ORTEC 264 User manual

i
r
s
r
INSTRUCTION
MANUAL
264
PHOTOMULTIPLIER
TIMING
DISCRIMINATOR
AND
PREAMPLIFIER

1
i
INSTRUCTION
MANUAL
MODEL
264
PHOTOMULTIPLIER
TIMING
DISCRIMINATOR
AND
PREAMPLIFIER
Serial
No.
33
Purchaser
Date
Issued
P.
0.
BOX
C
OAK
RIDGE,
TENNESSEE
37830
Telephone
(615)
483-8451
TWX
810-572-1078
)
ORTEC
Incorporated
1966
Printed
in
U*S*A«

TABLE
OF
CONTENTS
Page
WARRANTY
PHOTOGRAPH
1.
DESCRIPTION
1-1
2.
SPECIFICATIONS
2-1
3.
INSTALLATION
INSTRUCTIONS
3-1
3.1
Det-ector
Mounting
3-1
3.2
Photomultlplier
Insertion
3-1
3.3
System
Connection
and
Initial
Adjustment
3-1
4.
OPERATING
INSTRUCTIONS
4-1
4.1
Timing
with
Photomultipliers
4-1
4.2
Applications
4-2
5.
CIRCUIT
DESCRIPTIONS
5-1
5.1
Voltage
Divider
Network
5-1
5.2
Linear
Preamplifier
5-1
5.3
Discriminator
Circuit
5-1
6.
MAINTENANCE
INSTRUCTIONS
6-1
6.1
Divider
Network
6-1
6.2
Preamplifier
6-1
6.3
Fast
Crossover
Discriminator
6-1
BLOCK
DIAGRAM
AND
SCHEMATICS
264-0101-Bl
ORTEC
264
Block
Diagram
264-0101-Sl
ORTEC
264
Schematic
264-0201-Sl
ORTEC
264
Schematic
264-0301-Sl
ORTEC
264
Schematic

LIST
OF
FIGURES
Page
Figure
1.
Suggested
Cutaway
Drawing
of
PM
—
Scinti
I
lator
Mounting
2-3
Figure
2.
Incorrect
Adjustment
of
R-26
4-2
Figures.
Correct
Adjustment
of
R-26
4-2
Figure
4.
Incorrect
Adjustment
of
Crossover
Zero
R6
4-3
Figures.
Correct
Adjustment
of
Crossover
Zero
R6
4-3
Figure
6.
Simple
Sub-Nanosecond
Coincidence
System
4-6
Figure
7.
Wide
Dynamic
Range
Timing
Spectrum
Using
"Fast"
Crossover
4-7
Figure
8.
Narrow
Dynamic
Range
Timing
Spectrum
Using
"Fast"
Crossover
4-8
Figure
9.
Timing
Resolution
(Leading
Edge
vs.
"Fast"
Crossover)
4-9
Figure
10.
Open
Unit
Showing
Picture
of
Controls
6-2
SYSTEM
BLOCK
DIAGRAMS
200058
Gamma-Ray
Pair
Spectrometer
4-10
200057
Fast
Timing
System
for
(Semiconductor
Detector-Photo-
multiplier
Tube)
Coincidence
Using
Two
Crossover
Pickoff
Techniques
4-11
200067
Associated
-
Particle
Neutron
Time
of
Flight
System
with
Energy
Selecting
Side
Channels
4-11
200068
Pulse
Shape
Discrimination
System
4-12
200056
Fast-Fast
Coincidence
(Photomultiplier
Tube)
with
Pulse
Shape
Discrimination
4-12
200060
Sub-Nanosecond
Timing
System
(Semiconductor-Photomultiplier
Tube)
4-13

