Prestige medical Anima User manual

User Manual
Vacuum &
Non-Vacuum Autoclave
www.dlc.com.au

304278B
Please read these instructions before using the
autoclave.
Keep these “Instructions for use” in a safe place
close by the unit for future reference.
Service Helpline: Telephone: 01254 844 116
e.mail: [email protected]
Prestige Medical Service staff are available to provide
advice and assistance during normal office hours. To avoid
delays when making contact please have the unit’s Model
and Serial Numbers at hand. These can be found on the
rating plate located at the rear of the unit or on the right
hand side of the casing when the door is open.
For additional information visit
www.prestigemedical.co.uk
UK Customers
Prestige Medical Limited
East House, Duttons Way,
Shadsworth Business Park,
Blackburn BB1 2QR
Tel : +44 (0) 1254 682 622
Fax : +44 (0) 1254 682 606
www.prestigemedical.co.uk
Contents.
Section Page
1 Introduction 3
2 Operating Symbols, Displays
and Controls 4
3 Operation 4
4 Essential Information 8
5 Routine Care and Maintenance 9
6 Trouble Shooting 10
7 Recovery Sequences 11
8 Additional Information 12
9 Warranty 13
10 Printer Instructions (Optional) 13
11 Technical Specification 14
12 Water Quality 15
13 Loading the Autoclave 15
14 Daily Record Sheet 16
Instructions for use
Destination Kit Date of purchase
Vessel rating plate (duplicate)Model no.
Serial no.
Overseas
Customers
Contact your local
distributor. In case of
doubt contact Prestige
Medical.
0086
Prestige Medical Limited
East House, Duttons Way,
Shadsworth Business Park,
Blackburn BB1 2QR
Tel : +44 (0) 1254 682 622
Fax : +44 (0) 1254 682 606
www.prestigemedical.co.uk
Registered in England
Reg No. 2826793
2

304278B
Thank you for choosing the Prestige Medical Anima
autoclave.
The door remains closed throughout the cycle,
including drying.
An “Auto-cycle start” option is included which allows the
user to programme the time of day at which the selected
cycle should start.
Before removing the product from its packaging, refer to
the “Manual Handling” section on page 12.
Whilst unpacking check the unit for transit damage.
If damage is found, please report this to the shipping
agent immediately, in writing and then notify your dealer.
Product contents will contain the following:
•Autoclave with internal furniture
•Handbook and warranty card
•Performance Test Certificate and Certificate of
Conformance for pressure vessel (all UK models).
•TST test strips
•Waste water container
Types of load and loading.
This autoclave has been qualified to sterilize loads as
defined by EN13060.
Cycles P1 & P2 – Wrapped, pouched, solid, hollow
instruments (length to diameter ratio 750:1 Length may be
doubled for tubes that are open at both ends) and porous
items.
Cycle P3 – Unwrapped, solid, hollow instruments (length
to diameter ratio 750:1 Length may be doubled for tubes
that are open at both ends). A process challenge device
such as a Bowie Dick Helix can be used for validation of
cycles P1 and P2.
Cycle P4 – 121°C for wrapped, pouched, solid, hollow
instruments and porous items.
Cycle P5 – Unwrapped solid instruments only.
WARNING!
Refer to the instrument manufacturer to ensure the instruments
suitability for autoclaving and the maximum temperature they
can withstand.
•A “responsible person” must qualify other loads as
suitable. Refer to “Additional Information”.
•Refer to “Technical Specifications” for the maximum
instrument load for the autoclave.
•All instruments must be cleaned prior to sterilizing.
•Wrapped or pouched loads should not touch adjacent
loads when placed in the rack. Pouches must be used
for only one item.
•When placing on a tray, ensure items do not touch
each other and the load does not touch the other trays
or chamber in any way.
•Always use the lifting device when removing trays from
the autoclave, as they may be hot. Long trays should
be supported at their rear as they become free of the
tray carrier. Do not use an unprotected hand to hold
hot trays.
Section 1: Introduction
3

