8CELON ELITE QUICK REFERENCE GUIDE
CELON PROCEDURE
MODE OVERVIEW
Technique Effect Mode Max
Power
VPShort Explanation
Bipolar
Coag
Fine RFITT 40 W 250 Automatic end-of-procedure detection.
Audible feedback.
Pure RFITT 40 W 150 Automatic end-of-procedure detection.
Audible feedback.
Strong RFITT
(w. and w/o
RCAP)
40 W 260
Automatic deactivation when the tissue
resistance exceeds a limit value.
(Resistance-Controlled Automatic Power (RCAP)
to avoid premature tissue desiccation.)
Soft Coag 120 W 260 Little carbonization and adhesion.
Autostart selectable.
Cut
Fine Cut 120 W 470 Low spark intensity for small thermal effect.
Pure Cut 120 W 530 Medium spark intensity
for medium thermal effect.
Strong Cut 120 W 560 High spark intensity for large thermal effect.
Monopolar
Coag
Soft Coag 120 W 250 Little carbonization and adhesion.
Forced Coag 120 W
2000
Spark allows coagulation
also with relatively small electrodes.
Cut
Fine Cut 120 W
580 Low spark intensity for small thermal effect.
Pure Cut 120 W
630 Medium spark intensity
for medium thermal effect.
Strong Cut 120 W
680 High spark intensity for large thermal effect.
Pulse Cut 120 W
690 Intermittent cutting,
e.g. for endoscopic operations.