STANDARD
WARRANTY
FOR
ORTEC
ELECTRONIC
INSTRUMENTS
DAMAGE
IN
TRANSIT
Shipments
should
be
examined
immediately
upon
receipt
for
evidence
of
external
or
con
cealed
damage.
The
carrier
making
delivery
should
be
notified
immediately
of
any
such
damage,
since
the
carrier
is
normally
liable
for
damage
in
shipment.
Packing
materials,
waybills,
and
other
such
documentation
should
be
preserved
in
order
to
establish
claims.
After
such
notification
to
the
carrier,
notify
ORTEC
of
the
circumstances
so
that
we
may
assist
in
damage
claims
and
in
providing
replacement
equipment
when
necessary.
WARRANTY
ORTEC
warrants
its
electronic
products
to
be
free
from
defects
in
workmanship
and
materials,
other
than
vacuum
tubes
and
semiconductors,
for
a
period
of
twelve
months
from
date
of
ship
ment,
provided
that
the
equipment
has
been
used
in
a
proper
manner
and
not
subjected
to
abuse.
Repairs
or
replacement,
at
ORTEC
option,
wil
l
be
made
without
charge
at
the
ORTEC
factory.
Shipping
expense
will
be
to
the
account
of
the
customer
except
in
cases
of
defects
discovered
upon
initial
operation.
Warranties
of
vacuum
tubes
and
semiconductors,
as
made
by
their
manufacturers,
will
be
extended
to
our
customers
only
to
the
extent
of
the
manufacturers'
liabi
lity
to
ORTEC.
Specially
selected
vacuum
tubes
or
semiconductors
cannot
be
warranted.
ORTEC
reserves
the
right
to
modify
the
design
of
its
products
without
incurring
responsibility
for
modification
of
previously
manufactured
units.
Since
instal
lation
conditions
are
beyond
our
control,
ORTEC
does
not
assume
any
risks
or
liabilities
associated
with
the
methods
of
installation,
or
instal
lation
results.
QUALITY
CONTROL
Before
being
approved
for
shipment,
each
ORTEC
instrument
must
pass
a
stringent
set
of
quality
control
tests
designed
to
expose
any
flaws
in
materials
or
workmanship.
Permanent
records
of
these
tests
are
maintained
for
use
in
warranty
repair
and
as
a
source
of
statistical
information
for
design
improvements.
REPAIR
SERVICE
ORTEC
instruments
not
in
warranty
may
be
returned
to
the
factory
for
repairs
or
checkout
at
modest
expense
to
the
customer.
Standard
procedure
requires
that
returned
instruments
pass
the
same
qual
ity
control
tests
as
those
used
for
new
production
instruments.
Please
contact
the
factory
for
instructions
before
shipping
equipment.

P3Z
H"®"

1-1
ORTEC
264
PHOTOMULTIPLIER
TIMING
DISCRIMINATOR
AND
PREAMPLIFIER
1.
DESCRIPTION
The
ORTEC
264
Photomultiplier
Timing
Discriminator
and
Preamplifier
consists
of:
1)
a
base
structure
providing
a
mechanical
assembly
and
resistive
voltage
divider
network
with
appropriate
capacity
decoupling
for
the
operation
of
the
RCA
8575
photomultiplier
tube;
2)
a
linear
preamplifier
providing
integration
and
line
drive
capability
for
the
linear
dynode
signal;
and
3)
a
walk-free
timing
discriminator
that
is
remotely
controlled.
This
complex
is
designed
to
operate
with
fast
organic
phosphors,
e.g.,
plastic
and
fast
decay
liquid
phosphors.
It
may
be
readily
altered
to
accommodate
phosphors
having
a
fast
decay
component
time
constant
that
is
rel
atively
long,
i.e.,
5
to
20
nsec.
It
is
not
meant
for
use
with
Nal(TI)
or
Csl(TI).
I

J

2-
1
2.
SPECIFICATIONS
(NOTE:
All
photomultiplier
specifications
ore
given
by
monufocturer.)
Base
Structure:
Hi-Voltage:
Negative,
3
kV
maximum
Bleeder
Current:
2
mA
maximum
(last
4
dynodes
available
externally
for
added
stabilization
when
desired)
Controls:
Focus
electrode,
2nd
Dynode
(D2)
Connectors:
Discriminator
Output:
BNC
Preamplifier
Output;
BNC
Auxiliary;
MS
3112E12-10S
or
Bendix
PT02E12-10S
Hi-Voltage:
Kings
KV-79-13;
mating
connector
KV-59-22,
compatible
with
RG-59/U
coaxial
cable,
available
as
an
accessory.
Preamplifier:
Input:
Signal
current
from
9th
dynode
Input
Impedance:
2.2M
shunted
by
100
pF
Voltage
Gain
(Ay):
2
Output
Signal
Dynamic
Range:
lOV
open
circuit,
5V
on
100
ohms
Integral
Nonlinearity:
=±0.1%
Output
Rise
Time
(Tp):
=50
nsec
Temperature
Stability:
±0.015%/°C
Output
Impedance
(Z^):
100
ohms
Power:
Dc
from
403*or
403A
Time
Pickoff
Control
Unit,
via
power
and
bias
cable
supplied.
Walk-Free
Timing
Discriminator:
Input
Signals:
Anode
Signal,
Dynode
12
signal
Output
Signal:
12
mA
on
50
ohms,
Tp,
<2
nsec,
width»ilO
nsec
*Refer
to
modification
note
on
403
to
403A.

