFI-201. To engage operation in this fashion, press 3 times
on the panel button. Repeating this operation will return the module to WiFi
mode.
Operation with NeuLog Graphic Display Module (viewer)
Materials needed:
NUL-223 Drop Counter Sensor
VIEW-101 Graphic Display Module
BAT-200 Battery Module
Procedure:
1. Connect the drop counter sensor to the left side of the viewer.
2. Connect the BAT-200 module to the right side of the viewer.
3. The VIEW-101 will auto detect the sensor and you will see it appear on the left
side of the screen.
4. Once the sensor appears it will be monitoring data in real time.
5. To run an experiment and collect data press the run icon (little green person).
Off-line experiments
(Off-line experiments are when you do not have a sensor connected directly to a
computer, tablet, smartphone, or NeuLog viewer).
Materials needed:
NUL-223 Drop Counter Sensor
BAT-200 Battery Module
NEULOG DROP COUNTER LOGGER SENSOR GUIDE
Materials needed to configure your offline experiment:
USB-200 Module, WIFI-201 module or VIEW-101 Module
USB to mini USB cable (included with the USB-200 module)
Procedure:
1. To determine the experiment settings, first connect to your chosen device (PC,
Mac, tablet, smart device) and click on the “Off-line experiment” icon after the
sensor search has ended. (For more information on how to change the
experiment settings, view the quick start procedure section for your chosen
device in this document.)
2. Connect the drop counter sensor directly to the left side of a charged BAT-200
module.
3. When ready to collect data press the “Start/Stop” button on the drop counter
sensor; a red light will turn on for the duration of the experiment.
4. After the experiment has concluded (the “Start/Stop” button is pressed again or
the data collection period ends) connect the sensor to your chosen device (you
can find full instructions in each device’s quick start procedure section).
5. Open NeuLog application.
6. Click the “Off-line experiment” icon.
7. Click the “Load data from sensors” icon.
8. Select which experimental data to upload (5 experimental runs can be stored at
one time on the NUL-223 drop counter sensor).
Restoring sensor’s factory default settings:
Procedure:
1. Connect the USB-200 to a PC, Mac, tablet, or smart device.
2. Connect the drop counter sensor to a USB-200 module (they directly plug
together).
3. Open the NeuLog application.
4. Click the “Tools” icon.
5. Click “Restore sensor’s factory defaults” on the menu.
Included with the sensor:
NeuLog NUL-223 Drop Counter Sensor Guide (this document)
Durable plastic drop counter probe attached directly to the sensor's body by a
rubber-coated wire
Threaded rod for easy connection to retort stands or other labware
Sensor's specifications
Drops
Drop Counter
signal
Range and
operation modes
0 to 6,500 drops
0 to 6,500 drops X
vol. in mL
ADC resolution
Digital
Resolution
1 drop
Max Sample Rate
(S/sec)
100
Experiment Duration: 1 second to 31 days.
Sensor's features:
Fully digital data
Rugged plastic ergonomic case
Push button switch for Start/Stop experiments in off line mode
LED indicator of experiment status (blinks while collecting data)
Pre-calibrated sensing equipment
Note: NeuLog products are intended for educational use
Videos and experiment examples:
Videos, literature and other probes can be found at www.NeuLog.com.
In order to access the drop counter sensor's page, choose "Products" on the
main menu and then "Drop counter logger sensor".
In order to find the science experiments, choose "Experiments" on the main
menu of the NeuLog website, then choose "Physics", "Chemistry", "Biology" or
"Environmental Science".
Chemistry Experiment:
oTitration of a Strong Acid and a Strong Base (C-10)
NEULOG DROP COUNTER LOGGER SENSOR GUIDE
Technical background:
The philosophy behind NeuLog’s plug and play technology is based on each sensor’s
ability to store its own data due to an internal flash memory chip and micro-controller
in each plastic NeuLog body. This technology allows the sensor to collect and then
store the digital data in the correct scientific units (oC, oF, Lux, %, ppm, for example).
The sensor is pre-calibrated at the factory. The built-in software in the logger can be
upgraded for free at any time using the provided firmware update.
The NeuLog drop counter sensor uses an infrared beam which passes between the
two endpoints of the sensor probe to count drops. As liquid passes through the
infrared beam, infrared light is scattered and not noticed by the infrared detector on
the opposing side.
Regardless of the liquid's opacity or color, the infrared light bends away from the
detector. Each time there is a break in the infrared beam, a drop is counted.
Maintenance and storage:
Never submerge the NeuLog plastic body in any liquid.
Do not allow liquid into the drop counter sensor’s body and probe.
After use, gently wipe away any foreign material from the drop counter sensor.
Store in a box at room temperature out of direct sunlight.
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship. The
warranty is for a period of 3 years from the date of purchase and does not cover
damage of the product caused by improper use, abuse, or incorrect storage. Sensors
with a shelf life such as ion selective probes have a warranty of 1 year. Should you
need to act upon the warranty, please contact your distributor. Your sensor will be
repaired or replaced.
Thank you for using NeuLog!
Flexible, simple, fast, forward thinking.
W: www.neulog.com
E: info@neulog.com
A: 850 St Paul Street, Suite 15, Rochester, NY 14605
P: 1.866.553.8536
V11042014

