Onset HOBO UX100-011A User manual

HOBO® Temp/RH 2.5% Data Logger (UX100-011A) Manual
22931-A
The HOBO Temp/RH data logger records temperature and relative humidity (within 2.5%
accuracy) in indoor environments with its integrated sensors. Using HOBOware®, you can easily
configure the logger alarm to trip for specific high or low sensor readings. Or, you can set up burst
logging in which the logger records data at a different interval during certain conditions. The
logger can also calculate minimum, maximum, average, and standard deviation statistics. This
compact data logger has a built-in LCD screen to monitor the current temperature, relative
humidity, logging status, battery use, and memory consumption in between readouts.
Specifications
Temperature Sensor
Range -20° to 70°C (-4° to 158°F)
Accuracy ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
Resolution 0.024°C at 25°C (0.04°F at 77°F), see Plot A
Response Time 4 minutes in air moving 1 m/s (2.2 mph)
Drift <0.1°C (0.18°F) per year
RH Sensor
Range 1% to 95% (non-condensing)
Accuracy ±2.5% from 10% to 90% typical to a maximum of ±3.5% including
hysteresis at 25°C (77°F); below 10% and above 90% ±5% typical
Resolution 0.01%
Response Time 10 seconds to 90% in airflow of 1 m/s (2.2 mph)
Drift <1% per year typical
Logger
Logger Operating Range Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing)
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification
Logging Rate 1 second to 18 hours, 12 minutes, 15 seconds
Logging Modes Fixed interval (normal), burst, or statistics
Memory Modes Wrap when full or stop when full
Start Modes Immediate, push button, date & time, or next interval
Stop Modes When memory full, push button, or date & time
Restart Mode Push button
Time Accuracy ±1 minute per month at 25°C (77°F), see Plot B
Power Source One 3V CR2032 lithium battery and USB cable
Battery Life 1 year, typical with logging rate of 1 minute and sampling interval of
15 seconds or greater
Memory 128 KB (63,488 measurements, maximum)
Download Type USB 2.0 interface
Full Memory Download Time 20 seconds
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react
slowly or go blank in temperatures outside this range
Size 3.66 x 8.48 x 2.29 cm (1.44 x 3.34 x 0.9 in.)
Weight 30 g (1.06 oz)
Environmental Rating IP50
The CE Marking identifies this product as complying with all relevant
directives in the European Union (EU).
HOBO Temp/RH 2.5%
Data Logger
UX100-011A
Included Items:
•Command™ strip
•Double-sided tape
•Hook & loop strap
Required Items:
•HOBOware 3.7.16 or later
•USB cable
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

HOBO Temp/RH 2.5% Data Logger (UX100-011A) Manual
2
Specifications
(continued)
Logger Components and Operation
Start/Stop Button: Press this button for 3 seconds to start or
stop logging data, or to resume logging on the next even logging
interval. This requires configuring the logger in HOBOware with a
push button start or stop, and with “Resume logging on next
button push” selected (see Setting up the Logger). You can also
press this button for 1 second to record an internal event (see
Recording Internal Logger Events) or to turn the LCD screen on if
the option to turn off the LCD has been enabled (see Setting up
the Logger).
Battery Tray: Remove the battery tray on the top of the logger
to access the logger battery (see Battery Information).
Alarm/Stats Button: Use this button to clear a tripped alarm
(see Setting up Alarms) or to switch between statistics, alarm
readings, and the current sensor readings.
Mounting Loops: Use the two mounting loops to mount the
logger with the hook-and-loop strap (see Mounting the Logger).
Temperature Sensor: This sensor is located to the right of the RH
sensor.
RH Sensor: The RH sensor is housed in the cap to the right of the
LCD screen.
USB Port: Use this port (not visible in the diagram) to connect
the logger to the computer or the HOBO U-Shuttle via USB cable
(see Setting up the Logger and Reading Out the Logger).
LCD Screen: This logger is equipped with an LCD screen that
displays details about the current status. This example shows all
symbols illuminated on the LCD screen followed by definitions of
each symbol in the table below.
LCD Symbol Description
The logger is waiting to be launched. Press and
hold the Start/Stop button for 3 seconds to launch
the logger.
The logger has been launched with a push button
stop enabled; press and hold the Start/Stop
button for 3 seconds to stop the logger. Note: If
you also launched the logger with a push button
start, this symbol will not appear on the display for
30 seconds.
The battery indicator shows the approximate
battery power remaining.
The logger has been configured to stop logging
when memory fills. The memory bar indicates the
approximate space remaining in the logger to
record data. When first launched, all five segments
in the bar will be empty. In this example, the
logger memory is almost full (only one segment in
the memory bar is empty).
The logger has been configured to never stop
logging (wrapping). The logger will continue
recording data indefinitely, with newest data
overwriting the oldest data. When first launched,
all five segments in the memory bar will be empty.
In this example, the memory is full (all five
segments are filled in) and new data is now
overwriting the oldest data. This will continue until
the logger is stopped or the battery runs out.
The logger is currently logging.
Plot A: Temperature
Accuracy and Resolution
Plot B: Time Accuracy
Alarm/Stats Button
Start/Stop Button
USB Port
Mounting
Loop
Battery Tray
Mounting
Loop
LCD Screen
RH Sensor
Temperature
Sensor

