
1
1 DESCRIPTION
1.1 GENERAL
The ORTEC 570 Spectroscopy Amplifier is a
singlewidth NIM module that features a versatile
combination of switch-selectable pulse-shaping
characteristics. The amplifier has extremely low
noise, a wide gain range, and excellent overload
responseforuniversalapplicationinhigh-resolution
spectroscopy. It accepts input pulses of either
polarity that originate in germanium or silicon
semiconductor detectors, in scintillation counters
with either fast or slow scintillators, in proportional
counters, inpulsedionizationchambers,inelectron
multipliers, etc.
The 570 has an input impedance of approximately
1000
S
andacceptseitherpositiveornegativeinput
pulses with rise times <650 ns and fall times >40
:
s. Six integrate and differentiate time constants
are switch-selectable to provide optimum shaping
for resolution and count rate. The differentiation
networkhasvariablepole-zerocancellationthatcan
beadjustedtomatchpreamplifierswithdecaytimes
>40
:
s. The pole-zero cancellation drastically
reduces the undershoot after the differentiator and
greatly improves overload and count rate
characteristics.Inaddition,theamplifiercontainsan
active filter shaping network that optimizes the
signal-to-noise ratio and minimizes the overall
resolving time.
The output is unipolar and is used for spectroscopy
in systems where dc coupling can be maintained
from the 570 to the analyzer. A BLR (baseline
restorer) circuit is included in the 570 for improved
performance at all count rates. Baseline correction
is applied during intervals between input pulses
only, and a front panel switch selects a
discriminator level to identify input pulses. The
unipolar output dc level can be adjusted in the
range from -100 mV to +100 mV. This output
permits the use of the direct-coupled input of the
analyzer with a minimum amount of interface
problems.
The 570 can be used for constant-fraction timing
when operated in conjunction with an ORTEC 551,
552, or 553 Timing Single-Channel Analyzer. The
ORTEC Timing Single-Channel Analyzers feature
a minimum of walk as a function of pulse amplitude
and incorporatea variable delay time on the output
pulse to enable the timing pick-off output to be
placed in time coincidence with other signals.
The 570 has complete provisions, including power
distribution, for operating any ORTEC solid-state
preamplifier. Normally, the preamplifier pulses
should have a rise time of 0.25
:
s or less to
properly match the amplifier filter network and a
decay time greater than 40
:
s for proper pole-zero
cancellation. The 570 input impedance is 1000
S
.
Whenlong preamplifier cablesare used, the cables
can be terminated in series at the preamplifier end
or in shunt at the amplifier end with the proper
resistors. The output impedance isabout 0.1
S
,and
the output can be connected to other equipment by
a single cable going to all equipment and shunt
terminated at the far end of the cabling. See
Section 3 for further information.
1.2 POLE-ZERO CANCELLATION
Pole-zero cancellation is a method for eliminating
pulse undershoot after the differentiating network.
In an amplifier not using pole-zero cancellation
(Fig.1.1), the exponential tail on the preamplifier
output signal (usually 50 to 500
:
s) causes an
undershoot whose peak amplitude is roughly
determined from:
undershoot amplitude
differentiation time
=differentiated pulse amplitude
preamplifier pulse decay time
For a 1-
:
s differentiation time and a 50-
:
s pulse
decay timethemaximum undershoot is2%andthis
decays witha50-
:
s time constant. Under overload
conditions this undershoot is often sufficiently large
to saturate the amplifier during a considerable
portion of the undershoot, causing excessive dead
time. This effect can be reduced by increasing the
preamplifier pulse decay time (which generally
reduces the counting rate capabilities of the
preamplifier)orcompensatingfortheundershootby
causing pole-zero cancellation.