Pintek DP-25 User manual

DP-25
差動測試棒
HIGH VOLTAGE DIFFERENTIAL PROBE
DP-50
DP-100
INSTRUCTION MANUAL
使用說明書
1400Vp-p/25MHz
7000Vp-p/50MHz
7000Vp-p/100MHz


DP-25
High Voltage Differential Probe
1. FEATURES
2. SPECIFICATIONS
3. PANEL DESCRIPTION
4. OPERATING ENVIRONMENTAL
CONDITIONS
5. OPERATING PROCEDURE
6. MAINTENANCE
7. CLEANING
8. STORAGE
9. WARRANTY
10. REPAIR
11. ACCESSORIES
DP-50
High Voltage Differential Probe
1. FEATURES
2. SPECIFICATIONS
3. PANEL DESCRIPTION
4. OPERATING ENVIRONMENTAL
CONDITIONS
5. OPERATING PROCEDURE
6. MAINTENANCE
7. CLEANING
8. STORAGE
9. WARRANTY
10. REPAIR
11. ACCESSORIES
DP-100
High Voltage Differential Probe
1. FEATURES
2. SPECIFICATIONS
3. PANEL DESCRIPTION
4. OPERATING ENVIRONMENTAL
CONDITIONS
5. OPERATING PROCEDURE
6. MAINTENANCE
7. CLEANING
8. STORAGE
9. WARRANTY
10. REPAIR
11. ACCESSORIES
1
1
2
3
3
4
5
5
5
5
6
TABLE OF CONTENTS
8
8
9
10
10
11
12
12
12
12
13
17
17
18
19
19
19
19
20
16
15
15
DP-25 / DP-50 / DP-100
VOLTAGE DERATING CURVE 40

DP-25 / DP-50 / DP-100
耐壓曲線參考圖
目 次
40
DP-25 差動測試棒
1. 簡 述
2. 規 格
3. 測試棒面板說明
4. 操作環境及狀況
5. 操作程序
6. 維護
7. 清潔
8. 保固
9. 維修
10.附件
DP-50 差動測試棒
1. 簡 述
2. 規 格
3. 測試棒面板說明
4. 操作環境及狀況
5. 操作程序
6. 維護
7. 清潔
8. 保固
9. 維修
10.附件
DP-100 差動測試棒
1. 簡 述
2. 規 格
3. 測試棒面板說明
4. 操作環境及狀況
5. 操作程序
6. 維護
7. 清潔
8. 保固
9. 維修
10.附件
23
23
24
25
25
26
27
27
27
27
29
29
30
31
31
32
33
33
33
33
35
35
36
37
37
38
39
39
39
39

! Differential Voltage Probe,
Read the instructions before using the instrument:
1.Must acquire a differential voltage probe & get the best service
from instrument.
2.Read carefully the Instruction Manual.
3.Respect the safety precautions.
! SAFETY PRECAUTIONS
WARNING: Risk of Electric Shock,
1.Do not use the probe in damp environment or where there is
risk of explosion.
2.Do not use the probe with its case open.
3.Disconnect the inputs and outputs of the probe before
opening the case.
4.The probes are for indoor use only.
Respect the max input voltages:
DP-25:
1.Max differential voltage: 1400V (DC + AC peak) or 450 Vrms
2.Max voltage between each input terminal and ground:600 Vrms
DP-50 & DP-100:
1.Max differential voltage: 7000V (DC + AC peak) or 2200 Vrms
2.Max voltage between each input terminal and ground: 6500 Vrms
! TO ORDER Differential Voltage Probe and Accessories:
An Insulated BNC/BNC lead, length 100cm.
Supplied a Adapter preset 9 V DC (115 V or 230 V)
2 x high voltage IC clips
2 x Banana to Banana high voltage plug
2 x Alligator plug

