
HOBO U24 Conductivity Logger (U24-002-C) Manual
1-800-LOGGERS 4 www.onsetcomp.com
ranges (the calibrated ranges are also printed on the logger
housing). The logger will not record readings outside of the
extended range selected. (For the U24-002-C, readings
below the extended range are reported as 0.) If in doubt on
the range needed for your deployment, or for environments
with wide fluctuations, select both ranges. This will shorten
the deployment duration from 18,500 samples to 14,400
samples per parameter (not logging battery voltage). See
the HOBOware User’s Guide or online help for details on
launching.
Note: Logging battery voltage is not essential because you can
check the battery using the Status screen at launch or readout
of the logger. Logging the battery voltage will reduce the
number of conductivity and temperature readings you can log.
Taking Calibration Readings
It is important to take temperature and conductivity calibration
readings with a portable conductivity meter at both the
beginning (launchtime) and end of a deployment (readout)
because these readings are necessary for data calibration and
to compensate for any measurement drift during deployment.
The conductivity calibration readings should be the actual
conductivity without temperature compensation (not in specific
conductance at 25°C), and should be recorded in a notebook
with the time and location of the reading. You will use these
readings in the HOBOware Conductivity Assistant to calibrate
the readings for the corresponding data series offloaded from
the logger.
The calibration reading should be taken in the middle of the
expected range in the environment. For example, if the range
where the logger will be deployed is 20,000 µS/cm to 40,000
µS/cm, then the calibration reading should be taken at 30,000
µS/cm, which is the midpoint of that range (see Method 2 at
right for advice on how to get readings at the midpoint).
There are three methods for obtaining accurate calibration
readings. The first method involves placing the meter’s probe
into the water next to the logger. The second method involves
placing the logger and meter probe in a field water sample in a
jar. In both methods, the conductivity meter probe must be
close to the data logger—but not touching—so that it is
measuring water at the same conductivity and salinity as the
logger. The third method involves taking a sample back to the
office to measure with a meter there.
If the conductivity and salinity in the water where the logger is
deployed is stable and it is easy to reach the logger, then you
can obtain calibration readings by placing the probe directly
into the water next to the logger. However, taking calibration
readings in areas that have tidal and freshwater mixing is more
challenging due to the rapidly changing salinities. Similarly,
taking calibration readings in wells can also be difficult because
it may be hard to get the meter probe next to the logger. In
these instances, you should fill a jar with a water sample from
where the logger is deployed to take the calibration readings.
To obtain the water sample from a well or stilling well, you can
use a bailer with a diameter that is small enough to fit down
the well.
Note: Some salt residue may remain on the logger from factory
calibration. Carefully rinse the logger in distilled water or clean
freshwater to remove any residual salt before taking your first
calibration readings.
Method 1:
Taking readings directly in the water (recommended for
locations with access for the field meter probe and with
conductivity that is stable)
1. If you have just deployed the logger, allow enough time for
the logger temperature to stabilize for the best accuracy
(approximately 15 minutes).
2. Gently tap the logger to remove any bubbles from the
surface. Tug the cable if you cannot reach the logger itself.
3. Measure the temperature and actual conductivity with the
field meter, making sure the meter probe readings stabilize
per the meter’s specification. Record the values, time, and
location of the readings in a field notebook for use later in
the HOBOware Conductivity Assistant.
Method 2:
Taking readings in a jar (recommended for readings in wells or
in water with rapidly changing conductivity, such as areas
with saltwater and freshwater mixing)
1. Take a sample of water in a jar that is large enough to hold
both the logger and the probe from a portable conductivity
meter, leaving an inch of space between the probe and the
logger. For wells, use a bailer to obtain the water sample.
IMPORTANT: For the best accuracy, use a calibration
reading at the midpoint of the expected conductivity
range. This can be done either by taking a sample in the
middle of a tide cycle or by taking a sample at high tide
and then adding freshwater to reach the expected
midpoint. Always be sure to stir thoroughly to fully mix
the fresh and saltwater.
2. Leave the logger and the meter probe submerged in this jar
of water long enough so they reach temperature
equilibrium and the logger has logged at least three
readings (allow at least 15 minutes for the best accuracy).
(Three readings are necessary because this will help you
identify which readings were taken while the logger is in the
jar.)
3. Measure the temperature and actual conductivity with the
field meter. Record the values, time, and location of the
reading in a field notebook for use later in the HOBOware
Conductivity Assistant.
4. When using the Conductivity Assistant, look for the spot in
the data where there are three similar readings in a row
and link the last of those readings to the meter reading.
(The time you noted may be slightly different than the
logger time so looking for the three similar readings will
help identify the correct reading.) The Conductivity
Assistant uses that value to calibrate the specific
conductance and salinity readings for that data series.
Method 3:
Taking a sample back to the office in a sealable jar to measure
there (recommended for locations with conductivity that is
stable when you do not have a field meter or it is not
convenient to access the logger)
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524