Power Plate Pro7M User manual

Instructions for Use
Pro7M Model
Medical Device

Contents
Introduction 01
Correct Use is Essential for Optimal Results 01
Medical Indications 02
Important Safety Instructions 04
Health Warnings 04
Setup and Handling 04
Position and Balance 05
proMOTION™Dynamic Vibration Technology 06
Technical Specications 07
Maintenance 08
Troubleshooting 08
Box Contents 09
EMC Declaration and Precautions 10
Symbol Descriptions 14
Safety Classications 15
Appendix – FCC / IC statement 16
Touchscreen Controls 17
Parameters 19
Contact and Support 21
Warranty 21
©2021 Performance Health Systems, LLC. All rights reserved. Power Plate, the Power
Plate device/logo, pro7, Pro7M, Pro7HC, pro6+, pro5, Pro5M, pro5HP, My7M, my7, my5,
my3, Personal, MOVE, Pro4PlusM and proMOTION, are registered trademarks and/or
trademarks of Peformance Health Systems and/or their aliates. All other trademarks
are the property of their respective owners. Power Plate machines are protected under
patents and design rights in numerous countries around the world. Performance Health
Systems retains all rights (including copyright, trademark and all other intellectual
property rights) in relation to all information provided in this manual. You may not
copy, publish or distribute any of the information contained in this manual, or in any
other documents published by Performance Health Systems without the prior written
consent of Performance Health Systems.

Introduction
The Pro7M Power Plate is a commercial medical device, not intended for home healthcare.
Pro7M Power Plate model may be used by almost everyone. Power Plate allows you to
adjust training to your own level and reduce the burden on your joints, tendons and
ligaments. As with every form of training, the correct use of exercises, adjusted to your
personal abilities, will determine the benets and eects of completing a training session
on Power Plate, while at the same time, reduce the risk of injury to the body.
We recommend consulting your physician before beginning any new exercise program.
For your safety, please review the product instructions before operating Power Plate. If at
any time you feel faint, dizzy, nauseous, short of breath or in pain, stop exercising immedia-
tely and consult with your physician.
Body posture, muscle stiness and muscle tension are important contributing factors in
your training session. If muscles are tensed, or contracted, they will absorb vibrations to
help strengthen and tone those muscles. Passive or active exercises, such as stretching and
massage, can be performed on a more frequent basis.
Exercise should be performed on a regular basis, starting with low intensity, which means
low frequency and amplitude settings for short sessions. The body should be gently stimu-
lated in a way that will allow you to adjust to vibration training, but will not overload your
body. Over time, the intensity and duration can be increased in the same manner as other
progressive training programs. Once the body has adapted to vibration, training can be
changed or intensied to keep improving performance.

Medical Indications
Fall Prevention
Claim: Power Plate training can be a
benecial tool/intervention in reducing
the risk of falls in the elderly population.
Denition: Fall prevention is dened as
a variety of actions to help reduce the
number of accidental falls suered by
older people.
Muscular Strength and Power
Claim: Power Plate can be a benecial
tool/intervention to help preserve
muscular strength and power.
Denition: Physical strength is the ability
of a person or animal to exert force
on physical objects using muscles.
Increasing physical strength is the goal
of strength training. Power is the amount
of work done or energy transferred per
unit of time.
Pain
Claim: Power Plate training can be a
benecial tool/intervention in reducing
and/or alleviating chronic pain.
Denition: Pain is an unpleasant sensory
and emotional experience associated
with actual or potential tissue damage, or
described in terms of such damage.
Bone Density
Claim: Power Plate training can be a
benecial tool/intervention to increase
bone density and prevent bone mineral
density loss related to aging.
Denition: Bone density, or bone
mineral density, is the amount of bone
mineral in bone tissue. The concept is
of mass of mineral per volume of bone,
although clinically it is measured by
proxy according to optical density per
square centimetre of bone surface upon
imaging.
Weight Loss
Claim: Power Plate training can
be benecial in helping to reduce
percentage body fat in overweight
individuals.
Denition: Weight loss, in the context
of medicine, health or physical tness,
is a reduction of the total body weight,
due to a mean loss of uid, body fat or
adipose tissue and/or lean mass, namely
bone mineral deposits, muscle, tendon
and other connective tissue.
Circulation and Cardiovascular
Claim: Power Plate training can be a
benecial tool/intervention in reducing
arterial stiness and improving the
function of the cardiovascular system.
Denition: The circulatory system is
an organ system that moves nutrients,
gases and wastes to and from cells,
helps ght diseases and helps stabilize
body temperature and pH to maintain
homeostasis. Two types of uids move
through the circulatory system: blood
and lymph. The blood, heart and blood
vessels form the cardiovascular system.
The lymph, lymph nodes and lymph
vessels form the lymphatic system. The
cardiovascular system and the lymphatic
system collectively make up the
circulatory system. Pulmonary circulation
is the portion of the cardiovascular
system which transports oxygen-
depleted blood away from the heart, to
the lungs, and returns oxygenated blood
back to the heart.
Claimed medical
indications include
use in relation to the
prevention, treatment
and/or alleviation
of disease.
Based on standards published by the
European Medical Device Directive,
Pro7M Power Plate medical device is not
intended for home healthcare use. Safe
use is partially dependent on a properly
grounded power supply.
2of 22

