Nickel-Electro Clifton NE1B Series User manual

Contents
Aboutthis Manual 3
GeneralNotes 3
Amendments 3
Symbols 3
Location 4
Unpacking 4
Safety 4
Assembly 4
PowerSupplyLeadandConnectionto ElectricalSupply 4
LiquidLevel 4
SuitableLiquids 4
NE1BSeriesProductOverview 5
NE1SeriesProductOverview 6
OperatingInstructions NE1BSeries 7
OperatingInstructions NE1Series 7
Explanationof TemperatureControlTerms 8
GeneralCleaning 9
CleaningexternalpaintedsurfaceswithAnti-bacterialfinish 9
Cleaningthestainless steeltank 9
Descalingthestainless steeltank 10
Decontaminationofequipment 10
VirucidalDisinfectant 10
Disinfectant/Sterilant 10
3YearWarranty 11
Outof Warranty 11
Endof Life 11
PortableApplianceTesting 11
Accessories 12
FittingAccessories 13
ServiceandPartsList 14
SpecificationNE1BSeries 15
SpecificationNE1Series 16
SpecificationNE1ShallowForm 17
Declarationof Conformity 18
FinalInspectionandElectricalSafety TestReport EnclosedRear
TheCliftonRange-3YearWarranty 2HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

Symbols
AboutthisManual
This userManualcontains instructions whichmustbefollowedin orderthattheproductis operated
correctly.
General Notes
Pleaseobservethefollowingsafety precautions:
1. Fill thetankpriorto connectionto powersupply.
2. Connectonlyto apowersupplywiththecorrespondingvoltageto thatspecifiedontherating
labelpositionedontherearof theunit.
3. Ensurethepowersupplyhas asafety earth(ground) terminal.
4. Ensurethemains switchandpowersupplyconnectorareaccessibleduringuse.
5. Themains supplycord fittedto theseproductsis aheatresistanttypeandshould bereplaced
byan equivalenttype.
6. Donot blockventilationslotsduringuseandfollowinstallationinstructions.
7. Alwaysfollowgoodlaboratorypracticebyensuringsubstancesbeingheatedpresentnoriskof
ahazard (explosion,implosionorreleaseof toxicorflammablegases)orthatthesehavebeen
addressed. Whenheatingsubstanceswhereliberationof gasesoccurssuitableextraction
should beused.
8. Useonlyliquidsspecifiedin this InstructionManualwithin theirspecifiedtemperaturerange.
9. Usecautionwhentoppingupordrainingthetankas theliquidin thetankmaybeveryhot or
cold. Allowtheliquidto coolto 40°Cbeforedraining.
10. Drain beforemovingthebath.
11. RecommendusingaLidabove60°C.Takecarewhenliftingthelid-steamandhot vapours
can causescalding.
12. Useathermometerto checkthetemperature-donot touchliquid.
13. Ifthis productis not usedin accordancewiththeseinstructions,thenbasicsafety protection
affordedbythewaterbathmaybeaffected.
14. Beforeusinganycleaningordecontaminationmethodexceptthoserecommended, checkwith
yourdistributorthattheproposedmethodwill not damagetheequipment.
Amendments
Issue1 July 1991 Initialissueinstructionbook.
Issue2 February 1993 RevisedaddedNE1-28 version.
Issue3 June 1993 NE1Brangeintroduced.
Issue4 August 1994 Instructionbookrevisedlayout.
Issue5 May 1995 CEversion.
Issue6 September1995 Multi-lingualversion.
Issue7 May 1996 Updateaddress details.
Issue8 November 1997 NE1-22/28,NE1B-14 fuserevision.
Issue9 September1999 F to T.
Issue10 July 2000 Completeupdate.
Issue11 July 2001 Productupdatesfor2002.
Issue12 September2004 Revisedinstructionbook.
Issue13 July 2005 NewNE1-9,NE1BrangeupdatedconstantlevelandMCB’s.
Issue14 July 2006 NE1B-14 1000Wto 1200WandNotefrontcoverfillingbath.
Issue15 November 2006 Improveddisinfection/sterilisationnotes,explanationof
accuracy,sensitivity etc,Anti-bacterialpaintfinish.
HOTSURFACES
Paragraphs markedbythis symbolindicatethatapotentialhazard to yourpersonal
safety existsfromheatedsurfacesorotherappendagesontheoutsideorinsideof
theequipment.
CAUTION
This iconaccompaniestextand/orotherinternationalsymbolsdealingwithpoten-
tialdamageto equipment.Whenpresent,itindicatesthatthereis apotential
dangerof equipmentdamagemayoccurifinformationstatedwithin the"CAUTION"
paragraphis not adheredto orproceduresareexecutedincorrectly.
PROTECTIVEEARTH ORGROUND TERMINAL
Protectiveearthconductorterminal.
TheCliftonRange-3YearWarranty 3HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

