Ohmeda 3000 User manual

EDB/
ECI
Infant
Warmer
Systems
Model
3000
and
3300
Service
Manual
BOC
Health
Care

Ohmeda
Service
Manual
Infant
Warmer
System
Model
3000
and
3300
Copyright
1986
Ohmeda
This
document
is
not
to
be
reproduced
in
any
manner,
nor
are
the
contents
herein
to
be
disclosed
to
anyone,
without
the
express
authorization
of
the
_Service
Department,
Ohmeda,
Madison,
Wisconsin.

Table
of
Contents
Εις
ιο
ee
eee
ee
eee
Technical
Competence
.
5/Disassembiy
and
Repair
_..................
5.1
Heater
Module
Repairs
........
…
Repair
Policy
.....................44........
A.
Heater
Housing
Disassembly
...
..
Precautions
iv
B.
Heater
Replacement
....................
Torres
assesses
cee
secs sess
teres
C.
Heater
Housing
Assembly
...............
1/Functional
Description
....................
1-1
D.
Alarm
Lamp
Replacement
........
A.
Power
Supply
Board
....................
1-1
E.
Examination
Lamp
Replacement
.. ..
B.
Control
Board
.........................
1-2 5.2
Control
Module
Repairs
.................
C.
Display
Board
.........................
1-8
A.
Control
Module
Removal
................
B.
Control
Board
and
Power
Supply
Board
2/Specifications
に
で
に
た
と
た だ
とまと すすま
たま
たま
と
た
まま
まま
まま
2-1
Replacement
..
ον
εντ
νωνν
ενω
κε
ων
εως
2.1
Electrical...
-.
24
C.
Display
Module
Disassembly
2.2
Controller
.............................
2-1
D.
Display
Module
Assembly
...............
2.3
Alarms
...............................
2-1
E.
Control
Module
Replacement
...
2.4
Environmental
................
2-2
F.
Battery
Replacement
.........
2.5
Mechanical
(without
accessories)
22
G.
Circuit
Breaker
Reset
..
2.8
Accessories
...................
2-2
5.3
Bed
Platform
Repairs
....
3/Setup
and
Checkout
Procedure
....
.
34
A.
Side
Panel
Replacement.
3.1
Setup......................
...
34
3.
Side
Panel
Repairs
.......
3.2
Checkout
Procedure
....................
34
5
Bed
Flatform
Disassembly
-
i .
Be
form
Asse
yo...
μα
B.
Control
Unit
Checks
....................
31
,F.
Hydraulic
System
Installation
5.4
Caster
Replacement
................-..
5-16
4/Calibration
and
Adjustments
..............
4-1
5.5
Yoke
Manifold
Repairs
...:.............
5-16
A.
Control
Unit
Access
..............
.
41
A.
General
.............
5-16
B.
Power
Supply
Board
Voltage
Checks
...
4-1
B.
Gauge
Replacement
.....
5-16
C.
Display
Brightness
Check
...........
.
4-1
C.
Gauge
Lens
Replacement
.
5-16
D.
Alarm
Volume
Adjustment
......
.
41
D.
Strainer
Replacement
.....
5-19
E.
Analog
to
Digital
Converter
(ADC)
.
41
E.
Check
Valve
Replacement
.....
5-19
F.
Line
Voltage Sensing
...........
4-2
F.
High
Pressure
Regulator
Repair
.
5-19
G.
Triac
Safety
Circuit
Test
.
42
G.
Pneumatic
Troubleshocting
...
5-20
H.
Test
Loop
............................
4-3
。
。
.
L
Oxygen
Regulator
Check
and
Adjustment
.
4-3
6/Control
Unit
Troubleshooting
Guide
....
.
6-1
y.
Electrical
Safety
Check
..................
7/Illustrated
Parts
and
Parts
List
.............
71
K.
Ground
Resistance
Check
...
L.
Leakage
Current
Tests
..................
8/Schematics
and
Wiring
Diagrams
...........
31
Appendix..................
„A1
A.
Test
Eguipment
and
Special
Tools.
.
А-1
B.
Temperature
Conversion
Chart
.....
.
A-1
C.
Patient
Probe
Characteristics
..
-
A-1
D.
Drawer
Kit
Accessory
.......
..
A2
E.
O2
Yoke
and
Regulator
Assembly
Kit
......
A3

Important
The
information
contained
in
this
service
manual
pertains
only
to
those
models
of
products
which
are
marketed
by
Ohmeda
as
of
the
effective
date
of
this
manual
or
the
latest
revision
thereof.
This
manual
‘was
prepared
for
exclusive
use
by
Ohmeda
service
personnel
in
light
of
their
training
and
experience
and
the
availability
to
them
of
proper
tcols
and
test
equipment.
Consequently,
Ohmeda
provides
this
manual
to
its
customers
purely
as
a
business
con-
venience
and
for
the
customer's
general
information
only
without
warranty
of
the
results
with
respect
to
any
application
of
such
information.
Furthermore,
because
of
the
wide
variety
of
circumstances
under
which
maintenance
and
repair
activities
may
be
per-
formed
and
the
unique
nature
of
each
individuals
own
experience,
capacity,
and
qualifications,
the
fact
that
customer
has
received
said
information
from
Ohmeda
does
not
imply
in
any
way
that
Ohmeda
deems
said
individual
to
be
qualified
to
perform
any
such
maintenance
or
repair
service.
Moreover,
it
should
not
be
assumed
that
every
ac-
ceptable
test
and
safety
procedure
or
method,
pre-
caution,
tool,
equipment
or
device
is
referred
to
within,
or
that
abnormal
or
unusual
circumstances
may
not
warrant
or
suggest
different
or
additional
procedures
or
requirements.
This
manual
is
subject
to
periodic
review
and
cus-
tomers
are
cautioned
to
obtain
and
consult
the
latest
revision
thereof
and
suggestions
are
invited
from
our
customers
for
consideration
by
Ohmeda
with
these
periodic
reviews.
WARNING:
After
completing
a
repair
of
the
Infant
Warmer
System
the
appropriate
calibration
proce-
dure
must
be
performed
to
make
sure
the
Infant
Warmer
System
is
in
proper
operating
condition.
In
addition
a
final
electrical
safety
check
and
leak-
age
current
test
must
be
performed.
Record
the
in-
formation
for
future
reference.
WARNING:
After
completing
any
portion
of
the
calibration
and
adjustments
procedure
for
the
In-
fant
Warmer
System
the
checkout
procedure
must
be
performed
to
make
sure
the
Infant
Warmer
System
is
in
proper
operating
condition.
In
addi-
tion
a
final
electrical
safety
check
and
leakage
cur-
rent
test
must
be
performed. Record
the
informa-
tion
for
future
reference.
CAUTION:
Servicing
of
this
product
in
accordance
with
this
service
manual
should
never
be
underta-
ken
in
the
absence
of
proper
tools,
test
equipment
and
the
most
recent
revision
of
this
service
man-
ual
which
is
clearly
and
thoroughly
understood.
This
static
control
precaution
symbol
appears
throughout
this
manual.
When
this
symbol
ap-
pears
next
to
a
procedure
in
this
manual,
static
control
precautions
must
be
observed.
Use
the
static
control
work
station
(Part
No.
0175-2311-
000)
to
help
ensure
that
static
charges
are
safely
conducted
to
ground
and
not
through
sta-
tic
sensitive
devices.
This
document
is
not
to
be
reproduced
in
any
man-
ner,
nor
are
the
contents
herein
to
be
disclosed
to
anyone, without
the
express
authorization
of
the
Ohmeda
Product
Service
Department,
Madison,
Wis-
consin.
iii
Repair
Policy
and
Procedure
Note:
Service
must
be
performed
by
a
“Technically
Competent”
individual.
Do
not
use
malfunctioning
equipment.
Make
all
nec-
essary
repairs,
or
have
the
equipment
serviced
by
an
Authorized
Ohmeda
Service
Representative.
After
repair
test
the
equipment
to
ensure
that
it
is
functioning
properly,
in
accordance
with
the
man-
ufacturer's
published
specifications.
To
ensure
full
reliability,
have
all
repairs
and
service
done
by
an
authorized
Ohmeda
Service
Representa-
tive.
If
this
cannot
be
done,
replacement
and
maintenance
of
those
parts
listed
in
this
manual
may
be
undertaken
by
a
competent,
trained
indi-
vidual
having
experience
in
the
repair
of
this
type
of
equipment.
CAUTION:
No
repair
should
ever
be
undertaken
or
attempted
by
anyone
not
having
such
qualifica-
tions.
Replace
damaged
parts
with
components
manufac-
tured
or
sold
by
Ohmeda.
Then
test
the
unit
to
as-
certain
that
it
complies
with
the
manufacturer's
published
specifications.
Contact
the
nearest
Ohmeda
service
office
for
ser-
vice
assistance.
If
you
send
the
unit
to
the
Ohmeda
Service
Center,
package
it
securely
in
the
original
shipping
container,
if
possible,
and
ship
it
prepaid.
Enclose
a
letter
with
the
unit
describing
in
detail
any
difficulties
experienced
and
the
repairs
felt
nec-
essary.
In
all
cases,
other
than
where
Ohmeda's
warranty
is
applicable,
repairs
will
be
made
at
Ohmeda's
current
list
price
for
the
replacement
part(s)
plus
a
reasonable
labor
charge.
CAUTION:
Detailed
drawings
and
procedures
for
more
extensive
repairs
are
included
herein
solely
for
the
convenience
of
users
having
proper
knowl-
edge,
tools,
and
test
equipment,
and
for
service
representatives
specially
trained
by
Ohmeda.
Technical
Competence
The
procedures
described
in
this
service
manual
should
be
performed
by
trained
and
authorized
per-
sonnel
only.
Maintenance
should
only
be
underta-
ken
by
competent
individuals
who
have
a
general
knowledge
of
and
experience
with
devices
of
this
nature.
Genuine
replacement
parts
manufactured
or
sold
by
Ohmeda
must
be
used
for
all
repairs.
Read
completely
through
each
step
in
every
proce-
dure
before
starting
the
procedure;
any
exceptions
may
result
in
a
failure
to
properly
and
safely
com-
plete
the
attempted
procedure.
Copyright
1986
by
Ohmeda,
Ohmeda
Drive,
Madi-
son,
Wisconsin
53707-7550.

