Using the PlatoWork headset
Important
Never take the headset off or adjust it during an active session, as this could lead to a short but
uncomfortable ‘sting’ under the active electrodes (
more info below
).
Keep the usage to a maximum of one full session, or 30 minutes total, per day per user (
more info
below
).
Ensuring good contact
For neurostimulation to work, the electrodes need to have good contact to the head. Good
contact is ensured by a combination of 1. Wet sponges, 2. Pressure, and 3. Sponge surface
touching the head. When starting a session, the headset will test if the relevant electrodes
have sufficient contact, and ask you to improve contact if they don’t. If asked to improve the
contact by adjusting the headset, please make sure that:
1. The sponges are moist.
2. The headset is fitted tightly to your head.
3. The angle of the electrodes is aligned with your head so the whole surface of the sponge is
touching the head.
4. There is not too much dry hair directly underneath the electrodes.
Tip: If you are having trouble, look at the instructions in the Getting Started section and try
adjusting the headset again.
Select mode
In the app you need to choose between four modes: Learn, Create, Concentrate and Rethink.
You have to decide if you need to focus on a task at hand or create new ideas. We suggest you
try out each mode with different tasks to figure out what works best for you.
Red marks under the electrodes
Some users experience temporary red marks on the skin under the electrodes after
stimulation; this is normal and always disappear again quickly after the session. The marks are
caused by a combination of heat, current and the salt water used to moist the sponges. These
marks can be reduced by rinsing the sponges under running tap water often, thereby lessening
the amount of salt building up in the electrodes.
Adjusting the intensity slider
During stimulation, the app offers a slider to adjust the intensity of the stimulation. The slider
is there to enable users to adjust the sensation on the skin under the electrodes to their pain
6