2-2
Discriminator
Control
Range:
1
mA
to
10
mA
of
anode
current
(Specified
with
403A
Time
Pickoff
Control)
Pulse
Pair
Resolving
Time:
Threshold
discriminator
recovers
to
=5%
error
in
25
nsec
Discriminator
Temperature
Stability:
=±0.01
m^°C
(Anode
Current)
Walk
(Time
Shift
vs.
Pulsje
Height):
Typically
^±0.1
nsec
for
100:1
dynamic
range
Power:
Dc
from
403*
or
403A
Time
Pickoff
Control,
via
15
foot
cable,
264-Cl,
supplied
with
base.
The
timing
or
logic
signal
from
the
discriminator
output
will
normally
connect
to
the
input
of
the
403A
which
provides
fan-out
capabilities.
The
linear
signal
from
the
preamplifier
will
normally
connect
to
a
shaping
amplifier
such
as
the
410
or
435.
The
ORTEC
216
(for
56
AVP,
etc.)
or
217
(for
58
AVP
or
XP
1040)
Magnetic
Shield
is
a
recommended
accessory
to
reduce
the
effect
of
the
earth's
magnetic
field
and
of
stray
fields
from
other
equipment.
Power
Required:
(From
403A):
+24V
-
15
mA
+
12V
-
30
mA
-24V-20
mA
-12V-40
mA
*Refer
to
modification
note
on
403
to
403A.
L

2-3
AL
iJM/NUM
30Cf<ET
MAGNLTIC
MYLAR
ALUMINUM
£POy.Y
-^J/^ULAT/QN
^C£MI£N7^
£l£CTR/CAL
TAPE
AAULTIPHER
/0£Y
PER
14
Y£R
TUBE.
^-srvporoAM
socmtet
P^/NC
SCINTILIATINQ
crystal
ALUM/NUM
CLPCUir
Figure
1.
Suggested
Cutaway
Drawing
of
PM
—
Scinti
I
later
Mounting

3-1
3.
INSTALLATION
INSTRUCTIONS
3.1
Detector
Mounting
The
264
is
designed
for
the
best
in
pulse
fidelity
and
requires
that
the
anode
be
operated
at
ground
potential.
This
means
that
the
photocathode
is
at
negative
high
voltage.
Care
must
then
be
taken
to
assure
that
this
high
voltage
is
not
dropped
across
the
glass
envelope
of
the
photomultiplier.
Care
should
be
taken
to
prevent
the
scintiMotor
also
from
imposing
a
ground
at
the
front
surface
of
the
photocathode
to
cause
the
above.
A
drawing
of
a
suggested
method
of
mount
ing
a
simple
detector
is
shown
in
Fig.
1
(Drawing
264-0000).
3.2
Photomultiplier
Insertion
The
218
Magnetic
Shield,
if
used,
should
be removed.
The
tube
may
now
be
inserted
directly
into
the
socket.
CAUTION:
The
socket
end
of
this
photo
multiplier
tube
should
be
made
light
tight
with
the
boot
furnished
by
RCA
or
by
ORTEC.
Place
the
felt
washers
around
the
photomultiplier
and
re-mount
the
magnetic
shield.
3.3
System
Connection
and
Initial
Adjustment
This
unit
is
tuned
up
with
a
specific
photomultiplier
at
the
factory.
The
design
of
the
photomultiplier
ensures
that
there
is
little
variation
from
unit
to
unit
so
that
it
may
not
be
necessary
to
readjust
even
when
changing
from
one
photo
multiplier
to
another
as
long
as
the
high
voltage
for
which
the
tube
is
operated
is
in
the
region
of
-2200
volts
±200
volts.
One
may
assume
that
this
is
the
case
and
skip
Section
3.3.2
in
the
initial
adjustments
procedure.
3.3.1
System
Connection
1.
Connect
the
power
and
bias
control
cable,
264-Cl,
from
the
assoc
iated
403A
to
the
auxiliary
connector
of
the
264.
2.
Connect
the
signal
from
the
discriminator
out
connector
by
way
of
a
50-ohm
coaxial
cable
to
the
input
CN-1
of
the
403A.
3.
Connect
the
preamp
output
signal
to
the
input
of
a
linear
shaping
amplifier
(if
side
channel
energy
selection
is
to
be
used).
4.
Connect
a
high
voltage
power
supply
to
the
high
voltage
input
con
nector
and
ENSURE
that
the
power
supply
is
set
to
deliver
negative

