Nickel-Electro NE4-T Series User manual

THE CliftonRANGE
StirredBaths
NE4-TSeries
NE4-ESeries
This series of water baths features a low liquid level float switch and if the liquid drops below the
recommended level heating is automatically switched OFF.
- top up liquid level, then switch off and back on again using the mains switch located on the rear of the
thermostirrer.
IMPORTANT: ALWAYS FILL the water bath before connecting to the power supply.

Contents
About this Manual 3
General Notes 3
Symbols 3
Location 4
Unpacking 4
Safety 4
Assembly 4
Power Supply Lead and Connection to Electrical Supply 4
Liquid Level 5
Suitable Liquids 5
Front, Tank and Rear Views 6
Operating Instructions 7
Control Panel 7
General Use 7
Low Liquid Float Switch 7
General Cleaning 8
Cleaning external painted surfaces with Anti-bacterial finish 8
Cleaning the stainless steel tank 8
Descaling the stainless steel tank 9
Decontamination of Equipment 9
VirucidalDisinfectant 9
Disinfectant/Sterilant 9
3 Year Warranty 10
Out of Warranty 10
End of Life 10
Portable Appliance Testing 10
Accessories 11
Fitting Accessories 12
Spares/Service Diagram 13
Specification 14
Declaration of Conformity 15
Final Inspection and Electrical Safety Test Report Enclosed Rear
The Clifton Range - 3 Year Warranty 2High Quality Products at Affordable Prices

About this Manual
This user Manual contains instructions which must be followed in order that the product is operated
correctly.
General Notes
Please observe the following safety precautions:
1. Fill the tank prior to connection to power supply.
2. Connect only to a power supply with the corresponding voltage to that specified on the rating
label positioned on the rear of the unit.
3. Ensure the power supply has a safety earth (ground) terminal.
4. Ensure the mains switch and power supply connector are accessible during use.
5. The mains supply cord fitted to this products is a heat resistant type and should be replaced by
an equivalent type.
6. Do not block ventilation slots during use.
7. Always follow good laboratory practice by ensuring substances being heated present no risk of a
hazard (explosion, implosion or release of toxic or flammable gases) or that these have been
addressed. When heating substances where liberation of gases occurs suitable extraction should
be used.
8. Use only liquids specified in this Instruction Manual within their specified temperature range.
9. Use caution when topping up or draining the tank as the liquid in the tank may be very hot or
cold.
10. Drain before moving the bath. Allow the liquid to cool to 50°C or heat up to 10°C or above
before draining.
11. Use caution when refilling as a hot heating element can create scalding steam.
12. Always use a lid or layer of polypropylene spheres when operating at temperatures above 60°C.
Take care when lifting the lid. Steam and hot vapours can cause scalding.
13. Always use the display or a thermometer to check the temperature - do not touch liquid.
14. If this product is not used in accordance with these instructions, then basic safety protection
afforded by the water bath may be affected.
15. Check the operation of the low liquid level float switch regularly.
16. Before using any cleaning or decontamination method except those recommended, check with
your distributor that the proposed method will not damage the equipment.
Amendments
Issue 1 February 2004 New book dedicated to NE4T’s.
Issue 2 December 2004 Amendment to Service Diagram.
Issue 3 December 2005 Updated to service diagram.
Issue 4 August 2006 Update for WEE, DOC, accessories, overview, cleaning
Issue 5 November 2006 Anti-bacterial paint finish
Issue 6 May 2013 Updatedwith E version instructions.
Symbols
HOT SURFACES
Paragraphs marked by this symbol indicate that a potential hazard to your personal
safety exists from heated surfaces or other appendages on the outside or inside of
the equipment.
CAUTION
This icon accompanies text and/or other international symbols dealing with potential
damage to equipment. When present, it indicates that there is a potential danger of
equipment damage may occur if information stated within the "CAUTION" paragraph
is not adhered to or procedures are executed incorrectly.
The Clifton Range - 3 Year Warranty 3High Quality Products at Affordable Prices
PROTECTIVE EARTH OR GROUND TERMINAL
Protective earth conductor terminal.

