OML LINQ 1.0 Operator's manual

Odstock Medical Limited
To be used in conjunction with the
OML Stimulator Instruction Manual
www.odstockmedical.com
USER INSTRUCTION MANUAL
OML LINQTM
Model: OML LINQTM
Model Version: 1.0
Software: V1.2

2
Table of Contents
Introduction and Key Features 3
Precautions 4
Attaching the Holder 6
Removing the Holder 7
Installing a Battery 8
Joining Instructions 10
Installing a Wired Footswitch 12
Use and Care Instructions 13
Miscellaneous 14
Warranty 15
Components and Accessories 16
Specification and Operating Environment 17
Fault Finding 18
Description of Symbols 19
Manufacturer Details 20

3
Introduction
The OML LINQTM enables a wireless link to be set up between an
Odstock Medical Wired Footswitch and an Odstock Medical wireless
enabled stimulator such as the ODFS®Pace XL. Uniquely the OML
LINQTM can have two Wired Footswitches attached allowing for
additional triggering options (refer to the OML LINQTM: Addendum for
Clinicians). The OML LINQTM has been designed to allow the user
to set the device up with their wireless stimulator. However some
functions identified in this manual require a clinician to set them up.
Key Features
Introduction and Key Features
Holder
Battery
Cover
Coin Cell
Battery
OML
LINQTM
Wired
Footswitch

4
Precautions
1. Warning: keep the OML LINQTM components out of reach of
babies and children. The OML LINQTM is not a toy!
• Components of the OML LINQTM are a choking hazard.
• Battery is extremely hazardous if swallowed. Seek
immediate medical attention.
• Wires of the Wired Footswitch are a strangulation hazard.
2. Ensure that a suitable Wired Footswitch length is used so that
the user does not catch excess wire on themselves or other
objects.
3. Replace Wired Footswitch if lead becomes stiff or damaged.
4. Warning: battery may explode if mistreated. Do not recharge,
disassemble or dispose of in fire.
5. The OML LINQTM should be securely mounted especially if
used for jogging/running.
6. The OML LINQTM is splashproof but must not be submerged.
7. The OML LINQTM must not be dis-assembled by anyone other
than Odstock Medical Limited. Doing so will invalidate the
warranty.
8. The OML LINQTM must not be modified by anyone other than
Odstock Medical Limited. Doing so will invalidate the warranty.
9. The OML LINQTM must only be used with Odstock Medical Ltd
products.
10. Before use check that the OML LINQTM has no visual signs of
damage.

5
11. The Wired Footswitch should be securely mounted to either the
insole provided or within the shoe.
12. Like all wireless devices the OML LINQTM should not be used
in areas where wireless is prohibited (e.g. aircraft). The OML
LINQTM should be carried in luggage with the battery removed
from the device. Revert to using the stimulator with a Wired
Footswitch.
13. Whilst driving ensure that the OML LINQTM is not in a position
where it could be knocked or dislodged or affect foot pedal
operation.
14. Do not over bend the clip of the Holder. Remove the Holder
from the shoe by sliding upwards only.

6
1. Ensure that the OML LINQTM is positioned correctly in the
Holder.
2. Push the OML LINQTM into the Holder ensuring that both arms
click into place into the slots on the side of the device.
The Holder allows the OML LINQTM to be attached to the side of
a shoe or sock. Ensure that it is securely attached before use!
When removing the Holder from the shoe be careful not to over
bend the clip. Slide the Holder upwards, off the rim of the shoe.
The OML LINQTM can be used without the Holder and placed directly
in a sock.
Attaching the Holder
Arms
Slots

7
1.Pull either arm of the Holder away from the OML LINQTM.
2.Gently lift the Holder away from the OML LINQTM.
Removing the Holder

8
Installing a Battery
WARNING: replace the battery with a CR2430 coin cell only.
Remove battery if the device is not to be used for an extended
period of time, such as 1 month.
1. Turn the stimulator off and then back on.
2. Remove OML LINQTM from the Holder if attached (see page 7).
3. Slide open the battery cover. The cover slides away from the
OML LINQTM. Press down on the rear portion of the cover to
release the clip (please refer to image A and B).
4. If present then remove the old battery.
5. Install the new battery with the positive side of the battery
facing out from the OML LINQTM. If installed correctly the blue
light will flash once (please refer to image C).
6. Slide the battery cover back on to the main body. The battery
cover is fully inserted when there are no gaps between the
casing and the battery cover (please refer to image D). Do not
use excessive force.
7. Re-insert the OML LINQTM into the Holder (see page 6), insert
the Wired Footswitch and check that the stimulator unpauses
and responds to a press of the footswitch (see page 12).

