Opti-Med JM026-001 User manual

Opti-Med ®
LED Examination Light
JM026-001
USER MANUAL
Hangzhou Jupiter Medical Equipment Co., Ltd
1116, Gudun Road, West Lake District, Hangzhou, China
Tel: +86-15021634329 Web: www.opti-med.net
Email: [email protected]

Contents
I. Brief introduction
II. Structure and working theory
III. Features and technical data
IV. Installation and Modulation
V. Operating instruction
VI. Service
VII. Safety
VIII. Trouble shooting and Maintenance
IX. Circumstance for transportation and storage
X. Packing List
Opti-Med ®
www.opti-med.net

www.opti-med.net
1
Opti-Med ®
I.
I. Brief
Brief Introduction
Introduction
Thanks for using LED examination light. Because of the international advanced LED cold
illuminant, novel design and advanced structure. JM026-001 LED examination light are
widely used for various kinds of operation. To ensure correct and safe use of our products,
please read this manual carefully before installation and operation.
This manual introduces the using ways of product and the technical specification.
Operating requirement:
Equipment Safety Type: Category I. Type B
Working system: continuous working
Circumstance requirement:
Ambient relative temperature: 10℃ ~ 30℃
Ambient relative humidity: 10% ~ 80%
Normal working conditions:
Mains voltage: ~220V ± 10% Frequency: 60HZ ± 1%
Voltage of Bulb: DC 24V
II.
II. Structure
Structure and
and W
Working
orking T
Theory
heory
The operation lamp is divided into three parts: Base; Balancer; Lamp head.
The balancer works by means of the elastics force that arises from the compression on the
spring.
The elastic force then becomes the balancing force to balance the lamp head at different
positions, thus the lamp head can be moved up and down with great flexibility. At the distance
of 2.2 m to 2.3m between the lamp head and the ground, the operation lamp are used for
various kinds of surgical operations as thoracic surgery, abdomen surgery and cerebral surgery
department, etc. And at the distance of 1.1m, used for gynecology department, etc. needs
assistant illumination.
The light source adopts high luminous efficiency, good color rendering, long service life and
low thermal LED light emitting diode.And the lamp head of JM026-001 series consist of
more light lamp, and each light source adopts multiple LED light emitting diode to produce
cold light beam which reflect to surgical wound from every Angle to make the operative site
to the effect of illumination
III. Features
Features and
and T
Technical
echnical D
Data
ata
A. Features
With the development of three-dimensional technical design, the structure and model tend to be more

www.opti-med.net
2
Opti-Med ®
excellent, more reasonable.
Advanced lighting system of integral reflection makes the luminance of the lamp reach 35,000LUX.
Illumination can be adjusted steplessly and be memorized. The focal depth can be adjusted at will.
B. Technical data
Input voltage
~ 220V
Voltage of bulb
DC 24V
Luminance
≥35000Lux
Input power
60VA
Color temperature
3500~5000K
Spot diameter
150~300(mm)
Circumstance for
Storage and transportation
Temperature
-40℃ ~ +55℃
Humidity
10% ~ 90%
Fuse
RT1-20 T2A L250V RF1-20 F5A L250V
Rated power of bulb
3W×19
Our company reserves the right of the change of product. Any models and specification are
subject to change without notice.
IV.
IV. Installation
Installation and
and Modulation
Modulation
Please install and modulate the lamp according to Diagram.
Check items and Check ways:
1. Turn on the switch to check if the lamp is on normal condition or not.
2. Turn the light-adjusting knob on the lamp head to check if the brightness is normal.
3. Turn round each joint to check whether flexible
4. Turn the focus knob on the lamp head to check if the focus is on normal condition or not.
V.
V. Operating
Operating instructions
instructions
1. Preparation and Check
Turn on the power to check whether the light source works well.
2. Procedure
Connect the electric wire plug with the electric wire socket and turn on ‘Power’
switch on the base of the operation lamp.
Adjust the positions of the base, balancing arm and lamp head to ideal using
positions.
Focus the light source to the operation field by rotating the focus knob
Turn the light-adjusting knob on the operation lamp head to desired illumination.
Switch off the power on the base of the lamp after operation.