Precautions
Warnings
After
completing
a
repair
of
the
Infant
Warmer
Sys-
tem
the
appropriate
calibration
procedure must
be
performed
to
make
sure
the
Infant
Warmer
System
is
in
proper
operating
condition.
In
addition
a
final
electrical
safety
check
and
leakage
current
test
must
be
performed.
Record
the
information
for
fu-
ture
reference.
After
completing
any portion
of
the
calibration
and
adjustments
procedure
for
the
Infant
Warmer
Sys-
tem
the
checkout
procedure
must
be
performed
to
make
sure
the
Infant
Warmer
System
is
in
proper
operating
condition.
In
addition
a
final
electrical
safety
check
and
leakage
current
test
must
be
per-
formed.
Record
the
information
for
future
reference.
Overloading
the
shelves
can
affect
the
stability
of
the
unit.
Do
not
perform
the
Check-Out
Procedure
while
a
patient
occupies
the
Infant
Warmer
System.
Never
oil
or
grease
oxygen
equipment
unless
a
lu-
bricant
that
is
made
and
approved
for
this
type
of
service
is
used.
Oils
and grease
oxidize
readily,
and
in
the
presence
of
oxygen,
will
burn
violently.
Vac
Kote*
is
the
oxygen
service
lubricant
recommended
(Order
No.
0220-0091-300).
When
replacing
gauges,
be
sure
to
use identical
pressure
ranges.
Do
not
use
oil
or
oil
bearing
materials
on
or
near
the
regulator.
Oils
and
greases
oxidize
readily
and,
in
the
presence
of
oxygen,
they
will
burn
violently.
All
metallic
parts
of
the
regulator
must
be
discarded
if
contaminated
with
oil
or
grease.
*
Vac
Kote
is
a
trademark
of
the
Ball
Corporation.
Cautions
Servicing
of
this
product
in
accordance
with
this
service
manual
should
never
be
undertaken
in
the
absence
of
proper
tools,
test
equipment
and
the
most
recent
revision
of
this
service
manual
which
is
clearly
and
thoroughly
understood.
This
static
control
precaution
symbol
appears
throughout
this
manual.
When
this
symbol
ap-
pears
next
to
a
procedure
in
this
manual,
static
control
precautions
must
be
observed.
Use
the
static
control
work
station
(Part
No.
0175-2311-
000)
to
help
ensure
that
static
charges
are
safely
conducted
to
ground
and
not
through
sta-
tic
sensitive
devices.
No
repair
should
ever
be
undertaken
or
attempted
by
anyone
not
having
such
qualifications.
Detailed
drawings
and
procedures
for
more
exten-
sive
repairs
are
included
herein
solely
for
the
con-
venience
of
users
having
proper
knowledge,
tools,
and
test
equipment,
and
for
service
representatives
specially
trained
by
Ohmeda.
For
the
Model
3000
Infant
Warmer
System
the
height
of
the
bed
should
be
27
inches
+2
inches
from
the
bottom
of
the
heater
module
for
proper
heating
and
temperature
control.
Insulation
on
electrical
wiring
can
deteriorate
with
age.
Check
for
brittle
or
deteriorated
insulation
on
power
cord
and
all
other
electrical
wiring.
Use
the
Static
Control
Work
Station
(Part
No.
0175-
2311-000)
to
help
ensure
that
static
charges
are
safely
conducted
to
ground.
The
Velostat
material
is
conductive.
Do
not
place
electrically
powered
circuit
boards
on
it.
The
back
panel
and
display
panel
may
drop
down
when
the
bottom
cover
mounting
screws
are
re-
moved.
Be sure
to
secure
the
panels
with
tape
be-
fore
disassembly.
Disconnect
the
Infant
Warmer
System
power
cord
and
allow
the
unit
to
cool
before
replacing
the
examination
or
alarm
light.
Disconnect
the
Infant
Warmer
System
power
cord
and
allow
the
unit
to
cool
before
replacing
the
examination
light.
The
lamp
normally
operates
at
a
high
temperature.
Take
static
precautions
for
these
procedures.
For
safety
have
at
least
2
people
available
to
re-
place
a
caster.
Remove
all
accessory
equipment.
iv


1/Functional
Description
A.
Power
Supply
Board
This
is
a
functional
description
for
the
Infant
Warmer
System
Power
Supply
Board
Part
No.
0631-
5032-700.
Réfer
to
Schematic
No.
0676-0327-000
for
a
detailed
circuit
diagram.
The
power
supply
board
contains
circuitry
for
the
control
and
monitoring
of
voltage
devices.
The
board
also
provides
power
to
the
control
board
and
the
display
board.
Also
found
on
the
board
is
a
line
voltage
sensing
circuit
that
provides
an
indication
of
line
voltage
to
the
microcontroller.
.
The
control
circuits
for
each
line
voltage
device
on
the
power
supply
board
are
functionally
identical
with
a
logic
“High”
signal
from
the
control
board
switching
"On”
the
desired
device.
This
is
per-
formed
with
an
opto-isolator
triac
driver
so
low/line
voltage
circuits
can
interact
but
remain
electrically
isolated
(2500
volt
dielectric).
The
heater
is
controlled
from
the
supply
board
with
an
opto-isolator
triac
driver
and
triac.
There
is
also
a
relay
contact
connected
in
series
with
the
neutral
to
the
heater
triac.
This
is
used
to
switch
“Off"
the
heater
if
the
heater
triac
fails
or
there
is
a
failure
on
the
control
board.
The
regulator
circuits
provide
a
+5
Vdc
supply
to
the
display
board
and
+5
Vdc,
and
+9
Vdc,
supplies
to
the
control
board.
A
ni-cad
battery
supplies
the
§
Vdc
supply
and
a
de-rated
9
Vdc
sup-
ply
for
standby
power,
in
the
case
of
a
power
loss.
Standby
power
of
9
volts
is
used
to
activate
the
transducer
alarm,
while
the
5
volt
supply
provides
power
to
the
microcontroller
and
associated
IC's
for
memory
retention
purposes.
5
Volt
LEDS
A
nominal
8
Vac
is
input
to
the
power
supply
board
at
J11
pins
3
and
4.
The
line
frequency
is
also
con-
nected
to
the
control
board
via
J12
pin
2.
The
bridge
rectifier
CR2
and
capacitor
C11
provide
a fil-
tered
unregulated
8
Vdc
to
the
relay,
opto-isolators,
and
the
regulator
VR12.
The
8
Vdc
unregulated
sup-
ply
can
be
measured
at
TP-1.
The
unregulated
sup-
ply
must
be
a
minimum
of
7.32
volts
for
proper
op-
eration
of
the
relay
circuit.
The
output
of
regulator
VR2
is
nominally
+5
Vde
and
supplies
power
to
drive
the
LED
displays
on
the
display
board.
The output
is
measurable
at
J12
pin
12
(TP-10).
When
the
supply
voltage
is
within
10%
of
nominal,
the
output
voltage
should
be
between
4.8
and
5.2
volts
dc
with
a
maximum
load
of
500
ma.
The
maximum
allowable
ripple
voltage
is
150
millivolts.
Line
Voltage
Sensing
A
voltage
of
approximately
11
Vac
from
the
trans-
former
secondary
is
input
to
the
board
at
J11
pins
1
and
2.
Bridge
rectifier
CR1
and
capacitor
C12
pro-
vide
a
full
wave,
filtered
voltage
of
approximately
12
Vdc.
Resistor
R3
is
preset
to
produce
an
output
of
approximately
0.6
volts
at
J12
pin
11
(TP-11)
when
the
line
voltage
is
at
the
nominal
value
for
the
unit.
The
analog
voltage
signal
at
J12
Pin
11
connects
to
the
control
board
and
is
fed
into
the
A/D
Converter,
ADC
3711
(U6),
via
the
multiplexer,
MC14051B
(U-
13).
The
digital
output
of
the
A/D
converter
is
input
to
the
microcontroller
where
the
software
then
ad-
justs
the
power
to
the
heater
to
compensate
for
var-
iations
in
line
voltage.
9
Volt
Standby
The
output
of
regulator
VR3
is
adjusted
by
R4
to
provide
9.0
+
0.2
volts
(TP12).
This
voltage
is
used
for
charging
the
NI-CAD
battery,
and
supplying
the
input
voltage
to
the
+5
standby
regulator.
5
Volt
Standby
When
line
voltage
is
available,
current
flows
from
the
output
of
VR3
and
through
CR5
to
provide
9.0
+
0.2
volts
to
the
input
of
VR4,
and
to
J12
pin
3
(TP-12).
In
tum,
regulator
VR4
outputs
a
voltage
of
5.0
+
0.2
volts
to
J12
pin
14
(TP-9)
with
a
maximum
ripple
voltage
of
150
millivolts.
If
power
loss
occurs
with
the
unit
switched
ON,
the
7.2
volt
NI-CAD
battery
maintains
a
de-rated
output
voltage
of
approximately
6.5
volts
to
pin
3
of
J12
(TP-12).
It
also
provides
input
to
VR4.
The
output
of
VR4
only
regulates
to
approximately
5.0
volts
as
the
input
voltage
drops
below
7.0
volts.
Observation
Light,
Alarm
Lights,
Motor
Control
The
control
circuits
for
the
observation
light,
and
the
alarm
light
are
similar.
The
alarm
light
is
a
resis-
tive
load
and
does
not
have
a
snubber
circuit
in
parallel
with
the
load.
The
snubber
circuit
on
the
observation
light
supply
protects
the
triac
from
large
voltage
spikes
characteristic
of
inductive
loads.
If
the
control
lines
are
logic
“Low",
less
than
0.45
volts,
this
keeps
the
FET,
triac
driver,
and
triac
switched
“Off”.
The
triac
acts
as
a
switch
to
the
line
voltage
circuit,
removing
voltage
from
the
load.
When
a
device
should
be
switched
“On”,
a
logic
“High”
of
2.4
volts
minimum
is
output
to
the
corres-
ponding
FET.
The
FET
switches
“On”
causing
the
LED
ot
the
isolator-driver
to
switch
on.
The
isolator/
driver
output
drives
sufficient
current
to
the
triac
gate,
switching
the
triac
"On"
allowing
the
selected
device
to
switch
“On”.
Heater
Status
and
Control
The
heater
circuitry
consists
of
a
monitoring
circuit,
a
controller
for
the
heater,
and
à
relay
to
switch
“Off"
the
heater
in
the
event
of
a
triac
or
system
failure.
The
full
wave
bridge
rectifier
CR6
takes
a
sample
(through
R 13)
of
the
ac
signal
supplied
to
the
hea-
ter
and
provides
rectified
dc
to
the
opto-isolator
U3.
If
the
heater
is
“On”
the
de
output
switches
“On”
the
LED
in
the
opto-isolator,
except
at
voltage
levels
below
the
forward
bias
voltage.
When
the
LED
is
“On”
the
transistor
goes
into
saturation
causing
the
output
at
J12
pin
1
to
go
“Low”
(about
0.3
volts).
When
the
heater
is
“Off”
the
de
signal
is
in
the
re-
gion
of
zero
potential
and
there
is
insufficient
for-
ward
bias
voltage
for
the
LED.
This
switches
“Off”
the
transistor
allowing
capacitor
C10
to
charge
and
causes
J12
pin
1
to
go
high
(5
volts),
When
the
hea-
ter
is
switched
“On"
the
LED
switches
the
transis-
tor
“On”
again,
and
the
capacitor
discharges.
The
low
output
shows
small
glitches
caused
by
the
charge/discharge
of
the
capacitor
at
every
half
cycle.
The
glitches
are
acceptable
provided
they
do
not
exceed
the
trigger
voltage
of
1.4
volts
for
the
74LS132
on
the
Control
Board.
1-1

