1Introduction
1.2 SAFETY CONSI ERATIONS
This device is mar ed with the international caution symbol. It is important to read
this manual before installing or commissioning this device as it contains important
information relating to Safety and EMC (Electromagnetic Compatibility).
This instrument is a panel mount device protected in accordance with Class I of
EN 61010 (115/230 AC power connections), Class III for the DC Voltage Power Option
(10-32 Vdc). Installation of this instrument should be done by Qualified personnel. In order to
ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall
be included in the building installation as a disconnecting device. It shall be mar ed to
indicate this function, and it shall be in close proximity to the equipment within easy
reach of the operator. The switch or circuit-brea er shall not interrupt the Protective
Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and
IEC 947-3 (International Electrotechnical Commission). The switch shall not be
incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the
mains supply in case of a fault in the equipment, an overcurrent protection device
shall be installed.
• Do not exceed voltage rating on the label located on the top of the instrument housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a wor bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
• Unit mounting should allow for adequate ventilation to ensure instrument does not
exceed operating temperature rating.
• Use electrical wires with adequate size to handle mechanical strain and power
requirements. Install without exposing bare wire outside the connector to minimize
electrical shoc hazards.
EMC Considerations
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
•
Install Ferrite Bead(s) on signal wires close to the instrument if EMC
problems persist.
Failure to follow all instructions and warnings may result in injury!