www.opti-med.net
3
Opti-Med ®
Notes
The power should be switched off after operation, otherwise, the transformer will burnt
out due to the condition of power on for a long time.
JM026-001 integral reflection operation lamp, its tilt degree must be less than or equal
to 10°.
VI.
VI. Services
Services
1. Annually periodic inspection should be conducted by specialized personnel to see
whether the connection of each component is loose. If so, please tighten them
immediately.
2. If the lamp is out of order, please contact our company as soon as possible, so that our
specialized maintenance workers can provide timely service, otherwise, users will bear
responsibility for damages arising therefore.
3. Our company guarantees the lamp for a period of one year from date of purchase
VII.
VII. Safety
Safety
1. Periodically ( one time per year is recommended ) check the connection status of each
joint of the lamp.
2. When install, connect Input wire or Output wire according to color code strictly to
avoid confusing. ( Operator must possess the basic professional knowledge ).
VI.
VI. Troubleshooting
Troubleshooting and
and Maintenance
Maintenance
A. Common trouble, Causes, Dealing way
Trou ble
Ca u se
De al in g wa ys
Light goes out
Source power fails to supply electricity
Wait for supplying electricity
Fuse burns out
Cut off power supply&replace fuse
Transformer burns out
Cut off power supply &replace
transformer
Printed circuit board burns out
Replace board
Indicator glows on the
control panel
Halogen bulb burns out
Replace a new bulb
Position fails to lock
Uneven base
Adjust the wheel place
Wrong installation way
Consult specialized personnel
Mechanical parts injured
Consult specialized personnel
Light source fails to focus
Focusing device loose
Consult specialized personnel
Balance device fails to lock
up / down position
Wrongly adjust the balance device
Consult specialized personnel
Spring of the balance device lose effect
Consult specialized personnel
Opti-Med ®

www.opti-med.net
4
Opti-Med ®
B. Repair
How to replace a fuse:
Cut off the power. Remove the cover by screwing out screws by a “ + ”screwdriver. Take the
burnt fuse out of the connector at the base and replace a new one with the
same specification.
Specialist is needed to replace LED bulb and switch power.
C. Maintenance
Periodically inspect the connection status of each screw of the lamp ( including joint between
the base and the balancer; the lamp head and the balancer ) and tighten the loose screws in time.
One -every-two-month –inspection is recommended.
Keep cleaning the lamp with soft cloths to prevent dust accumulation.
The glass cover on the lamp head should be cleaned with the glass detergent.
IX. Circumstance
Circumstance for
for transportation
transportation and
and storage
storage
Relative humidity: 10%~90%
Ambient temperature: -40℃-70℃
Room with good ventilation and without corrosive gas
Marks for transportation: Keep dry; Handle with care; Do not turn over
X.
X. Packing
Packing list
list
Lamp head ----------------------------------- 1pc
Base ------------------------------------ 1pc
Balancer --------------------------------------1pc
Fuse ------------------------------------ 2pcs
Spanner ----------------------------------- 1pc
Upper pole --------------------------------- 1pc
Lower pole --------------------------------- 1pc
USER’S MANUAL --------------------------1pc
Certificate of test -----------------------------1pc
Before installation and modulation, firstly unpack and inspect if the outward appearance of the
lamp is good; secondly check Model, Components, Quantity according to packing list. If there
is anything missing, please contact us immediately.

1 2 3 4
A
B
C
D
4321
D
C
B
ATitle
Nu mber R ev i sio nS ize
A4
Date: 2 8-Sep-2 0 17 Sheet o f
File: E:\说明 书\2 01 1 版说 明书中文 \1 60 简易 式\JSL160 D1 .DDBDrawn By:
D1RD1
2A
L
AC 220V
Control Circuitry Board
DC 24V
JSL2009-160D
_
+
LAMP
LAMP
2 4V/ 75 W
+
-
K
Opti-Med ®
www.opti-med.net

Hangzhou Jupiter Medical Equipment Co., Ltd
1116, Gudun Road, West Lake District, Hangzhou, China
Tel: +86-15021634329 Web: www.opti-med.net
Email: [email protected]
Opti-Med ®
Table of contents
Popular Lighting Equipment manuals by other brands

Lightmybricks
Lightmybricks LEGO UCS RAZOR CREST installation guide