9
A
C
D
B

10
Joining Instructions
Use this process if the OML LINQTM is not joined to the stimulator
or when the OML LINQTM is provided separately from the
stimulator.
NB. Ensure that the OML LINQTM does not contain a battery prior
to starting this process.
1. Insert a battery into the OML LINQTM and close the battery cover.
2. Ensure that the stimulator is turned on.
3. Select the joining option from the wireless menu on the stimulator
(refer to stimulator user instruction manual or OML LINQTM:
Joining to ODFS®Pace XL quick start card supplied).
4. When the stimulator indicates that it is waiting to join (this will
time out after 1 minute), press and release the joining button on
the OML LINQTM. A successful press of the button will result in
the blue light flashing for up to 1 minute.
5. Bring the stimulator and OML LINQTM close to each other.
6. After a few seconds the stimulator will indicate that joining has
been successful and the blue light on the OML LINQTM will stop
flashing.
7. Your OML LINQTM is now joined.
If unsuccessful repeat this process.

11
Flashing Blue Light
‘Joining in Progress’
Joining Button
‘Single Press’

12
Installing a Wired Footswitch
Prior to inserting the footswitch plug into the socket make sure
that there is no debris in the socket of the OML LINQTM.
Insert the Odstock Medical Wired Footswitch into either socket of the
OML LINQTM. Your clinician will set up the position of the footswitch
in your shoe/insole for reliable operation.
If using a single footswitch it does not matter which socket the
footswitch is inserted into.
NB. Two Odstock Medical Wired Footswitches can be used on
the OML LINQTM. This replicates the function of the OML double
footswitch setup. Please refer to the OML LINQTM: Addendum for
Clinicians card.
If this functionality is required then this can be setup by your
clinician. Warning: do not attempt to set this up yourself. This
should only be set up by a trained clinician.

13
Use and Care Instructions
Normal Use
Normal use of the OML LINQTM is defined as wearing the device in
the Holder mounted to either a shoe or sock. Alternatively, the OML
LINQTM can be used without the Holder and placed directly in a sock.
Warning: if you have latex or rubber allergies do not use against
the skin.
Care Instructions
The OML LINQTM can be cleaned using a damp cloth. Do not
immerse in water or use solvents.
If the OML LINQTM becomes wet please remove the battery cover
and battery and allow to air dry thoroughly for 24 hours before re-
use. If unresponsive contact Odstock Medical Limited.
Expected Service Life
The OML LINQTM has an expected service life of 2 years. After this
period it may be necessary to replace the device. The sockets of the
OML LINQTM may require replacement during the life of the product
due to daily wear.
FES Support Group
There is a web based discussion group for users of Odstock Medical
FES devices. Visit www.odstockmedical.com for information on how
to join the group.

14
Miscellaneous
Activity and Inactivity Detection
To save battery power the OML LINQTM goes into a sleep mode
when no activity is detected for 5 minutes or more. The device is
woken up by any small motion of the OML LINQTM or by pressing
the Wired Footswitch. If no movement occurs before pressing the
pause on the stimulator to begin walking, the OML LINQTM may not
be detected by the stimulator. If this occurs a small movement of the
foot or force on the Wired Footswitch will wake up the OML LINQTM.
Turning o the OML LINQTM
There is no off switch on the OML LINQTM. The device will enter a
low power state if it is not being used but if it is not to be used for an
extended period (i.e. 1 month), then remove the coin cell battery.
Close Proximity Joining
Close proximity joining is used if many devices are being joined at
the same time in a confined space (e.g. a clinical area). If a Wired
Footswitch is inserted and pressed down at the same time that the
joining button is pressed then close proximity joining will start. This is
indicated with the blue light flashing quickly. The two devices should
be positioned touching each other for joining to be successful.
Transportation
Odstock Medical Limited recommend transporting the OML LINQTM
without a battery installed to turn off wireless communication (e.g.
on an aircraft). In this case use the stimulator in Wired Footswitch
mode. Warning: Please check with local authorities/regulations
on the requirements for transportation of lithium batteries.

15
Warranty
The OML LINQTM is warranted for a period of one year from the date
of sale or tting.
The following exceptions apply:
• Footswitches are warranted for a period of one month
• Leads and batteries
Should the unlikely event of any failure of the device occur during the
warranty period, the device should be returned to Odstock Medical
Limited for inspection. Should the failure be due to manufacturing or
material defect the device will be repaired or a replacement supplied
free of charge.
The warranty is valid providing that:
1. The failure can not be attributed to misuse or improper tting
2. The warranty registration card has been completed and
returned to Odstock Medical Limited as soon as possible from
the date of sale or tting.
The warranty card is supplied inside the product packaging and
should be returned to Odstock Medical Limited.
This warranty is in addition to any statutory rights avaliable to the
purchaser.

16
Components and Accessories
Kit Contents
Description Product Code Qty
OML LINQTM 01-001-0029 1
User Instruction Manual 11-003-0027 1
Clinician Addendum Card 11-003-0031 1
User Quick Start Card 11-003-0049 1
OML Warranty Card 11-003-0067 1
Black Wired Footswitch 22cm 01-005-0044 1
CR2430 Varta Coin Cell Battery 05-001-0004 1
Universal Foam Insole 01-005-0043 1
Optional Extras and Spares
Description Product Code
Footswitch length options
Black Wired Footswitch 22cm 01-005-0044
for standard footwear
Black Wired Footswitch 30cm 01-005-0041
for short boots
Black Wired Footswitch 60cm 01-005-0042
for high boots
User Replaceable Parts
OML LINQTM Battery Cover 12-001-0031
OML LINQTM Holder 12-001-0032
Consumables
Varta Coin Cell CR2430 Battery105-001-0004
1 Coin cell batteries should meet IEC 60086. We advise to use VARTA batteries.