Location
The product must be placed on a smooth, level and sturdy work surface, preferably near a sink or drain
for emptying. Use in a ventilated room. Suitable for use in ambient temperatures 5°C to 40°C with a
maximum humidity 80% (temperature 31°C) decreasing to 50% (temperature 40°C).
Unpacking
Remove the product from its packaging. Any damage please notify your dealer immediately. Retain the
packaging over the warranty period. Contents consist of a thermostirrer mounted on a stainless steel
bridge unit to fit the slected tank, tank, stainless steel shelf, black nylon propellor, fasteners for securing
the thermostirrer to the tank, power lead and an instruction manual.
Assembly NE4-T series
Turn the thermostirrer upside down and screw the nylon propellor hand tight onto the spindle. The spindle
is threaded in the opposite direction to the propellor to ensure the propellor stays secure in use. Place
the stainless steel shelf into the tank. Fit the thermostirrer onto the tank aligning the slots in the bridge
unit with the holes in the side of the tank. Secure the thermostirrer to the tank using the the M4 screws
and washers supplied.
Turn the thermostirrer upside down and screw the nylon propellor hand tight onto the spindle. The spindle
is threaded in the opposite direction to the propellor to ensure the propellor stays secure in use. Place
the stainless steel shelf into the tank. Place digital thermostirrer mounted on stainless steel bridge onto
right hand end of the clear tank. Place clamping plate inbetween the tank rim and stainless steel bridge
- marked as “A” below. Fit screws into the bridge and clamping plate and tighten to secure the thermostirrer
to the tank.
Safety
Do not touch any electrical contacts or open any closure panels.
RISK OF ELECTRICAL SHOCK!
NE4-T series is Class 1 (IEC519 - Part 2) = low liquid level protection,
reference to over temperature condition providing product over
temperature protection.
Power Supply Lead and Connection to Electrical Supply
Fit the power lead by plugging it into rear of the Thermostirrer and then to mains supply.
Before connecting the product to the electrical supply, check the
information on the rating label is compatible.
IF IN DOUBT CONSULT AN ELECTRICIAN. THE PRODUCT MUST BE EARTHED!
Where the mains supply or plug connection differs refer to local
regulations or qualified electrician.
The Clifton Range - 3 Year Warranty 4High Quality Products at Affordable Prices
Assembly NE4-E series

Liquid Level
8,14, 22 Litre Baths Minimum liquid level - must cover the top of the false base by 70mm
Maximum liquid level - must not exceed the ridge in the tank.
28 Litre Baths Minimum liquid level - must cover the top of the false base by
130mm.
Maximum liquid level - must not exceed the ridge in the tank.
When filling or emptying the bath disconnect the unit from the
mains electrical supply.
Suitable Liquids
Operating temperatures and recommended liquid options:
Ambient to 99°C Distilled water.
Heat transfer liquid. The LB range is formulated for temperatures
from -45°C to 90°C and provides complete protection from algae
growth and safeguards against corrosion. See accessories for the full
range available.
Above 60°C or below room temperature it is recommended that to
achieve optimum performance the bath should be covered with
SL4 lid or polypropylene spheres.
The Clifton Range - 3 Year Warranty 5High Quality Products at Affordable Prices

Mains power switch “I” ON, “O” OFF.
IEC connector, for power supply lead and
plug. Fuse draw fitted underneath.
Rear View
NE4-T and NE4-E Series Stirred Thermostatic Baths
Front View
Splash proof controls, all over smooth
wipe clean surfaces.
Stainless steel construction for
durability.
Thermostirrer
Water bath: NE4-T series stainless
steel tank with false base shelf fitted
inside. NE4-E series clear tank with
shelf fitted.
Indicators: heating, temperature alarm,
timer and set temperature.
Tank View
Low liquid Safety cut-out Float switch
Heating element coil.
False base/shelf for standing work/
samples onto - removable for cleaning.
Stirrer shaft and prop.
Thermostat control.
Serial number and electrical supply details.
The Clifton Range - 3 Year Warranty 6High Quality Products at Affordable Prices

