Olympus USG-410 User manual

Other Olympus Portable Generator manuals

Olympus OME-L200 User manual

Olympus

Olympus OME-L200 User manual

Olympus USG-400 User manual

Olympus

Olympus USG-400 User manual

Olympus ESG-410 User manual

Olympus

Olympus ESG-410 User manual

Olympus ESG-150 User manual

Olympus

Olympus ESG-150 User manual

Olympus ESG-400 User manual

Olympus

Olympus ESG-400 User manual

Olympus CLH-SC User manual

Olympus

Olympus CLH-SC User manual

Olympus USG-400 User manual

Olympus

Olympus USG-400 User manual

Olympus CLV-160 User manual

Olympus

Olympus CLV-160 User manual

Popular Portable Generator manuals by other brands

Comdel CLF5000/PLL manual

Comdel

Comdel CLF5000/PLL manual

Outsunny 84F-006V80 Using instruction

Outsunny

Outsunny 84F-006V80 Using instruction

Icel Manaus GV-2015 User's operation manual

Icel Manaus

Icel Manaus GV-2015 User's operation manual

Stat-X 30E owner's manual

Stat-X

Stat-X 30E owner's manual

Metallwarenfabrik Gemmingen Eisemann T 18000 E operating instructions

Metallwarenfabrik Gemmingen

Metallwarenfabrik Gemmingen Eisemann T 18000 E operating instructions

Alpha ACX3100i quick start guide

Alpha

Alpha ACX3100i quick start guide

HERON 8896140 Translation of the original user manual

HERON

HERON 8896140 Translation of the original user manual

SEERS STF-V11 Operator's manual

SEERS

SEERS STF-V11 Operator's manual

JOYZIS BS100 Faq

JOYZIS

JOYZIS BS100 Faq

LIPOWER G1600 user manual

LIPOWER

LIPOWER G1600 user manual

Panasonic YD-350GL4 operating instructions

Panasonic

Panasonic YD-350GL4 operating instructions

RUNHOOD HE1200-UK user manual

RUNHOOD

RUNHOOD HE1200-UK user manual

Kohler 20EORZ Operation

Kohler

Kohler 20EORZ Operation

Powerhorse 83171 owner's manual

Powerhorse

Powerhorse 83171 owner's manual

ETQ TG4000 owner's manual

ETQ

ETQ TG4000 owner's manual

Briggs & Stratton 40265 Illustrated parts list

Briggs & Stratton

Briggs & Stratton 40265 Illustrated parts list

evertz 7700 Series manual

evertz

evertz 7700 Series manual

Firefly Cygnus CYG2400/460/24 user manual

Firefly

Firefly Cygnus CYG2400/460/24 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