304278B
A
D
F
G
H
L1
L2
L3
B
C
S1
P2 P3
P4P1
P5
N
Pouched instruments should be placed on pouch trays for
best drying results
Do not use an unprotected hand to hold hot trays.
The following descriptions refer to the symbols,
controls and displays shown on this page.
Symbols:
S1 IMPORTANT: Read Operating Instructions
before use.
S2 Do not use tap water.
S3 Maximum water fill line.
S4 Fill water tank.
S5 Hot parts (behind) Do Not Touch.
Also behind door (L).
General Controls / Operations:
Note: The function is selected when the LED
adjacent to the button is illuminated.
AStandby / Ready button.
BCycle start button.
CDoor open button.
DSelect time/date/cycle counter.
EAdjustment time/date settings.
FAccept button time/date/settings.
GSelect vacuum cycles.
HSelect Non-vacuum cycle.
IDoor moulding.
JFuses (located behind access panel).
KPrinter
LFresh water tank drain.
MFresh water tank fill
Section 2: Operating Symbols, Displays and Controls
IMPORTANT: Read
Operating Instructions
before use.
Hot parts.
Do Not
Touch.
Do not use tap water
(S2 opposite).
For acceptable water
quality refer to Section 13.
Fill water tank when
LED illuminated
(Maximum fill level is
S3 opposite).
S2
S5
I
L
S5
S1
M
S3
J
K
4

304278B
Please take time to read these instructions
before using the autoclave. It is essential that the
operator is correctly trained and a “Responsible
Person” has been assigned for the management of
the autoclave. By following these simple step-by-step
instructions you can ensure your instruments are correctly
sterilized every time.
1. Installation.
Ensure the unit
is placed on a
strong, flat and level
surface.
To check if the unit
is level, pour half a
cup of water into
the chamber. The
water should flow
towards the rear of
the chamber, not
out of the front.
WARNING!
The autoclave is heavy; at least two people will be needed
to lift it.
2. Connection.
Plug the unit into the mains outlet socket of the correct
type & rating. Refer to the duplicate rating plate located
inside the door on the bezel.
After a few seconds the LED next to button (A) illuminates.
WARNING!
The mains outlet MUST BE EARTHED (GROUNDED).
The mains plug should always be easily accessible as it is to be
relied upon as “the means of disconnection”.
3. Single Use Water System.
This unit has a “single use” water
system designed to prevent the
recycling of any contaminants, which
may be present on the instruments
being sterilized.
Ensure the “waste water container”
is checked regularly to prevent
overflowing. The container should be
emptied when the wastewater reaches
the “Max” line.
To empty the container, undo the screw cap and carefully
remove the steam-condensing coil, which passes through
the cap.
Cycle controls.
Press G to select the vacuum cycles these are as follows:
P1: 134°C/3min.20sec Vacuum cycle
for porous loads, wrapped, pouched,
solid / hollow instruments, with 25
minutes drying.
P2: 134ºC /3min.20sec. Vacuum cycle
for porous loads, wrapped, pouched,
solid / hollow instruments with
20 minutes drying.
P3: 134ºC /3min.20sec. Vacuum
cycle for unwrapped, solid / hollow
instruments without drying. Also
Process challenge device cycle.
P4: 121ºC /15min. Vacuum cycle for
porous loads, wrapped, pouched,
solid / hollow instruments with
25 minutes drying.
Press H to select non-vacuum cycles.
P5: 134ºC /3min.20sec. Non Vacuum
cycle for unwrapped, solid instruments
without drying.
AFTER EVERY CYCLE THE UNIT RETURNS TO THE
DEFAULT CYCLE P1.
Display Options. (Figures shown are for a vacuum unit)
L1 Display for “Time” “Date”
& “Cycle counter”.
The normal display mode
is time. The date or cycle
counter can be selected
by pressing the button D
adjacent to the display.
The selection is valid until
a cycle is completed or
until the unit is returned to
standby mode, at which
point the display will revert
to the current time.
L2 Display for “Temperature”
L3 Display for “Pressure”.
The following show the cycle status: (Vacuum unit)
L4 Cycle started.
L5 Heating and air bleed.
L6 Sterilizing.
L7 Depressurisation / drying.
L8 Cycle complete.
L1
L2
L3
L8
L7
L5
L4 L6
Section 3: Operation
5

