
3
approximately 4 ft (see Fig. 3.1). The input
impedance of the 427A is approximately 1100
S
. It
is recommended that RG-62/U or RG-63/U coaxial
cable be used because of their relatively high
impedanceof 93
S
and125
S
respectively.Baseline
restoration will normally be accomplished in the
linear amplifier. The input dc level furnished into
the 427A should be zero with no signal present, and
thiswill normally be adjusted in the linear amplifier
output. The 427A has a dc-coupled output with a
dc-level adjustment for a range of ±1 V. This
permits the input to the analyzer to be normalized
in order to establish its zero-energy crossover
calibration.
3.4. LINEAR OUTPUT SIGNAL
CONNECTIONS AND TERMINATING
IMPEDANCE
The source impedance of the 0- to 10-V standard
linear outputs of most ORTEC instruments is
furnished through a series impedance of either
0.1
S
or 93
S
, depending on the connector that is
used for the connection to the next module. When
the interconnecting cable is short, such as
maximum of 4 ft, this will not usually result in any
interference problems even though impedance
matching is disregarded. However, if acablelonger
than approximately 4 ft. is necessary in a linear
output, it should be terminated in a resistive load
equal to the impedance in order to prevent
oscillations.
Therearethreegeneral methodsof terminationthat
are used. The simplest of these is shunt termination
at the receiving end of the cable. A second method
is series termination at the sending end. Thethirdis
a combination of series and shunt termination,
where the cable impedance is matched both in
series at the sending end and in shunt at the
receiving end. The most effective method is the
combination, but termination by this method
reduces the amount of signal strength at the
receiving end to 50% of that which is available in
the sending instrument.
To use shunt termination at the receiving end of
the cable, connect the 1
S
output of the sending
device through 93
S
cable to the input of the
receiving instrument. Then use a BNC tee
connector accept both the interconnecting cable
and 100
S
resistive terminator at the input
connector of the receiving instrument. Since the
input impedance of the receiving instrument is
normally 1000
S
or more, the effective instrument
input impedance with the 100
S
terminator will be
of the order of 93
S
, and this correctly matches the
cable impedance.
For series termination, use the 93
S
output of the
sending instrument for the cable connection. Use
93
S
cable to interconnect thisinto the input of the
receivinginstrument. The1000
S
(or more) normal
input impedance at the input connector represents
an essentially open circuit, and the series
impedance inthesendinginstrumentnowprovides
the proper termination for the cable.
For the combination of series and shunt
termination, use the 93
S
output of the sending
instrument for the cable connection and use 93
S
cable. Attheinput for thereceivinginstrument,use
a BNCtee to accept both the interconnecting cable
and a 100
S
resistive terminator. Note that the
signal span at the receiving end of this type of
receiving circuit will always be reduced to 50% of
the signal span furnished by the sending
instrument.
For your convenience, ORTEC stocks the proper
terminators and BNC tees, or you can obtain them
from a variety of commercial sources.