3-2
voltage.
5.
Turn
on
dc
power
to
403A
and
linear
amplifier.
6.
Turn
on
high
voltage
power
supply
and
set
the
power
supply
to
-2200
volts
(or
to
the
voltage
at
which
the
tube
is
to
be
operated).
7.
For
operating
instructions
see
Section
4.0
of
this
manual.
3.3.2
Initial
Adjustments
(Perform
when
necessary)
Before
connecting
into
the
system,
remove
the
high
voltage
divider
cover
to
obtain
access
to
the
controls.
WARNING:
The
voltages
used
in
this
network
are
dangerous,
so
caution
in
adjusting
the
controls
should
be
exercised.
The
controls
for
this
unit
are
trimmed
for
optimum
operation
with
a
specific
photomultiplier
at
the
factory;
however,
if
when
inserting
a
new
photomultiplier,
the
unit
may
need
trimming,
the
following
steps
are
given
as
a
guideline
for
performing
this
operation:
1.
Connect
the
power
and
bias
control
cable,
264-Cl,
from
the
assoc
iated
403A
to
the
auxiliary
connector
of
the
264.
2.
Connect
the
signal
from
the
discriminator
out
connector
by
way
of
50-ohm
coaxial
cable
to
the
input
CN-1
of
the
403A.
3.
Connect
the
preamp
output
signal
to
the
input
of
a
linear
amplifier
such
as
the
410
or
435
with
100
ohm
coaxial
cable.
4.
Turn
on
dc
power
to
403A
and
amplifier.
5.
Connect
a
high
voltage
power
supply
to
the
high
voltage
input
con
nector
and
ENSURE
that
the
power
supply
is
set
to
deliver
negative
voltage.
6.
Turn
on
high
voltage
power
supply
and
set
the
power
supply
to
-2200
volts
(or
the
voltage
at
which
the
tube
is
to
be
operated).
7.
Place
a
radiation
source
appropriate
to
the
chosen
scintiMotor
near
the
detector.
8.
Observe
the
output
waveform
from
the
shaping
amplifier
associated
with
the
linear
output
signal
and
adjust
the
two
bleeder
string
controls.