DP-25
High Voltage Differential Probe

DP-25 HIGH VOLTAGE DIFFERENTIAL PROBE
1
1. FEATURES
The DP-25 differential probe provides a safety means for
measuring differential voltage to all models of oscilloscopes.
The DP-25 converts the high differential voltage (≦1400Vpeak)
into a low voltage (≦7.0V, with reference to the earth) and display
on the oscilloscopes.
The DP-25 is designed to operate with the 1MΩimpedance
oscilloscopes. When combine with the 50Ω load, the attenuation
will be 2 times.
We recommend to use PINTEK PL-10 with DP-25 to expand the
measuring with DMM to observe more accurate measurement.
The accuracy of oscilloscope is 3% and the DMM is less than 1%.
NOTE: If you connect DP-25 to the DMM without PL-10, the
accuracy will be higher than 10%.
2. SPECIFICATIONS
(1) Bandwidth:
DC - to 25 MHz (-3 dB) for x 50, x 200
DC - to 15 MHz (for attenuation x 20)
(2) Attenuation: x 20, x 50, or x 200
(3) Accuracy: +/- 2%
(4) Voltage Input Ranges (DC + AC peak to peak)
≦ 140 Vp-p for x 20, (i.e about 45 Vrms or DC)
≦ 350 Vp-p for x 50, (i.e about 110 Vrms or DC)
≦ 1400 Vp-p for x 200, (i.e about 450 Vrms or DC)
(5) Permitted Max Input Voltage
Max differential voltage: 1400 V (DC + AC peak to peak) or
450Vrms
Max voltage between each input terminal and ground:
600 Vrms

2
DP-25 Instruction Manual
(6) Input Impedance:
Differential: 4 MΩ / 1.2 pF
Between terminals and ground: 2 MΩ / 2.3 pF
(7) Output: ≦ +/- 7.0 V
(8) Output Impedance: 50 Ω
(9) Rise Time: 14 ns for x 50, and x 200; 23.4 ns for x 20
(10) Rejection Rate on Common Mode:
60 Hz: > 80 dB ; 100 Hz: > 60 dB ; 1 MHz: > 50 dB
(11) Power Supply: Only External 9 V DC power supply .
(12) Consumption: 35 mA max (0.4 WATT)
3. PANEL DESCRIPTION
Power ON/OFF Indication
Attenuation Function Switch
Low Power Indication
Input Connector
Output Connector
External Power Source

3
5. OPERATING PROCEDURE
!Connect the leads to the input and place the wire-grip on the
circuit to be tested.
!Connect the probe to the oscilloscope with the insulated
BNC/BNC lead.
!Adjust the vertical zero adjustment of the oscilloscope if
necessary.
!Select the attenuation ratio* and the vertical deviation of the
oscilloscope in accordance with the conversion table below.
!NB: The POWER light must come on.
The conversion table gives the real vertical deviation.
4. OPERATING ENVIRONMENTAL CONDITIONS
(1) Dimensions and Weight:
195 x 55 x 30 mm; 250g
(2) Electrical Safety to IEC 1010-1
!Dual Insulation
!Installation Category III
!Degree of Pollution 2
!Rated Voltage or Max Line-Earth: 600 Vrms
(3) CE Mark
Conforms to EN 50081-1 and 50082-1 standards
(4) Indoor use only.
DP-25 Instruction Manual
X 200 X 50 X 20
1400Vp-p 350Vp-p 140Vp-p
(±700VDC) (±175VDC) (±70VDC)
Voltage Input Range
(DC+AC Peak)
Attenuation
Temperature
Relative Humidity
Reference Use Storage
+20°C … +30°C
≦70 % RH
0°C … +50°C
10 % … 85 % RH
-30°C … +70°C
10 % … 90 % RH

4
[N.B]
The real vertical deviation in V/div is equal to the attenuation
factor multiplied by the range of vertical deviation selected on
the oscilloscope. It will be doubled in the case of use of a 50 Ω
load.
Example:
With the probe on factor x 200, the oscilloscope on 0.5 V/div,
the real vertical deviation is 200 x 0.5 = 100 V/div.
With a 50Ω load on the input of the oscilloscope the deviation
becomes 200 V/div.
200
Real Deviation In V/div
1 50 20
1000.5 25 10
400.2 10 4
200.1 5 2
1050 m 2.5 1
420 m 1 0.4
210 m 0.5 0.2
15 m 0.25 0.1
0.42 m 0.1 40 m
DP-25 Instruction Manual
Vertical Deviation on the
Oscilloscope in V/div
6. MAINTENANCE
For maintenance, only use specified spare parts.
The manufacturer can not be held responsible for any accident
arising following a repair made other than its after sales service or
approved repairers.
x 200 x 50 x 20

