OfiTE 175-25 User manual

Aging Cells
#175-25 Stainless Steel Grade 303, 260 mL
#175-30 Stainless Steel Grade 303, 500 mL
#175-50 Stainless Steel Grade 316, 500 mL
Instruction Manual
Updated 6/14/2019
Ver. 4
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ote.com
©Copyright OFITE 2014

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 1
Intro..................................................................................................2
Components ...................................................................................3
Parts and Accessories ...................................................................4
Specications .................................................................................5
Operation.........................................................................................6
Maintenance....................................................................................8
Warranty and Return Policy ..........................................................9
Table of
Contents

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 2
The OFITE Aging Cell is a patented (U.S. Patent No. 4,805,443) pressure
vessel that enables samples to be subjected to temperatures higher than the
boiling point of water and still be maintained in a liquid state. The cells may
be used for static temperature exposure or in a dynamic mode in a roller oven
with a preset minimum aging time of 16 hours.
OFITE Aging Cells are available in 260 mL and 500 mL sizes and utilize both
Viton® and Teon®o-ring seals. Aging cells are typically constructed of stain-
less steel (grade 303 or grade 316) and are used for high-temperature testing
(up to 500°F / 260°C). For prolonged exposure to elevated salinity at high
temperatures (such as 20,000 mg/L chlorides at 350°F / 176.6°C), cells con-
structed from premium metals, such as Hastelloy®or Inconel®600 are avail-
able. The aging cell walls may be protected against corrosive uids by using
the popular Teon®liner (part no. 175-60), designed by OFITE. Refer to the
Teon®liner instruction manual for more details.
For corrosion testing, a special 500 mL, 303 alloy stainless steel Corrosion
Test Cell is available with a modied inner cap designed to hold a corrosion
coupon. Also available in Hastelloy®upon request.
An optional calibrated rupture disk is available which can be installed in the
inner cap of the Aging Cells to rupture and release pressure at a predeter-
mined set point. OFITE can retrot the rupture disk to existing cells.
Intro

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 3
Components The following components are included with OFITE Aging Cells:
#170-17 Valve Stem O-ring, Viton, Qty: 6
#175-05 Thrust Washer
#175-09-1 Teon®O-ring for Inside of Aging Cell Body (For Tempera-
tures Above 400°F), Qty: 3
#175-09-2 Viton®O-ring for Inside of Aging Cell Body (For Temperatures
Below 400°F), Qty: 3
#175-14 Set Screw, ⅜" Diameter, Qty: 3
#175-15 Wrench for ⅜" Set Screw
#175-16 Valve Stem
#175-47 O-ring for Outside of Aging Cell Body, Viton 90, Qty: 2
#175-30-SP Spare Parts Kit
#170-17 Valve Stem O-ring, Qty: 36
#175-09-1 Teon®O-ring for Inside of Aging Cell Body (For Tempera-
tures Above 400°F), Qty: 12
#175-09-2 Viton®O-ring for Inside of Aging Cell Body (For Temperatures
Below 400°F), Qty: 12
#175-14 Set Screw, ⅜" Diameter, Qty: 6
#175-15 Wrench for ⅜" Set Screw
#175-16 Valve Stem, Qty: 2
#175-46 Teon®O-ring for Outside of Aging Cell, Qty: 4

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 4
Parts and
Accessories
Caps
#175-13 Outer Cap for Pressurized Aging Cells, 303 Stainless Steel
For tests below 400°F:
#175-18 Inner Cap for Pressurized Aging Cells, 303 Stainless Steel
#175-18-1 Inner Cap for Pressurized Aging Cells, 316 Stainless Steel
For tests above 400°F:
#175-00-3 Inner Cap, Accepts Rupture Disk, 316 Stainless Steel
#175-56 Rupture Disk, ¼", 2,000 psi
#175-57 Rupture Disk, ¼", 1,500 psi
For tests above 400°F, always use an inner cap with a rupture disk.
O-rings and Gaskets
For tests below 200°F:
#175-54 Buna N O-ring for Outside of Aging Cell
For tests up to 400°F:
#170-17 Viton®O-ring for Valve Stem
#175-09-2 Viton®O-ring for Inside of Aging Cell Body
#175-47 Viton®O-ring for Outside of Aging Cell
#175-62 Viton® O-ring for Teon®Liner Plug
#175-63 Viton® O-ring for Teon®Liner Lid (Piston)
For tests above 400°F:
#175-09-1 Teon®O-ring for Inside of Aging Cell Body
#175-46 Teon®O-ring for Outside of Aging Cell
#175-03 Peek O-ring for Inside of Aging Cell Body
Buna N O-rings should only be used for temperatures below 200°F.
Viton®O-rings can be used at temperatures up to 400°F.
Teon®O-rings should be used for temperatures above 400°F.
Set Screws and Wrenches
#175-14 Set Screw for Pressurized Aging Cells, ⅜"
#175-15 Wrench for ⅜" Set Screw
#175-60 Teon®Liner, for 500 mL Aging Cells with T-screw
#175-60-1 Liner
#175-60-2 Piston
#175-60-3 Plug
#175-60-4 T-screw
#175-62 Viton® O-ring for Teon®Liner Plug
#175-63 Viton® O-ring for Teon®Liner Piston