Flexibility/Range of Motion
Claim: Power Plate training can be a
benecial tool/intervention to improve
exibility and range of motion.
Denition: Flexibility is the absolute
range of movement in a joint or series of
joints and muscles that is attainable in
a momentary eort involving a partner
or a piece of equipment. The exibility
of a joint depends on many factors,
particularly the length and looseness
of the muscles and ligaments due to
normal human variation, and the shape
of the bones and cartilage that make up
the joint.
Pathology Studies
Claim: Power Plate training can be a
benecial tool/intervention to improve
general well being, tness and daily life
functioning in patient populations. Power
Plate might have a positive impact on
general tness, muscle strength, daily life
performance and well being but does not
have a direct impact on the state of the
disease or injury. It will improve patient’s
life quality.
Denition: No denition possible for this
very diverse group. Examples of patients:
Multiple Sclerosis, Parkinson’s Disease,
Fibromyalgia, Cerebral Palsy, Spinal Cord
Injury, Diabetes, Cardiac Rehabilitation.
Medical Indications
3of 22

Important Safety Instructions
Before using the Power Plate®machine,
it is essential that you read the entire user
manual, including all warnings and safety
instructions. You should also convey
all such warnings and instructions to
any other person using the Power Plate
machine. Retain this user manual for
future reference.
! Health Warning
Before beginning any exercise program,
you should consult a physician for a
physical examination and clearance to
engage in the program, or personal injury
could result.
If you have any known medical condition,
or any physical limitation on your
ability to exercise, Power Plate strongly
recommends that you seek the advice
of a physician before using the Power
Plate machine, in order to avoid possible
personal injury.
If while using the Power Plate machine,
you experience any dizziness, faintness,
shortness of breath or pain, you must
stop using the machine immediately
and consult a physician. Failure to do so
could result in personal injury.
The Power Plate machine is a medical
device and is designed for therapeutic
purposes within specic medical
indications and supervision integrated
into an closely monitored exercise
program only and as part of an exercise
program.
Always follow the directions on the
machine’s console for proper operation.
Close supervision is required when the
machine is used by or near, children or
disabled persons.
Always take care when getting on and
o the machine. Use the handles on the
machine as needed, to maintain stability
when getting on and o the machine.
Never reach into or under the machine,
or tip the machine on its side, while it is in
operation.
Use the machine only for the purposes
described in this user manual and only
with attachments or accessories that
come with the machine or which Power
Plate has specically approved for use
with the machine, or personal injury could
result.
Setup and Handling
! Danger: To reduce the risk of electrical
shock, always unplug the Power Plate
machine before cleaning or servicing it.
! Warning: To reduce the risk of electrical
shock, re, burns or other injury, always
plug the Power Plate machine into a
properly-grounded electrical outlet.
! Warning: To ensure safe use of the
Power Plate machine, it must be regularly
examined for damage and wear. The
machine, however, contains no user-
serviceable parts. Thus, with the
exception of the maintenance tasks
described later in this manual, the owner/
user should always retain an authorized
Power Plate service professional to
perform maintenance and/or service on
the machine.
!Warning
The device is not suitable for use in
the presence of ammable anesthetic
mixtures with air or with oxygen or nitrous
oxide.
The device requires no calibration.
The device is not repairable and contains
no user serviceable parts.
The user must check that the equipment
functions safely and see that it is in proper
working condition before being used.
4of 22