1/Functional
Description
The
heater
control
circuit
uses
a
zero
crossing
opto-
isolator
triac
driver
to
isolate the
line
voltage
from
the
low
voltage
circuits.
Operation
of
the
heater
control
and
other
line
voltage
controls
differ
only
in
the
type
of
isolator
used
and
the
use
of
snubber
cir-
cuits.
When
a
logic
“High”
signal
is
sent
to
the
hea-
ter
control
circuit
from
J12 pin
9
the
output
of
the
isolator
will
not
switch
“On”
until
the
ac
signal
of
the
heater
crosses
the
zero
potential
from
a
nega-
tive
voltage.
After
the
input
line
from
the
microcon-
troller
goes
“Low”,
heat
will
not
switch
“Off”
until
the
first
zero
crossing
preceded
by
the
negative
half
cycle.
This
provides
zero
crossing
control
of
the
hea-
ter
switching.
The
time
that
the
heater
is
“On"
de-
pends
on
the
percent
heat
desired
(controllable
in
5%
increments).
The
microcontroller
also
monitors
the
line
voltage
and
adjusts
the
number
of
ac
cycles
that
the
heater
is
switched
“On”.
This
provides
heater
compensa-
tion.
If
the
line
voltage
is
not
at
the
nominal
value
these
two
functions
result
in
60
levels
of
heat.
Relay
The
relay
circuit
is
used
to
switch
“Off"
the
heater
in
the
event
of
a
triac
or
microcontroller
failure.
Under
normal
conditions
the
input
line
from
J12 pin
10
is
a
logic
“High”,
2.4
volts
minimum.
A
logic
"High"
signal
on
the
input
from
the
control
board
switches
“On”
the
FET
causing
the
relay
coil
to
en-
ergize
and
close the
switch.
If
the
FET
input
is
a
“Low”
from
the
control
board,
(0.5
volts
max.)
the
FET
switches
“Off”
and
the
relay
contacts
open.
The
signal
at
J12
pin
10
comes
from
U1
on
the
con-
trol
board
which
is
a
part
of
a
logic/timing
circuit
in-
dependent
of
the
microprocessor.
A
minimum
voltage
of
7.2
volts
is
reguired
to
ener-
gize
the relay
coil.
Therefore
the
minimum
allowable
voltage
for
the
8
volt
unregulated
supply
is
7.32
volts
since
the
FET
has
an
internal
voltage
drop
of
0.12
volts.
B.
Control
Board
This
is
a
functional
description
for
the
Infant
Warmer
System
Control
Board
Part
No.
0631-5033-
700.
Refer
to
Schematic
No.
0676-0326-000
for
a
de-
tailed
circuit
diagram.
The
control
board
contains
electronic
circuitry
invol-
ved
with
the
measurement,
control,
computation,
memory,
logic,
and
decision
making
functions
of
the
Infant
Warming
System.
The
principle
IC
on
this
board
is
the
8031
single
component,
8-bit
microcon-
troller.
The
8031
has:
an
internal
read/write
memory
(RAM)of
128
bytes,
32
I/O lines
configured
as
four
8-
bit
parallel
ports,
two
16-bit
timers,
a
five
source
two
priority
nested
interrupt,
a
programmable
serial
VO
port,
and
an
on-chip
oscillator
with
clock
cir-
cuitry.
The
program
memory
is
stored
in
a
2764
64k
bit
(8k
x
8)
UV
EPROM.
An
octal
transparent
latch
(74LS373)
is
connected
to
address
inputs
of
the
EPROM
to
permit
the
use
of
the
bi-directional
data
bus
port
of
the
microcontroller
for
addressing
the
EPROM
and receiving
program
instructions.
Four
ICs
with
a
network
of
precision
resistors
are
used
to
interface
the
microcontroller.
The
tempera-
ture
sensor,
calibration
resistors,
or
line
voltage
1-2
scaler
are
selected
by
an
MC14051B
8
Channel
Mul-
tiplexer.
An
LM-10
precision
reference
with
adjusta-
ble
reference
buffer,
and
on-board
operational
amplifier
furnishes
a
stable
reference
supply.
This
is
reguired
by
the
temperature
measurement
circuits
and
the
ADC
3711
Analog
to
Digital
Converter.
An
8243
I/O
expander
is
used
to
interface
the
microcon-
troller
with
the
multiplexer
and
the
A/D
converter.
The
control
board
is
also
equipped
with
several
IC's
that
form
the
triac
watchdog
circuit,
watch-dog
timer,
and
the
audio
alarm
tone
generator.
The
audio
transducer
for
the
alarm
signals
and
its
driver
circuit
are
also
included
on
the
control
board.
The
operation
of
the
circuits
listed
in
the
preceding
paragraphs
will
be
explained
in
detail
below.
Analog
to
Digital
Converter
Temperatures
are
measured
using
a
negative
tem-
perature
coefficient
thermistor
that
is
calibrated
for
specific
resistance
values
and
interchangeability.
Analog
voltage
signals
inversely
proportional
to
temperature
are
derived
from
a
voltage
divider
net-
work
consisting
of
a
5.76k
+
0.1%
resistor
in
series
with
the
temperature
sensor.
The
voltage
source
for
the
measuring
circuit
is
obtained
from
the
LM-10's
internal
precision
reference
source
of
200
mV
amplified
to
a
nominal
1.0
volts
by
the
reference
buffer
of
the
LM-10.
The
Op-amp
portion
of
the
LM-
10
provides
an
adjustable
reference
of
2.0
volts
nominally,
which
is
required
by
the
A/D
converter,
U6.
In
addition
to
the
patient
probe,
there
are
three
other
voltage
divider
networks
on
the
control
board.
Two
have
fixed
output
and
are
used
for
calibration
check
points
of
the
A/D
system
at
25.0
and
37.9
de-
grees
C.
The
third
divider
network
is
unused.
A
separate
input
to
the
control
board
A/D
circuit
comes
from
the
line
voltage
monitor
network
lo-
cated
on
the
power
supply
board.
The
outputs
of
all
the
voltage
dividing
networks
are
connected
to
individual
switch
input
terminals
of
U13,
the
MC14051B
Analog
Multiplexer.
The
MC14051B
contains
eight
normally
open
switches
with
a
common
output
terminal.
The
common
out-
put
of
the
MUX
(pin
3)
is
tied
directly
to
the
analog
input
(pin 9)of
the
A/D
converter.
The
microcontrol-
ler
selects
which
sensor
is
to
be
measured
by
toggl-
ing
the
control
lines,
pin
11A,
pin
10B,
and
pin
9C
of
the
MUX
via
the
8243
#2,
US.
The
following
table
shows
the
digital
codes
used
to
select
the
individual
switches
of
the
MUX:
Control
On
Pin
Inputs
Switches
Number
ABC
000
χο
13
calibration
value
25C
001
xt
14
calibration
value
37.9C
010
X2
15
line
voltage
monitor
011
X3
12
unused
100
X4
01
patient
probe
101
X5
05
unused
110
X6
02
unused
111
X7
04
unused
Note:
Inhibit
terminal
(pin
6)
of
the
MUX
has
no
ef-
fect
on
the
switch
selection
because
it
is
tied
"Low"
through
R19
(200
ohms).

1/Functional
Description
The
ADC
3711,
U6,
uses
a
pulse
modulation
analog
to
digital
conversion
technique.
The
conversion
rate
is
set
by
the
frequency
of
an
internal
oscillation
whose
frequency
is
determined
by
the
external
com-
ponents
R4
and
C14.
The
exact
oscillator
frequency
is
not
critical
and
may
vary
by
+
15%
from
the
nominal
400
kHz.
The
oscillator
frequency
may
be
measured
on
pin
18
of
U6.
With
a
nominal
400
kHz
clock
frequency,
conversions
within
the
ADC
3711
will
take
place
at
an
approximate
rate
of
3
per
sec-
ond.
The
ADC
3711
will
output
BCD
data
on
demand
in
accordance
with
the
coded
digital
signals
applied
to
the
digit
select
inputs
DO
and
D1,
pins
20
and
21
re-
spectively.
The
data
latch
enable
is
tied
“Low”,
therefore,
the
BCD
data
of
the
A/D
converter
will
be
output
to
the
microcontroller
through
8243
#2
in
conformance
to
the
following
codes
that
are
applied
to
the
digit
select
inputs:
Do D1
Selected
Digit
L
L
Digit
0
LSD
L
H
Digit
1
H
L
Digit
2
H
H
Digit
3
MSD
Note:
The
magnitude
of
the
selected
digit
is
present
at
pins
23
and
24,
The
ADC
3711
is
continuously
converting
the
analog
voltage
present
at
its
input
to
a
number
of
counts
between
0
and
3989
(BCD
format).
Therefore,
the
start
conversion,
input
at
pin
7,
and
the
conversion
complete,
output
at
pin
6,
are
misnomered.
The
start
conversion
input
only
controls
the
transfer
of
information
from
the
internal
counter
to
the digital
latches.
The
conversion
complete
output
goes
to
a
logic
“Low”
on
the
rising
edge
of
the
start
conver-
sion
pulse
which
is
issued
by
the
microcontroller.
The
conversion
complete
will
go
to
a
logic
“High”
sometime
later
when
the
new
conversion
informa-
tion
has
been
transferred
to
the
display
latches.
The
start
conversion
pulse
may
occur
at
any
time
in
the
conversion
cycle
because
the
microcontroller
is
run-
ning
asynchronously
to
the
A/D
clock.
Therefore,
the
amount
of
time
from
the
start
to
finish
will
vary.
The
maximum
time
difference
between
the
start
conversion
and
conversion
complete
pulses
in
this
application
is
about
300
msec.
The
operation
of
the
temperature
and
line
voltage
Measurement
circuits
can
be
summarized
as
fol-
lows:
The analog
voltage
signal
derived
from
a
vol-
tage divider
network
and
a
precision
reference
source
is
directed
to
the
input
of
the
A/D
converter
through
an
eight
channel
analog
multiplexer.
For
the
line
voltage
measurement,
the
voltage
source
is
obtained
from
the
rectified,
filtered,
and
unregulated
output
of
the
power
transformer.
Switch
selection
is
software
controlled
by
the
microcontroller
which
toggles
the
A,
B,
and
C
input
lines
of
the
multi-
plexer.
The
analog
voltage
is
converted
in
the
ADC
3711
to
a
digital
signal
in
four
digit
BCD
format
(0
to
3999
counts).
The
microcontroller sends
a
start
conver-
sion
pulse
to
the
ADC
3711
which
then
starts
to
up-
date
the
digital
data
in
the
output
latches.
When
all
of
the
counts
have
been
internally
transferred,
the
A/D
converter
toggles
the
conversion
complete
out-
put
line.
The
microcontroller
then
reads
the
indi-
vidual
BCD
digits
using
coded
signals
to
the
digit
select
lines
of
the
A/D
converter.
ADC
Calibration
The
A/D
converter
is
calibrated
by
connecting
a
5900
+
0.1%
ohm
resistor
to
the
patient
probe
jack
and
placing
the
DIP
switch
on
the
control
board
in
the
following
position:
Swritch 41
Open
(Off)
Switch
42
Open
(Off)
Switch
#3
Open
(Off)
Switch
#4
Closed
(On}
Potentiometer
R44
on
the
control
board
is
then
ad-
justed
until
the
elapsed
time
display
reads
exactly
1122.
With
the
DIP
switches
in
the
given
position,
the
patient
temperature
display
will
read
out the
ac-
tual
patient
temperature,
even
if it
is
outside
of
the
normal
range
and
the
control
temperature
display
will
read
out
the
percent
of
nominal
line
voltage.
During
operation,
the
calibration
of
the
A/D
conver-
sion
system
may
be
checked
by
pressing
and
hold-
ing the
hidden
switch
located
above
the
alarm
si-
lence
switch
on
the
control
panel.
After
2
seconds,
the
patient
temperature
display
should
read
25.0
and
the
control
temperature
display
should
be
37.9.
The
elapsed
timer
should
read
the
nominal
line
vol-
tage
+
2%.
Microcontroller
The
control
system
is
located
in
the
8031
microcon-
troller.
It
operates
at
a
clock
speed
of
6MHz
and
can
he
verified
by
measuring
the
frequency
at
the
Ad-
dress
Latch
Enable
(ALE)
pin
to
be
1
MHz
("On"
=
0.33
usec
and
“Off”
=
0.67
usec).
Grounding
the
EA
pin
enables
the
8031
to
execute
instructions
from
an
external
memory
device.
When
the
microcontroller
performs
a
read
instruc-
tion
from
EPROM,
the
low
order
address
(8
bits)
is
output
from
Port
0
while
the
high
order
address
(6
bits)
outputs
from
Port
2.
(Note:
Bit
6 is
configured
only
to
provide
expansion
compatibility
with
a
27128
EPROM).
The
ALE
pin
goes
“High”
allowing
the
LS373
to
appear
transparent
between
the
EPROM
and
the
microcontroller.
After
the
ALE
out-
put
goes
“Low”,
the
low
order
address
is
latched
to
the
outputs
of
the
D
flip
flops
within
the
LS373.
This
allows
the
2764
to
remain
addressed
by
the
microcontroller,
and
return
8
bits
of
data
while
using
only
two
ports.
Port
1
of
the
8031
is
used
to
communicate
to
the
three
8243
I/O
expanders.
Bits
5-7
are
connected
to
the
Chip
Select
(CS)
line
of
the
first,
second,
and
third
respective
1/O
expanders.
Providing
a
“Low”
signal
on
one
and
only
one
of
the
outputs
activates
the
corresponding
IC.
Bits
0-3
hold
the
instruction
to
be
carried
out
by an
8243
when
the
enable
bit
4
transitions
between
“High”
and
“Low".