HOBO Temp/RH 2.5% Data Logger (UX100-011A) Manual
3
LCD Symbol Description
A sensor reading is above or below the high or low
alarm that you configured. Press and release the
Alarm/Stats button until the “alm” symbol
(described below) is displayed on the screen. This
symbol at left will clear depending on how alarms
were configured in HOBOware. If the alarm was
configured to clear when the logger is relaunched,
this symbol will remain on the LCD. Otherwise, it
will clear when the sensor reading is back within
the alarm limits or by pressing the Alarm/Stats
button for 3 seconds.
An alarm is ready to be cleared. This will only
appear if “Cleared with button press” was selected
in the HOBOware alarm settings. Press the
Alarm/Stats button for 3 seconds to clear the
alarm.
These symbols show the maximum, minimum,
average, and standard deviation values most
recently calculated for the logger (if the logging
mode has been set to Statistics in HOBOware).
Press the Alarm/Stats button for 1 second to cycle
through each of the available statistics and then
back to the current sensor reading (or to the alarm
value if applicable).
This is the sensor reading that tripped the alarm.
Press the Alarm/Stats button to view this reading.
Press the Alarm/Stats button again to cycle
through any statistics (defined above) and
ultimately back to the current sensor reading.
This is an example of a temperature reading.
Temperature units are determined by the settings
in HOBOware. To switch between Celsius and
Fahrenheit, change the Display Preferences in
HOBOware before launching the logger.
This is an example of an RH reading.
The logger has been configured to start logging on
a particular date/time. The display will count down
in days, hours, minutes, and seconds until logging
begins. In this example, 5 minutes and 38 seconds
remain until logging will begin.
The launch settings are being loaded onto the
logger from HOBOware. Do not disconnect the
USB cable during this process.
An error occurred while loading the launch
configurations onto the logger from HOBOware.
Make sure the USB cable is connected to both the
logger and the computer and try launching again.
The logger has been stopped with HOBOware or
because the memory is full.
Notes:
•You can disable the LCD screen when logging. Select “Turn
LCD off” when setting up the logger as described in the
next section. When this option is enabled, you can still
temporarily view the LCD screen by pushing the Start/Stop
button for 1 second. The LCD will then remain on for 10
minutes.
•The LCD screen refreshes every 15 seconds regardless of
the logging interval selected in HOBOware. If you choose a
logging interval less than 15 seconds, the data will be
recorded at the faster interval, but the sensor readings will
only be updated on the screen every 15 seconds.
•When the logger has stopped logging, the LCD screen will
remain on until the logger is offloaded to a computer or
HOBO U-Shuttle (unless launched with the “Turn LCD off”
option). Once the logger has been offloaded and
disconnected from the computer, the LCD will turn off
automatically after 2 hours. The LCD will turn back on the
next time the logger is connected to the computer.
Setting up the Logger
Use HOBOware to set up the logger, including setting alarms,
selecting the options to start and stop logging, and choosing a
logging mode.
1. Connect the logger and open the Launch Logger window. To
connect the logger to a computer, use the cable provided.
Click the Launch icon on the HOBOware toolbar or select
Launch from the Device menu.
Important: USB 2.0 specifications do not guarantee operation
outside the range of 0°C (32°F) to 50°C (122°F).
2. Select the sensors to log in this deployment. Choose
Temperature or Temperature and RH. Type a label for the
sensor(s) if desired. Note that both sensors are required to
calculate dew point, which is an additional data series
available for plotting after reading out the logger.
3. Set up alarms (optional). Click the Alarms button if you want
to configure an alarm to trip when the sensor reading is
above or below a value you specify. See Setting up Alarms for
details.
4. Configure filters (optional). Click the Filters button to create
additional filtered data series. Any filtered series will be
available automatically upon reading out the logger.
5. Select the Logging Interval. Select a logging interval from 1
second to a maximum of 18 hours, 12 minutes, and 15
seconds.
6. Select the Logging Mode:
•Fixed Interval. In Fixed Interval mode, data will always be
recorded at the regular logging interval set in the previous
step. This is the default setting.
•Burst. In Burst mode, logging will occur at a different
interval when a specified condition is met. See Burst
Logging for more information.