Operating Instructions
Control Panel
Temperature Dial
Low Liquid Level Alarm Indicator
Heater Indicator
Power Indicator
General Use
1. Ensure the bath has the required liquid level, then switch the unit
ON "I" using the mains switch at the rear of the unit.
Power indicator will illuminate.
2. Adjust the temperature control to the desired setting. Allow the
water temperature to stabilise and take an actual reading at the far end
of the bath to the Thermostirrer. Where necessary adjust the
temperature controller.
Heater indicator will illuminate.
Low Liquid Level Float Switch
The unit features a low liquid level float switch. This ensures heating
will be switched off if the liquid level falls below the recommended level.
The motor and heater are switched off in this instance. Use caution
when refilling the tank as hot elements can spit and create scalding
steam.
The Clifton Range - 3 Year Warranty 7High Quality Products at Affordable Prices

The Clifton Range - 3 Year Warranty 8High Quality Products at Affordable Prices
Cleaning
General
Important - please follow these instructions to avoid possible damage to
the unit, otherwise affecting its performance and warranty.
Always disconnect the product from the electrical supply before
cleaning.
Cleaning External Painted Surfaces featuring “Anti-bacterial Paint Finish”
The water bath should be cleaned at regular intervals wiping external surfaces with a cloth or sponge
soaked in warm soapy water with a mild detergent. All surfaces should be cleaned using a soft cloth or
sponge.
Do not under any circumstances use strong solvents or solutions containi
ng Chlorinated Hydocarbons, Esters, Ketones or abrasive cleaners or polish
on the paint finish otherwise it damages the built
in anti-bacterial properties.
All painted surfaces on Clifton range products features an “Anti-bacterial
paint finish” identified with this authenticating logo on the unit.
This “Anti-bacterial paint finish” inhibits the growth of bacteria. It has been
tested by independent specialist test houses such as the Law Laboratories (in
the UK) using internationally recognized test methods and proven to be effec-
tive versus a wide range of bacteria species including Escherichia coli and
Staphylococcus aureus (MRSA).
We recognise hygienic coatings are part of a controlled approach to a cleaner
working environment. Within its formulation an active ingredient with proven anti-bacterial properties
is bound into the paint finish. The efficacy of the paint finish applied to the Clifton range is maintained
over its lifetime, as the anti-bacterial agent is integral within the paint.
In a laboratory environment it makes this one less source of contamination, contributing to essential
clean working practices. A benefit of such a paint finish can lead to a reduction in cleaning schedules
because surfaces are more protected and improves protection between cleaning. Unlike detergents
“Anti-bacterial paint finish” does not offer an instantaneous action, but is intended for long-term
general protection against bacterial growth.
Moisture on the painted surface is necessary for the bacterium to absorb the agent and be affected
by it. The coating is therefore less active in very dry conditions, but dependent upon relative humid-
ity, moisture in the atmosphere maintains activity. Areas where moisture is trapped are also areas
that normally are difficult to clean and where bacteria proliferate but these areas are most active for
the anti-bacterial coating improving the defence against bacterial growth.
Cleaning the Stainless Steel Tank
The stainless steel crevice free tank with smooth corners should provide years of valuable service and is
resistant to chloride containing solutions it is however important to avoid high concentrations of halo-
gens - especially chloride. With such a high quality and resistant tank it may show symptoms of these
halogens as rust, which are deposits from external sources in the water supply.
We recommend always empty the bath of liquid after use and wipe out the internal faces of the tank with
a non-abrasive cloth and allow to dry. Any deposits can be removed with nitric acid (10%) on a cloth.
WEAR PROTECTIVE EQUIPMENT!
It is also recommended to use an accessory lid to prevent contaminates landing in bath liquids.