The manufacturer does not require such
preventive inspections by other persons.
The plug is used to disconnect from
the main supply. Do not position
the machine so that it is dicult to
disconnect the main plug.
!Warning: To avoid the risk of electric
shock, this equipment must only be
connected to a supply main with
protective earth grounding.
The machine should be set up on a
hard, level surface in an area free of
obstructions within at least three feet
of the vibration platform.
The machine should not be used
outdoors, near a pool, or near any
source of water or extreme humidity.
Contact with water could cause a
short-circuit, which could cause
personal injury or damage the machine.
Unplug the machine when not in use.
Never attempt to lift or move the
machine without assistance.
Never operate the machine if it has
been dropped, tipped over, damaged,
or even partially immersed in water,
unless an authorized Power Plate
service professional has examined the
machine and cleared it for use.
Never insert objects into any opening
on the machine. If an object falls into
the machine, shut the machine o
and remove the power plug before
attempting to retrieve the object.
If the object cannot be reached,
contact an authorized Power Plate
service professional.
Position and Balance
In order to avoid possible injury from
the machine’s vibrations, Power Plate
recommends that the user not lean back
on his/her heels or “lock” his/her joints
or straighten his/her legs when standing
upright on the machine.
It is important at all times to maintain
balance while on the machine.
The user, however, should not “hang”
on the machine’s handles in order to
maintain balance but should (unless the
directions for a particular exercise state
otherwise) keep his/her knees directly
above the toes.
Proper position and balance are
especially important if the user employs
weights during any exercise on the
machine. Power Plate strongly
recommends that users not employ
weights while using the machine,
unless they are being coached by an
exercise professional or unless the user
has extensive experience with weight
training or with the Power Plate machine.
The three illustrations on the right
demonstrate the incorrect ways to stand
on the machine.
The illustration below demonstrates the
correct way to stand on the machine.
Incorrect
You always need to be
balanced when
standing on the Power
Plate machine. Never
hang on to the handles,
use them only to
maintain balance.
Incorrect
Don’t lean on your
heels too much.
Balance your weight
predominantly on the
front of your feet.
Incorrect
When training with the
Power Plate machine,
do not “lock” any joints,
such as your knees and
elbows, but keep them
slightly bent.
Legs slightly bent.
Correct
5of 22

proMOTION
™
Dynamic Vibration
Technology
How Does it Work?
proMOTION™ Dynamic Vibration
Technology uses high strength Vectran®
cables, which have been specically
chosen for their unique ability to optimize
the transfer of vibrations generated by the
platform directly to the targeted muscle
Specically upper body muscles. The
Vectran cables, which pound-by-pound
are ve times stronger than steel, transfer
vibration to the upper body at high-
speed frequency rates between 25 and 50
times per second. Vectran®is a registered
trademark of Hoechst Celanese Corp
Each cable extends to more than two
meters, allowing for movement in all
planes and directions, so you can enjoy
full range of movement and innite
dynamic exercise options. WARNING:
Do not extend the cables more than 215
cm/84.64 inches (passed the colored line).
Doing so will void your warranty.
Not only does the proMOTION optimize
the transmission of vibrations to the
targeted muscle but it also oers variable
resistance. Changing the resistance puts
more load on the muscles during the
most intense part of the movement,
oering another way to progress your
Shift lever towards you to
increase; shift away from you to
decrease
Vectran®is a registered trademark of Hoechst Celanes Corp.
workouts by increasing the intensity of
your training. The unique gel dampening
system also increases the resistance
as you work harder.
Usage and Guidelines
As with any form of exercise, cease your
session immediately if you feel faint, dizzy
or ill while working out on the Power Plate®
machine. Always consult a doctor or spe-
cialist before restarting your training.
As with any training routine, it is important
to start using the proMOTION on the
lowest level of resistance and with the
Power Plate machine on a low setting.
Establish a good technique before
increasing the setting on the proMOTION
from low to high.
Always pay special attention to your
posture, especially the position of your
wrists and back, and how close you are
to the Power Plate machine. proMOTION
cables should be used within the normal
range and angle to maximize training
eect and prevent damaging the device.
The cables should be used between
90 degrees and 25 degrees from the
horizontal base.
6of 22