HOBO Temp/RH 2.5% Data Logger (UX100-011A) Manual
4
•Statistics. In Statistics mode, maximum, minimum,
average, and standard deviation statistics are calculated
for all enabled sensors during logging, sampling at an
interval you specify. See Statistics for more information.
7. Choose when to start logging:
•Now. Logging begins immediately.
•At Interval. Logging will begin at the next even interval as
determined by the selected logging interval.
•On Date/Time. Logging will begin at a date and time you
specify.
•Push Button. Logging will begin once you press the
Start/Stop logging button for 3 seconds.
8. Choose when to stop logging:
•When Memory Fills. Logging will end once the logger
memory is full.
•Never (Wrap When Full). The logger will continue
recording data indefinitely, with newest data overwriting
the oldest. This option is not available if Burst is selected
for the Logging Mode.
•Push Button. Logging will end once you press the
Start/Stop logging button for 3 seconds. Note that if you
also choose Push Button to start logging, then you will not
be able to stop logging until 30 seconds after logging
begins.
If you select the Push Button setting, then you also have
the option to select “Allow button restart.” This allows you
to stop and then restart logging during the deployment by
pushing the Start/Stop button on the logger for 3 seconds.
Important: When “Allow button restart” is selected and you
use the Start/Stop button to stop and restart logging, logging
will restart on the next even logging interval, not at the time
the button was pushed. For example, a logger started logging
at 7:00 AM with a logging interval set to 1 hour. If you press
the Start/Stop button to stop the logger at 8:45 AM and then
press the button again at 10:15 AM, logging will not begin
immediately at 10:15. Instead, logging will begin again at
11:00 AM, which is the next even interval time based on your
1-hour logging interval. Therefore, depending on the logging
interval, the gap between the time you press the button to
resume logging and the time actual logging begins could be
significant. The faster the logging interval, the less time will
elapse before logging resumes.
•Specific Stop Time. Logging will end at a date and time you
specify. Note that if you also configure the logger for a
Push Button stop and to “Allow button restart,” then the
logger will stop logging at the date you select regardless of
how many times you stop and restart the logger with the
Start/Stop button.
9. Choose whether to keep the LCD on or off. By default, the
LCD will always remain on while logging. If you select the
“Turn LCD off” checkbox, the LCD will not show the current
readings, status, or other information while the logger is
logging. You will, however, be able to temporarily turn the
LCD screen on by pressing the Start/Stop button for 1 second
if you select this option.
10. Click the Start button to launch the logger. Disconnect the
logger from the computer and deploy it using the mounting
materials (see Mounting the Logger). After logging begins,
you can read out the logger at any time (see Reading Out the
Logger for details).
Setting up Alarms
You can set an alarm to trip on the logger when a sensor reading
rises above or falls below a specified value. To set an alarm:
1. Click the Alarms button from the Launch Logger window. If
the Alarms button is disabled, make sure the Logging Mode is
not set to Burst. (Alarms can only be configured if the logger
is in Normal or Statistics mode.)
2. Select a sensor. In this example, the temperature sensor was
selected.
3. Select the High Alarm checkbox if you want an alarm to trip
when the sensor reading rises above the high alarm value.
Type the reading next to the High Alarm checkbox or drag
the red upper slider in the Configure Alarms window.
4. Select the Low Alarm checkbox if you want an alarm to trip
when the sensor reading falls below the low alarm value.
Type the reading next to the Low Alarm checkbox or drag the
blue lower slider.
5. Set the duration before an alarm is tripped.
6. Select either Cumulative or Consecutive. If you select
Cumulative, then the alarm will trip when the time the
sensor is out of range over the course of the deployment is
equal to the selected duration. If you select Consecutive,
then the alarm will trip when the time the sensor is
continuously out of range is equal to the selected duration.
For example, the high alarm for temperature is set to 85°F
and the duration is set to 30 minutes. If Cumulative is
selected, then an alarm will trip once a sensor reading has
been at or above 85°F for a total of 30 minutes since the
logger was configured; specifically, this could be 15 minutes
above 85°F in the morning and then 15 minutes above 85°F
again in the afternoon. If Consecutive is selected, then an
alarm will trip only if all sensor readings are 85°F or above for
a continuous 30-minute period.
7. Repeat steps 2 through 6 for the other sensor if desired.
8. Choose how long the logger should maintain a sensor alarm
once it has tripped. Select “Host has relaunched logger” if
you want the alarm to remain visible on the LCD until the