7. CLEANING
This probe does not require any particular cleaning. If necessary,
clean the case with a cloth slightly moistened with soapy water.
8. STORAGE
If the probe is not use more than 60 days, please store the probe in
a dehumidified environment to keep dry.
9. WARRANTY
Unless notified to the contrary, our instruments are guaranteed
against any manufacturing defect or material defect. They do not
bear the specification known as the safety specification. Our
guarantee, which may not under any circumstances exceed the
amount of the invoiced price, goes no further than the repair of our
faulty equipment, carriage paid to our workshops.
It is applicable for normal use of our instruments, and does not
apply to damage or destruction caused, notably by error in
mounting, mechanical accident, faulty maintenance, defective
use, overload or exceed voltage.
Our responsibility being strictly limited to the pure and simple
replacement of the faculty parts of our equipment, the buyer
expressly renounces any attempt to find us responsible for
damages or losses caused directly or indirectly.
Our guarantee is applicable for twelve (12) months after the date
at which the equipment is made available. The repair, modification
or replacement of a part during the guarantee period will not result
in this guarantee being extended.
10. REPAIR
Maintenance, repairs under or out of guarantee. Please return the
product to your distributor.
5
DP-25 Instruction Manual

11. ACCESSORIES
ADP-110V or ADP-220V: AC Adapter.
BP-250: BNC Plug to BNC Plug; 50ΩResistance , RG58C UL,
Length 100cm.
BP-356N: Banana Plug to Banana Plug Silicon Wire; UL 6KV,
18AWG, Length 60cm.(Red x 1pc , Black x 1pc)
BP-256N: IC Clip, UL 1000V CAT III.(Red x 1pc , Black x 1pc)
BP-276N: Alligator Clip, UL 1000V CAT II, 10A.(Red x 1pc , Black
x 1pc)
Instruction Manual(TINSE0004S4).
6
DP-25 Instruction Manual

High Voltage Differential Probe
DP-50

DP-50 HIGH VOLTAGE DIFFERENTIAL PROBE
1. FEATURES
The DP-50 differential probe provides a safety means for
measuring differential voltage to all models of oscilloscopes.
The DP-50 converts the high differential voltage (≦7000Vpeak)
into a low voltage (≦7.0V, with reference to the earth) and display
on the oscilloscopes.
The DP-50 is designed to operate with the 1MΩimpedance
oscilloscopes. When combine with the 50Ω load, the attenuation
will be 2 times.
We recommend to use PINTEK PL-10 with DP-50 to expand the
measuring with DMM to observe more accurate measurement.
The accuracy of oscilloscope is 3% and the DMM is less than 1%.
NOTE: If you connect DP-50 to the DMM without PL-10, the
accuracy will be higher than 10%.
2. SPECIFICATIONS
(1) Bandwidth:
DC - to 50 MHz (-3 dB) for x 200, x 500 and x 1000
DC - to 25 MHz (for attenuation x 100)
(2) Attenuation: x 100, x 200, x 500, x1000
(3) Accuracy: +/- 2%
(4) Voltage Input Ranges (DC + AC peak to peak)
≦ 700 Vp-p for x 100, (i.e about 230 Vrms or DC)
≦ 1400 Vp-p for x 200, (i.e about 460 Vrms or DC)
≦ 3500 Vp-p for x 500, (i.e about 1140 Vrms or DC)
≦ 7000 Vp-p for x 1000, (i.e about 2300 Vrms or DC)
(5) Permitted Max Input Voltage
Max differential voltage: 7000 V (DC + AC peak to peak)
Max voltage between each input terminal and ground:
6500 Vrms
8

9
DP-50 Instruction Manual
(6) Input Impedance:
Differential: 54 MΩ / 1.2 pF
Between terminals and ground: 27 MΩ / 2.3 pF
(7) Output: ≦ +/- 7.0 V
(8) Output Impedance: 50 Ω
(9) Rise Time: 7 ns for x 200, x 500, and x 1000; 14 ns for x 100
(10) Rejection Rate on Common Mode:
60 Hz: > 80 dB ; 100 Hz: > 60 dB ; 1 MHz: > 50 dB
(11) Power Supply: Only External 9 V DC power supply .
(12) Consumption: 35 mA max (0.4 WATT)
3. PANEL DESCRIPTION
Power ON/OFF Indication
Attenuation Function Switch
Low Power Indication
Input Connector
Output Connector
External Power Source