The Clifton Range - 3 Year Warranty 9High Quality Products at Affordable Prices
Descaling the Stainless Steel Tank
Descale the stainless steel tank regularly to maintain it in as new condition ensuring the corrosion
resistance and normal operating conditions are maintained throughout its working life. Descale by add-
ing 1 litre of vinegar to water and gently heating to 50°C for an hour, empty and brush the lime away.
Rinse thoroughly afterwards.
Decontamination of Equipment
Clifton laboratory equipment can be decontaminated after spillage or contact with potentially HIV and
Hepatitis infected blood samples during analysis using following recommended rapid disinfectants.
Virucidal Disinfectant
We recommend Virkon tablets for the safe and rapid disinfection of equipment in a wide variety of
situations available from your distributor or contact Day-Impex Ltd. for more details. Telephone: 44+(0)1787
223232 or http://www.day-impex.co.uk
The ultimate high level surface disinfectant, dissolve VIRKON in water, providing a
safe working solution with a faint lemon odor. It has proven efficacy against bacte-
ria (including mycobacteria), viruses, spores and fungi in a variety of independent
tests using different protocols. Presents no serious long term health risks to staff –
obviating the need for costly ventilation equipment and health monitoring. Also
provides high level disinfection of laboratory equipment and instruments where
autoclaving is neither practical nor necessary. For more detailed information relat-
ing to how Virkon should be used with access to test reports www.relyon.dupont.com
Is Virkon solution corrosive? Virkon solution requires only 10 minutes contact
time to be effective so long-term exposure is not necessary and therefore will not corrode most
materials. Care should be taken with Stainless steel water bath tanks, these surfaces should not be
affected however, it is important that generally you do not leave Virkon solution in contactwith
metal surfaces “FOR LONGER THAN IS NECESSARY”.
Virkon is Registered in accordance with the requirements of the Medical Devices Directive, (93/42/EEC)
as a Medical Device.
Disinfectant/Sterilant
We recommend PeraSafe a powder product for the safe and rapid chemical sterilant of equipment in a
wide variety of situations available from your distributor or contact Day-Impex Ltd for more details.
Telephone: 44+(0)1787 223232 or http://www.day-impex.co.uk
PeraSafe has a proven safety profile for end-users with none of the undesirable properties of skin
sensitisation, toxic fumes or unpleasant odours that are associated with aldehyde solutions.
Leading UK and USA microbologists have proven PeraSafe to be active
against viruses, mycobacteria and fungi. It is microbiologically superior to
glutaraldehyde, destroying sporing bacteria in one minute. It has also
been independently proven that PeraSafe sterilises in just 10 minutes.
For more detailed information relating to how PeraSafe should be used with
access to test reports www.relyon.dupont.com

This symbol indicates that this product should not be disposed of with your waste.
Instead, dispose waste electrical equipment by handing it over to a designated collec-
tion point for the recycling of waste electrical and electronic equipment. The separate
collection and recycling of your waste equipment at the time of disposal will help to
conserve natural resources and ensure that it is recycled in a manner that protects
human health and the environment. For more information about where you can drop
off your waste equipment for recycling, in UK please contact Service Department, rest
Europe contact your Distributor.
Health & Safety, unless in receipt of a Decontamination Notice or Report the unit
cannot be returned or accepted for disposal.
Clifton electrical and electronic equipment has been designed for recycling and takes
into account the dismantling and recovery its components and materials. Clifton prod-
ucts are easily recycled with majority of the product constructed from stainless or
mild steels, which can readily be re-used or recycled.
In excess of 78% of this product range can be easily re-cycled economically.
Portable Appliance Testing
When conducting testing, ensure it is conducted by a qualified person.
DO NOT PAT TEST THE BATH UNLESS IT CONTAINS WATER.
THIS EQUIPMENT MUST NOT BE FLASH TESTED!
3 Year Warranty
Our service engineers are fully trained in the assembly, calibration and servicing of all Clifton instru-
mentation. Products can be returned to our comprehensively equipped service centre where a fast
and efficient turnaround is guaranteed:
Service Department, Nickel Electro Limited, Oldmixon Crescent, Weston-super-Mare, North Somerset
BS24 9BL, UK. Tel +44 (0)1934 626691 Fax +44 (0)1934 630300.
Out of Warranty
Our Service Department has comprehensive stock of charegeable spare parts maintaining working life of
equipment or units can be returned for quotation before repairs are undertaken.
End of Life
The Clifton Range - 3 Year Warranty 10 High Quality Products at Affordable Prices