HOBO Temp/RH 2.5% Data Logger (UX100-011A) Manual
5
next time you relaunch the logger. Select “Sensor reading
within limits” if you want the alarm to clear once the sensor
reading returns to the normal range between the high and
low alarm limits. Select “Cleared with button press” if you
want the alarm to remain on until you press the Alarm/Stats
button on the logger.
9. Click OK to save alarm settings.
Notes:
•Once the logger is launched, alarms will trip as determined
by these settings. Logger alarms will display on the LCD
screen. Note that the alarm limits are only checked when
the logger’s LCD screen refreshes every 15 seconds.
•The actual values for the high and low alarm limits are set
to the closest value supported by the logger. For example,
the closest value to 85°F that the UX100 series logger can
record is 84.990°F and the closest value to 32°F is
32.043°F. In addition, alarms can trip or clear when the
sensor reading is within the logger specifications of 0.02°C
resolution. This means the value that triggers the alarm
may differ slightly than the value entered. For example, if
the High Alarm is set to 75.999°F, the alarm can trip when
the sensor reading is 75.994°F (which is within the 0.02°C
resolution).
•When you read out the logger, high and low alarm levels
will be displayed on the plot along with “Chan <#> Alarm
Tripped” and “Chan <#> Alarm Cleared” events showing
when the sensor alarm tripped and cleared. The “Chan <#>
Alarm Cleared” event contains the value that was furthest
out of range for the sensor before the alarm cleared (see
the Points table for the actual value).
Burst Logging
Burst logging is a logging mode that allows you to set up more
frequent logging when a specified condition is met. For example,
let’s say the logger is recording data at a 5-minute logging
interval and burst logging is configured to log every 10 seconds
when the temperature goes above 85°F (the high limit) or falls
below 32°F (the low limit). This means the logger will record data
every 5 minutes as long as the temperature remains between
85°F and 32°F. Once the temperature reaches 90°F, for example,
the logger will switch to the faster logging rate and record data
every 10 seconds until the temperature falls back below the high
limit (or 85°F in this case). At that time, logging then resumes
every 5 minutes at the normal logging interval. Similarly, if the
temperature falls to 30°F, for example, then the logger would
switch to burst logging mode again and record data every 10
seconds. Once the temperature rises back to 32°F, the logger will
then return to normal mode, logging every 5 minutes.
To set up burst logging:
1. Select Burst for Logging Mode in the Launch Logger window.
If Burst has already been configured for this logger, click the
Edit button in the Launch Logger window. Note that you
cannot set up alarms with burst logging. You also cannot
select the Stop Logging option “Never (wrap when full)” if
burst logging is configured.
2. In the Burst Logging window, select a sensor. In the example
at right, the temperature sensor was selected.
3. Select the High Limit checkbox if you want to set up a
condition in which burst logging will occur when the sensor
reading rises above the high limit value. Type in the value or
drag the red upper slider (see example below).
4. Select the Low Limit checkbox if you want to set up a
condition in which burst logging will occur when the sensor
reading falls below the low limit value. Type in the value or
drag the blue lower slider.
5. Repeat steps 2 through 4 for the other sensor, if desired.
6. Set the burst logging interval, which must be less than the
logging interval. Select either a preset burst logging interval
or select Custom and enter your own interval. Keep in mind
that the more frequent the burst logging rate, the greater
the impact on battery life and the shorter the logging
duration.
7. Click OK when done. This will return you to the Launch
Logger window. Click the Edit button next to Logging Mode in
the Launch Logger window to make additional changes.
Notes:
•Once the logger is launched, the high and low burst
logging limits are only checked when the logger's
LCD screen refreshes once every 15 seconds. Therefore, if
you set the logging interval to less than 15 seconds and
the sensor reading falls outside the limits, the burst
logging will not begin until the next 15-second refresh
cycle.
•If high and/or low limits have been configured for more
than one sensor, then burst logging will begin when any
high or low condition goes out of range. Burst logging will
not end until all conditions on all sensors are back within
normal range.
•The actual values for the burst logging limits are set to the
closest value supported by the logger. For example, the
closest value to 85°F that the logger can record is 84.990°F
and the closest value to 32°F is 32.043°F.
•Burst logging mode can begin or end when the sensor
reading is within the logger specifications of 0.02°C
resolution. This means the value that triggers burst logging
may differ slightly than the value entered. For example, if
the high limit for a temperature alarm is set to 75.999°F,
burst logging can start when the sensor reading is
75.994°F (which is within the 0.02°C resolution).
•Once the high or low condition clears, the logging interval
time will be calculated using the last recorded data point
in burst logging mode, not the last data point recorded in