Technical Specications
Power Plate®Pro7M™Technical Specications
Color (standard) Graphite or Silverstone
Dimensions (W x D x H) 96cm x 116cm x 152cm / 38in x 46in x 60in
Base surface area 0.50sq.m
Weight 202kg / 445lb (Est.)
Power Supply 100-240V, 50/60 Hertz, Universal Voltage
Nominal Power 160-265VA
Maximum Load 227.2kg / 500lb
Operation User-friendly interactive touch screen computer
DualSync Twin Motor System DualSync Twin Motor System maintains precise balance at any
frequency and amplitude level, allowing perfect synchronization of
vibration for maximum muscle response and eciency.
PrecisionWave Technology High-delity harmonic vibration system that provides
uncompromising performance for unsurpassed results.
Frequencies 25 - 50 Hertz
Time Selections 0 - 9 minutes (15 second increments)
Amplitude High or Low
G-Factor 0 25Hz/Low Amplitude G-Factor 5 35Hz/High Amplitude
G-Factor 1 30Hz/Low Amplitude G-Factor 6 40Hz/High Amplitude
G-Factor 2 35Hz/Low Amplitude G-Factor 7 45Hz/High Amplitude
G-Factor 3 40Hz/Low Amplitude G-Factor 8 50Hz/High Amplitude
G-Factor 4 30Hz/High Amplitude
Certications MDD, CE2797 and EMC (TUV certied); RoHS / WEEE compliant, NRTL,
C-NRTL, FDA listed as Class 1 device, 510k exempt
Computer & Software Specications
Operating System Windows CE 6.0 Professional
Software Power Plate proprietary Pro7M software
Graphics Built in Graphic Card
Total Memory Mobile-DDR SDRAM (133Hz) : 256MB
Hard Drive (Storage) 32GB micro-SD card
Screen/Monitor 10.1’ LCD Display (800 x 480 pixel)
USB Port/Hub USB 1.0 Port
LAN port 10/100M Ethernet
Audio 2 Recessed Speakers
proMOTION™ Dynamic Vibration Technology Specications
Resistance Levels 7 resistance levels. Pull toward front of the machine to increase.Push
toward back of the machine to decrease.
Cable Maximum 2.2m / 98.4in
proMOTION Embedded
Computer & Software Specications
Exercise Modalities Program, Express Workout, Single Exercise, Manual
3 Goals Feel Better, Look Better, Play Better
Number of programs 25
Levels/Durations 5
User proles Thousands
Usage Control Yes; Password enabled locking system
7of 22

WARNING: THE MACHINE
CONTAINS NO USERSERVICEABLE
PARTS. PLEASE CONTACT AN
AUTHORIZED POWER PLATE®
SERVICE PROFESSIONAL
FOR ANY MAINTENANCE OR
TROUBLESHOOTING NOT
OTHERWISE DESCRIBED BELOW,
OR PERSONAL INJURY OR DAMAGE
TO THE MACHINE COULD RESULT.
CAUTION: Any changes, modications
or unauthorized maintenance performed
to or on the machine could void the
product warranty.
Always unplug the machine and then
wait for at least one minute before
performing any maintenance.
Clean the machine only with a moist
cloth. Do not use sharp objects, bristles,
scrubs or acid-based detergents, which
will damage the lacquer nish.
Do not spray any cleaning solution
directly onto the machine. Instead,
moisten a cloth, then apply to the
machine.
On plastic components, use only
polishes specically designed for
plastic. Use a soft brush, not a cloth,
to clean the platform, including the
contoured mat.
Clean electrical components only with
a dry cloth, in order to avoid the risk of
shock or damage to the components.
Troubleshooting
Always unplug the machine and then
wait for at least one minute before
performing any troubleshooting.
If the machine is not operational and the
console display is not illuminated, check
the power connections to the machine.
If all power connections are working,
please check the circuit breaker or fuse
for the electrical outlet that is supplying
power to the machine in order to ensure
that the outlet is receiving electrical power.
If the console display is illuminated, but
the platform will not vibrate, unplug the
machine, wait at least one minute, then
check the connection to the machine’s
motor, located under the base of
the machine.
If the machine makes a rattling noise,
unplug the machine, wait at least one
minute, then check to make sure that
the feet are properly tightened.
DANGER:The user not try to resolve
any electrical issues regarding the power
source to the machine. Instead, Power
Plate requests that the user contact
a licensed, professional electrician to
conduct any examination and make any
necessary repairs. Otherwise, serious
personal injury or property damage
could result. No modication for this
equipment is allowed.
APPLIED PARTS: applied parts are not
intended to deliver heat during normal
use. The maximum temperature for
applied part may be up to 43degrees
Celsius / 109.4 degree Fahrenheit.
!
!
•
•
•
•
•
•
•
•
•
•
!
Maintenance
Applied Parts:
• Platform
• Handlebars
Power On / O:
• To Power On the Machine: Press the
rocker switch located on lower back of
the column to indicate “I” position.
• To Power O the Machine: Press the
rocker switch located on lower back of
the column to indicate “0” position.
• To Power O the Main Control Panel:
While power to the machine is on, press
and hold the green power button located
on the lower half of the main control
panel. This will power o only the main
control panel for standby mode. Press
and hold the green power button again
to power on the main control panel.
8of 22