Accessories for NE4 Series
GRP Gable Lids
GL1-14 GRP Gable Lid to suit 14 Litre Baths
GL1-22 GRP Gable Lid to suit 9, 22 and 28 Litre Baths
Stainless Steel Gable Lids
SL4-8 Stainless Steel Gable Lid to suit 8 Litre Baths
SL4-14 Stainless Steel Gable Lid to suit 14 Litre Baths
SL4-22 Stainless Steel Gable Lid to suit 22 and 28 Litre Baths
SL4-38 Stainless Steel Gable Lid to suit 38 Litre Baths
Stainless Steel Flat Lids
LD4-8 Flat One Piece Stainless Steel Lid to suit 8 Litre Capacity Bath
LD4-14 Flat One Piece Stainless Steel Lid to suit 14 Litre Capacity Bath
LD4-22 Flat One Piece Stainless Steel Lid to suit 22 and 28 Litre Capacity Bath
LD4-38 Flat One Piece Stainless Steel Lid to suit 38 Litre Capacity Bath
Stainless Steel Raised Shelves
RS4-14 Stainless Steel Raised Shelf to suit 14 Litre Baths
RS4-22 Stainless Steel Raised Shelf to suit 22 and 28 Litre Baths
Stainless Steel Test Tube Racks - Dimensions 270 x 70 x 138mm (L x W x H)
6870 Stainless Steel Test Tube Rack 26 Hole x 17mm Diameter
6871 Stainless Steel Test Tube Rack 16 Holes x 26mm Diameter
6872 Stainless Steel Test Tube Rack 36 Holes x 13mm Diameter
6873 Stainless Steel Test Tube Rack 18 Holes x 19mm Diameter/suitable for 1.5ml microtubes
6900 Stainless Steel Test Tube Rack 12 Holes x 32mm Diameter
Note: 8 Litre = 1 Rack, 14 Litre = 2 Racks, 22 and 28 Litre = 5 Racks
Misceellaneous
LB-2.5 2.5 Litres - Lab Bath 4590 - Heat transfer fluid - Temperature range -15° to +90°C
LB-5.0 5 Litres - Lab Bath 4590 - Heat transfer fluid - Temperature range -15° to +90°C
TC-1 Thermometer Clip Complete With Bent Stem Spirit Filled Thermometer
BX0616 Draining Syphon
NE4-MB Mounting Bracket allows Thermostirrers or Thermocirculators to be mounted on
any type of separate tank.
BP0368 Polypropylene Spheres
For more information on Accessories www.nickel-electro.co.uk
The Clifton Range - 3 Year Warranty 11 High Quality Products at Affordable Prices

FITTING ACCESSORIES FOR NE4 SERIES
GRP Gable Lids
Position the lid over the bath and lower into place.
The lids are designed to fit snugly so that condensate runs into the bath.
Stainless Steel Gable Lids
Position the lid over the bath and lower into place.
These lids have bo drip points, provide full working height across the bath when fitted and have a
ventilation gap.
Stainless Steel Flat Lids
Position the lid over the bath and lower into place.
Stainless Steel Raised Shelves
The raised shelf provides a reduced liquid depth for shallow vessels in NE4 series water baths. With
recommended minimum water level in NE4 range of 70mm with this in place it will provide a working
depth of 5mm or 35mm.
Stainless Steel Test Tube Racks
Place the loaded test tube rack into the bath on top of either the stainless steel perforated shelf or the
raised shelf.
Thermometer Clip
This can be positioned anywhere on the perimeter of the bath by cliping over the edge. The thermom-
eter then lies flush with the top edge of the bath to prevent accidental damage.
Draining Syphon
Allows the easy and quick emptying of any water bath. Water temperature must be below 45°C before
emptying can commence.
Mounting Bracket
This Mounting Bracket is designed to fit the NE4 series Clifton Thermostirrers for mounting on any
number of vessels or tanks.
Maximum tank thickness for the Bracket = 30mm.Thumb wheel adjuster secures the Bracket and
Thermostirrer to tank.
Polypropylene Spheres
Place on top of the water/oil to provide a floating lid reducing evaporation and insulating the bath
reducing heat losses.
The Clifton Range - 3 Year Warranty 12 High Quality Products at Affordable Prices