Location
Thesurfaceonwhichtheproductis to belocatedshould besmooth,levelandsturdy.Usein aventi-
latedroom.Productis designedforlaboratoryuse.
Unpacking
Removetheproductfromitspackaging.Anydamagetotheproduct notifyyourdistributorimmedi-
ately.Retain packagingoverwarranty period. Contentsconsistof abath,stainless steelfalsebase,
powerleadandinstructionmanual.
Safety
Donottouchanyelectrical contactsoropenanyclosurepanels
RISKOF ELECTRICALSHOCK!
Assembly
Placethefalsebaseinsidethebath.Fitthepowerleadinto thesocketatrear.
PowerSupply
Beforeattemptingtoconnecttheproducttoanelectrical supply check
theinformation on theratinglabeliscompatible.
IFINDOUBTCONSULTANELECTRICIAN.
THEPRODUCTMUST BEEARTHED!.
Wherethemainssupply orplugdiffersrefertolocal regulationsor
qualifiedelectrician.
LiquidLevel-NE1range
Minimumliquidlevel-mustcoverthetopof thefalsebaseby40mm
Maximumliquidlevel-mustnot exceedtheridgein thetank.
Alwaysensuretheproductisdisconnectedfromtheelectricalsupplywhen
emptyingandfilling.
SuitableLiquids-NE1range
Operatingtemperaturesfromambient+5°Cto 99°C,forgeneralusewerecommendeither:
- Distilledwater.
- Heattransferliquid [TheLBrangeis formulatedfortemperaturesfrom-45°Cto 90°Candpro-
videscompleteprotectionfromfreezingandalgaegrowthandsafeguardsagainstcorrosion.See
accessoriesforthefull rangeavailable].
- VIRKONdissolvedindistilledwater has provenefficacyagainstbacteria (includingmycobacteria),
viruses,sporesandfungiin avariety of independenttestsusingdifferentprotocols.
LiquidLevel-NE1Brange
Minimumandmaximumwaterlevelisgovernedby theconstantlevel
whichmaintainswaterlevelduringboiling.Usingstandardlaboratory
tubing,connecttothebottomoutletoftheconstantleveltoaregulated
watersupply/tap, adjustthewatersupply toprovideaslowfeedof
waterandmaintainingboilingandthisautomatically maintainssafe
waterlevel. Thesideoutletactsasanoverflowtoasinkordrain.
SuitableLiquids-NE1Brange
Normaltapwateris suitablesinceamaintainedwatersupplyis required.
TheCliftonRange-3YearWarranty 4HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

NE1BSeries ProductOverView
FrontView
RearViews
Splash proof controls.
Clifton“Low”heightshelfneedsless
waterforsame worksavingenergy
andcosts.Alsoallowstall samplesto
beimmersedto eitheragreater
workingdepth.
Indicator: heating.
Drain outletforaccessorytap:
28 litremodel.
Mains powerswitch“I”ON,“O”OFF.
Serialnumberandelectricalsupply
details.
Powersupplyleadandplug.
Constantleveldevice,maintains liquid
levelduringboiling.
Outlet,connectwithlaboratorytubing
to adrain orwasteoutlet.
Inlet,connectwithlaboratorytubing
to fill theBoilingbaththenadjustthe
flowwatersupply,to maintain liquid
level
MCBfuses
Normaluserequiresnoaction.
Faultsituation,buttons will popout
disconnectingelectricalsupply.
SEEKGUIDANCEbeforeresetting-
push to reset.
Dial: adjustslevelof boiling.
TheCliftonRange-3YearWarranty 5HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