Other NeuLog Accessories manuals

NeuLog NUL-221 User manual

NeuLog

NeuLog NUL-221 User manual

NeuLog NUL-236 User manual

NeuLog

NeuLog NUL-236 User manual

NeuLog NUL-228 User manual

NeuLog

NeuLog NUL-228 User manual

NeuLog NUL-205 User manual

NeuLog

NeuLog NUL-205 User manual

NeuLog NUL-243 User manual

NeuLog

NeuLog NUL-243 User manual

NeuLog NUL-227 User manual

NeuLog

NeuLog NUL-227 User manual

NeuLog NUL-236 User manual

NeuLog

NeuLog NUL-236 User manual

NeuLog NUL-209 User manual

NeuLog

NeuLog NUL-209 User manual

NeuLog NUL-212 User manual

NeuLog

NeuLog NUL-212 User manual

NeuLog NUL-246 User manual

NeuLog

NeuLog NUL-246 User manual

NeuLog NUL-212 User manual

NeuLog

NeuLog NUL-212 User manual

NeuLog NUL-215 User manual

NeuLog

NeuLog NUL-215 User manual

NeuLog NUL-216 User manual

NeuLog

NeuLog NUL-216 User manual

NeuLog NUL-214 User manual

NeuLog

NeuLog NUL-214 User manual

NeuLog NUL-219 User manual

NeuLog

NeuLog NUL-219 User manual

NeuLog NUL-202 User manual

NeuLog

NeuLog NUL-202 User manual

NeuLog NUL-231 User manual

NeuLog

NeuLog NUL-231 User manual

NeuLog NUL-208 User manual

NeuLog

NeuLog NUL-208 User manual

NeuLog NUL-216 User manual

NeuLog

NeuLog NUL-216 User manual

NeuLog NUL-209 User manual

NeuLog

NeuLog NUL-209 User manual

NeuLog NUL-245 User manual

NeuLog

NeuLog NUL-245 User manual

NeuLog NUL-240 User manual

NeuLog

NeuLog NUL-240 User manual

NeuLog NUL-211 User manual

NeuLog

NeuLog NUL-211 User manual

NeuLog NUL-231 User manual

NeuLog

NeuLog NUL-231 User manual

Popular Accessories manuals by other brands

Lanaform HEATING BLANKET FOR BACK manual

Lanaform

Lanaform HEATING BLANKET FOR BACK manual

poly-planar MRD-60 Specifications

poly-planar

poly-planar MRD-60 Specifications

Gardena 1188 operating instructions

Gardena

Gardena 1188 operating instructions

United Technologies Interlogix NX-451-N Installation sheet

United Technologies

United Technologies Interlogix NX-451-N Installation sheet

Velux KRD 100 manual

Velux

Velux KRD 100 manual

EastPoint Sports Go Time Gear Equator 1-1-51711 Assembly instructions

EastPoint Sports

EastPoint Sports Go Time Gear Equator 1-1-51711 Assembly instructions

nedis PEBL110CWT2 quick start guide

nedis

nedis PEBL110CWT2 quick start guide

Skywalker SWFB100 user manual

Skywalker

Skywalker SWFB100 user manual

Nanometrics Trillium Compact user guide

Nanometrics

Nanometrics Trillium Compact user guide

Philips HF3650 quick start guide

Philips

Philips HF3650 quick start guide

Emerson MICRO MOTION T-SERIES installation manual

Emerson

Emerson MICRO MOTION T-SERIES installation manual

Byron 776 instructions

Byron

Byron 776 instructions

ICARO paragliders Energy SP manual

ICARO paragliders

ICARO paragliders Energy SP manual

IFM UGT203 operating instructions

IFM

IFM UGT203 operating instructions

R.M. Young 61302V instructions

R.M. Young

R.M. Young 61302V instructions

HEIDENHAIN SE 660 Mounting instructions

HEIDENHAIN

HEIDENHAIN SE 660 Mounting instructions

Maxcess FIFE DSE-11 operating instructions

Maxcess

Maxcess FIFE DSE-11 operating instructions

Baseus PPCX010005 user manual

Baseus

Baseus PPCX010005 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.