Pro7M Box Contents
Straps Set (2pcs)
Exercise Mat
Protection Cover
Power Cord
Handgrips (2pcs)
Box Contents
9of 22

Guidance and manufacture’s declaration – electromagnetic emissions-
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The Pro7M is intended for use in the electromagnetic environment specied below. The customer or the
user of the Pro7M should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
Group 1 The Pro7M use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interfe rence in nearby electronic
equipment.
RF emission CISPR 11 Class B Class A with IEC61000-3-2
Complies with IEC61000-3-3
Harmonic emissions Class A The Pro7M is suitable for use in all establishments,
IEC 61000-3-2 including domestic establishments and those directly
connected to the public low-voltage power supply
Voltage uctuations/ Comply network that supplies buildings used for domestic
icker emissions purposes.
IEC 61000-3-3
Wi-Fi Specications Network Standard: IEEE 802.11b/g/n
Band: 2.400-2.4835GHz
Cover Range: 20meter
Wi-Fi Power: 10dBm
EMC Declaration
10 of 22

Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro7M is intended for use in the electromagnetic environment specied below. The customer or the user of
Pro7M should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or
discharge (ESD) ±8 kV air ±8 kV air ceramic tile. If oor are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical fast ±2 kV for power ±2kV for power Mains power quality should be that of
transient/burst supply lines supply lines a typical commercial or hospital IEC
61000-4-4 environment.
Surge ± 1 kV line(s) t ±1 kV dierential Mains power quality should be that of
IEC 61000-4-5 o line(s) mode a typical commercial or hospital
environment.
Voltage dips, short <5% UT<5% UTMains power quality should be that
interruptions and (>95% dip in UT) (>95% dip in UT) of a typical commercial or hospital
voltage variations for 0.5 cycle for 0.5 cycle environment. If the user of the Pro7M
on power supply requires continued operation during
input lines 40% UT40% UTpower mains interruptions, it is IEC
61000-4-11 (60% dip in UT) (60% dip in UT) recommended that the Pro7M be powered
for 5 cycles for 5 cycles from an uninterruptible power supply or a
battery.
70% UT70% UT
(30% dip in UT) (30% dip in UT)
for 25 cycles for 25 cycles
<5% UT<5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec
Power frequency
3A/m
3A/m
Power frequency magnetic elds should
(50Hz) magnetic be at levels characteristic of a typical
61000-4-8
location in a typical commercial or
hospital environment.
NOTE UTis the a.c. mains voltage prior to application of the test level.
11 of 22

Guidance and manufacture’s declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro7M is intended for use in the electromagnetic environment specied below. The customer or the user of
Pro7M should assure that it is used in such an environment.
Immunity test IEC 60601 test Compliance Electromagnetic environment
level level - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Pro7M, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF 3 Vrms 3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz
Radiated RF 3 V/m 3 Vrms 80 MHz to 800 MHz
IEC 61000-4-3 80 MHz to 2.5 GHz 800 MHz to 2.5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption
and reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site
survey should be considered. If the measured eld strength in the location in which the Pro7M is used exceeds
the applicable RF compliance level above, the Pro7M should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the Pro7M.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
Pd 167.1=
Pd 167.1=
Pd 167.1=
12 of 22

