Quick Start for the HOBO® MX pH and Temperature Logger (MX2501)
22510-J MAN-MX2501-QSG
IMPORTANT: The logger is shipped with the pH electrode installed and submerged in storage solution. The pH sensor must be kept in
storage solution when not being calibrated or deployed in water. The hydration layer surrounding the sensor starts to break down after
10 minutes if the pH sensor is not submerged in storage solution or calibration solution, or deployed in water.
IMPORTANT: The pH electrode is uncalibrated. You must calibrate the logger as described in step 2 before configuring and deploying it.
1 Download and open HOBOconnect.
a. Download HOBOconnect™ to your phone or tablet.
b. Open HOBOconnect. If prompted, enable Bluetooth® in your device settings.
c. Wake up the logger by pushing the switch on the mounting end of the logger into the gap beneath the mounting lanyard
(see diagram on back). Tap Devices and make sure the logger appears. If it does not, make sure it is within range of your
phone or tablet.
2 Calibrate the logger.
Before you begin: To complete these steps, you will need
deionized or distilled water, a squirt bottle, calibration
solution (also known as buffer solution), and beakers.
Note: pH values of calibration buffers and samples will vary
depending on their temperature. As a result of this, it is
important that the logger is calibrated in buffers that have
the same temperature as that of the sample. Refer to the
chart on the calibration buffer bottles or in the logger manual
for the effect of temperature on the buffer pH values.
a. Pour each calibration solution into its own beaker (pH
7.00 solution and either one or both of pH 4.01 and pH
10.00 solutions). You will need enough solution in the
beaker to cover the sensor end cap and closure cap on
the logger.
b. Unscrew the clear storage cap filled with storage solution
and remove it from the logger. Keep the cap filled with
storage solution nearby in case you need to reinstall it
after calibrating and before deploying the logger.
c. Screw the closure cap onto the logger.
d. Use a squirt bottle to rinse the pH sensor with deionized
or distilled water.
e. Tap Devices (if necessary) and tap the logger in the app
to connect to it.
f. Tap .
g. Place the sensor end of the logger in the pH 7.00
solution. Make sure the sensor end cap, temperature
sensor, and closure cap are submerged in the solution as
shown in the illustration. Tap Start. Follow the
instructions on the screen. Once the pH reading is stable,
tap Confirm.
h. Rinse the pH sensor with deionized or distilled water.
i. If you are using the pH 4 buffer, place the sensor end in
the pH 4 solution and tap Start. If you are not using the
pH 4 buffer, tap Skip and go to step k.
j. Once the pH reading is stable, tap Confirm. Rinse the pH
sensor with deionized or distilled water.
k. If you are using the 10 buffer, place the sensor end in the
pH 10 solution and then tap Start. Once the pH reading is
stable, tap Confirm. If you are not using the pH 10 buffer,
tap Skip.
l. Verify that the slope and offset is acceptable and tap
Save. If the slope and offset are not acceptable, then tap
the back arrow and repeat the calibration steps in this
section as needed.
m. Rinse the pH sensor with deionized or distilled water.
If you will not be deploying the logger immediately, place
the pH electrode back into the storage solution. Unscrew
the sensor end cap you installed in step c. Make sure the
storage solution reaches the marked fill line on the clear
storage cap; pour more into the cap if necessary. Carefully
screw the clear cap with storage solution onto the logger.
Only use storage solution in the clear cap; do not use water.
Do not use the copper guard when the pH electrode is in
storage solution.
Unscrew the clear
cap filled with
storage solution
from the logger
Screw the closure
cap onto the logger
before calibrating
Closure cap
Sensor end cap Temperature sensor
1-800-LOGGERS (564-4377) • 508-759-9500
www.onsetcomp.com/support/contact
© 2018–2020 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOconnect are trademarks or
registered trademarks of Onset Computer Corporation. App Store is a service mark of Apple Inc. Google Play is a trademark
of Google LLC. Bluetooth is a registered trademark of Bluetooth SIG, Inc. All other trademarks are the property of their
respective companies.
This product has been manufactured by Onset Computer Corporation and in compliance with Onset’s ISO 9001:2015
Quality Management System.
Patent #: 8,860,569 22510-J MAN-MX2501-QSG
3
Configure the logger.
a. Tap Devices and then tap the logger in the app to connect to to it (push the switch on the logger to wake it up if necessary).
b. Tap
c. Choose the logger settings. Note that the default Bluetooth setting for the logger is Bluetooth Always On, which means it will
“advertise” or regularly send out a Bluetooth signal for the app to receive. This setting uses battery power. You can change
this to one of the following power saving modes:
•Bluetooth Always Off. The logger will only advertise when you wake it up by pushing the
switch on the mounting end of the logger.
•Bluetooth Off Water Detect. The logger will only advertise when it is out of water. When it is
in water, the Bluetooth signal is automatically disabled.
d. Tap to save the settings to the logger. The logger will begin logging data based on the
settings you selected. If you configured the logger to start or stop logging with the On
Button Push option, push the switch on the mounting end of the logger for 3 seconds.
4
Deploy the logger.
Deploy the logger to the location where you will be monitoring the conditions. Remove the clear
storage cap from the logger if installed and replace it with the closure cap. If your water sample is
prone to biofouling, insert the optional copper anti-biofouling guard around the pH sensor as shown
at right. Gently twist the copper guard so that the holes line up with the holes in the sensor end cap
and pH electrode. Do not use the copper guard when the pH sensor is being calibrated or in storage
solution, or if the sensor is placed in a very small container with still water or in water where copper
can affect the environment.
5
Download data from the logger.
a. Open HOBOconnect.
b. If the logger was configured with Bluetooth Always Off, push the switch on the logger for 1 second to wake it up. If the logger
was configured with Bluetooth Off Water Detect and it is deployed in water, remove it from the water.
c. Tap Devices, connect to the logger, and then tap .
d. To view, export, and share the data, tap HOBO Files, tap , and then tap .
For information on using this logger, refer to the complete product manual. Scan the code at left or go to
www.onsetcomp.com/support/manuals/22511-mx2501-manual.
Copper guard
installed to
protect against
biofouling
Push switch into the gap beneath the mounting lanyard
to start or stop logging or to wake up the logger
Make sure the closure cap is installed
before deploying the logger