NE1Series ProductOverView
FrontView
RearView
Splash proof controls.
Clifton“Low”heightshelfneedsless
waterforsame worksavingenergy
andcosts.Alsoallowstall samplesto
beimmersedto eitheragreater
workingdepth.
Indicator: heating.
Drain outletforaccessorytap:
22,28 and56 litremodels.
Mains powerswitch“I”ON,“O”OFF.
Serialnumberandelectricalsupply
details.
IECconnector,forpowersupplylead
andplug.Fusedrawfittedunderneath.
Dial, thermostaticcontrol, setto desired
temperature.
TheCliftonRange-3YearWarranty 6HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

NE1BSeries:OperatingInstructions
1.Fill thebathwithwaterfollowinginstructions onPage4
LiquidLevel-NE1Brange anddiagramonPage5.
2.Connectthemains plugto an electricalsupply.
3.Turnunitonusingmains switchlocatedonthefrontof theunit.
4.Heaterindicatorwill illuminateto showheateractivity.
5.Adjusttheblueknobto controlthedegree of boilfromgentlesimmerto
vigorous boil.
6.Allowthewatertemperatureto stabilisein thebath,andtakean actual
temperaturereading.
NE1Series:OperatingInstructions
1.Fill thebathwithwaterobservingrecommendedlevelsonPage4
LiquidLevel-NE1range
2.Connectthemains plugto an electricalsupply.
3.Turnunitonusingmains switchlocatedonthefrontof theunit.
4.Adjustthetemperaturecontrolto thedesiredsettingusingtheblueknob.
5.Heaterindicatorwill illuminateto showheateractivity.
6.Allowthewatertemperatureto stabilisein thebath,andtakean actual
temperaturereading.Adjustthetemperaturecontrolleraccordingly.
TheCliftonRange-3YearWarranty 7HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

Explanation ofTemperatureControlTerms
Accuracy
Wedonotprovide,claimorassureanyformofaccuracy.Accuracyis definedas "theability ofa
measurementtomatchtheactualvalueof thequantity beingmeasured".Foraccuracywerecom-
mendusingacalibratedreferenceprobeattheactualsetpointtemperatureandwherenecessary,
adjustthesetpointaccordingly.
Sensitivity
Foran explanationofsensitivity consideraunstirreddigitalwaterbath,thePIDtemperature
controlsystemmeasuresanddisplaystheactualtemperatureofthewaterandthencomparesit
withthe‘setpoint’temperature.It automaticallycalculatesandadjuststherequiredquantityof
heatintothebathto makethemeasuredtemperatureequalto thesettemperature.Aswithany
process thereis atimedelaybetweenmeasuringthetemperatureandtheheatenteringthewater,
whichcausesminorfluctuations in thetemperatureof thebath.
Heatis alsodistributedinan unstirredbathbyconvectionandconductionandthereareheatlosses
fromthesurfaceof theliquidwhichcan causetemperaturelosses.Theselossesandheatdistribu-
tionproducesmallfluctuationsin temperatureacross thewaterin thebath.
Thesesmall temperaturefluctuationsatanyonepointaredefinedas “sensitivity”andvary
betweenan upperandlowerlimit,howeveroccasionallyalargervariationcan beobserved.
Sensitivity as statedin DIN58966 isthetemperaturedifferencebetweentheupperandlower
temperaturelevelover100 cyclesafterremovingthelargest25%of readings.
Wedeterminesensitivity byrecordingtheactualupperandlowertemperaturesof thebathusing
temperatureloggersandis statedasplusorminus onehalfof themeasuredvalue.
Uniformity
Uniformity iscalculatedbymeasuringthetemperatureinopposingendsofthewaterbathandis
thedifferencebetweenthemean temperaturesatthesetwopointsandstatedas plus orminus
halfthisvalue.
TheCliftonRange-3YearWarranty 8HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