1/Functional
Description
Port
3
is
used
to
perform
remaining
tasks
required
by
the
control
system.
Connections
3.0
and
3.1,
(re-
ceive
and
transmit
respectively),
are
used
in
con-
junction
with
the
serial
interface
chips
so
that
com-
munication
to
an
external
microcomputer
is
possi-
ble.
Connection
INTO/P3.2
is
a
line
frequency
inter-
rupt
line
that
is
used
to
aid
in
timing
subroutines
found
within
the
system
software.
Connection
T0/
P3.4
sends
serial
data
to
the
display
driver
while
connection
T1/P3.5
provides
clocking
to
the
driver.
Line
Frequency
The
line
frequency
circuit
converts
a
60
or
50
Hz
sinusoidal
signal
into
a
square
wave
signal.
The
output
of
the
circuit
is
used
to
clock
the
4020B
counter
and
to
provide
a
low
frequency
clock
source
for
the
system
software.
The
1N4001
diode
haif-
wave
rectifies
the
8
Vac
(nominal)
signal
for
use
with
the
Schmitt
trigger
NAND
gate.
With
one
line
tied
“High”,
the
output
of
the
trigger
will
be
inver-
ted.
Since
the
gate
will
not
respond
until
the
input
exceeds
1.9
volts
minimally,
the
duty
cycle
of
the
output
will
be
slightly
more
than 50%.
Heater
Status
The
Heater
Status
function
signals
the
microcontrol-
ler
and
the
safety.
circuitry
as
to
whether
or
not the
heater
is
“On"
or
“Off”.
The
input
to
the
Schmitt
trigger
is
“High”
if
the
heater
is
“Off”
and
“Low”
if
the
heater
is
“On”.
Small
glitches
appear
when
the
heater
is
“On".
Consult
the
Functional
Description
of
the
power
supply
board
for
further
explanation.
The output
of
the
NAND
gate
is
inverted
because
one
input
is
tied
“High”.
Hardware
Triac
Test
The
4020B
14
bit
binary
counter,
U9,
counts
at
a
rate
equal
to
the
line
frequency
and
responds
to
the
negative
edge
of
the
clock
pulse.
The
clock signal
is
received
from
a
Schmitt
trigger
NAND
gate,
pin
6
of
U8.
The counter
resets
when
the
7418123
retrigger-
able
one
shot
flip
flop
outputs
a
“High”
level
pulse
on
the
O
output
Ime.
With
CLR
tied
“High”
and
A
tied
“Low”,
the
counter
will
reset
when
B
of
the
7418123,
US,
is
“High”
at
a
time
equal
to
(04)
+(013)
+(014)
or
after
12296
counts
(04
=
8,
013
=
4096,
014
=
8192).
Approximately
6.19
usec.
later
the
output
of
the
one shot
will
return
to
its in-
itial
“Low”
state.
Q13
and
Q14
of
the
4020B
are
tied
to
a
2
input
“And”
gate
which
will
go
“High”
after
12288
counts.
On odd
numbered
counts
Q1
of
the
counter
goes
“High”.
Q1
is
tied
to
the
CLR
pin
of
D
flip
flop
Ui,
which
when
“High”
allows
the
output
O
to
equal
the
input
D
on
the
next
positive
edge
of
the
clock
pulse.
Therefore
the
output
at
pin
5
will
up-
date
on
even
counts.
After
12288
counts
(3.4133
minutes
on
60Hz
units,
or
4.096
minutes
for
50Hz
units)
the
signal
at
the
D
input
of
the
flip
flop
goes
“High”.
This
signal
is
also
input
to
the
microcontrol-
ler
through
the
1/0
expander
U4.
The
software
will
then
switch
“Off”
the
heat.
Two
counts
later
the
“High"
input
on
D
is
clocked
to
the
output
Q.
The
heater
status
(“Off"-“Low",
“On"-“High")
sent
from
the
Schmitt
trigger
NAND
gate
pin
11
of
U8
is
al-
ways
present
at
the
input
of
U2
pin
5. If
the
heater
is
still
“On”
after
the
2
counts,
the
output
of
the
“And”
gate
pin
6
of
U2
will
clock
the
second
D
flip
flop.
The
outputs
of
the
flip
flops
switch
~O
goes
“High”
and
“Not
Q”
goes
“Low”.
A
“Low"
on
“Not
O"
sets
off
the
audio
alarm
and
drops
out the
non-
1-4
resettable
safety
relay
causing
the
heater
to
switch
“off”.
Heater
Status
LED
A
heater
status
LED
is
located
on
the
control
board
for
troubleshooting.
The
LED
can
be
seen
through
the
rear
of
the
controller
assembly
cover.
When
the
status
line
from
the
Schmitt
trigger
is
“High”,
(hea-
ter
“On”)
the
transistor
O2
switches
“On”
causing
the
LED
to
emit
light.
If
heat
is
“Off”,
the
LED
is
“Off”.
Watchdog
Timer
A
watch
dog
timer
is
used
to
“check”
that
the
microcontroller
is
working
properly.
After
every
cycle
through
the
system
software
the
microcontrol-
ler
sends
a
“Low”
pulse
to
the A
input
of
U3,
a
74L$123.
The
RC
network
connected
to
the
RxCx
and
Cx
pins
create
a
time
constant,
t
=
0.45xRxC
=
0.263
seconds.
If
a
pulse
is
not
received
at
the
input
before
the
time
constant
expires,
the
output
will
go
“Low".
The
high
priority
alarm
will
then
be
activated
due
to
the
microcontroller
failure.
Note
when
the
microcontroller
detects
a
high
priority
alarm
condition,
pulses
to
the
watch
dog
circuit
stop.
Alarm
Tone
Generator
and
Control
Circuits
The
alarm
circuit
consists
of
an
alarm
tone
generator
and
control
circuitry
for
high
or
low
prior-
ity
alarm
conditions.
Under
a
no
alarm
condition
the
7556
timers
are
both
inactive,
(reset
lines
low).
Low
Priority
Alarm
Under
normal
operating
conditions
the
input
to
U8
pin
9 is
“High”.
When
the
microcontroller
detects
a
low
priority
alarm
a
1
Hz
square
wave
is
output
to
U8
pin
9.
The
timer
activates,
causing
a
2
kHz
audio
output.
This
results
in
a
one
second
“On”,
one
sec-
ond
“Off"
audio
alarm.
The
2
kHz
signal
is
adjusted
within
+
100
Hz
by
R38.
The
volume
of
the
audio
alarm
is
adjusted
by
R37. This
should
be
adjusted
fully
CCW
for
maximum
volume.
High
Priority
Alarm
The
high
priority
alarm
is
activated
if
the
microcon-
troller
quits
sending
pulses
to
the
watchdog
timer.
This
occurs
when
a
high
priority
alarm
condition
is
detected
or
if
the
microcontroller
fails.
The
high
priority
alarm
is
also
activated
if
the
hardware
triac
test
circuitry
detects
a
failed
triac.
Both
timers
be-
come
active
with
one
timer
feeding
a
1
Hz
signal
to
the
control
line
of
the
second.
The
1
Meg
resistor
changes
the
output
frequency
of
the
second
timer
to
produce
a
warbling
effect
(two
tone
alternating
alarm).
If
high
and
low
priority
alarms
are
both
“On",
the
output
of
the
“And”
gate
overrides
the
low
priority
signal,
keeping
both
timers
active.

