Press the air release button
to release the air in the arm
cuff until the deflation symbol
is no longer displayed.
END
Manual Inflation Blood
Pressure Monitor
Model HEM-4030
Instruction Manual
Thank you for purchasing the OMRON HEM-4030 Manual Inflation Blood
Pressure Monitor.
The OMRON HEM-4030 is a compact manual inflation blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body
when taking measurements at night. This could result in injury when the air
pressure in the air tube is increased.
•Do not leave the cuff wrapped on the arm if taking measurements during the
night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only two “AAA” alkaline or manganese batteries with this unit. Do not use
other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace both batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull the air tube with holding near the air jack.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
Important Safety Information
Introduction
1. Overview
Main unit:
A. Display
B. M (Memory) button
C. ON/OFF (Power switch) button
D. Air jacks for cuff and inflation bulb
E. Battery compartment
F. Arm cuff
(Medium cuff: arm
circumference 22-32 cm)
G. Air tube
H. Air release button
I. Air tube
J. Air inflation bulb
Display:
Arm cuff:
O. Systolic blood pressure
P. Diastolic blood pressure
Q. Pulse display
R. Memory No. symbol
S. Deflation symbol
K. Heartbeat symbol
1. Flashes during measurement
2. If flashing after measurement
complete indicates blood
pressure out of recommended
range
L. Reinflation symbol
M. Battery low symbol
N. Memory symbol
2.1 Installing/Replacing the Batteries
2. Preparation
1.
2.
Remove the battery cover.
Insert two “AAA”batteries
as indicated in the battery
compartment and then
replace the batterycover.
Notes:
• If the battery low symbol ( ) appears on the display, turn the unit off
then replace both batteries at the same time.
•The measurement values continue to be stored in memory even after
the batteries are replaced.
Disposal of used batteries should be carried out in accordance with the
national regulations for the disposal of batteries.
The bottom edge of the cuff should be 1 to 2 cm above the elbow.
Marker (arrow under tube) is centred on the middle of your inner arm.
Close the fabric fastener FIRMLY.
3. Using the Unit
4. Troubleshooting and Maintenance
1. Insert the air tubes into the air jacks.
2. Put your arm through the cuff loop.
3. Position the arm correctly.
1.
2. Pump the inflationbulb to inflate the arm cuff.
This criteria is for home blood pressure measurement.
Note: If the memory is full, the monitor will delete the oldest readings.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
3.1 Applying the Arm Cuff
The monitor automatically stores the result up to 21 sets.
3.3 Taking a Reading 6. Technical Data
5. Optional Parts
4.1 The Icons and Error Messages
Blink
Lit
4.2 Troubleshooting
4.3 Maintenance
Problem
3.4 Using the Memory Function
Calibration and Service
To protect yourunit fromdamage, pleaseobserve the following:
Cause Remedy
Notes:
• When you take a measurement on the right
arm, air tube will be at the side of your
elbow. Be careful not to rest your arm on
the air tube.
Note:
If the deflation symbol does not disappear soon, press the air release
button to release any air in the arm cuff.
To delete all the values stored in memory
• Press the ON/OFF button and repeat measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
•Do not subject the main unit, cuff and inflation bulb to
extreme temperatures, humidity, moisture or direct
sunlight.
•Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
•Do not disassemble the unit.
•Do not subject the unit to strong shocks or vibrations
(for example, dropping the unit on the floor).
•Do not use volatile liquids to clean the main unit. The
unit should be cleaned with a soft, dry cloth.
•Use a soft, moistened cloth and soap to clean the arm
cuff.
•Do not wash the arm cuff or immerse it in water.
•Do not use petrol, thinners or similar solvents to clean
the arm cuff.
•The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
•It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your local authorised
OMRON distributor or dealer.
•Do not carry out repairs of any kind yourself. If a defect
occurs, consult your local authorised OMRON
distributor or dealer.
Manual Inflation Blood PressureMonitor
HEM-4030
LCDDigital Display
Oscillometric method
Pressure: 0 mmHg to 299mmHg
Pulse: 40 to 180/min.
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Manual by inflationbulb
Automatic pressure release valve
21Measurements
2 “AAA”batteries 1.5V
Capacity of newmanganese batteries is approx.
1500measurements
= Type B
Internally powered ME equipment
+10°C to +40°C / Maximum: 30 to 90% RH
-20°C to +60°C / Maximum: 10 to 95% RH /
700-1060 hPa
Approximately75g without batteries
Approximately120g
Approximately64(w)mm x35 (h)mm x 105(I)mm
Approximately146 mm x 446 mm