Cleaning
General
Important-please followthese instructionstoavoidpossibledamage to
theunit,otherwise affectingitsperformanceandwarranty.
Alwaysdisconnect theproductfromtheelectrical supply before
cleaning.
CleaningExternal PaintedSurfaces featuring“Anti-bacterial PaintFinish”
Thewaterbathshould becleanedatregularintervalswipingexternalsurfaceswithaclothorsponge
soakedin warmsoapywaterwithamild detergent.All surfacesshould becleanedusingasoft clothor
sponge.
Donotunderanycircumstances use strongsolventsorsolutionscontaini
ngChlorinatedHydocarbons,Esters,Ketonesorabrasivecleanersorpolish
on thepaintfinishotherwise itdamages thebuiltinanti-bacterial
properties.
AllpaintedsurfacesonCliftonrangeproductsfeaturesan “Anti-bacterial
paintfinish”identifiedwiththis authenticatinglogoontheunit.
This“Anti-bacterialpaint finish” inhibitsthegrowthofbacteria.It has been
testedbyindependentspecialisttesthousessuchas theLawLaboratories(in
theUK)usinginternationallyrecognizedtestmethodsandproventobe
effectiveversus awiderangeofbacteriaspeciesincludingEscherichia coli
andStaphylococcusaureus (MRSA).
Werecognisehygieniccoatingsarepartofacontrolledapproachto acleaner
workingenvironment.Within itsformulationanactiveingredientwithprovenanti-bacterialproperties
is boundintothepaintfinish.Theefficacyof thepaintfinishappliedtotheCliftonrangeis maintained
overitslifetime,as theanti-bacterialagentis integralwithinthepaint.
Inalaboratoryenvironmentitmakesthisonelesssourceofcontamination,contributingto
essentialclean workingpractices.Abenefitof suchapaintfinish can leadto areductionin
cleaningschedulesbecausesurfacesaremoreprotectedandimprovesprotectionbetween
cleaning.Unlikedetergents“Anti-bacterialpaintfinish”doesnotofferaninstantaneousaction,
butis intendedforlong-termgeneralprotectionagainstbacterialgrowth.
Moistureonthepaintedsurfaceis necessaryforthebacteriumto absorbtheagentandbeaffected
byit.Thecoatingis thereforelessactiveinverydryconditions,butdependentuponrelative
humidity,moisturein theatmospheremaintainsactivity.Areaswheremoistureistrappedarealso
areas thatnormallyaredifficultto cleanandwherebacteriaproliferatebuttheseareas aremost
activefortheanti-bacterialcoatingimprovingthedefenceagainstbacterialgrowth.
CleaningtheStainless SteelTank
Thestainless steelcrevicefree tankwithsmoothcornersshould provideyearsof valuableserviceand
is resistantto chloridecontainingsolutions itis howeverimportantto avoidhighconcentrations of
halogens -especiallychloride.Withsuchahighquality andresistanttankit may showsymptomsof
thesehalogens as rust,whicharedepositsfromexternalsourcesin thewatersupply.
Werecommendalwaysempty thebathof liquidafteruseandwipeouttheinternalfacesof thetank
withanon-abrasiveclothandallowtodry.Anydepositscan beremovedwithnitricacid(10%)ona
cloth. WEARPROTECTIVEEQUIPMENT!
It is alsorecommendedto usean accessorylidto preventcontaminateslandingin bathliquids.
TheCliftonRange-3YearWarranty 9HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