! Electromagnetic Compatibility (EMC)
Precautions
The Pro7M model is suitable for use in
commercial medical establishments,
including those directly connected to
the public low-voltage power supply
network that supplies buildings used for
domestic purposes. Both models have
been tested against, and have passed
the applicable requirements of relevant
electromedical standards, including EN
60601-1-2:2007.
The Pro7M model emits electromagnetic
energy to perform its intended function.
Nearby electronic equipment may be
aected by this emission. Similarly
the Pro7M model may be aected by
electromagnetic emissions from other
equipment in the vicinity.
The Pro7M model is intended for use
in the electromagnetic environment
specied below. The customer or user of
the Pro7M model should ensure that it is
used in such an environment:
– Floors should be wood, concrete or
ceramic tile. If the oor is covered with
syntheticmaterial, therelative humidityin
the room should be at least 30%.
– Power quality should be that of a typical
domestic, commercial or hospital
environment as appropriate.
– The Pro7M model should not be used
during power interruptions.
– Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
The Pro7M model is intended for use in an
electronic environment in which radiated
radio-frequency (RF) disturbances
are controlled.
The customer or user of the Pro7M model
can help prevent electromagnetic
interference by maintaining a minimum
distance between portable and
mobile RF communications equipment
(transmitters) and the Pro7M model
as recommended in this user manual,
according to the maximum output power
of the communications equipment:
The adequate separation distance
to provide base immunity to RF
disturbances is 1.0m.
• USB Port - USB port supplies 5Volt DC
current. This port is only used for soft-
ware updates and uploading personal
pictures to user proles. The operator
must take care to not touch the USB port
and a patient simultaneously.
• LAN port - the Pro7M has included a
LAN port on the lower back side of
the machine. This LAN port is used
to connect a LAN cable to access the
internet.The operator must take care to
not touch the LAN port and a patient
simultaneously.
• Phone jack adapter- On the Control
Panel of the Pro7M, there is a phone
jack adapter for sound playback. The
operator must take care to not touch
the Phone jack adapter and a patient
simultaneously.
Retain this user manual for your future
reference.
13 of 22

The following symbols may appear in this manual or on the machine. Some of
the symbols represent standards and compliances associated with the machine
and its use.
Caution: Consult accompanying documents
CE Mark: conforms to essential requirements of the Medical Device Directive
93/42/EEC.
Follow instructions for use
Date of manufacture.
Manufacturer
Type B applied part
DISPOSAL: Do not dispose this product as unsorted municipal waste. Collection
of such waste separately for special treatment is necessary.
CB Certication Scheme
C-TICK Certication from the Australian Communications Authority
GOST - Standard of Russia
Compliant with RoHS Directive 2002/95/EC
CQC Safety and EMC Product Certication Mark
Compliance with DENAN law Japan
Symbol Descriptions
CE 2797
14 of 22

Safety Classication
1. Class I equipment;
2. Type B applied part;
3. IPX0;
4. Not category AP / APG equipment;
5. For 60601-1 2nd edition, Mode of operation: Short Time Operation : Max. of 4
minutes / This is for North America and Canada
6. For 60601-1 3rd edition, Mode of operation: Non-continuous Operation : 9 mins
on/ 3 hours o / This is for Europe
Environmental Requirements
OPERATING CONDITIONS
Temperature: 10ºC to 40ºC
Humidity: 10% to 95% RH
Pressure altitude: Normal atmospheric
conditions/ 860-1060hPa
STORAGE AND SHIPPING CONDITIONS
Temperature: 10ºC to 40ºC
Humidity: 10% to 95% RH
Pressure altitude: Normal atmospheric
conditions/ 860-1060hPa
WARNING
The device is not suitable for use in the presence of ammable anesthetic mixtures
with air or with oxygen or nitrous oxide.
The device requires no calibration.
The device is not repairable and contains no user serviceable parts.
The user must check that the equipment functions safely and see that it is in
proper working condition before being used.
The manufacturer does not require such preventive inspections by other persons.
The plug is used to disconnect from the main supply. Do not position the machine
so that it is dicult to disconnect the main plug.
Warning: To avoid the risk of electric shock, this equipment must only be
connected to a supply main with protective earth grounding.
Users
The machine is intended for use by a diverse patient population including:
• Adult men and women who are medically cleared and able to undertake
physical exercises such as weight bearing and ground reaction force type
exercise programs. In general this applies to healthy adults (16+ years old).
• All others users should be cleared by their physician and only use Power Plate
under medical supervision.
Facilities
The device is intended to be used in all types of facilities including, but not limited
to:
• Domestic • Hospitals, Clinics
• Hotels, Spas, Resorts • Rehabilitation Centers
• Health Clubs • Active Aging Communities
15 of 22