1/Functional
Description
Heat
Control
Routine
Proportional
control
of
the
heater
power
is
obtained
by
varying
the
number
of
full
heat
cycles
of
ac
cur-
rent
delivered
to
the
heater.
To
allow
for
line
vol-
tage
compensation
and
stili
have
at
least
20
dis-
creet
levels
of
heat,
a
proportioning
range
of
0
to
60
full
heat
cycles
was
selected.
In
other
words,
at
very
low
line
voltages,
100%
heat
will
be
output
by
having
the
heat
“On”
for
60
full
cycles
out
of
a
pos-
sible
maximum
of
60.
Similarly,
at
this
low
line
vol-
tage
90%
heat
is
obtained
by
having
the
heat
"On"
for
54
out
of
60
cycles.
In
the
“manual”
mode
of
operation,
the
heat
output
is
determined
by
the
bargraph
setting
selected
by
the
operator.
There
are
20
steps
on
the
bargraph
so
each
step
represents
a
5%
heat
increment.
To
accomplish
the
desired
compensation
for
line
vol-
tage
variations,
the
maximum
number
of
heat
cycles
is
calculated
based
on
the
last
measurement
of
the
power
line
voltage.
For
115v
nominal
units,
at
106
volts
or
less,
the
maximum
number
of
heat
cycles
is
60.
At
greater
than
125
volts
the
maximum
number
of
heat
cycles
is
40
cycles
out
of
a
possible
60.
Therefore,
the
number
of
cycles
of
current
furnished
to
the
heater
in
the
manual
mode
is
determined
by
multiplying
the
maximum
for
the
line
voltage
pre-
sent
by.the
bargraph
setting.
For
example:
assume
the
line
voltage
is
115v
(maximum
number
of
cycles
“On"
=
50)
and
the
bargraph
setting
is
30%;
the
number
of
heat
cycles
to
be
output
will
be
0.3
times
50
=
15
cycles.
Under
these
conditions
the
heat
will
be
“On”
for
15
cycles
and
“Off”
for
45
cycles,
this
sequence
will
continue
until
the
line
voltage
changes
or
the
setting
is
changed
on
the
bargraph.
In
the
“servo”
mode,
the
heater
power
is
controlled
by
comparing
the
patient's
skin
temperature
to
the
selected
value
of
control
temperature.
The
differ-
ence
between
the
control
temperature
and
the pa-
tient
temperature
is
referred
to
as"PTG"
(patient
temperature
gradient).
A
positive
PTG
indicates
a
patient
is
cooler
than
the
control
temperature
and
a
negative
PTG
occurs
when
the
patient
temperature
is
higher
than
the
control
temperature.
Based
on
the
magnitude
and
sign
of
the
PTG,
a
software
look-
up
table
is
used
to
find
the
percent
heat
required.
The percent
heat
is
then
converted
to
the
appropri-
ate
number
of
bargraph
steps
and
then
the
selected
amount
of
heat
is
output
by
the
same
process
used
in
the
manual
mode.
A
hardware
circuit
is
used
to
interrupt
the
micro-
controller
once
every
cycle
of
the
ac
power
line.
During
the
interrupt
routine,
two
registers
are
de-
cremented
to
keep
track
of
the
heater
"On"
and
“Off”
cycles.
One
register
is
used
for
counting
the
total
interval
(60)
and
another
register
is
loaded
on
every
sixtieth
count
with
the
number
of
heat
cycles
to
be
output.
A
flag
is
set
whenever
this
register
is
not
zero,
the
heat
is
“On”
only
when
this
flag
is
set.
The
operation
of
the
heat
control
software
and
the
heat
output
hardware
are
repeatedly
tested
during
operation
of
the
warmer.
An
opto-isolator
connected
with
a
series
resistor
directly
actoss
the
heater
ter-
minals
is
used
to
monitor
heater
power.
The
output
of
the
opto-isolator
is
fed
into
a
Schmitt
trigger,
which
outputs
directly
to
an
input
port
of
the
micro-
controller.
Therefore,
the
microcontroller
can
verify
if
the
heat
is
actually
on
when
it
is
supposed
to
be
on.
If
not,
a
system
fail
alarm
will
be
activated.
Approxi-
mately
every
three
minutes,
an
external
hardware
network
(safety
circuit)
signals
the
microcontroller
to
switch
“Off"
the
heat.
This
hardware
also
moni-
tors
the
output
of
the
Schmitt
trigger
(heater
status
line).
If
the
heater
power
is
not
switched
“Off”
after
a
short
delay,
the
hardware
circuit
will
de-energize
the
“safety”
relay
to
switch
“Off”
heater
power
and
algo
initiate
an
alarm
which
cannot
be
silenced
without
switching
the
power
“Off".
Service
Features
The
electronic
controller
assembly
is
easily
removed
for
servicing
or
calibration.
This
controller
contains
all
the
circuitry
and
components
except
for
the
hea-
ter,
alarm
lamps,
and
observation
lamps.
All
indicators
and
the
audio
alarm
are
activated
on
power-up
for
operator
verification
of
proper
display
operation.
These
can
also
be
activated
by
depres-
sing
the
alarm
silence
switch
for
2
seconds.
In
addi-
tion
the
software
revision
number
and
the
line
fre-
quency
are
displayed.
Test
points
on
the
printed
circuit
boards
are
acces-
sible
for
troubleshooting
and
calibration
without
re-
moval
of
the
boards.
In
addition
integrated
circuits
with
24
pins
or
more
have
sockets
to
aid
in
troub-
leshooting
and
repair.
Software
routines
are
built
into
the
warmer
to
pro-
vide
test
functions,
to
aid
in
troubleshooting,
calib-
ration,
and
operation
verification.
These
test
routines
are
activated
using
à
DIP
switch
located
on
the
control
board.
Some
of
the
test
routines
can
be
activated
using
the
display
panel.
Calibration
may
be
verified
on
the
controller
display
without
disassembly.
A
high
calibration
point
and
a
low
calibration
point
are
displayed
when
the
service
test
switch
is
pressed
for
2
seconds.
Line
voltage
is
monitored
by
the
warmer
and
fluctu-
ations
of
+
10%
from
nominal
voltage
are
compen-
sated
for
so
that
heat
output
is
held
constant.
If
the
voltage
exceeds
+
17.5%
from
nominal
an
alarm
is
activated
and
the
heater
switches
off.
Self
Test
Functions
The
following
text
is
a
description
of
the
self
test
functions
performed
by
the
infant
warmer.
If
an
error
results
on
any
of
the
power-up
or
on-line
tests
then
the
error
number
will
be
displayed
on
the
elapsed
time
display
in
the
format
E
##.
The
high
priority
alarm
(System
Failure
LED)
will
be
“On"
and
cannot
be
silenced.
Power
must
be
switched
“Off”
to
reset
this
alarm.
1-5

1/Functional
Description
Power
Up
Testing
On
power
up
the
following
tests
are
performed.
1.
Instruction
Test
(Error
#01)
Selected
instructions
are
tested
and
verified
opera-
tional.
2.
Checksum
(Error
#04)
The
hex
values
of
Eprom
locations
from
0000
to
1FFD
are
added
together
and
a
2
byte
sum
is
stored.
Eprom
locations
1FFE
and
1FFF
contain
a
2
byte
number
which
when
added
to
the
calculated
checksum
should
total zero.
3.
RAM
Test
(Error
#05)
Rams
10
through
7F are
tested
with
patterns
of
00,
FF,
AA,
and
55.
4.
Test
Port
1
Lines
(Error
#06)
The
port
one
I/O lines are
tested
to
verify
they
can
be
toggled.
On
Line
Testing
The
following
tests
are
run
during
the
normal
oper-
ation
of
the
software.
An
error
on
any
of
these
tests
results
in
a
System
Failure alarm.
1.
ADC
Calibration
Test
(Cal
High
Error
#02,
Cal
Low
Error
#03)
Verifies
that
readings
of
the
precision
calibration
re-
sistors
are
within
0.3
degrees
of
the
nominal
values.
These
readings
can
be
checked
by
depressing
the
hidden
switch
on
the
display
panel
(located
directly
above
the
alarm
silence
switch)
for
2
seconds.
After
2
seconds
the
displays
should
indicate
as
fol-
lows:
‘Patient
Temperature
is
25.0
+
0.3
degrees.
Control
Temperature
is
37.9
+ 0.3
degrees.
2.
Hardware
Triac
Test
A
circuit
independent
of
the
microcontroller
moni-
tors
that
the
micro
can
switch
the
heat
“Off”.
Every
3
minutes
24
seconds
in
60
Hz
operation
(4
minutes
and
5
seconds
for
50
Hz
operation)
a
request
is
made
to
the
micro
to
switch
the
heat
“Off”.
If
the
heat
does
not
go
“Off",
a
hardware
latch
is
latched
and
a
relay
contact
is
opened
so
thera
is
no heat.
This
verifies
that
the
triac
is
not
shorted
and
that
the
micro
is
still
able
to
control
the
heat.
This
failure
does
not
display
an
error
number
because
it
is
not
controlled
by
the
micro
but
will
cause
the
software
triac
test
to
fail
when
heat
is
called
for
by
the
pro-
gram.
3.
ADC
Converter
not
Converting
(Error
#07)
Verifies
that
the
ADC
conversion
complete
occurs
within
1.second.
4.
Hardware
Triac
Timer
Not
Running
(Error
#08)
Verifies
that
the
request
from
the
hardware
triac
test
circuit
occurs
within
256
seconds.
5.
Software
Triac
Test
(Error
#09)
The
heater
status
line
is
checked
to
verify
that
the
heat
is
“On”
when
the
micro
is
switching
it
“On”.
This
verifies
that
the
triac
is
not
failed
open.
1-6
6.
Line
Voltage
Out
of
Range
(Error
#10)
Verifies
that
the
line
voltage
is
within
the
range
of
82.6%
to
117.4%
of
nominal
input
voltage.
(95v
to
135v,
for
115v
units)
Diagnostic
Testing
Diagnostic
testing
can
be
accessed
by
one
of
the
following:
1.
Depressing
and
holding
the
Apgar
Tones
switch
while
powering
up
unit.
This
causes
the
unit
to
cycle
in
the
self
test
loop
until
power
is
removed.
See
Self
Test
Loop
in
step
2e.
2.
Selecting
one
of
the
test
positions
on
the
4
posi-
tion
“Dip”
switch
located
on
the
control
board.
Following
is
a
description
of
the
functions
of
the
“Dip”
positions:
a.
Switches
All
Open
(“Off”)
Normal
Operating
Posi-
tion
(00).
b.
Switches
2,3,4,
Closed
(“On”)
and
Switch
1
Open
(“Off”)
Hardware
Triac
Test
(0E)
.
This
mode
can
be
used
to
test
the
hardware
triac
test
circuit.
The
heat
is
switched
“On”
all
the
time
to
simulate
a
failed
triac.
The
elapsed
time
display
will
start
at
zero
on
power
up
and
display
the
elapsed
time.
At
about
3
minutes
24
seconds
for
60Hz
operation
(4
minutes
and
5
seconds
for
50
Hz
operation)
a
failed
triac
should
be
detected.
The
high
priority
audio
alarm
should
come
“On”
and
the
heat
should
go
“Off".
The
heat
indicator
LED
lo-
cated
on
the
control
board
should
be
checked
to
verify
that.
the
heat
is
“Off”.
c.
Switches
1,2,3,
Open
(“Off”)
and
Switch
4
Closed
(“On")
ADC
Calibration
(08)
The
system
displays
the
actual
ADC
counts
on
the
elapsed
time
display,
the
patient
temperature
on
the
patient display
even
if
outside
of
the
normal
dis-
played
range,
and
the
%
of
nominal
line
voltage
on
the
control
display.
This
position
is
used
for
calibrat-
ing the
analog
to
digital
converter
and
the
line
vol-
tage
compensation
circuit.
d.
Switches
1,2,4
Open
(“Off”)
and
Switch
3
Closed
(“On”)
Alarm
Calibration
(04)
All
segments
of
all
LEDS
are
lit.
The
heater,
over-
head
alarm
lamps,
and
the
observation
lamp
are
on.
The
audio
alarm
emits
a
steady
low
priority
alarm
sound.
The
2
kHz
alarm
frequency
can
be
adjusted
using
this
mode.
e.
Switches
All
Closed
(“On”)
Self
Test
Loop
(0F)
In
this
mode
the
unit
cycles
through
a
display
test,
checks
ADC
calibration,
cycles
the
heater,
alarm
lights,
and
observation
lights,
and
steps
through
the
tests
described
in
power
up
testing.
It
also
monitors
the
touch
switches
and
sounds
the
critical
alarm
while any
switch
is
depressed.
If
any
error
occurs
the
error
number
will
be
displayed
on
the
elapsed
time
display
and
the
critical
alarm
will
sound
for
two
seconds, The
program
will
then
continue
to
loop
through
this
test,
even
if
the
4
DIP
switches
are
retumed
to
“Open”
(“Off”).