Spares/Service Diagram
The Clifton Range - 3 Year Warranty 13 High Quality Products at Affordable Prices

BATHS
NE4-T series, thermostat control
Stirred Clifton water baths are designed for immersion of
flasks/bottles/racks in a highly stable temperature
environment throughout the water bath. Available in a wide
range of sizes suiting a variety of applications.
This entry level thermostatically controlled analogue
Thermostirrer is housed in a durable corrosion resistant stainless
steel body. Immersed components are corrosion resistant. The
Thermostirrer features a safety cut-out float switch, illuminated
on/off switch, heater and mains indications. The Thermostirrer is
supplied with a stainless steel mounting bridge to fit the selected
tank. Each bath consists of a 304 spec stainless steel tank
contained within a sturdy outer case with chemical resistant paint
finish and is complete with a stainless steel shelf, designed for
good temperature stability.
STIRRED
Features Include:
• Sensitivity: ±0.2°C
• Uniformity: ±0.05°C
• Temperature range:
ambient +5°C - 99°C š
• Thermostatic control - easy to
use
• Totally unobstructed, crevice
free,
stainless steel tank
• Powerful stirring within tank
- for stable temperature control
• Stainless steel tank
- corrosion resistant and easy to
clean
• Sturdy construction
- powder coated for a chemical
resistant exterior
• Safety cut-out float switch
-low liquid level protection
• Removable sturdy stainless
steel
shelf
• Lids optional extra
- recommended above 60°C
Cat No NE4-8T NE4-14T NE4-22T NE4-28T
Capacity Litres 8l 14l 22l 28l
Working Dims 129x332x150mm 219x298x150mm 395x298x150mm 395x298x200mm
Overall Dims 271x332x170mm 361x332x170mm 537x332x170mm 537x332x240mm
Heater Watts 1250W 1250W 1250W 1250W
Voltage 230V 230V 230V 230V
The Clifton Range - 3 Year Warranty 14 High Quality Products at Affordable Prices

EC Declaration of Conformity
We herewith confirm the following product
NE4 Stirred Waterbath Range
Conforms with the requirements outlined by following European Directives.
Low Voltage Directive (2006/95/EC)
EMC Directive (89/336/EEC)
We confirm the declaration
Conforms with the requirements of following Standards
BS EN 61010:1
BS EN 61010:2.010
Safety requirements for electrical equipment for measurement, control and laboratory use.
BS EN 61326
Electrical equipment for measurement control and laboratory use - EMC requirements.
NICKEL-ELECTRO LTD
Manufacturers of laboratory, medical and clinical equipment.
Oldmixon Crescent, Weston-super-Mare,
North Somerset, BS24 9BL, United Kingdom.
Tel: +44 (0)1934 626691
Fax: +44 (0)1934 630300
Email: clifton@nickel-electro.co.uk
www.nickel-electro.co.uk
The Clifton Range - 3 Year Warranty 15 High Quality Products at Affordable Prices
Nickel-Electro Ltd is also registered ISO9001 refer-
enceNo.Q09820

BX0844
BX0845 www.nickel-electro.co.uk
This manual suits for next models
1
Table of contents
Other Nickel-Electro Laboratory Equipment manuals

Nickel-Electro
Nickel-Electro Clifton 1000ER Series User manual

Nickel-Electro
Nickel-Electro Clifton Range NE4-14D User manual

Nickel-Electro
Nickel-Electro Clifton Range NE9S Series User manual

Nickel-Electro
Nickel-Electro Clifton Range SM-1 User manual

Nickel-Electro
Nickel-Electro Clifton NE1B Series User manual

Nickel-Electro
Nickel-Electro Clifton RANGE NE2D Series User manual