HOBO Temp/RH 2.5% Data Logger (UX100-011A) Manual
6
“normal mode.” For example, let’s assume the logger has
a 10-minute logging interval and logged a data point at
9:05. Then, the high limit was surpassed and burst logging
began at 9:06. Burst logging then continued until 9:12
when the sensor reading fell back below the high limit.
Now back in normal mode, the next logging interval will be
10 minutes from the last burst logging point, or 9:22 in this
case. If burst logging had not occurred, the next data point
would have been at 9:15.
•A New Interval event will appear on the plot (if you select
events for plotting in the Plot Setup window) each time
the logger enters or exits burst logging mode.
Statistics
Statistics is a logging mode in which the logger calculates
maximum, minimum, average, and standard deviation statistics
during logging, recording the results at each logging interval
based on samples taken at a rate you specify. This will result in
up to four additional series per sensor that record the following
information at each logging interval:
•The maximum, or highest, sampled value,
•The minimum, or lowest, sampled value,
•An average of all sampled values, and
•The standard deviation from the average for all sampled
values.
For example, let’s say both the temperature and RH sensors
have been enabled, the logging interval is set to 5 minutes and
the sampling interval is set to 30 seconds (with maximum,
minimum, average, and standard deviation all enabled). Once
logging begins, the logger will measure and record the actual
temperature and RH sensor values every 5 minutes. In addition,
the logger will take a temperature and RH sample every 30
seconds and temporarily store them in memory. The logger will
then calculate the maximum, minimum, average, and standard
deviation using the samples gathered over the previous 5-
minute period and log the resulting values. When reading out
the logger, this would result in 10 data series (not including any
derived series, such as dew point): two sensor series (with
temperature and RH data logged every 5 minutes) plus eight
maximum, minimum, average, and standard deviation series
(four for temperature and four for RH with values calculated and
logged every 5 minutes based on the 30-second sampling).
To set up statistics:
1. Select Statistics for Logging Mode in the Launch Logger
window. If Statistics has already been configured for this
logger, click the Edit button in the Launch Logger window.
2. Click the Maximum, Minimum, Average, and Standard
Deviation checkboxes for each of the statistics you want to
calculate during logging. Note that Average is automatically
enabled when selecting Standard Deviation. Important:
Statistics apply to all enabled sensors; every selected statistic
will be calculated for all sensors (except battery voltage). For
example, if both the temperature and RH sensors have been
selected in the Launch Logger window and you select
Average, then the average will be calculated for both
temperature and RH. In addition, the more statistics you
record, the shorter the logger duration and the more
memory is required.
3. Set the sampling interval, which must be less than and a
factor of the logging interval. Choose either a preset
sampling interval or select Custom and enter your own
sampling interval. Keep in mind that the more frequent the
sampling rate, the greater the impact on battery life.
4. Click OK when done. This will return you to the Launch
Logger window. Click the Edit button next to Logging Mode in
the Launch Logger window to make additional changes.
Once logging begins, click the Alarm/Stats button on the logger
to cycle through the current maximum, minimum, average, and
standard deviation data on the LCD screen. You can plot the
statistics series once you read out the logger.
Reading Out the Logger
There are two options for reading out the logger: connect it to
the computer with a USB cable and read out it with HOBOware,
or connect it to a HOBO U-Shuttle (U-DT-1, firmware version
1.18m030 or higher) and then offload the data files from the
U-Shuttle to HOBOware. Refer to the HOBOware Help for more
details.
Recording Internal Logger Events
The logger records the following internal events to track logger
operation and status. You can plot these events in HOBOware
after reading out the logger and opening the data file.
Internal Event Name Definition
Host Connected The logger was connected to the computer.
Started The Start/Stop button was pressed to begin
logging.
Stopped The logger received a command to stop
recording data (from HOBOware or by
pushing the Start/Stop button).
Button Up/Button
Down
The Start/Stop button was pressed for
1 second.
Chan <#> Alarm
Tripped
A sensor alarm has tripped; # is the channel
or sensor number as shown in the Launch
Logger window in HOBOware.
Chan <#> Alarm
Cleared
A sensor alarm has cleared; # is the channel
or sensor number as shown in the Launch
Logger window in HOBOware. This event
also contains the value that was furthest
out of range for the sensor before the
alarm cleared.
New Interval The logger has entered or exited burst
logging mode.
Safe Shutdown The battery level dropped below 2.5 V; the
logger performs a safe shutdown.

