Pall 4821A User manual

User Guide
Laboratory Shaker
Part Number 4821A, 115 Volt CE
Part Number 4822A, 230 Volt CE

2
Tab e of Contents
Des ription ......................................................................................................................................................3
General Information ........................................................................................................................................3
Ordering Information........................................................................................................................................3
Spe ifi ations ..................................................................................................................................................3
Installation of the Laboratory Shaker ............................................................................................................4
Instru tions for Use ........................................................................................................................................4
Maintenan e ....................................................................................................................................................6
Repla ement Parts ..........................................................................................................................................6
Warning ............................................................................................................................................................7
Note: The proced res herein are intended only as a g ide. Users sho ld always verify prod ct performance with
their specific applications nder act al se conditions. If yo have q estions abo t the information presented in
this g ide, please contact Pall Life Sciences Technical Service.

Description
• Designed to hold capsule filters in the horizontal orientation for elution of particulate and organisms.
• Easily detachable arms equipped with eight cork-lined clamps.
• Process up to eight Envirochek®Sampling Capsules at one time.
• Programmable digital speed and timer.
Genera Information
1. Handle the parcel with care. Unpack and check that the contents coincide with the packing list.
If any part is damaged or missing, please advise the distributor immediately.
2. Do not install or use the equipment without first reading the Product Information Insert.
3. The manual should be made available for all users.
4. If you have any inquiries, please contact your supplier or Pall Life Sciences Technical Service.
5. Do not use the apparatus with liquids which can give off vapors capable of making explosive mixtures.
Ordering Information
Laboratory Shaker
Part Number Description Pkg
4821A 115V 1/pkg
4822A 230V 1/pkg
Pa king List
The standard equipment consists of the following components:
Product Qty
Laboratory Shaker (115 or 230) 1
Arms 2
C amps 8
Power Cord - 115V/230V 1/2
Product Information Insert 1
A en Wrenches (3 mm, 5/32, 3/32) 3
Specifications
Dimensions
Height: 37.21 cm
Length: 82.21 cm
Width: 34.72 cm
Maximum Capa ity
4 kg (8.8 lb.)
Os illations/Minute
100-900 RPM
Power Consumption
150 W
www.pall.com/lab 3

4
Weight
P 4821A: 40 kG (88.2 lb.)
P 4822A: 41 kG (90.4 lb.)
Voltage
P 4821A: 115V, 50/60 Hz
P 4822A: 230V, 50/60 Hz
Insta ation of the Laboratory Shaker
1. Place the apparatus on a level surface with sufficient stability to resist its own vibrations.
2. Fit the arms with clamps vertical at the shaker side shaft so that the screws tighten against the side of the
shaft.
3. If the shaker moves during operation, this can be alleviated by putting a piece of cushioned rubber between
the table and the shaker.
CAUTION!!!! IMPORTANT FOR YOUR SAFETY
Be sure that the voltage supply is the same as the one indicated on the characteristics plate.
Do not use the apparatus to shake products capable of making an explosive mixture.
Do not use the apparatus if it is not grounded.
Instructions for Use
Control Panel
1. Power Switch
2. Display – time, speed
3. D-Pad
4. Enter button
5. Cursor indicator
Starting Up
1. Plug Laboratory Shaker into 115V 50/60 Hz or 230V 50/60 Hz line as indicated on the characteristics plate
located on the back of the instrument.
2. Turn on the Power switch located on the front of the instrument. Follow steps outlined below for Continuous
Operation or Timed Operation.
FRONT PANEL DISPLAY
1. Run
2. Timed Run
*
5
1234

Continuous Operation Example
To run Laboratory Shaker in continuous (non-timed) operation, move the Cursor using the D-Pad to the 1. Run
from the Main Menu and press the Enter button. The Shaker will engage at the last remembered RPM.
The RPM can be adjusted by increments of +/- 50 using the D-Pad. The clock will keep track of the total run
time. Selecting Pause will stop the clock and disengage the Shaker.
Selecting Resume will start the clock count and re-engage the Shaker. Selecting Exit will disengage the Shaker
and return the user to the Main Menu.
Timed Operation Example
To run the Laboratory Shaker in timed run operation, select the 2. Timed Run from the Main Menu and press the
green button to navigate to the Timed Run Set-up Menu.
The first line will display Run Time, which controls the total run time before the Shaker stops. Hours, minutes,
and seconds can be selected using the left and right arrows on the D-Pad. Once the time has been set, press
the Enter button. The second line RPM can be adjusted by increments of +/- 50 using the D-Pad. Once the pa-
rameters have been set, select Start and press Enter to begin the timed run.
www.pall.com/lab 5
RUN TIME 00: 00: 01
RPM 100
*PAUSE EXIT
*PAUSE*
RUN TIME 00: 00: 01
RPM 100
RESUME EXIT
*
(R n Men )
(Pa sed R n Men )
1. RUN
2. TIMED RUN
*
(Main Men )
TIMED RUN SETUP
RUN TIME HR: MN: SC
RPM 100
START EXIT
*
(Timed R n Set- p Men )
REMAINING 99: 59: 59
RPM 960
*PAUSE EXIT
(Timed R n Men )