1/Functional
Description
If
the
test
loop
is
entered
on
power
up
by
depres-
sing
the
Apgar
Tones
switch
the
program
will
loop
until
an
error
is
detected.
If
an
error
is
detected
the
unit
will
then
stop
the
test
loop,
the error
code
will
be
displayed
in
the
elapsed
time
display,
and
the
critical
alarm
will
sound.
The
power
must
be
switched
“Off"
to
exit this
mode.
Self
Test
Loop
The
unit
cycles
in
the
following
loop
until
the
power
is
removed.
Power
up
tests
performed:
Instruction
test
(“Error"
#01)
Check
calibrate
high
(“Error"
+02)
Check
calibrate
low
(“Error"
#03)
Checksum
(“Error"
#04)
Ram
test
("Error"
#05)
Test
port
1
lines
(“Error”
#06)
Check
if
ADC
is
converting.
(“Error”
+07)
Display
loop
test:
Seven
Seg
Bar
Graph Alarm
LEDs
Display's
Segments
All
1's
1,11
Probe
fail
All
2's
2,12
Pat
temp
All
3's
3,13
Sys
fail
All
4's
4,14
Heater
“Off”
Ail
S's
5,15
Reset
timer
Al
6's
6,16
Spare
LED
All
7's
7,17
All
“Off”
All
8's
8,18
All
“Off”
Al
Ys
9,19
Al
"Off"
All
0's
10,20
All
“Off”
Al
"Off"
All
“Off"
Al
"Off"
The
unit
returns
to
start
of
self
test
loop.
Error
Codes
Error
Description
#01
Instruction.test
fails
#02
Calibrate
high
fails
#03
Calibrate
low
fails
#04
Checksum
fails
#05
Ram
test
fails
#06
Port
1
lines
#07
ADC
not
converting
#08
Hardware
triac
timer
#09
Heat
not
controlled
#10
Line
voltage
out
of
range.
Mode
LEDs
Heater
and
Lights
Servo
“On"
Servo
“Off”
Servo
“On"
Manual
“Off”
Manual
“On"
Manual
“off”
Apgar
“On”
Apgar
“Off”
Apgar
“On”
All
“Off”
"Off"
All
“Off”
“Off”
Possible
Cause
Microprocessor
8031
defective
ADC
calibration
Cal
high
resistor
defective
ADC
calibration
Cal
low
resistor
defective
Eprom
defective
Microprocessor
8031
defective
Microprocessor
8031
defective
VO expander
8243
defective
Microprocessor
8031
defective
A/D
Converter
ADC3711
defective
Voltage
Reference
LM10
defective
I/O
expander
8243
#2
defective
Logic
gate
4020B
defective
IC
in
triac
test
area
defective
Heater
triac
defective
Microprocessor
8031 defective
Heater
opto-isolator
or
driver
defective
Line
voltage
compensation
pot.
on
power
supply
board
not
calibrated.
1-7

1/Functional
Description
C.
Display
Board
This
is
a
functional
description
for
the
Infarit
Warmer
System
Display
Board
Part
No.
0631-5031-
700.
Refer
to
Schematic
No.
0676-0325-000
for
a
de-
tailed
circuit
diagram.
The
display
board
provides
the
interface
between
the
operator
and
the
control
system.
It
displays
the
status
of
the
unit,
the
patient
status,
and
can
also
be
used
as
a
diagnostic
aid.
The operator
controls
the
system
by
depressing
the
various
switches
on
the
front
display.
Operation
of
the
display
board
is
simplified
with
the
use
of
two
ICs:
the
8243
I/O
ex-
pander
which
is
used
in
conjunction
with
the
switches;
and
the
MM5451
(or
MM5450)
driver
used
in
conjunction
with
the
LED
display.
Switch
Decoding
The
I/O
expander,
U1,
is
always
enabled
in
the
read
mode
because
its’
sole
purpose
is
to
detect
switch
depressions.
The
8243
is
activated
by
the
microcon-
troller
sending
a
“Low”
signal
on
the
Chip
Select
(CS)
line.
A control
word
(4
bits)
is
latched
from
the
input
port
2
on
the
“High”
to
“Low”
transition
of
the
PROG
pin.
The
word
is
decoded
as
follows.
Instruction
Address
P23
P22
Code
P21
P20
Code
0 0
Read
0 0
Port
4
0
1
Write
0
1
Port
5
1
0
Or
1
0
Port
6
1
1
and
1
1
Port
7
As
soon
as
the
read
instruction
and
the
port
ad-
dress
are
decoded
the
corresponding
port
lines
are
set
to
a
“High”
impedance
state
and
the
input
buf-
fers
within
the
8243
are
switched
“On”.
When
a
switch
is
depressed
on
the
display,
the
respective
line
switches
“Low”
and
is
loaded
into
the
input
buffer.
The
“Low”
to
“High”
transition
on
the
PROG
line
terminates
the
read
instruction
and
transfers
in-
formation
back
to
port
2.
When
the
microprocessor
sets
the
CS
line
“High”
the
8243
is
disabled.
1-8
LED
Display
Driver
The
LED
display,
driver,
U2,
controls
the
LED
dis-
plays.
The
displays
are
multiplexed
with
a
duty
cycle
of
20%
and
a
refresh
rate
of
60
hertz.
Data
is
input
to
pin
22
synchronously
with
the
clock
(pin
21).
The
first
“1"
bit
activates
the
driver
and
35
data
bits
will
follow.
After
the
35th
bit
is
loaded
the
data
is
latched
to
provide
direct
output.
Note
that
a
logic
“High”
at
the
input
switches
the
output
“Low"
and
switches
“On”
the
LED
connected
to
the
output
(output
is
inverted).
Brightness
Adjust
R9
is
used
to
adjust
the
output
current
from
U2
and
in
tum
change
the
brightness
of
the
LEDs.
R9
is
ad-
justed
to
produce
3.30
+
0.10
volts
across
R10
(3.3V/221
ohms
=
15ma).
C6
is
used
to
prevent
os-
cillations
at
pin
19.
Multiplexing
of
Displays
Since
there
are
not
enough
data
bits
to
drive the
en-
tire
display,
the
displays
are
divided
into
four
sec-
tions.
Bits
1-28
are
used
to
supply
the
necessary
in-
formation
to
each
section.
Bit
29
is
unused.
Bit
30
is
tied
to
a
221
ohm
+
1%
resistor
which
is
used
for
calibration.
Bits
31-34
select
which
channel
of
the
display
is
activated
by
switching
ON
a
Darlington
transistor.
The
Darlington
provides
a
large
gain
so
that
a
small
drive
current
will
sustain
the
large
cur-
rent
draw
from
the
LEDs.
A
string
of
35
zeroes
are
sent
on
the
data
line
every
fifth
update
cycle.
The
driver
has
a
serial
input
and
does
not
have
a
master
reset.
This
string
of
zeroes
resets
the
driver
in
case
an
extra
pulse
was
entered
by
a
noise
spike.
The
basic
circuit
for
one
LED
segment
consists
of
the
5
volt
LED
supply
(reduced
to
4.3
volts
by
a
series
1N4001
diode,)
a
Darlington
switch
to
enable
the
supply
to
the
LED
group,
and
the
MM5451
driver
to
select
a
low
voltage
return
for
the
segment
(if
selected).

2/Specifications
All
specifications
are
subject
to
change
without
notice.
2.1
Electrical
Power
Requirements
Model
3000
0305-0402-910
120
V
50/60
Hz
Models
115V~
+
10%
5.7
Amps
0305-0402-911
220
V
50/60 Hz
Models
220V~
+
10%
3.0
Amps
0305-0402-912
240
V
50/60 Hz
Models
240V~
+
10%
2.7
Amps
0305-0402-913
100
V
50/60
Hz
Models
95V“
+
10%
6.4
Amps
Model
3300
0305-0403-910
120
V
50/60
Hz
Models
115V
~
+
10%
5.7
Amps
0305-0403-911
220
V
50/60 Hz
Models
220V~
+
10%
3.0
Amps
0305-0403-912
240
V
50/60 Hz
Models
240V
”
+
10%
2.7
Amps
0305-0403-913
100
V
50/60
Hz
Models
95V~
+
10%
6.4
Amps
These
models
were
designed
to
conform
to
IEC
601-
1
requirements,
Nominal
Power
Consumption
500
watts
at
maximum
%
power
setting.
Heater
Output
440
watts
+
5%
at
maximum
%power
setting.
Average
Energy
at
Mattress
Level
is
35
mw/cm2
at
Maximum
%
power
setting
Recommended
Bed
Level
(Model
3000)
27
inches
+
2inches
from
the
bottom
of
heater
mod-
ule.
CAUTION:
For
the
Model
3000
Infant
Warmer
Sys-
tem
the
height
of
the
bed
should
be
27
inches
+
2
inches
from
the
bottom
of
the
heater
module
for
proper
heating
and
temperature
control.
Line
Voltage
Compensation
Input
voltage
is
monitored
and
the
heat
output
is
adjusted
to
compensate
for
variations
of
line vol-
tage.
Circuit
Breaker
Rated
Current:
7
Amps
Trip
Point:
9.45
Amps
Minimum
Type:
Manual
Resetting
Model:
Airpax
Snapak
Chassis
Leakage
Current
Less
than
8
microamperes
on
100v
and
120v
units
(18
microamperes
on
220v and
240v
units)
measured
at
the
patient
probe
connection.
Less
than
90
microamperes
on
100v
and
120v
units
{180
microamperes
on
220v
and
240v
units)
mea-
sured
at
an
exposed
metal
surface.
2.2
Controller
Electronics
Microprocessor
based
control
system.
Self
test
func-
tions
are
performed
at
power
up
and
during
normal
operation.
Power
Control
Method
Proportional
heat
control
with
zero
voltage switch-
ing
to
minimize
radiated
and
conducted
EMI.
Examination
Light
Nominai
illuminance
output:
100
foot
candles
at
center
of
mattress.
Estimated
lamp
life:
3,000
hours
Temperature
Sensing
System
Probe
Model
Number:
LA003
Range:
22-42
degrees
C
Accuracy:
+
0.3
degrees
C
Resolution:
+ 0.1
degrees
C
Probe
interchangeability:
+
0.1
degrees
C
Elapsed
Timer
60
minute
elapsed
timer
with
hold
mode
and
Apgar
tones.
Manual
Mode
Manual
mode
heat
selector
range:
0
to
440
watts
in
20
increments
(5%
per
step).
Servo
Mode
Servo
control
range
36.0
to
37.5
degrees
C
in
incre-
ments
of
0.1
degrees
C.
2.3
Alarms
Multiple
audio
tones
1.
Operator
prompt
tone
2.
Alternating
single
tone
3.
Alternating
two
tone
Overhead
Alarm
Light
一
一
Large
alarm
light
located
on
the
front
of
the
heater
assembly
for
easy
visual
identification.
Probe
Failure
Alarm
Activates
when
the
skin
temperature
probe
fails
electrically
open
or
short,
or
is
disconnected
from
the
warmer.
The
alarm
is
only
active
in
the
servo
mode.
The
heater
is
tumed
off
and
the
patient
tem-
perature
display
flashes
HH.H
when
this
alarm
con-
dition
exists.