DescalingtheStainless SteelTank
Descalethestainless steeltankregularlyto maintain itin as newconditionensuringthecorrosion
resistanceandnormaloperatingconditions aremaintainedthroughoutitsworkinglife.Descaleby
adding1litreof vinegarto waterandgentlyheatingto 50°Cforan hour,empty andbrush thelime
away.
Rinsethoroughlyafterwards.
Decontamination ofEquipment
Cliftonlaboratoryequipmentcan bedecontaminatedafterspillageorcontactwithpotentiallyHIV
andHepatitis infectedbloodsamplesduringanalysis usingfollowingrecommendedrapiddisinfect-
ants.
Virucidal Disinfectant
WerecommendVirkontabletsforthesafeandrapiddisinfectionof equipmentin awidevariety of
situations availablefromyourdistributororcontactDay-ImpexLtd. formoredetails.Telephone:
44+(0)1787 223232 or http://www.day-impex.co.uk
Theultimatehighlevelsurfacedisinfectant,dissolveVIRKONin water,providing
asafeworkingsolutionwithafaintlemonodor.It has provenefficacyagainst
bacteria (includingmycobacteria), viruses,sporesandfungiin avariety of inde-
pendenttestsusingdifferentprotocols.Presentsnoserious longtermhealthrisks
to staff–obviatingtheneedforcostlyventilationequipmentandhealthmonitor-
ing.Alsoprovideshighleveldisinfectionoflaboratoryequipmentandinstruments
whereautoclavingis neitherpracticalnornecessary.Formoredetailedinforma-
tionrelatingtohowVirkonshouldbeusedwithaccesstotestreports
www.relyon.dupont.com
IsVirkonsolutioncorrosive?Virkonsolutionrequiresonly10 minutescontacttimetobeeffective
solong-termexposure isnotnecessaryandthereforewillnotcorrodemostmaterials.Careshould
betakenwithStainlesssteelwater bathtanks,thesesurfacesshould notbeaffectedhowever,it
isimportantthatgenerallyyoudonotleave Virkonsolutionincontactwithmetalsurfaces“FOR
LONGERTHANISNECESSARY”.
Virkonis Registeredin accordancewiththerequirementsof theMedicalDevicesDirective,(93/42/
EEC)as aMedicalDevice.
Disinfectant/Sterilant
WerecommendPeraSafeapowderproductforthesafeandrapidchemicalsterilantof equipmentin a
widevariety of situations availablefromyourdistributororcontactDay-ImpexLtdformoredetails.
Telephone:44+(0)1787 223232 or http://www.day-impex.co.uk
PeraSafehas aprovensafety profileforend-userswithnoneof the
undesirablepropertiesof skin sensitisation,toxicfumesorunpleasant
odoursthatareassociatedwithaldehydesolutions.
LeadingUK andUSAmicrobologistshaveprovenPeraSafeto beactive
againstviruses,mycobacteria andfungi.It is microbiologicallysuperior
to glutaraldehyde,destroyingsporingbacteria in oneminute.It has also
beenindependentlyproventhatPeraSafesterilisesin just10 minutes.
Formoredetailedinformationrelatingto howPeraSafeshouldbeusedwithaccessto test reports
www.relyon.dupont.com
TheCliftonRange-3YearWarranty 10 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

3YearWarranty
Ourserviceengineersarefullytrainedin theassembly,calibrationandservicingof all Cliftoninstru-
mentation.Productscanbereturnedto ourcomprehensivelyequippedservicecentrewhereafastand
efficientturnaroundis guaranteed:
ServiceDepartment,NickelElectroLimited, OldmixonCrescent,Weston-super-Mare,NorthSomerset
BS24 9BL,UK.Tel+44 (0)1934 626691 Fax +44 (0)1934 630300.
OutofWarranty
OurServiceDepartmenthas acomprehensivestockof charegeablesparepartsmaintainingworking
lifeof equipmentorunitscan bereturnedforquotationbeforerepairsareundertaken.
EndofLife
This symbolindicatesthatthis productshould notbedisposedof withyourwaste.
Instead,disposewasteelectricalequipment by handingit overtoadesignated
collectionpointfortherecyclingof wasteelectricalandelectronicequipment.
Theseparatecollectionandrecyclingofyourwasteequipmentatthetimeof
disposalwill help to conservenaturalresourcesandensurethatitis recycledin a
mannerthatprotectshumanhealthandtheenvironment.Formoreinformation
aboutwhereyoucandropoffyourwasteequipmentforrecycling:UKplease
contactServiceDepartment,restEuropecontactyourDistributor.
Health&Safety,unlessinreceiptofaDecontaminationNoticeorReporttheunitcannotbe
returnedoracceptedfordisposal.
Cliftonelectricalandelectronicequipmenthasbeendesignedforrecyclingandtakesintoaccount
thedismantlingandrecoveryitscomponentsandmaterials.Cliftonproductsareeasilyrecycledwith
majorityof theproduct constructedfromstainlessormildsteels,whichcan readilybere-usedor
recycled.
Inexcessof86%ofthisproductrange canbeeasily re-cycledeconomically.
PortableApplianceTesting
Whenconductingtesting,ensureitis conductedbyaqualifiedperson.
DONOTPATTESTTHEBATH UNLESS ITCONTAINSWATER.
THISEQUIPMENTMUSTNOT BEFLASHTESTED!
TheCliftonRange-3YearWarranty 11 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