HOBO Temp/RH 2.5% Data Logger (UX100-011A) Manual
7
Mounting the Logger
There are several ways to mount the logger using the materials
included:
•Use the four built-in magnets on the back of the logger to
mount it to a magnetic surface.
•Attach the Command strip to the back of the logger to
mount it a wall or other flat surface.
•Use the double-sided tape to affix the logger to a surface.
•Insert the hook-and-loop strap through the mounting
loops on both sides of the logger to mount it to a curved
surface, such as a pipe or tubing.
Protecting the Logger
The logger is designed for indoor use and can be permanently
damaged by corrosion if it gets wet. Protect it from
condensation. If the message FAIL CLK appears on the LCD
screen, there was a failure with the internal logger clock possibly
due to condensation. Remove the battery immediately and dry
the circuit board.
Note: Static electricity may cause the logger to stop logging.
The logger has been tested to 8 KV, but avoid electrostatic
discharge by grounding yourself to protect the logger.
Replacing the RH Sensor
The internal RH sensor may require replacement if it is damaged
mechanically or by exposure to condensation or chemicals. The
replacement sensor (HUM-RHPCB-3C) is available from Onset. To
replace the sensor:
1. Remove the cap from the logger case with a one quarter turn
counter-clockwise.
2. Pull to remove the RH sensor from the board inside the
logger. Discard the sensor.
3. Wearing gloves or finger cots to prevent contamination or
damage, hold the replacement sensor by the sides of the
board only. Do not touch the actual sensor (the small black
square). Touching the sensor may compromise accuracy.
Insert the connector on the sensor board into the pins on the
board in the logger. Make sure the RH sensor is oriented with
the flat part of the board pointed down as shown below and
gently insert the connector into the pins on the logger board.
4. Place the cap over the RH sensor lining up the notches on the
cap with the grooves on both sides of the sensor. Secure the
cap with a one quarter turn clockwise until it snaps back in
place.
5. Check logger status in HOBOware to verify the RH reading.
Battery Information
The logger contains a 3V CR2032 battery (HRB-TEMP). Expected
battery life varies based on the ambient temperature where the
logger is deployed, the logging or sampling interval, frequency of
offloading to the computer, number of channels that are active,
if burst or statistics logging modes are active, and battery
performance. A new battery typically lasts 1 year with logging
intervals greater than 1 minute. Deployments in extremely cold
or hot temperatures, a logging interval faster than 1 minute, or a
sampling interval faster than 15 seconds can impact battery life.
Estimates are not guaranteed due to uncertainties in initial
battery conditions and operating environment.
The logger can also be powered by the USB cable when the
remaining battery voltage is too low for it to continue logging.
Connect the logger to the computer, click the Readout button on
the toolbar, and save the data as prompted. Replace the battery
before launching the logger again.
Cap
Pull to
remove
old RH
sensor
WARNING: Wearing gloves or
finger cots, touch the sides of
the board only; do not touch
the sensor (small black square)
Insert this connector
into pins on logger Insert the RH sensor
connector into these pins
When installed correctly, the flat part
of the RH sensor board is pointed down
and the black sensor is facing up
Do not touch
the sensor
Touch the
sides of the
board only

