Ordering Information: Disposable BiPolar EHL Probes
9-195-371......1.9fr. (0.63mm) Biliary Probe (375cm)(Sterile) 1/Box
9-195-25 ……1.9fr. (0.63mm) Biliary Probe (250cm)(Sterile) 1/Box
9-196-12 ….. 1.9fr. (0.63mm) Ureteral Probe (120cm)(Sterile) 3/Box
9-900-54 ….. 9.0fr. (3.0mm) Urinary Bladder Probe (54cm)(Sterile) 3/Box
9-001-00 ….. Pneumatic Foot Switch
9-002-00 ….. Extender Cable
9-003-00 ….. U.S. Power Cord
The OUTPUT light will flash every time a shot has been triggered.
Note:
• Using higher power settings and a higher number of shots
may reduce probe life.
• The most effective stone fragmentation occurs with the end
of the EHL probe 1-2mm from the stones surface.
• Press the Standby button to avoid
unintentional probe firings. This feature is
useful if the scope and EHL probe are brought
outside of the body.
• Press Total Shots to see how many shots
have been discharged.
Frequency of Shots
This adjusts how fast the selected number of shots are
delivered. For stones that are free to move or bounce
away, select fewer shots at a higher frequency. A rate
of 30 shots per second is the highest (fast) frequency
setting. A rate of 1shot per second is the slowest
frequency setting. Suggested setting: 30/sec.
Number of Shots
Again, if the stone can move or bounce away, select
fewer shots. A maximum of 60 shots delivered over 2
seconds is the maximum number of shots that can be
fired with 1 foot-pedal depression.
Suggested setting: 3-5 shots per foot-pedal
depression.
Keeping the probe tip 1-2mm
from the stone surface is ideal.