Accessories forNE1andNE1BSeries
GRP GableLids
GL1-4 GRPGableLid tosuit4LitreUnstirredBaths
GL1-14 GRPGableLid tosuit14 LitreUnstirredBaths
GL1-22 GRPGableLid tosuit9,22 and28 LitreUnstirredBaths
Stainless SteelGableLids
SL1-4 Stainless SteelGableLid tosuit4LitreUnstirredBaths
SL1-8 Stainless SteelGableLid tosuit8LitreUnstirredBaths
SL1-14 Stainless SteelGableLid tosuit14 LitreUnstirredBaths
SL1-22 Stainless SteelGableLid tosuit9,22 and28 LitreUnstirredBaths
SL1-22H Stainless SteelHingedGableLid tosuit22 +28 LitreUnstirredBaths
Stainless SteelFlatLids
LD-4 FlatOne PieceStainless SteelLid tosuit4LitreCapacityBath
LD-8 FlatOne PieceStainless SteelLid tosuit8LitreCapacityBath
LD-14 FlatOne PieceStainless SteelLid tosuit14 LitreCapacityBath
LD-22 FlatOne PieceStainless SteelLid tosuit9,22 &28 LitreCapacityBath
SA01368 FlatOne PieceStainless SteelLid tosuit56 litrebath(tworequired)
Stainless SteelFlatLidswithConcentricRings
SLR1-4 4x105mm Stainless SteelRingedLid tosuit4LitreUnstirredBaths
SLR1-8 4x83mm Stainless SteelRingedLid tosuit8LitreUnstirredBaths
SLR1-14 4 x105mm Stainless SteelRingedLid tosuit14 LitreUnstirredBaths
SLR2-14 6 x83mm Stainless SteelRingedLid tosuit14 LitreUnstirredBaths
SLR1-22 6 x105mm Stainless SteelRingedLid tosuit9,22 and28 LitreUnstirredBaths
Stainless SteelTestTubeRacks-Dimensions 270 x70 x138mm (LxWxH)
6870 Stainless SteelTestTubeRack26 Holex17mm Diameter
6871 Stainless SteelTestTubeRack16 Holesx26mm Diameter
6872 Stainless SteelTestTubeRack36 Holesx13mm Diameter
6873 Stainless SteelTestTubeRack18 Holesx19mm Diameter/suitablefor1.5mlmicrotubes
6900 Stainless SteelTestTubeRack12 Holesx32mm Diameter
Note:4Litre=1Rack, 8Litre=2Racks,14 Litre=4Racks,22 and28 Litre=6Racks
Miscellaneous
LB-2.5 2.5Litres-LabBath4590 -Heattransferfluid -Temperaturerange-15° to+90°C
LB-5.0 5Litres-LabBath4590 -Heattransferfluid -Temperaturerange-15° to+90°C
TC-1 ThermometerClip CompleteWithBentStemSpiritFilledThermometer
BX0616 DrainingSyphon
BX0688 DrainTap
BP0368 Polypropylene Spheres
Formoreinformation onAccessories visit www.nickel-electro.co.uk
TheCliftonRange-3YearWarranty 12 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