HOBO Temp/RH 2.5% Data Logger (UX100-011A) Manual
© 2013–2018 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOware are
trademarks or registered trademarks of Onset Computer Corporation. All other trademarks are the property
of their respective companies.
22931-A
To replace the battery:
1. Holding the logger with the LCD screen facing up, pull the
battery tray out of the logger housing.
2. Remove the old battery from the tray.
3. Place the new battery in the tray with the positive side
facing down.
4. With the LCD screen still facing up, slide the tray back into
the logger. The LCD should display “HOBO” briefly after
correctly installing the battery.
WARNING: Do not cut open, incinerate, heat above 85°C
(185°F), or recharge the lithium battery. The battery may
explode if the logger is exposed to extreme heat or conditions
that could damage or destroy the battery case. Do not dispose
of the logger or battery in fire. Do not expose the contents of
the battery to water. Dispose of the battery according to local
regulations for lithium batteries.
HOBOware provides the option of recording the current battery
voltage at each logging interval, which is disabled by default.
Recording battery life at each logging interval takes up memory
and therefore reduces logging duration. It is recommended you
only record battery voltage for diagnostic purposes.
Battery tray removed
from logger
CR2032 battery being
placed in the tray,
positive side facing
down
Table of contents
Other Onset Data Logger manuals

Onset
Onset HOBO U20L User manual

Onset
Onset HOBO H22 User manual

Onset
Onset HOBO UX120-017 User manual

Onset
Onset MX2501 User manual

Onset
Onset MX1102A User manual

Onset
Onset HOBO U-DT-1 User manual

Onset
Onset HOBO User manual

Onset
Onset HOBO MX CO2 Logger User manual

Onset
Onset InTemp CX450 User manual

Onset
Onset HOBO U14 User manual

Onset
Onset HOBO RG3 User manual

Onset
Onset HOBO ZW-008 User manual

Onset
Onset Hobo UX100-023A User manual

Onset
Onset HOBO MX2001 User manual

Onset
Onset Hobo MicroRX RX2105 User manual

Onset
Onset U26-001 User manual

Onset
Onset HOBO U30 GSM User manual

Onset
Onset HOBO MX Series User manual

Onset
Onset HOBO U24 User manual

Onset
Onset EG4100 Series User manual