3-3
i.e.,
the
FOCUS
(R17)
and
DEFLECTOR
2
adjustment
(R15)
for
maximum
output
signal
(see
Fig.
10).
This
assures
that
the
input
optics
are
adjusted
properly
for
that
specific
photomultiplier
used.
3.3.3
Adjustment
of
the
Time
Derivation
Circuit
This
adjustment
has
been
performed
at
the
factory
and
will
not
be
affected
by
exchange
of
the
photomultiplier;
therefore,
it
wil
l
normally
not
re
quire
readjustment.
The
264
will
normally
be
shipped
with
the
clipping
line
(DL-1)
of
the
correct
length
for
fast
plastic
phosphors
such
as
Naton
136,
Pilot-B,
and
NE-102;
however,
if
a
slower
phosphor
such
as
NE-213,
Stilbene
or
Anthracene
is
used,
the
delay
line
(DL-1)
should
be
changed
to
the
longer
line
included
as
an
accessory
with
the
264
and
so
marked.
This
is
accomplished
merely
by
disconnecting
DL-1,
which
is
in
the
unit,
and
connecting
in
the
longer
length
delay
line.
To
perform
the
necessary
discriminator
zero
adjustment,
the
following
steps
should
be
followed:
1.
Perform
steps
1-4
of
Section
3.3.2.
2.
Connect
one
of
the
403A
outputs
to
on
oscilloscope.
3.
Set
the
DISC
LEVEL
on
the
403A
to
0.00.
4.
Adjust
DISC
ZERO
control
on
403A~clockwise
until
output
signals
appear
(unit
is
self-triggering)
and
back
off
until
signals
are
no
longer
present.
5.
Connect
a
fast
oscil
loscope,
preferably
a
sampling
type
oscilloscope,
to
the
coaxial
connector
on
the
discriminator
board
labeled
TP-1
with
50-ohm
coaxial
cable
making
sure
that
the
receiving
end
is
terminated
either
by
the
input
impedance
of
the
oscilloscope
or
by
an
external
tee
and
terminator.
6.
Attach
one
of
the
output
signals
from
the
403A
to
the
trigger
input
of
the
oscilloscope.
7.
Apply
high
voltage
(-2200
volts)
to
the
264.
8.
Place
a
radiation
source
near
the
scinti
1
lator,
e.g.,
Cobalt-60
or
Sodium-22.
9.
Set
the
DISCRIMINATOR
LEVEL
on
the
403A
to
50
dial
divisions.

3-4
10.
See
that
the
oscilloscope
Is
triggering
on
the
output
signal
from
the
403A.
11
.
Obtain
the
correct
time
delay
by
means
of
delay
line
lengths
be
tween
the
264
and
oscilloscope
input
to
allow
viewing
of
the
signal
from
TP-1
.
This
should
normally
require
only
five
feet
or
so
of
coble.
12.
Observe
the
signal
at
TP-1
and
adjust
R26
on
the
bleeder
network
(see
Fig.
10)
for
maximum
signal
amplitude
without
pulse
shape
distortion.
Examples
of
correct
and
incorrect
adjustment
of
this
potentiometer
are
shown
in
Figures
2
and
3.
13.
Observe
the
point
of
zero
volts,
and
adjust
the
CROSSOVER
ZERO
trim
control,
R6,
located
on
the
time
derivation
board
to
obtain
a
minimum
shift
in
time
of
the
zero
voltage
point.
This
assures
that
there
is
a
minimum
of
time
shift
vs.
pulse
height
associated
with
the
time
derivation
unit.
(Figures
4
and
5
show
examples
of
correct
and
incorrect
adjustment
of
this
control.)
14.
Disconnect
the
cables,
replace
the
high
voltage
cover
and
the
unit
is
ready
for
operation.

4-
1
4.
OPERATING
INSTRUCTIONS
Once
the
steps
outlined
in
Section
3
of
this
manual
are
performed,
the
unit
is
ready
for
use.
High
voltage
may
be
applied,
and
adjusted
for
appropriate
gain
associated
with
the
specific
experiment.
The
photomultiplier
gain
will
vary
approxi
mately
a
factor
of
two
with
the
high
voltage
change
of
100
volts.
The
time
derivation
unit
should
now
be
controlled
by
the
403A
and
the
output
of
the
403A
should
be
attached
to
the
appropriate
time
measuring
device,
e.g.,
time
to
pulse
height
converter.
The
threshold
adjustment
of
the
timing
device
may
be
set
by
the
fol
lowing
steps:
1.
Connect
the
output
of
a
single
trace
scope
such
as
a
545
or
585
to
the
output
of
a
shaping
amplifier
that
is
connected
to
the
linear
output
from
the
264.
2.
Connect
one
of
the
outputs
of
the
403A
to
the
trigger
input
to
the
oscilloscope
and
obtain
scope
trigger
with
high
voltage
supplied
and
associated
radiation
source
near
the
scintillator.
3.
Observe
the
linear
pulse
height
distribution
on
the
oscil
loscope.
This
distribution,
which
is
a
Compton
distribution,
is
essential
ly
white
over
the
pulse
height
range
associated
with
the
input
energy
from
the
source,
i.e.,
there
is
a
constant
dens
ity
of
pulses
from
zero
volts
to
the
maximum
volts
of
the
pulse
height,
except
for
the
region
below
the
threshold
level
set
by
the
DISCRIMINATOR
LEVEL
control
on
the
403A.
For
pulse
heights
below
this
level,
no
output
signals
wil
l
be
ob
served.
One
may,
in
this
manner,
precisely
locate
the
edge
of
the
discriminator
threshold
and
set
this
edge
to
any
desired
point
within
the
spectrum
of
energies
involved
in
the
timing
experiment.
Once
the
discriminator
level
is
adjusted
to
the
correct
point,
the
equipment
is
ready
to
perform
time
measurements.
4.1
Timing
with
Photomultipliers
Timing
or
time
measurement
implies
some
type
of
coincidence
measurement.
This
measurement
may
be
performed
with
standard
coincidence
circuits
such
as
pulse
overlap
types
which
are
essentially
single
channel
time
analyzers,
or
with
time
to
pulse
height
converters
which
are
differential,
or
multichannel
time
analyzers.
The
response
of
a
photomultiplier
coincidence
system
to
a
prompt
gamma
cascade
always
has
a
finite
width
which
comes
from
a
variety
of
sources.
The
most
important
of
these
are
as
follows:
1
.
The
variation
of
time
of
interaction
of
radiation
with
the
scintillator
and