17
Specification and Operating Environment
Specication
Battery VARTA, Lithium Coin Cell, CR2430, 3V
Battery Life >3 weeks
Dimensions 49 x 42 x 14mm 2
Weight 16g
2,3
Materials ABS, Polycarbonate, Rubber Coating
2Without Holder 3Without battery
Operating Environment
Operating Temperature 0 to +45oC
Storage and Transport Temperature -25 to 70oC
Operating Relative Humidity 15 to 93%
Storage and Transport Relative Humidity 15 to 93%
Pressure Range 700 to 1060hPa
Wireless Specication
Frequency Band 2.4 to 2.483GHz (2.4GHz ISM Band)
Contains FCC ID S4GEM35XA
The wireless module used has been designed to meet all national
regulations for worldwide use. The wireless module is EC-R&TTE
certified and complies with FCC CFR Part 15 (USA) meeting the
requirements for modular transmitter approval as detailed in the FCC
public notice DA00.1407 transmitter.

18
Fault Finding
Fault Problem Solution
No blue light on
battery insertion
1. Battery inserted incorrectly
2. Insucient battery power
Please refer to
instructions on page 8.
Stimulator indicates
that there is no
footswitch
1. Battery inserted incorrectly
2. Insucient battery power
3. OML LINQTM not joined
Please refer to instructions
on page 8. If not successful
please refer to instructions
on page 10.
OML LINQTM
does not stay
in the Holder
1. OML LINQTM not
properly inserted
2. Slots dirty
3. Holder broken
Please refer to
instructions on page 6.
Clean slots
Replace Holder
Stimulator
unpauses but is
not triggered by the
wired footswitch
1. Wired footswitch
not plugged in
2. Worn wired footswitch or lead
3. Wired footswitch incorrectly
inserted into the shoe
Plug in wired footswitch
Replace wired footswitch
Check that the wired
footswitch is correctly
positioned into the shoe
in the correct location.
Footswitch lead
does not fully insert
or stay in the socket
1. Dirt within the socket
2. Worn or damaged socket
Clear the socket.
Use alternative socket
Please contact your
clinician or supplier.

19
Description of Symbols
Read instruction manual before use.
CE mark declares that the product meets the
requirements of applicable EC directives.
Use by date in the format of year/month.
The OML LINQTM contains an RF transmitting device.
Protected against solid foreign objects of greater than
2.5mm and dripping water at an angle up to 15 degrees.
The OML LINQTM should not be disposed of with other
household waste.
CR2430
The OML LINQTM uses a CR2430 coin cell battery.
Manufactured by Odstock Medical Limited.
Not for use by children under the age of 3 years.
Do not use if packaging is broken or damaged.
Product code.
Unique product serial number.
Transport and storage temperature limit.
Transport and storage humidity limit.
0 3
REF
SN
IP32

20
Wireless Information - OML LINQ™
Description The OML LINQ™ contains an industry
standard ZigBee PRO compliant RF
telemetry module, the ETRX3.
Contains RF
transmitter
(FCC ID)
S4GEM35XA
Industry Canada
(IC-ID)
8735A-EM35XA
Frequency band 2.4 to 2.483 GHz (2.4GHz ISM band)
Radiated power +8dBm
Wireless
compliance
The wireless module used has been
designed to meet all national regulations
for worldwide use. The wireless module
complies with the requirements of the
Radio Equipment Directive (2014/53/EU)
and complies with FCC CFR Part 15(USA)
meeting the requirements for modular
transmitter approval as detailed in FCC
public notice DA00. 107. Transmitter.
Quality of
service
The OML LINQ™ is designed and tested to
have a response rate of less than 100ms
latency depending on system configuration
after the detection of a heel event.
Co-exsistence
and interference
The system has been stringently tested
in a range of scenarios showing that it is
not susceptible to interference. However,
there is no guarantee that other wireless
communication devices will not interfere
with wireless communications.
The module is also certied to European
Certication (ETSI) EN 300 328:V2.1.1
(radio) EN301 489-1 V2.1.1 (EMC) EN301
489-17:V3.3.1 (EMC) and IEC 60950-
1:2005+AMD2:2013
This manual suits for next models
1
Table of contents
Other OML Personal Care Product manuals
Popular Personal Care Product manuals by other brands

Silvercrest
Silvercrest SNS 45 B4 operating instructions

SeaLand
SeaLand VacuFlush 3000 Series owner's manual

Medel
Medel Smartskin WaterWear for BTE user manual

Home Skinovations
Home Skinovations Silk'n Reveal manual

Medel
Medel Bodytech user manual

BaByliss PRO
BaByliss PRO NANO TITANIUM BNTHB250A quick start guide