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 5
Specications Maximum Temperature: 500°F (260°C)
Maximum Pressure: 2,000 psi (13.8 MPa)
For tests above 200°F, refer to the chart below for the appropriate pres-
sure.
Do not use nitrous oxide cartridges as pressure sources for high-
temperature, high-pressure (HTHP) aging. Under high temperature and
pressure, nitrous oxide can detonate in the presence of grease, oil, or
carbonaceous materials. Nitrous oxide cartridges are to be used only
for Garrett Gas Train Carbonate Analysis.
Carbon dioxide and nitrous oxide cartridges are pressurized to approxi-
mately 900 psi at 1 atmosphere (sea level). Therefore, they should never
be transported by airplane without proper packing because cabin de-
pressurization may cause an explosion.
If the aging cells are going to be rolled in a roller oven during a test,
install o-rings on the outer perimeter on the top and bottom of the cells.
Failuretodosocandamagetherollersintheoven.Teon(#175-46),
Viton®(#175-47), and Buna N (#175-54) o-rings are available.
Cell Corrosion
Test uids under high tempreature and pressure can corrode the cell body
and caps. Carefully inspect the cell body and calls for corrosion before and
after each test.
Some materials are more susceptible to corrosion than other. Also, some
uids and additives are more corrosive than others. OFITE offers a variety of
cell materials for different levels of corrosion resistance and cost.
Mud Volume and Pressure for
High-Temperature Aging
Aging
Temp.
(°F / °C)
Water
Vapor
Pressure
(psi)
Coefcient
of
Expansion
of Water
Suggested
Applied
Pressure
(psi / kPa)
Mud
Volume in
260 mL
Cell (mL)
Volume
with
Teon
Liner
Mud
Volume
in 500
mL Cell
(mL)
Volume
with
Teon
Liner
212 / 100 14.7 1.04 25 / 172 225 130 450 326
250 / 121 30 1.06 50 / 345 225 130 450 326
300 / 149 67 1.09 100 / 690 200 116 425 308
350 / 176 135 1.12 150 / 1,034 200 116 400 289
400 / 204 247 1.16 250 / 1,724 - - 375 271
450 / 232 423 1.20 300 / 2,069 - - 375 253
500 / 260 680 1.27 375 / 2,586 - - 325 235

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 6
Operation 1. Carefully inspect the o-ring for defects and place it in the groove in the
cell body. Replace the o-ring if it is hard or has cuts and nicks. Blow air
through the valve stem to make sure it is not plugged.
O-ring
2. Determine a safe volume and safe initial pressure for the temperature at
which the sample will be tested (refer to the chart on page 5). Place
the correct volume of uid into the cell. Be careful not to get uid in the
o-ring groove or on the o-ring.
Donotoverllthecell.
3. Clean any spilled uid from the edge of the cell and place the inner cap
on top of the cell body o-ring so that it seats in place. Hand tighten the
outer cap in place. Using the Allen wrench, tighten the set screws in the
outer cap. Insert the valve stem with the o-rings in place into the inner cap
and tighten completely. Loosen the valve stem approximately one half
turn before pressurizing.
Inner Cap
Place O-rings
Here
Valve Stem