4-2
Figure
2
Incorrect-
Adjustment
of
R-26
RCA-8575
with
1"
x
1"
Naton
136
Source
—
®°Co
H.V.
=
-2200V
Time
Scale
(Horiz.)
=
2
nsec/cm
A
l-f
A
(\/
4-
t
noo
mV/cm
@
TP-1
Amphfude
(Vert.)
=(
Figure
3
Correct
Adjustment
of
R-26
RCA-8575
with
1"
x
1"
Noton
136
Source
—
®°Co
H.V.
=
-2200V
Time
Scale
(Horiz.)
=
2
nsec/cm
(100
mV/cm
@
TP-1
I
40
mA/cm
(Anode
Current)
Amplitude
(Vert.)

4-3
Figure
4
Incorrect
Adjustment
of
Crossover
Zero
R6
X
1"
Noton
136
RCA-8575
with
1
Source
—
®°Co
H.V.
=
-2200V
Time
Scale
(Horiz.)
Amplitude
(Vert.)
2
nsec/cm
20
mV/cm
@
TP-1
8
m4y/cm
(Anode
Current)
Figure
5
Correct
Adjustment
of
Crossover
Zero
R6
RCA-8575
with
1"
x
1"
Noton
136
r
60/—
bource
—
Co
H.V.
-
-2200V
Time
Scale
(Horiz.)
=
2
nsec/cm
@
TP-1
A
1-4.
^
i\/
4.
\
_
(20
mV/cm
@
TP-1
Amplitude
(Vert.)
-j
3

4-4
the
amount
of
energy
deposited
therein;
2.
Finite
decay
time
of
light
emitting
states
in
the
phosphor
and
variation
of
times
of
photon
arrival
at
the
photomultiplier
cathode;
3.
Variation
of
transient
times of
photoelectrons
in
the
photomultiplier
due
to
a)
different
path
lengths,
and
b)
variation
of
initial
energy
and
angle
of
secondary
electrons;
4.
Jitter
and
uncertainty
of
times
of
triggering
of
the
associated
electronics.
The
variation
in
the
time
of
interaction
can
be
minimized
by
appropriate
geom
etry
and
small
scintiIlators
at
a
corresponding
loss
in
efficiency
and
average
energy
deposition.
For
a
complete
discussion
of
timing
with
photomultipliers,
the
reader
is
referred
to
some
of
the
excellent
literature
available
on
the
general
subject
^
and
on
the
specific
method
employed
here.^
4.2
Applications
The
different
specific
applications
for
the
264
are
essentially
limitless,
but
since
the
unit
is
designed
primarily
for
timing
applications
with
a
wide
dynamic
range
of
input
energy,
a
number
of
system
diagrams
utilizing
this
unit
are
given.
Some
typical
resolution
curves
for
one
of
these
systems
wil
l
be
shown,
from
which
operational
characteristics
of
other
systems
may
be
applied.
4.2.1
Typical
Fast
Coincidence
System
Figure
6
is
a
block
diagram
of
a
system
one
might
use
to
perform
life
time
measurements
or
to
study
time
dispersion
associated
with
prescribed
^A.
Schwarzschild,
Nuclear
Instr.
&
Methods,
21
pp
1-16
(1963)
^G.
Present,
etal..
Nuclear
I
nstr.
cS;
Methods,
31,
No.
1
pp
71-76
(1964)
^E.
Gatti
and
V.
Svelto,
Nuclear
Instr.
&
Methods,
30
pp
213-223
(1964)
^D.
L.
Weiber
and
H.
W.
Lefevre,
IEEE
Trans,
on
Nucl
.
Sci.,
NS-13,
No.
1
pp
406-412