(Medium cuff: armcircumference22 to 32cm)
Nylonandpolyester
Mainunit, cuff, inflationbulb, instructionmanual,
storage case, battery set
Inflation bulb: Storage case:
Remove tight-fittingclothing or tight rolled up sleeve fromyour
upperarm.
Do not place the cuff over thick clothes.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortableroom temperature. No eating,
smoking or exercising30minutes before
taking a measurement.
•Sit on a chair with your feet flat on the floor.
•Sit upright with your back straight.
•The cuff should be at the same level as your
heart.
1 to 2 cm
•The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. Omron recommends to always use the same arm for
measurement. If the values between the two arms differ substantially,
please check with your doctor which arm to use for your
measurement.
Press the airrelease button to releaseanyair in the arm
cuff, thenpress the ON/OFF button to turn the unit on.
3. Measurement starts.
Measurement starts automatically after you stop inflating
the arm cuff.
4. Remove the arm cuff.
5. Press the ON/OFF button to turn off the monitor.
The monitor automatically stores the measurement in its
memory.
It will automatically turn off after fiveminutes.
Notes:
• If the reinflation symbol ( ) appears, squeeze the inflation bulb to
re-inflate the arm cuff.
•Do not inflate the arm cuff more than necessary.
Notes:
•To cancel a measurement, press the ON/OFF button to turn off the
unit and press the air release button to release the air in the arm cuff.
• Wait 2-3 minutes before taking another blood pressure measurement.
Waiting between readings allows the arteries to return to the condition
prior to taking the blood pressure measurement.
Note: You cannot partially delete the stored readings.
When the memory symbol ( ) appears, first press the M button.
Then while holding it down, press the ON/OFF button simultane-
ously for about 2-3 seconds.
Note: Remain still while taking measurement.
Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
1) Inflate the cuff until it is 30 to 40 mmHg
above your expected systolic blood
pressure value.
ex) If your expected blood pressure is
around 140mmHg, inflate the arm cuff to
between 170 and 180 mmHg. Inflate the
cuff rapidly so that the pressure is reached in about five seconds.
2) When the desired pressure has been achieved, release the inflation
bulb.
START DEFLATE COMPLETE
Warning:
•If your systolic or diastolic pressure is outside
the standard range, the heartbeat symbol will
blink when the measurement result is
displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Important:
Previous Reading
The systolic blood pressure
is above135mmHg or
diastolic bloodpressure is
above 85 mmHg.
The batteriesarelow.
The batteriesare
exhausted.
Cuff is under inflated.
Movement during
measurement.
Cuff over inflated.
Device error.
Pressure is too low.
Recent research suggests
these values can be used
as a guide to high blood
pressure for measurements
taken at home.
The reading is
extremely low (or
high).
Arm cuff pressure
does not rise.
Cannot measure or
readingsare too low
or too high.
The unit losespower
duringmeasurement.
Nothinghappens
when you press the
buttons.
Otherproblems.
Arm cuff not applied
correctly.
Movement or talking
duringmeasurement.
Clothing is interfering
with the arm cuff.
The air tube is not
securely connected
into the mainunit.
Air is leaking from the
arm cuff.
The arm cuff has not
been inflated
sufficiently.
The airrelease button
beingpressed during
inflation.
The batteriesare
empty.
The batteriesare
empty.
The batterieshave
beeninserted
incorrectly.
Apply the arm cuff
correctly.
Refer to section 3.1.
Remainstilland do not
talk during
measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.2.
Make sure that the air
tube is connected
securely.
Refer to section 3.1.
Replace the arm cuff
with a newone.
Refer to Chapter 5.
Inflate the cuff so that it
is 30 to 40 mmHgabove
yourprevious
measurement result.
Refer to section 3.3.
Be careful not to press
the airrelease button
duringmeasurement.
Replace the batteries
with newones.
Refer to section2.1.
Replace the batteries
with newones.
Refer to section2.1.
Insert the batteries with
the correct (+/-)polarity.
Refer to section2.1.
You shouldreplace them
with newonesahead of
time.
Refer to section2.1.
You shouldreplace them
with newones at once.
Refer to section2.1.
Carefully readandrepeat
the stepslisted under
section 3.3.
Contact your OMRON retail
outlet or distributor.
Refer to Chapter 6.
y d eme Re s u a CError Display
Press the inflationbulb to
inflate the arm cuff until the
reinflation symbolgoes out.
Or, deflate the arm cuff and
repeat measurement after
checking that the heartbeat
symbol ()has been
displayed.
Refer to section 3.3.
1. Unplug the air tube from the airconnector.
2. Gently fold the air tube into the arm cuff.
4.4 Storage
Keep the unit in its storagecasewhen not in use.
Do not keep the mainunit beingwound with the arm cuff.
Note:
Do not bend the air tube excessively.
Note:
Should you require a small cuff, please ensure that the small cuff and
small bulb are used together.
They may also be purchased as a combination set.
3.
Do not store the unit in the following situations:
•If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct
sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an
angle.
Place the arm cuff, inflation bulb and
main unit in the storage case.
Medium Arm Cuff
Arm circumference
22 - 32 cm
Large Arm Cuff
Arm circumference
32 - 42 cm
Small Arm Cuff
Arm circumference
17 - 22 cm
Small cuff and bulb combination
Arm circumference 17 - 22 cm
Regular bulb
Product Description
Model
Display
Measurement Method
Measurement Range
Accuracy
Inflation
Deflation
Memory
Power Source
Battery Life
Applied Part
Protection Against
Electric Shock
Operating
temperature/ Humidity
Storage temperature/
Humidity/ Air pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Cuff Material
Package Content
5325569-2B
Turn the Unit on View Previous
Reading
View Readings
Stored in Memory
•The Memory number
appears for a second
before the pulse rate
is displayed. The
newest set is
numbered “1”.
•Press repeatedly to
cycle through the
previous results.
•Hold down to cycle
rapidly.
Please read this instruction manual carefully before using the device.
•This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
•Thisbloodpressuremonitor is designed according to the European Standard
EN1060, Noninvasive sphygmomanometers Part 1: GeneralRequirements
and Part 3: Supplementaryrequirements for electromechanicalblood
pressure measuring systems.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronicdevicessuch as PC’sandmobile
(cellular)telephones, medicaldevices in use maybesusceptible to
electromagnetic interference from otherdevices.
Electromagnetic interference mayresult in incorrect operation of the medical
deviceand create a potentiallyunsafe situation.
Medicaldevicesshouldalso not interfere with otherdevices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standardhasbeenimplemented. This standard defines the levels of immunity to
electromagnetic interferences as wellasmaximum levels of electromagnetic
emissions for medical devices.
Thismedical devicemanufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity andemissions.
Nevertheless, specialprecautionsneed to be observed:
•Do not use mobile(cellular)telephonesand otherdevices, which generate
strongelectricalorelectromagnetic fields, near the medical device. Thismay
result in incorrect operation of the unit andcreate a potentiallyunsafe situation.
Recommendation is to keep a minimumdistance of 7 m. Verify correct
operation of the deviceincase the distanceisshorter.
Furtherdocumentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instructionmanual.
Documentationisalsoavailable at www.omron-healthcare.com.
Thismarking shown on the product or its literature, indicates that it
should not be disposed of, with otherhousehold wastes at the end
of its working life. To prevent possibleharm to the environment or
humanhealth fromuncontrolledwaste disposal, pleaseseparate
this from other types of wastes and recycle it responsibly to
promote the sustainablereuse of material resources.
Household usersshouldcontact either the retailerwhere theypurchased this
product, or theirlocal government office, for details of whereandhow they can
take this item for environmentally safe recycling.
Business users shouldcontact theirsupplierandcheck the termsandconditions
of the purchase contract. Thisproduct should not be mixed with othercommercial
wastes for disposal.
Thisproduct does not containanyhazardoussubstances.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
Notes:
•Subject to technical modification without prior notice.
•
Disposal of this product andused batteries should be carried out in accordance
with the nationalregulations for the disposal of electronic products.
Asia Pacific HQ
OMRON HEALTHCARE Co., Ltd.
24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto
615-0084 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Kruisweg577, 2132NA Hoofddorp THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08, Alexandra Technopark,
Singapore 119967
Other Omron Blood Pressure Monitor manuals

Omron
Omron T9P User manual

Omron
Omron HEM-7121 User manual

Omron
Omron HEM-630 User manual

Omron
Omron M3 HEM-7154-E User manual

Omron
Omron M3 Comfort User manual

Omron
Omron HEM--790IT User manual

Omron
Omron M7 Intelli IT User manual

Omron
Omron M3 Intellisense User manual

Omron
Omron BP652 User manual

Omron
Omron HEM-629 User manual

Omron
Omron HEM-7155T-ESL User manual

Omron
Omron HEM-SOLAR User manual

Omron
Omron R6 User manual

Omron
Omron 705IT User manual

Omron
Omron HEM-6052 User manual

Omron
Omron HEM-6121 User manual

Omron
Omron BP652 User manual

Omron
Omron M6 Comfort User manual

Omron
Omron HEM--670IT User manual

Omron
Omron RC212 User manual