Radio Frequency Interference Requirements-FCC
Note: This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However
there is no guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to radio or television reception, which
can be determined by turning the equipment o and on, the user is encouraged to
try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna
• Increase the separation between the equipment and receiver
• Connect the equipment into an outlet on a circuit dierent from that to which
the receiver is connected
• Consult the dealer or an experienced radio/TV technician for help.
Radio Transmitters (Part 15)
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) this device may not cause harmful interference, and (2)
this device must accept any interference received, including interference that may
cause undesired operation.
FCC RF Exposure Guidelines
Safety Information
Reducing RF Exposure - Use Properly
Only operate the device in accordance with the instructions supplied.
This device complies with FCC radiation exposure limits set forth for an
uncontrolled environment. In order to avoid the possibility of exceeding the FCC
radio frequency exposure limits, human proximity to the antenna shall not be less
than 20cm (8 inches) during normal operation.
Changes or modications not expressly approved by the party responsible for
compliance could void the user‘s authority to operate the equipment.
Radio Frequency Interference Requirements-Canada
This Class B digital apparatus complies with Canadian ICES-003.
Radio Transmitters
This device complies with RSS 210 of Industry & Science Canada. Operation is
subject to the following two conditions: (1) this device may not cause harmful
interference and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Label Marking: The Term “IC:” before the radio certication only signies that
Industry Canada technical specications were met.
Appendix – FCC / IC statement
16 of 22

Pro7M Touchscreen controls
The Pro7M software is organized in four modules: Complete Programs, Express workouts,
Single Exercises, Manual Mode.
Complete Program:
• Select your goal between Feel Better,
Look Better or Play Better
• Select a specic Program
• Select your level and duration of the
program between 10 and 30 min.
Express Workouts:
• Select a mini program
Single Exercise:
• Select an element between Stretch,
Core, Balance, Strength and Massage
• Select a specic exercise.
Manual Mode:
• Select a duration
• Select the G-Factor
This button takes you to the education
screen. It is broken down into 5 educa-
tional videos
3 Goals
5 elements
G-Factor
New to Power Plate
Welcome to the Pro7M
I.1 Home page
I.1.1 Education
17 of 22

This button takes you to the users’ login screen. It is broken down into 2 areas
• New Users • Existing User
The setting module is reserved for the
commercial facility manager. It will be ac-
cessible via a private button and login.
User name: admin
Password: Asyn086Si This button takes you to the setting
module. The settings are broken down
into 4 areas
• Machine Conguration
• Education/Help
• Service Menu
• Contact
I.1.2 User login
I.1.3 Settings
Set Login Settings
18 of 22
Table of contents
Other Power Plate Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Actia
Actia LEXIA3 EXPERT user guide

Direct Supply
Direct Supply Attendant SFP1 owner's manual

swissflex
swissflex uni 20_75RF operating instructions

Physio Control
Physio Control LIFEPAK 15 Service work instructions

Stryker
Stryker Secure Connect Operation & maintenance manual

Stryker
Stryker 6370 Operation manual

GENICON
GENICON REUSABLE MONOPOLAR X-SURGE SHAFT HANDLE Instructions for use

iCRco
iCRco VERTX Service manual

NanoVibronix
NanoVibronix PainShield quick guide

Boston Life Labs
Boston Life Labs H-Box user manual

Zimmer
Zimmer Soleo Sono Instructions for use

St. Jude Medical
St. Jude Medical CPS AIM 410140 Instructions for use