4-5
coincidence
events.
One
notes
immediately
the
simplicity
of
the
sys
tem.
It
obtains
its
discrete
advantages
from
the
fact
that
there
is
no
time
shift
associated
with
pulse
height
variation;
therefore,
wide
dy
namic
ranges
of
pulse
amplitude
may
be,
and
normally
are,
used
in
the
timing
experiment.
The
resolution
of
the
system
in
this
case
will
nor
mally
be
limited
by
the
statistical
spreading
due
to
the
scintillation
process
for
the
lower
energy
events.
It
is
important
to
remember
that
the
resolution
obtainable
varies
as
1//N,
where
N
represents
the
num
ber
of
photoelectrons
created
by
each
event,
and
is
therefore
represent
ative
of
the
amount
of
energy
deposited
in
the
scintillating
phosphor,
and
is
strongly
influenced
by
photomultiplier
optics.
The
total
time
spectrum
is
composed
of
the
summation
from
zero
to
of
the
time
dispersion
associated
with
each
unit
energy
interval
between
the
discrim
inator
threshold
level
and
maximum
energy.
Normal
ly,
this
system
will
require
no
energy
selection
side
channel
electronics;
however,
if
it
is
desirable
to
restrict
the
energy range
to
a
small
value
this
side
channel
electronics
may
be
imposed
in
the
nor
mal
manner
associated
with
a
typical
fast-slow
coincidence
system.
Figure
7
shows
a
typical
time
spectrum
taken
with
this
type
of
system
utilizing
all
signals
detected
which
deposited
greater
than
30
keV
of
gamma
ray
energy
into
the
scintillator.
A
comparison
of
this
spectrum
with
Fig.
8,
which
has
the
energy
range
restricted
by
side
channel
windows,
indicates
the
effect
of
resolution
broadening
as
a
function
of
the
energy
range
utilized.
4.2.2
Matching
the
Time
Derivation
System
to
the
Experiment
It
is
important
to
understand
when
the
fast
crossover
system
may
be
used
to
ah
advantage
over
that
system
utilizing
a
leading
edge
discrimi
nator.
In
general,
the
leading
edge
type
discriminator
may
be
used
on
signals
of
any
shape
and
therefore
it
may
be
used
with
photomuItipliers
in
conjunction
with
al
l
scintillators.
The
advantage
of
a
leading
edge
time
derivation
scheme
is
that
of
time
resolution
for
a
narrow
dynamic
range.
One
may
achieve,
for
a
very
narrow
dynamic
range,
time
reso
lution
in
the
order
of
a
factor
of
two
better
by
leading
edge
discriminator
than
may
be
achieved
by
a
fast
crossover
method.
The
disadvantages
of
leading
edge
discrimination
are
primarily
those
associated
with
walk,
i.e.,
time
shift
vs.
pulse
height.
Figure
9
presents
comparative
curves
on
fast
crossover
and
leading
edge
discrimination
vs.
dynamic
range.
It
is
readily
evident
from
Fig.
9
that
if
fast
phosphors
are
used
in
the
Table of contents
Other ORTEC Amplifier manuals
Popular Amplifier manuals by other brands

Mitsubishi Electric
Mitsubishi Electric CV-0MW7G21-92 Service manual

Eden
Eden World Tour WT800C Operation manual

Sherbourn
Sherbourn PRE-1 user manual

Zeck Audio
Zeck Audio STAC Force180 Service manual

Lyngdorf Audio
Lyngdorf Audio TDAI-2170 owner's manual

Altinex
Altinex Distribution Amplifier DA1322AT user guide