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 7
4. When the desired pressure is reached, close the valve stem by tightening
it with a wrench. To ensure there are no leaks, immerse the cell in water
and check for any bubbles coming from the valve stem or cell cap.
5. Place the Aging Cell inside the oven and adjust to the desired tempera-
ture.
6. After the desired aging time has elapsed, remove the cell from the oven
and allow it to air cool until the sample temperature is 300°F (149°C) or
less. The cell may then be either air or water cooled.
The sample temperature must be equal to the ambient temperature
before you release the pressure and open the cell.
7. Loosen the set screws, unscrew the outer cap, and remove the inner cap.
8. Clean out the valve stem with water. Blow air through the stem to remove
any residual water.
9. Observe the aged uid and record the condition as “uid”, “gelled”,
“plastic”, “hard”, etc. You may also want to test for viscosity, shear or gel
strength, or ltration control.
10. Thoroughly clean the entire cell with soap and water.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 8
Maintenance Regular maintenance will extend the useful life of the aging cells.
After every test:
1. Thoroughly clean and dry all components including the cell body,
inner cap, outer cap, valve stem, and o-rings.
2. Inspect all o-rings for damage or wear. Discard any that are nicked,
cracked, or distorted from their original shape. Replace them with new
o-rings.
3. Apply grease to the threads on the outer cap and cell body.
Every three years (yearly for highly corrosive conditions):
1. Sand blast the metal components (cell body, inner cap, outer cap).
2. Visually inspect the cell body and inner cap for pitting or other signs of
corrosion. If any material loss is present, replace the damaged com-
ponent.
3. Inspect the threads on the outer cap and cell body for burrs.
4. Replace all o-rings and apply grease to the threads on the outer cap
and cell body.
5. Pressure test with water up to 4,000 psi.
OFITE offers aging cell recertication that includes all of the above plus a
one-year certicate. Contact your sales representative for more information.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 9
Warranty and
Return Policy
Warranty:
OFI Testing Equipment, Inc. (OFITE) warrants that the products shall be free from liens and defects in
title, and shall conform in all respects to the terms of the sales order and the specications applicable
to the products. All products shall be furnished subject to OFITE’s standard manufacturing variations
and practices. Unless the warranty period is otherwise extended in writing, the following warranty shall
apply: if, at any time prior to twelve (12) months from the date of invoice, the products, or any part
thereof, do not conform to these warranties or to the specications applicable thereto, and OFITE is
so notied in writing upon discovery, OFITE shall promptly repair or replace the defective products.
Notwithstanding the foregoing, OFITE’s warranty obligations shall not extend to any use by the buyer
of the products in conditions more severe than OFITE’s recommendations, nor to any defects which
were visually observable by the buyer but which are not promptly brought to OFITE’s attention.
In the event that the buyer has purchased installation and commissioning services on
applicable products, the above warranty shall extend for an additional period of twelve
(12) months from the date of the original warranty expiration for such products.
In the event that OFITE is requested to provide customized research and development for the buyer,
OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided.
OFITE makes no other warranties or guarantees to the buyer, either express or implied,
and the warranties provided in this clause shall be exclusive of any other warranties
including ANY IMPLIED OR STATUTORY WARRANTIES OF FITNESS FOR PURPOSE,
MERCHANTABILITY, AND OTHER STATUTORY REMEDIES WHICH ARE WAIVED.
This limited warranty does not cover any losses or damages that occur as a result of:
• Improper installation or maintenance of the products
• Misuse
• Neglect
• Adjustment by non-authorized sources
• Improper environment
• Excessive or inadequate heating or air conditioning or electrical
power failures, surges, or other irregularities
• Equipment, products, or material not manufactured by OFITE
• Firmware or hardware that have been modied or altered by a third party
• Consumable parts (bearings, accessories, etc.)
Returns and Repairs:
Items being returned must be carefully packaged to prevent damage in
shipment and insured against possible damage or loss. OFITE will not be
responsible for equipment damaged due to insufcient packaging.
Any non-defective items returned to OFITE within ninety (90) days of invoice are subject to a
15% restocking fee. Items returned must be received by OFITE in original condition for it to
be accepted. Reagents and special order items will not be accepted for return or refund.
OFITE employs experienced personnel to service and repair equipment manufactured by us, as
well as other companies. To help expedite the repair process, please include a repair form with all
equipment sent to OFITE for repair. Be sure to include your name, company name, phone number,
email address, detailed description of work to be done, purchase order number, and a shipping address
for returning the equipment. All repairs performed as “repair as needed” are subject to the ninety (90)
day limited warranty. All “Certied Repairs” are subject to the twelve (12) month limited warranty.
Returns and potential warranty repairs require a Return Material Authorization (RMA)
number. An RMA form is available from your sales or service representative.
Please ship all equipment (with the RMA number for returns or warranty repairs) to the following address:
OFI Testing Equipment, Inc.
Attn: Repair Department
11302 Steeplecrest Dr.
Houston, TX 77065
USA
OFITE also offers competitive service contracts for repairing and/or maintaining your lab
equipment, including equipment from other manufacturers. For more information about
our technical support and repair services, please contact techservice@ote.com.
This manual suits for next models
2
Table of contents
Other OfiTE Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Konica Minolta
Konica Minolta Drypro 832HG Brochure & specs

Beckman Coulter
Beckman Coulter Microfuge 16 Instructions for use

Retsch
Retsch AS 400 control manual

Applied Biosystems
Applied Biosystems 7300 Getting started guide

Philips
Philips UV-C 9290024887 user manual

Hirschmann
Hirschmann Pipetus-Junior instruction manual