2/Specifications
Patient
Temp.
Alarm
The
Patient
Temp.
alarm
activates
in
the
servo
mode
when
the
difference
between
the
patient
tem-
perature
and
the
control
temperature
is
greater
than
one
degree
C.
The
alarm
cancels
when
the
patient
temperature
returns
to
within
0.8
degrees
C
of
the
control
temperature.
System
Failure
Alarm
The
system
failure
alarm
activates
and
turns
the
heater
off
if
the
analog
to
digital
converter
calibra-
tion
drifts
by
more
than
0.3
degrees
C.,
the
heater
triac
fails,
the
microprocessor
fails,
or
the.self
check
functions
fail
on
power-up.
This
alternating
two
tone
alarm
cannot
be
silenced.
Heat
Off
Alarm
The
LED
activates
whenever
the
heater
is
in
the
X-
ray
position.
The
audio
alarm
activates
after
5
min-
utes
in
the
X-ray
position.
Check
Patient
Alarm
Activates
in
the
manual
mode
if
the
heater
has
been
energized
at
greater
than
25%
heat
for
12
con-
tinuous
minutes.
In
the
servo
mode
the
alarm
acti-
vates
when
the
heater
has
been
at
full
power
for 12
continuous
minutes.
Power
Failure
Alarm
The
power
failure
alarm
activates
if
line
power
is
in-
terrupted.
A
rechargeable
maintenance
free
ni-cad
battery
powers
the
audio
alarm
and
the
micropro-
cessor.
If
power
is
restored
within
10
minutes
the
mode
of
operation
and
the
set
point
are
recalled.
2.4
Environmental
Operating
temperature
range:
10
to
40
degrees
C
Storage
temperature
range:
-25
to
60
degrees
C
Humidity:
0
to
95%
2.5
Mechanical
(without
accessories)
Dimensions:
Height:
73in.
(185cm.)
Depth:
39.5in.
(100cm.)
Width:
24.5in.
(62cm.)
Mattress:
18.2
x
25.2
inches
(46.2
x
64.0
cm)
2-2
Tilting
Positions
+
10
degrees
Weight:
Model
3300
—
approx.
210
pounds
(95
kg.)
Model
3000
—
approx.
165
pounds
(75
kg.)
Casters:
5
inch
diameter,
2
locking,
2
non-locking
2.6
Accessories
Oxygen
yoke
and
regulator:
Pin
indexed
yokes
accommodate
two
E
size
oxygen
cylinders
Diss
oxygen
fittings
52 +
2
psig
regulator
Cylinder
pressure
gauge,
0
to
3000
lbs.
(0
to
210
kg/
cm2)
Three
drawer
storage
accessory:
Drawers
15
X
15.5
X4
inches
(930
cubic
inches
per
drawer)
38
x
39
x
10
cm.
(14,820
cubic
centimeters
per
drawer)
Rail
mounted
accessories:
Monitor
shelf:
Dimensions
12
x
26
inches
(30
x
66
centimeters)
Load
limit:
50
pounds
(22
kg)
Instrument
shelf:
Dimensions
12
x
12
inches
(30
x
30
centimeters)
Load
limit:
20
pounds
(9
kg}
WARNING:
Overloading
the
shelves
can
affect
the
stability
of
the
unit.
Oxygen
flowmeter
w/DISS
fittings
(0
to
15
LPM)
Air
flowmeter
w/DISS
fittings
(0
to
15
LPM)
Manometer
(-20
to
+
100
centimeters
of
water)
TV
pole
Vacuum
manifold
w/DISS
fittings
Gas
manifold
w/
1/8”
npt
fitting
3.5
inch
utility
post
22
inch
utility
post
for
mounting
infusion
pumps,
humidifiers,
proportioners,
ventilators
etc.
Ventilator
mounting
accessory

3/Setup
and
Checkout
Procedure
3.1
Setup
Refer
to
the
setup
instructions
shipped
with
the
In-
fant
Warmer
System
for
initial
unpacking
and
setup
of
the
unit
after
shipment.
Inspect
the
Model
3000
and
3300
Warmer
Systems
and
all
accessory
items
after
removal
from
the ship-
ping
containers
for
any
signs
of
damage
that
may
have
occurred
during
shipment.
File
a
damage
claim
with
the
shipping
carrier
if
damage
has
occurred.
Also
confirm
the
presence
of
all
accessory
items
as
listed
on
the
packing
slip.
3.2
Checkout
Procedure
WARNING:
Do
not
perform
the
Check-Out
Proce-
dure
while
a
patient
occupies
the
Infant
Warmer
System.
Perform
the
Checkout
Procedure
before
each
use
on
a
patient.
Refer
servicing
to
qualified
service
person-
nel
if
the
unit
does
not
perform
as
specified.
Refer
to
the
Troubleshooting
Guide
and
the
Disassembly
and
Repair
Sections
if
the
unit
fails
any
steps
of
the
Checkout
Procedure.
A.
Mechanical
Checks
(Model
3000
and
3300)
Overall
Appearance
1.
Disconnect
the
power
cord
for
the
Infant
Warmer
System
for
the
mechanical
checks
portion
of
this
procedure.
2.
Check
the
overall
appearance
of
the
Infant
Warmer
System.
There
should
be
no
obvious
damage.
3.
Place
the
Infant
Warmer
System
on
a
level
sur-
face.
Check
that
all
four
casters
are
in
firm
con-
tact
with
the
floor
and
that
the
warmer
moves
freely.
4.
Lock
the
two
front
casters
and
check
that
the
warmer
is
held
in
place.
5.
Examine
the
power
cord
for
damage.
Replace
the
power
cord
if
damage
is
evident.
CAUTION:
Insulation
on
electrical
wiring
can
de-
teriorate
with
age.
Check
for
brittle
or
deteriorated
insulation
on
power
cord
and
all
other
electrical
wiring.
Heater
Rotation
1.
Rotate
the
heater
to
the
X-ray
position
and
back
to
the
normal
position,
Check
for
smooth
rotation.
Mechanical
Checks
(Model
3300)
1.
Check
the
operation
of
the
bed
sides.
The
bed
sides
should operate
smoothly.
2.
Check
the
operation
of
the
tilt
mechanism.
Verify
that
the
bed
platform
operates
smoothly
and
jocks
in
any
position.
Optional
Accessory
Checks
Check
that
all
accessories
are
mounted
securely
to
the
uprights.
Oxygen
Yoke
and
Regulator
Checks
.
Check
that
all
oxygen
cylinders
are
mounted
sec-
urely.
2.
Check
that
the
output
from
the
regulator
is
52
+
2
psig
with
a
500
cc
flow.
If
adjustment
is
re-
quired
refer
to
section
4H.
B.
Control
Unit
Checks
1.
Connect
the
Infant
Warmer
power
cord
to
an
ap-
propriate
power
source
(see
rating
plate
for
prop-
er
voltage
etc.).
Switch
the
power
“On”
and
ver-
ify
the
following:
a.
The
alternating
two
tone
audible
alarm
sounds
and
all
displays
and
indicators
are
lit
for
approxi-
mately
two
seconds.
Note:
During
this
time
the
controller
also
per-
forms
self
check
functions.
If
the
controller
de-
tects
a
failure
the
alarm
stays
on
and
service
is
required.
b.
The
manual
mode
indicator
is
lit.
с.
An
operator
prompt
tone
sounds
and
the
%
power
display
flashes.
2.
Adjust
the
heat
output
with
the
increase
and
de-
crease
touch
switches
to
the
high
and
low
limits
as
indicated
by
the
%
power
display.
3.
Connect
the
skin
temperature
probe
to
the
Infant
Warmer
System.
4.
Press
the
mode
touch
switch
to
place
the
warmer
in
the
servo
mode
and
verify
the
follow-
ing:
Note:
An
alternating
two
tone
alarm
and
a
flash-
ing
overhead
alarm
light
may
occur
here
if
the
skin
temperature
probe
is
below
30
degrees
C.
Warm
the
probe
with
your
fingers
or
silence
the
alarm.
.
The
servo
mode
indicator
is
lit.
.
An
operator
prompt
tone
sounds
and
the
control
temperature
display
flashes
36.5
degrees
C.
vp
5.
Press
the
increase
touch
switch
and
verify
that
the
maximum
servo
control
temperature
attaina-
ble
is
37.5
degrees
C.
Note:
A
patient
temperature
alarm
occurs
if
the
difference
between
the
patient
temperature
and
the
control
temperature
is
greater
than
one
de-
gree
C.
6.
Press
the
decrease
touch switch
and
verify
that
the
minimum
servo
control
temperature
attaina-
ble
is
35.0
degrees
C.
M
.
Disconnect
the
skin
temperature
probe.
Verify
the
following:
.
The
probe
failure
indicator
light
is
lit.
.
There
is
an
alternating
two
tone
alarm.
The
overhead
alarm
light
is
flashing.
.
The
patient
temperature
display
flashes“HH.H”.
m
ASFP
.
Press
the
alarm
silence
touch
switch
and
verify
the
following:
.
The
probe
failure
indicator
light
is
lit.
.
The
alternating
two
tone
alarm
is
silenced.
.
The
overhead
alarm
light
is
lit.
.
The
patient
temperature
display
indicates
"HH.H".
e.
After
one
minute
the
alternating
two
tone
alarm
sounds,
the
overhead
alarm
flashes
and
the
pa-
tient
temperature
display
flashes
HH.H.
ao
op
9.
Switch
to
the
manual
mode
and
set
the
heat
at
25%
power.

3/Setup
and
Checkout
Procedure
Elapsed
Timer
Check
1.
Press
the
start/hold
switch
to
activate
the
elapsed
timer.
Verify
that
the
timer
starts
opera-
tion.
2.
Press
the
on/off
switch
for
the
Apgar
tones.
Ver-
ify
that
the
indicator
light
for
the
Apgar
tones
is
not
lit.
3.
Press
the
on/off
switch
for
the
Apgar
tones
again.
Verify
that
the
indicator
light
for
the
Apgar
tones
is
lit.
4.
Press
the
start/hold
touch
switch.
Verify
that
the
present
elapsed
time
is
held.
5.
Press
the
start/hold
touch
switch
and
verify
that
the
timer
updates
to
the
current
elapsed
time
and
the
Apgar
tones
continue
to
sound
at
the
specified
times
(at
1
minute
and
at
every
5
min-
ute
interval
after
the
elapsed
timer
is
started).
6.
Press
the
reset
touch
switch
and
verify
that
the
timer
indicates
00:00.
If
the
elapsed
timer
is
not
used
for
one
minute
the
display
is
switched
off.
Examination
Light
Check
1.
Press
the
Light
“On"/“Off"
touch
switch.
Verify
that
the
examination
light
functions.
Interlock
Switch
Check
1.
Place
the
warmer
in
the
manual
mode
at
25%
power
output.
2.
Rotate
the
heater
assembly
to
the
X-ray
position.
Verify
that
the
heater
off
indicator
light
is
“On”
and
the
%
power
display
indicates
0%
heat.
3.
Rotate
the
heater
assembly
to
the
normal
operat-
ing
position.
Verify
that
the
heater
off
indicator
light
is
“Off”
and
the
%
power
display
indicates
25%.
Display
and
Alarm
Check
1.
Press
and
hold
the
alarm
silence
switch
for
more
than
2
seconds,
then
check
for
the
following:
a.
Every
segment
of
each
digital
display
should
be
lit.
All
segments
should
be
of
uniform
brightness
and
visible
under
ordinary
room
lighting
condi-
tions.
b.
All
LED
indicators
should
be
lit.
c.
The
warbling
two-tone
audio
alarm
should
be
on.
Battery
Test
and
Memory
Test
The
battery
is
charged
in
normal
operation
by
a
trickle
charge
current
from
the
regulated
9
volt
sup-
ply.
If
the
battery
is
discharged
it
must
be
re-
charged
before
allowing
a
patient
to
occupy
the
In-
fant
Warmer.
The
battery
may
be
recharged
by
placing
the
unit
in
the
manual
mode
at
a
0%
heat
setting.
If
the
battery
is
defective,
replace
it.
Re-
placement
of
the
battery
is
recommended
every
two
years.
There
is
no
maintenance
required
for
the
bat-
:
tery.
3-2
Note:
The
battery
must
be
fully
charged
to
pass
the
10
minute
test
or
partially
charged
to
pass
the
two
minute
test.
1.
Disconnect
the
patient
temperature
probe.
2.
Place
the
Infant
Warmer
in
“servo”
mode.
3.
Silence
the
probe
failure
alarm.
4.
Set
the
control
temperature
at
37.0
degrees
C.
5
.
Remove
the
Infant
Warmer
power
plug
from
the
power
source
for
two
minutes.
Do
not
switch
the
power
"Off".
The
power
failure
alarm
should
sound
for
two
minutes.
Note:
If
the
power
failure
alarm
is
tested
for
10
minutes,
the
Infant
Warmer
must
be
connected
to
the
correct
power
source
and
operated
for
24
hours
to
recharge
the
battery
before
allowing
a
patient
to
occupy
the
Infant
Warmer.
6.
Reconnect
the
Infant
Warmer
to
the
power
source.
Verify
the
following:
a.
The
Infant
Warmer
is
operating
in
the
servo
mode.
b.
The
control
temperature
is
37.0
degrees
C.
c.
The
audio
power
failure
alarm
is
off.
Calibration
Check
1.
Press
and
hold
the
hidden
control
panel
switch
(directly
above
the
alarm
silence
switch).
2.
After
2
seconds
the
displays
should
indicate
as
follows:
a.
Patient
Temperature
displays
25.0
+
0.1
degree.
b.
Control
Temperature
displays
37.9
+ 0.1
degree.
c.
Elapsed
Time
displays
%
nominal
line
voltage
+
2%.
Note:
Line
voltage
may
be
measured
on
terminals
marked
with
the
phase
symbol
and
“n”
on
the
power
supply
board.
Measure
the
voltage
with
the
heat
off.
The
%
of
line
voltage
can
be
calculated
by:
(Measured
Voltage
/
Nominal
Voltage)
x
100
=
%
of
Nominal
Line
Voltage
Nameplate
Voltage
-
Nominal
Voltage
100
95
120
115
220
220
240
240