FittingAccessories forNE1andNE1Series
GRP GableLids
Positionthelidoverthebathandlowerinto place.
Thelidsaredesignedto fitsnuglysothatcondensateruns into thebath.
Stainless SteelGableLids
Positionthelidoverthebathandlowerinto place.
Theselidshavebodrippoints,providefull workingheightacross thebathwhenfittedandhavea
ventilationgap.
For“H”suffix lidsplacethelidonto thebathwiththehingeshangingovertherearof thewater
bathaliigningwiththreadedbushes.Inserttwoscrewsprovidedinto eachhingeandtightenscrews.
Thelidcan beraisedandloweredusingthehandleontheside.
Stainless SteelFlatLids
Positionthelidoverthebathandlowerinto place.
Stainless SteelFlatLidswithConcentricRings
Positionthelidoverthebathandlowerinto place.Eachsetof ringscan beusedto varythediam-
eterof apertureto accommodatedifferentsizesof vesselsof evaporatingdishesandglassware,
includinground-bottomflasks.
Stainless SteelTestTubeRacks
Placetheloadedtesttuberackinto thebathontopof eitherthestainless steelperforatedshelfor
theraisedshelf.
Note:4Litre=1Rack,8Litre=2Racks,14 Litre=4Racks,22 and28 Litre=6Racks
HeatTransferFluid
Useas a‘neat’liquidsolutionforbestresults.
ThermometerClip
This can bepositionedanywhereontheperimeterof thebathbyclipingovertheedge.Thether-
mometerthenliesflush withthetopedgeof thebathto preventaccidentaldamage.
DrainingSyphon
Allowstheeasyandquickemptyingof anywaterbath.Watertemperaturemustbebelow45°C
beforeemptyingcan commence.
Drain Tap
Adrain tapoutletis featuredonthe22,28 and56 litresizes.Makesurethebathis empty,hold the
outernut,thenusingasocket/allenkeyundothethreadedinsert.Onceremovedandstill holding
theouternut,screwin theaccessorydrain tapinto positionandpinchtight.
PolypropyleneSpheres
Placeontopof thewater/oilto provideafloatinglidreducingevaporationandinsulatingthebath
reducingheatlosses.Therangeof Unstirredwaterbaths:4litrebathrequires1pack;8litrebath
requires2pack;14 litre2packsand9,22 and28 litrerequires4packs.
TheCliftonRange-3YearWarranty 13 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

NE1Brange Clifton BoilingBaths,water,unstirred
Designedforapplications thatrequirecontinuous boiling.This rangealsohas theflexibility to allow
thetemperaturetoberegulatedfromagentlesimmertoavigorous boil.
Comprisingofacorrosionresistantstainlesssteelworkingenvironment,withheaterandsafety cut
outconcealedunderneathprovidingatotallyclear‘clutterfree’workingarea,allhousedinadurable,
chemicalresistant,powdercoatedpaintfinishedoutercase.
Integralconstantleveldeviceforconnection towatersupplytocombat evaporation ofwater
andmaintainboiling.
FeaturingClifton “Low”heightremovablestainless steelshelfincreasingworkingvolume.
Drainoutleton 28litremodelforconvenientbathemptying
Heatingelementwithsafetycut-outconcealedunderneaththebath
CatNo. NE1B-8 NE1B-14 NE1B-28
Capacity,litres 8 14 28
Heaterpower,W 800 1200 1500
Drain Outlet No No Yes
Safety Fixedcutout Fixedcutout Fixedcutout
InternalDims,mm(wxdxh) 240x300x150 325x300x150 500x300x200
OverallDims,mm (wxdxh) 332x270x290 361x332x290 538x332x290
Voltage 230Vor120V
TheCliftonRange-3YearWarranty 15 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