QUICK REFERENCE GUIDE
USG-410
Ultrasonic Bipolar Generator
USG-410
Get started
1. Connect THUNDERBEAT or SONICBEAT via a transducer to USG-410.
- Align the symbols on plug and socket and fully insert the transducer
plug in the socket of USG-410.
- The USG-410 recognizes the connected instrument and displays the
corresponding set screen (see next page).
2. Verify the ultrasonic output.
3. Verify the RF bipolar output (THUNDERBEAT only).
Power switch
Transducer socket
Touch screen
Pre-Procedure Device Checks
The activation test for ultrasonic output needs to be done every time
a THUNDERBEAT or SONICBEAT instrument is connected.
Activation test for ultrasonic output
1. Connect THUNDERBEAT or SONICBEAT and tap the “Test” button.
Caution: Do not apply the instrument on the patient during activation test.
2.
Activate instrument with the jaws opened following the instructions on the screen.
3. When the activation test result is displayed, release the hand switch and confirm
with OK.
4. After a successful test, the instrument can be used safely.
Verification of the RF bipolar output (THUNDERBEAT only)
1. Fill a sterile container with saline.
2. Open the THUNDERBEAT instrument jaws and submerge only the distal half
of both jaws in saline.
3. Press and hold the SEAL button (blue).
4. Confirm the following during the SEAL
mode output:
- The output screen is displayed.
- The output tone sounds.
5. After the confirmation, release the SEAL
button.
6. Wipe any remaining saline off the jaws
with a sterile, lint-free cloth.
Touch Screen: Status and Settings
Settings
Procedures: Save, select, overwrite
or delete a procedure. You can save up
to 40 procedures.
Setup: Change device settings such
as language, hand switch activation or
automatic mist and smoke evacuation.
Maintenance: Get access to your error
log, transfer data, set the maintenance
reminder and more.
Procedure name is displayed
here, if a saved procedure is
selected.
Foot switch indicator
Instrument name
Activation test for
ultrasonic output
Mode name
Intelligent Tissue Monitoring (ITM)
detects when the tissue is cut and
stops the ultrasonic and RF bipolar
output automatically.
If ITM is OFF, the activation needs to
be stopped manually by releasing the
hand switch.
ITM can be used with THUNDERBEAT
SEAL & CUT and SONICBEAT MAX.
Adjust the output level by tapping plus/minus.
When reconnecting an instrument, the last-used
output level and ITM are recalled automatically.
Instrument Mode Energy Level Standard
Settings Characteristic Vessel sealing ITM
THUNDERBEAT
SEAL & CUT Ultrasonic
RF bipolar 1–3 1Fastest cutting speed at level 1
Constant sealing performance over all levels up to 7 mm available
SEAL RF bipolar 1 1 Shorter sealing time by improved SEAL algorithm compared
to ESG-400 & USG-400 up to 7 mm –
SONICBEAT
VAR Ultrasonic 1–3 1Fastest cutting speed at level 3 up to 5 mm (level 1 only) –
MAX Ultrasonic 3 3 Fastest cutting speed –available
This sheet is for quick reference only. Please refer to the relevant User’s Manual(s) for
instructions, warnings and cautions. Nothing in this presentation is meant to supersede
or replace the instructions for use applicable to each specific device, or the processes
and procedures in place at your facility.
USG-410 – Quick Reference Guide
Troubleshooting
Error Possible causes Solution
U504/U601
Probe damage
Instrument has not been replaced. Replace the instrument.
Plug is not connected properly to the socket. Connect plug and socket firmly.
Transducer temperature is too high. Stop output until the temperature cools down.
Transducer malfunction or probe damage/malfunction. Replace the transducer/instrument.
U508/U511
Short circuit
(THUNDERBEAT only)
Output is generated while a metallic object is grasped. Confirm that no metal object is grasped between jaw and probe tip.
Output in body fluids. Remove excess fluids.
Grasping section wear out. Replace instrument.
U509
Ultrasonic level error
Excessive load is applied to the distal end. Reduce the force applied to the distal end or to the probe tip.
Withdraw body fluid and tissue on the grasping section, probe tip, and shaft surface.
Transducer plug is not connected properly to the THUNDERBEAT / SONICBEAT
socket at USG-410.
Confirm that the transducer plug is connected to the THUNDERBEAT / SONICBEAT socket at USG-410.
If not, connect it properly.
Transducer malfunction or probe damage/malfunction. Replace the transducer/instrument.
U510
Ultrasonic frequency
error
Connection of transducer and instrument is loose. Connect transducer firmly with instrument.
Excessive load is applied to the distal end.
Reduce the force applied. Remove body fluid and tissue on the grasping section, probe tip, and shaft surface.
Transducer socket and plug are not connected properly. Check if plug and socket are connected properly.
Transducer temperature is too high. Stop output until the temperature cools down.
Transducer malfunction/probe damage or malfunction. Replace transducer/instrument.
U512
Open circuit
(THUNDERBEAT only)
Instrument is activated without holding tissue. Re-grasp tissue and confirm activation.
Transducer and instrument or transducer socket and plug are not connected properly.
Check if transducer and instrument or plug and socket are connected properly.
Transducer/instrument malfunction. Replace transducer/instrument.
U514
Seal incomplete
(THUNDERBEAT only)
The specified output time period has expired (maximum output time: 15 s). Withdraw tissue or fluids from jaw and probe tip. Re-grasp tissue and confirm activation.
W7083361_00 · 12/22 · GMA