4/Calibration
and
Adjustments
CAUTION:
Use
the
Static
Control
Work
Station
(Part
No.
0175-2311-000)
to
help
ensure
that
static
charges
are
safely
conducted
to
ground.
The
Velos-
tat
material
is
conductive.
Do
not
place
electrically
powered
circuit
boards
on
it.
Note:
The
audio
alann
will
sound
for
about
2
sec-
onds
whenever
powering
up
the
unit.
Note:
Warm
up
the
unit
for
5
minutes
before
mak-
ing
these
adjustments.
A.
Control
Unit
Access
1..
Disconnect
the
power
cord
for
the
Infant
Warmer
System
from
the
wall
outlet.
2.
Remove
the
mounting
screws
for
the
back
panel.
B.
Power
Supply
Board Voltage
Checks
1.
Set
the
test
switch
(S1)
located
on
the
control
board
to
the
following
test
positions:
Switch
#1,#2,#4
Open
(“Off”)
Switch
#3
Closed
(“On")
2.
Connect
the
controller
assembly
to
the
power
source
(listed
voltage
+
10%),
and
switch
the
In-
fant
Warmer
“On”.
3.
Check
that
all
display
segments
are
“On”,
the
observation
lamp
is
“On”,
the
alarm
light
is
“On”,
the
heater
radiates
heat,
and-a
continuous
alarm
tone
sounds.
4.
Check
that
the
following
D.C.
voltages
are
pre-
sent
at
the
test
connector
(T1)
located
on
the
control
board.
Voltages
should
be
within
the
to-
lerances
specified:
TP7
Ground
(common)
TP3
(9.0
Vdc
ST.)
+9V
+
0.2V
(adjust
R4 on
P.S.
Bd.)
TP4
(5.0
Vde
LEDS)
+5V
+
0.2V
(replace
P.S.
Bd.)
TP5
(5.0
Vdc
ST.)
+5V
+
0.2V
(replace
P.S.
Bd.)
C.
Display
Brightness
Check
Note:
The
display
brightness
is
precalibrated
at
the
factory
and
should
only
require
adjustment
if
replac-
ing
a
component
on
the
display
board.
With
the
test
switches
set
as
follows:
Switch
#1,#2,#4
Open
(“Off”)
Switch
#3
Closed
(“On”)
Check
that
all
the
displays
are
lit
and
are
of
uniform
brightness.
If
the
displays
are
acceptable
proceed
to
Section
D.
If
the
displays
are
not
illuminated
adequately
proceed
with
the
adjustment
procedure.
Adjustment
Procedure
CAUTION:
The
back
panel
and
display
panel
may
drop
down
when
the
bottom
cover
mounting
screws
are
removed,
Be
sure
to
secure
the
panels
with
tape
before
disassembly.
1.
Set
the
test
switch
($1)
located
on
the
control
board
to
the
following
test
positions:
Switch
#1,#2,#4
Ореп
("ОН")
Switch
#3
Closed
(“On")
2.
Tape
the
display
panel and back
panel
to
the
top
cover.
3.
Remove
the
four
outside
corner
mounting
screws
from
the
bottom
of
the
display
panel.
Do
not
re-
move
the
two
inside
mounting
screws.
4.
Connect
a
digital
voltmeter
across
R10 located
on
the
bottom
edge
of
the
display
board.
5.
Adjust
R9 on
the
display
board
until
the
voltage
across
R10
is
3.30V
+
0.2V.
6.
Verify
that
all
segments
of
all
displays
are
lit
and
are
of
uniform
brightness.
D.
Alarm
Volume
Adjustment
Note:
The
alarm
volume
and
frequency
are
precalib-
rated
at
the
factory
and
should
only
require
adjust-
ment
if
replacing
a
component
on
the
control
board.
1.
With
the
test
switches
set
as
follows:
Switch
#1,#2,#4
Open
(“Off”)
Switch
#3
Closed
(“On”)
2.
Ensure
the
alarm
tone
and
volume
are
adequate.
Verify
that
the
audio
alarm
level
is
adequate
ina
lecation
with
a
background
noise
level
of
55
dBA
max.
If
the
audio
alarm
level
is
acceptable
go
to
step
E.
If
the
audio
alarm
level
is
unacceptable
proceed
with
the
adjustment
procedure.
Adjustment
Procedure
3.
Verify
the
frequency
output
at
U7
pin
9
is
2
kHz
+
0.1
kHz.
Adjust
R38
on
the
control
board
as
re-
quired.
Note:
If
test
equipment
is
not
available
for
check-
ing
the
2
kHz
frequency,
adjust
R38
for
maximum
sound
level.
4.
Verify
that
R37
on
the
control
board
is
set
fully
CCW
(maximum
volume).
E.
Analog
to
Digital
Converter
(ADC)
Note:
The
following
resistance
values
are
available
on
the
Temperature
Simulator
and
the
switch
posi-
tions
are
listed
in
parentheses.
1.
Switch
the
power
switch
“Off”.
2.
Set
the
control
board
test
DIP
switch
(located
on
the
upper
edge)
to
the
calibration
positions
as
follows:
Switch.#1,
#2,
#3
Open
(“Off”)
Switch
#4
Closed
(“On”)
3.
Connect
a
resistance
of
5900
ohms
+
0.1%
(17)
to
the
patient
jack
connector.
4-1

4/Calibration
and
Adjustments
4.
Switch
the
power
switch
“On”
and
allow
the
3.
The
front
panel should
display
the
following:
unit
to
stabilize
for
5
minutes.
-
ーーーー
..
~
-
.
.
Pat.
Temp...
EEE
5.
Verify
that
the
ADC
counts
displayed
on
the
Cont.
Temp.
EEE
elapsed
time
display
is
1122
+
2
counts.
Slowly
Elapsed
Time
Running
in
Stop
Watch
Mode
adjust
R44
on
control
board
as
required.
6.
Input
resistance
values
into
patient
probe
con-
4.
Confirm
that
the
heat indicator
LED
on
the
con-
nector
and
verify
patient
temperature
readings
trol
board
is
lit.
are
within
tolerances
specified.
、 :
P
5.
Use
a
stop
watch
and
verify
that
elapsed
time
Resistance
Input
Patient
Temperature
display
is
accurate
within
+
1
second
per
min-
5900
ohms
+
0.1%
(17)
37.3
+
0.1
degrees
ute.
7060
ohms
+
0.1%
13,
33.0
=
0.1
4.
6190
ohms
+
0
1%
fay)
36.2
+
0.1
degrees
6.
After
approximately
3
minutes
and
20
seconds
4
h +
0.19
12
39.0
+
0.
(60
Hz
models)
or
approximately
4
minutes
(50
5496
ohms
»
62)
9.0
0.1
degrees
Hz
models)
a
warbling
alarm
which
cannot
be
si-
F.
Line
Voltage
Sensing
lenced
occurs.
1.
Use
a
DVM
and
measure
the
line
voltage
at
the
7.
The
heat
indicator
LED
on
the
control
circuit
wall
outlet.
board
(viewed
from
the
rear
of
unit)
should
be
2.
Calculate
the
displayed
%
variance
of
the
supply
off.
voltage
from
the
rated
nominal
voltage
using
the
8.
Switch
the
power
“Off”
for
the
Infant
Warmer
following
formula:
and
restore
the
test
switch
to
the
original
config-
(Actual
line
voltage
/
Nominal
Voltage)
x
100
=
uration
(all
switches
"Open").
Displayed
%
H.
Test
Loop
3.
Slowly
adjust
R3
on
power
supply
board
as
re-
Complete
Unit
Testing
quired
until
the
control
temperature
display
1.
Switch
the
power
“Off".
als
the
calculated
value
for
th
i:
It:
red,
catonlated
value
lor
ine
supply
voltage
2.
Place
the
test
switches
on
the
control
board
in
.
the
test
loop
position.
All
4
switches
Closed
Note:
For
domestic
115v
units
and
an
input
vol-
(“On”).
tage
of
115
volts
the
reading
on
the
control
tem-
3.
Switch
th
itch
“On"
perature
display
should
be
100
%
+
2%.
-
Switch
the
power
switc!
п.
4.
Switch
the
power
switch
“Off”.
4.
Verify
that
the
following
sequence
occurs:
G.
Triac
Safety
Circuit
Test
a.
For
the
first
second:
y
All
segments,
LEDS,
and
high-low
alternating
1.
Place
the
individual
test
switches
in
the
follow-
tone
audible
alarm
are
“On”.
ing
positions
b.
For the
next
second:
、
High-low
alternating
tone
audible
alarm
“On”.
Switch
#1
Open
(“Off”)
000
Patient
display
—
60H
for
60
Hz
(50H
for
50
Hz)
Switch
#2,
#3,
#4
Closed
(“On”)
Elapsed
time
display
-
software
revision
number.
2.
Switch
the
Infant
Warmer
power
“On".
c.
The
unit
should
then
loop
in
the
following
order
until
the
power
is
removed.
Seven
Seg
Bar
Graph
Alarm
LEDs
Mode
LEDs
Heater
and
Display's
Segments
Lights
AN
Us
-
- 1.11
Probe
fail
Servo
“On"
AN
2's
2,12
Pat.
temp.
Servo
"off"
All
3’s
3,13
Sys.
fail
Servo
“On”
All
4’s
4,14
Heater
OFF
Manual
“Off”
All
5’s
5,15
Reset
timer
Manual
“On”
All
6's
6,16
Spare
LED
Manual
“Off”
All
7's
7
7,17
Al
"Off"
Apgar
“On”
All
3's
8,18
All
"Off"
Apgar
“Off”
All
9's
9,19
AN
“Off”
Apgar
"On"
All
0’s
10,20
Al
"Off"
All
“Off" "Off"
All
“Off"
АП
“Ой”
Al
"Off"
All
“Off”
“Off”
4-2
This manual suits for next models
1
Table of contents
Other Ohmeda Food Warmer manuals