NE1range Clifton Baths,water, thermostatic,unstirred
Abroadrangeof unstirredanaloguebathsdesignedforgeneraluseandeconomicallypriced
withoutcomprisingquality.Comprisingofacorrosionresistantstainlesssteelworkingenvironment
usinga304 spec(18/8)stainless steeltankwithheaterandsafety cutoutconcealedunderneath
providingatotallyclear‘clutterfree’workingarea,all housedinadurable,chemicalresistant,
powdercoatedpaintfinishedoutercase.Easyto usethermostaticcontrol,graduatedin 5°C
increments,withilluminatedon/off switchand“heateron” indicator.
FeaturingremovableuniqueClifton “Low”heightremovablestainlesssteelshelf,watersaving
featureandreducingenergyconsumption orincreasingworkingvolume.
Sideliftinghandles forease oftransportation.
Drainoutleton largervolumes 22,28and56 litremodelsforsafeandconvenientbath
emptying.
Temperaturefactorycalibrated@45°Cforconsistency.
Complementaryrangeofaccessories includingracksandlids.
Cat No.NE1-2.5 NE1-4 NE1-8 NE1-14 NE1-22 NE1-28 NE1-56
Capacity,L2.5 4 8 14 22 28 56
Temperature Ambient +5°C-100°C
Sensitivity ±0.25°C
Uniformity ±0.1°C
Heater,W 250 400 800 1000 1500 1500 2500
DrainOutlet No No No No Yes Yes Yes
Internal,mm 230x130x100 300x150x150 300x240x150 325x300x150 500x300x150 500x300x200 620x500x200
Overall,mm 270x190x240 332x185x290 332x270x290 361x332x290538x332x290538x332x290 660x540x290
Voltage 230V or120V
TheCliftonRange-3YearWarranty 16 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

NE1range Clifton Baths, shallowform, water,thermostatic, unstirred
Ashallowheightunstirredwaterbathdesignedforoperatingwithlowliquidlevelssuitableforsmall
bottles,tubesandmicroplates.
PerfectforheatingsamplesinMicrocentrifuge tubesandmolecularbiology
Lowprofiledesignallowingeasyaccess andvisibilityofworkinthebath
Unliketraditionalhighsidedbaths,smallcontainersandsamplesarenotobscured
CatNo. NE1-9
Capacity,litres 9
Temperaturerange Ambient +5°C-100°C
Sensitivity ±0.25°C
Uniformity ±0.1°C
Heaterpower,W 1000
Drain Outlet No
InternalDims,mm(wxdxh) 500x300x65
OverallDims,mm (wxdxh) 535x332x135
Voltage 230Vor120V
TheCliftonRange-3YearWarranty 17 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk

ECDeclarationofConformity
Weherewithconfirmthefollowingproduct:
NE1BandNE1Series ofUnstirredWaterBaths
ConformwiththerequirementsoutlinedbythefollowingEuropean Directives:
LowVoltageDirective(73/23/EEC)
EMCDirective(89/336/EEC)
Weconfirmthedeclaration
Conformswiththerequirementsof followingstandards:
BSEN61010:1
BSEN61010:2.010
Safety requirementsforelectricalequipmentfor
measurement,controlandlaboratoryuse.
BSEN61326
Electricalequipmentformeasurement
controlandlaboratoryuse-EMCrequirements.
Nickel-ElectroLtd, OldmixonCrescent,
Weston-s-Mare,NorthSomerset,
BS24 9BL,England
t01934 626691
f01934 630300
NICKEL-ELECTROLtd.
Manufacturers oflaboratory,medicaland clinicalequipment.
OldmixonCrescent, Weston-super-Mare,
NorthSomerset, BS24 9BL,United Kingdom.
Tel:+44(0)1934626691
Fax:+44(0)1934630300
Email: clifton@nickel-electro.co.uk
www.nickel-electro.co.uk
TheCliftonRange-3YearWarranty 18 HighQualityProductsatAffordablePrices
PDFcreated withpdfFactoryProtrialversion www.software-partners.co.uk
This manual suits for next models
5
Table of contents
Other Nickel-Electro Laboratory Equipment manuals

Nickel-Electro
Nickel-Electro Clifton 1000ER Series User manual

Nickel-Electro
Nickel-Electro Clifton Range NE4-14D User manual

Nickel-Electro
Nickel-Electro Clifton RANGE NE2D Series User manual

Nickel-Electro
Nickel-Electro NE4-T Series User manual

Nickel-Electro
Nickel-Electro Clifton Range NE9S Series User manual

Nickel-Electro
Nickel-Electro Clifton Range SM-1 User manual