Other manuals for MX2501

1

Other Onset Data Logger manuals

Onset EG4 Series User manual

Onset

Onset EG4 Series User manual

Onset HOBO ZW-008 User manual

Onset

Onset HOBO ZW-008 User manual

Onset HOBO U30 Series User manual

Onset

Onset HOBO U30 Series User manual

Onset HOBO RG3 User manual

Onset

Onset HOBO RG3 User manual

Onset InTemp CX40 Series User manual

Onset

Onset InTemp CX40 Series User manual

Onset HOBO U24 User manual

Onset

Onset HOBO U24 User manual

Onset Hobo MicroRX RX2105 User manual

Onset

Onset Hobo MicroRX RX2105 User manual

Onset HOBO UX100-023 User manual

Onset

Onset HOBO UX100-023 User manual

Onset HOBO MX Series User manual

Onset

Onset HOBO MX Series User manual

Onset HOBO U20L User manual

Onset

Onset HOBO U20L User manual

Onset MX2501 User manual

Onset

Onset MX2501 User manual

Onset Hobo UX90-001 Series User manual

Onset

Onset Hobo UX90-001 Series User manual

Onset HOBO MX CO2 Logger User manual

Onset

Onset HOBO MX CO2 Logger User manual

Onset HOBO UX100-001 User manual

Onset

Onset HOBO UX100-001 User manual

Onset HOBO UX90-002 User manual

Onset

Onset HOBO UX90-002 User manual

Onset HOBO Pendant UA-001 Series User manual

Onset

Onset HOBO Pendant UA-001 Series User manual

Onset InTemp CX450 User manual

Onset

Onset InTemp CX450 User manual

Onset HOBO H22 User manual

Onset

Onset HOBO H22 User manual

Onset HOBO UX120-017 User manual

Onset

Onset HOBO UX120-017 User manual

Onset HOBO User manual

Onset

Onset HOBO User manual

Onset HOBO UX100-011A User manual

Onset

Onset HOBO UX100-011A User manual

Onset HOBO User manual

Onset

Onset HOBO User manual

Onset EG4100 Series User manual

Onset

Onset EG4100 Series User manual

Onset HOBO Micro Station User manual

Onset

Onset HOBO Micro Station User manual

Popular Data Logger manuals by other brands

HW Group HWg-PWR 3 manual

HW Group

HW Group HWg-PWR 3 manual

Zebex Z-1160 user manual

Zebex

Zebex Z-1160 user manual

Leviton A8810 quick start guide

Leviton

Leviton A8810 quick start guide

Huawei SmartLogger2000 series user manual

Huawei

Huawei SmartLogger2000 series user manual

RST Instruments C109 instruction manual

RST Instruments

RST Instruments C109 instruction manual

Axiometrix Solutions imc CRONOS-XT Getting started

Axiometrix Solutions

Axiometrix Solutions imc CRONOS-XT Getting started

GRAPHTEC midi LOGGER GL240 user manual

GRAPHTEC

GRAPHTEC midi LOGGER GL240 user manual

Crane TorqueStar Lite Operator's manual

Crane

Crane TorqueStar Lite Operator's manual

EO ITAG 4 user manual

EO

EO ITAG 4 user manual

EasyLog EL-USB-RT quick start guide

EasyLog

EasyLog EL-USB-RT quick start guide

APAR AR207 operating instructions

APAR

APAR AR207 operating instructions

METER EM60 Series manual

METER

METER EM60 Series manual

SICK TIC102 master operating instructions

SICK

SICK TIC102 master operating instructions

Hioki LR5051 instruction manual

Hioki

Hioki LR5051 instruction manual

Seneca Z-LOGGER3 user manual

Seneca

Seneca Z-LOGGER3 user manual

Thies CLIMA DLU Instructions for use

Thies CLIMA

Thies CLIMA DLU Instructions for use

MadgeTech X Series Product user guide

MadgeTech

MadgeTech X Series Product user guide

Perfect Prime TH0165 manual

Perfect Prime

Perfect Prime TH0165 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.