6
Selecting Pause will stop the clock and disengage the Shaker. Selecting Resume will start the clock count down
and re-engage the Shaker.
Selecting Exit will disengage the Shaker and return the user to the main menu.
Maintenance
Before removing the casing, always be sure the Laboratory Shaker is unplugged. The manipulation of the
internal electronic circuits of the apparatus by unauthorized personnel can cause irreparable damage to
the apparatus.
Cleaning
Cleaning of plastic: Use alcohol with cotton cloth.
Painted metal case: Use damp cloth and mild soap.
Rep acement Parts
Part Number Description
89051 C amp with extension co ar
*PAUSE*
REMAINING 99: 59: 59
RPM 960
RESUME EXIT
*
(Pa sed Timed R n Men )
1. RUN
2. TIMED RUN
*
(Main Men )

WARNING
Employment of the products in applications not specified, or failure to follow all instructions contained in this
product information insert, may result in improper functioning of the product, personal injury, or damage to
property or the product. See Statement of Warranty in our most recent catalog.
ATTENTION
L’utilisation de nos produits dans des applications pour lesquelles ils ne sont pas spècifiès ou le non-respect
du mode d’emploi qui figure sur ce document, peut entrainer un disfonctionnement du produit, endommager le
produit ou d’autres biens matèriels ou reprèsenter un risque pour l’utilisateur. Se rèfèrer à la clause de garantie
de notre catalogue le plus rècent.
ACHTUNG
Der Einsatz dieses Produktes in Anwendungen für die es nicht spezifiziert ist, oder das ichtbeachten einiger,
in dieser Bedienungsanleitung gegebenen Hinweise Kann zu einem schlechteren Ergebnis, oder Zerstörung des
Produktes oder anderer Dinge oder gar zu Verletzungen führen. Beachten Sie auch unsere Garantiebedingungen
im aktuellen Katalog.
ADVERTENCIA
El uso de este producto en applicaciones no especificadas o el no considerar las instrucciones indicadas en la
hoja de informaciòn del producto puede ocasionar un mal funcionamiento del producto, daños en las instala-
ciones o en el producto y riesgo para el personal del laboratorio. Consulte el apartado de Garantía en nuestro
ultimo catàlogo.
ATTENZIONE
L’impiego dei prodotti in applicazioni non specificate, o il manacato rispetto di tutte le istruzioni contenute nel
presente bollettino tecnico, potrebbero portare ad un utilizzo improprio del prodotto, ferire gli opertori, o dan-
naeggiare le caratteristiche del prodotto stesso. Consultare la dichiarazione di garanzia pubblicata nel nostro
più recente catalogo.
Pa Life Sciences
25 Harbor Park Drive
Port Washington, ew York 11050
Internationa Offices
Pall Corporation has offices and plants throughout the world in: Argentina, Australia, Austria,
Belgium, Brazil, Canada, China, France, Germany, India, Indonesia, Ireland, Italy, Japan,
Korea, Malaysia, New Zealand, Norway, Philippines, Poland, Russia, Singapore, South Africa,
Spain, Sweden, Switzerland, Taiwan, Thailand, nited Kingdom, and Vietnam. Distributors in
all major industrial areas of the world. To locate the Pall office or distributor nearest you, visit
www.pall. om/ onta t.
The information provided in this literature was reviewed for accuracy at the time of publica-
tion. Product data may be subject to change without notice. For current information consult
your local Pall distributor or contact Pall directly.
© 2015, Pall Corporation. Pall, and Envirochek are trademarks of Pall Corporation. ® indicates a
trademark registered in the SA. ENABLING A GREENER F T RE and Filtration. Separation. Solution.
are service marks of Pall Corporation.
2/15, PDF, GN14.1031 PN91048 Rev B
Visit us on the Web at www.pa .com/ ab
E-mai us at LabCustomerSupport@pa .com
To see how Pall is helping enable a greener, safer and more sustainable
future, visit www.pall.com/green.
This manual suits for next models
1
Table of contents
Other Pall Laboratory Equipment manuals

Pall
Pall Jet Mixer User manual

Pall
Pall Envirochek User manual

Pall
Pall T Series Guide

Pall
Pall Minimate TFF User manual

Pall
Pall Allegro 1000 L User manual

Pall
Pall WandMixer RDUA003A User manual

Pall
Pall Biotech RDUA007BT User manual

Pall
Pall Allegro User manual

Pall
Pall USD3295 User manual

Pall
Pall Minimate EVO User manual
Popular Laboratory Equipment manuals by other brands

Thermo Scientific
Thermo Scientific Thermo Scientific Orion Star T900 Series user manual

MOGlabs
MOGlabs MSA manual

Hanil
Hanil Supra R30 quick guide

PASCO
PASCO TD-8570A instruction sheet

D&H
D&H Lifelabs AquaSolvo Eco AAE-5005 Series instruction manual

Grant-bio
Grant-bio Platform Rocker PMR-100 operating instructions