10
5. OPERATING PROCEDURE
!Connect the leads to the input and place the wire-grip on the
circuit to be tested.
!Connect the probe to the oscilloscope with the insulated
BNC/BNC lead.
!Adjust the vertical zero adjustment of the oscilloscope if
necessary.
!Select the attenuation ratio* and the vertical deviation of the
oscilloscope in accordance with the conversion table below.
!NB: The POWER light must come on.
The conversion table gives the real vertical deviation.
DP-50 Instruction Manual
4. OPERATING ENVIRONMENTAL CONDITIONS
(1) Dimensions and Weight:
240 x 80 x 30 mm; 280g
(2) Electrical Safety to IEC 1010-1
!Dual Insulation
!Installation Category III
!Degree of Pollution 2
!Rated Voltage or Max Line-Earth: 6500 Vrms
(3) CE Mark
Conforms to EN 50081-1 and 50082-1 standards
(4) Indoor use only.
X 1000 X 500 X 100
7000Vp-p 3500Vp-p 700Vp-p
(±3500VDC) (±1750VDC) (±350VDC)
Voltage Input Range
(DC+AC Peak)
Attenuation X 200
1400Vp-p
(±700VDC)
Temperature
Relative Humidity
Reference Use Storage
+20°C … +30°C
≦70 % RH
0°C … +50°C
10 % … 85 % RH
-30°C … +70°C
10 % … 90 % RH

[N.B]
The real vertical deviation in V/div is equal to the attenuation
factor multiplied by the range of vertical deviation selected on
the oscilloscope. It will be doubled in the case of use of a 50 Ω
load.
Example:
With the probe on factor x 200, the oscilloscope on 0.5 V/div,
the real vertical deviation is 200 x 0.5 = 100 V/div.
With a 50Ω load on the input of the oscilloscope the deviation
becomes 200 V/div.
1000
Real Deviation In V/div
1 500 200
5000.5 250 100
2000.2 100 40
1000.1 50 2
5050 m 25 10
2020 m 10 4
1010 m 5 2
55 m 2.5 1
22 m 1 0.4
DP-50 Instruction Manual
100
50
20
10
5
2
1
0.5
0.2
Vertical Deviation on the
Oscilloscope in V/div
11
6. MAINTENANCE
For maintenance, only use specified spare parts.
The manufacturer can not be held responsible for any accident
arising following a repair made other than its after sales service or
approved repairers.
x 1000 x 500 x 200 x 100

DP-50 Instruction Manual
12
7. CLEANING
This probe does not require any particular cleaning. If necessary,
clean the case with a cloth slightly moistened with soapy water.
8. STORAGE
If the probe is not use more than 60 days, please store the probe in
a dehumidified environment to keep dry.
9. WARRANTY
Unless notified to the contrary, our instruments are guaranteed
against any manufacturing defect or material defect. They do not
bear the specification known as the safety specification. Our
guarantee, which may not under any circumstances exceed the
amount of the invoiced price, goes no further than the repair of our
faulty equipment, carriage paid to our workshops.
It is applicable for normal use of our instruments, and does not
apply to damage or destruction caused, notably by error in
mounting, mechanical accident, faulty maintenance, defective
use, overload or exceed voltage.
Our responsibility being strictly limited to the pure and simple
replacement of the faculty parts of our equipment, the buyer
expressly renounces any attempt to find us responsible for
damages or losses caused directly or indirectly.
Our guarantee is applicable for twelve (12) months after the date
at which the equipment is made available. The repair, modification
or replacement of a part during the guarantee period will not result
in this guarantee being extended.
10. REPAIR
Maintenance, repairs under or out of guarantee. Please return the
product to your distributor.

DP-50 Instruction Manual
13
11. ACCESSORIES
ADP-110V or ADP-220V: AC Adapter.
BP-250: BNC Plug to BNC Plug; 50ΩResistance, RG58C UL,
Length 100cm.
BP-276N: Alligator Clip, UL 1000V CAT II, 10A.
(Red x 1pc , Black x 1pc)
BP-266: HV IC Clip, MAX. 6500V(DC+ACp-p).
(Red x 1pc , Black x 1pc)
BP-366: HV Banana Plug to Banana Plug Silicon Wire, 18AWG,
UL 20KV, Length 60cm.(Red x 1pc , Black x 1pc)
Instruction Manual(TINSE0004S4).

High Voltage Differential Probe
DP-100
This manual suits for next models
2
Table of contents
Other Pintek Measuring Instrument manuals

Pintek
Pintek DP-35 User manual

Pintek
Pintek HVP-18HF User manual

Pintek
Pintek HVP-40M User manual

Pintek
Pintek PA-622 User manual

Pintek
Pintek DP-15K User manual

Pintek
Pintek DP-65 pro User manual

Pintek
Pintek DP-60HS User manual

Pintek
Pintek DP-16K Series User manual

Pintek
Pintek DP-16VF User manual

Pintek
Pintek DP-5205A User manual