304278B
Place on a heat resistant surface whilst emptying the
container.
Use the carry handle to support the container when emptying.
Replace the steam-condensing coil and cap; ensuring it is
screwed on securely.
Note: Before using for the first time and after
emptying, always add water until the level reaches
the minimum level mark.
The wastewater container can be located in any
appropriate place although care should be taken
that it cannot be knocked over.
WARNING!
Waste water and steam condensing coil may be hot.
Care should betaken at all times.
4. Setting Date and Time.
Date and time are set in the following sequence.
Note: The 24 hour clock is used.
Year (tens): Year (hundreds): Month: Day: Minutes: Hours.
•Set the autoclave in “Ready Mode” by pressing ready
button (A). The LED next to button (A) will then go out.
•Press and hold button (D) for 5 seconds.
•Set year (tens) by pressing button (D) up and (E) down.
Press button (F) to accept.
•Set year (hundreds); Month; Day; Minutes; hours; by
pressing button (D) up and (E) down. Press button (F)
to accept.
•The unit returns to “Ready Mode” when the hours are
accepted.
5. Ready.
Press button (A) to set the autoclave into “Ready
Mode”.
The LED next to button (A) will go out.
Note: When in “Ready Mode”, the boiler and chamber
are kept warm.
6. Water fill – DO NOT USE TAP WATER.
Before using the autoclave for the first time fill with water.
Press button (C) to open the door. Pour water into the fill
spout (M) until it reaches the “Maximum line” (S3). The
water capacity is 3.7 litres.
When the low water indicator illuminates (S4), top up with
water.
Always use de-ionised, distilled or sterile water as
recommended. Never use tap water.
7. Loading.
Refer to “Type of load and loading” on page 3
“Specifications” on page 15 for the maximum permissible
load.
Failure to follow instructions may cause the unit to
malfunction and result in an unsuccessful cycle.
Before loading, ensure instruments are cleaned and rinsed.
Always use the instrument trays or racks, which are supplied.
Load instruments so that they do not touch other
instruments or the chamber and are resting on the ribs of
the tray.
Only one item should be placed in a pouch.
Linen loads should be formed into a cube of size
150mm.x 120mm. x 100mm. (maximum), weighing no
more than 1 kg. Place on an instrument tray in the middle
of the chamber.
Linen loads should not be sterilized in combination with
other loads.
8. Door closing.
Once the trays are in place close the door by pushing until
a “click” is heard. The LED next to the door open button
(C) illuminates.
9. Cycle Options.
After each cycle the unit returns to the default cycle “P1”
134ºC /3min.20secs. Vacuum cycle for porous loads,
wrapped, pouched, solid / hollow instruments with drying.
Cycle options are:
Press G to select the vacuum cycles these are: -
134ºC /3min.20scs. Vacuum cycle for
porous loads, wrapped, pouched, solid /
hollow instruments with 25 minutes drying.
134ºC /3min.20sec. Vacuum cycle for porous
loads, wrapped, pouched, solid / hollow
instruments with 20 minutes drying.
134ºC /3min.20sec. Vacuum cycle for
unwrapped, solid / hollow instruments
without drying. Also Process challenge
device cycle.
121ºC /15min Vacuum cycle for porous
loads, wrapped, pouched, solid / hollow
instruments with 25 minutes drying.
Press H to select the non vacuum 134 /
3 min cycle.
6

304278B
10. Start Cycle.
Press button (B) to start a fully automatic cycle. A visual
display shows the stages of the cycle:
Stage 1 Cycle started (L4)
Stage 2 Heating and air bleed (L5)
Stage 3 Sterilizing (L6)
Stage 4 Depressurisation / drying (L7)
Stage 5 Cycle completed (L8)
11. Opening door.
At the end of a cycle a buzzer sounds 3 times. Press
button (C) to open the door, allowing access to the load.
Additional Operations.
Steam Penetration Test.
The “Steam penetration test” should be performed on a
daily basis to confirm that the unit is operating correctly.
A Bowie Dick Test Pack or a Test Helix MUST be
used with this test (available from Prestige Medical).
The Pack or Helix should be placed on an instrument
tray in the middle of the chamber towards the front.
Albert Brown Test Helix Bowie Dick Test Pack
Run the P3 for the steam penetration test.
Following this test, the Test Pack or Helix TST should
change to a uniform purple colour. If the TST is not a
uniform purple colour check the door gasket and vessel.
Repeat the test.
If the test fails again seek technical assistance.
Vacuum system leak test.
The “vacuum system leak test” checks the integrity of the
vacuum system. NB. A printer MUST be fitted.
The test MUST be undertaken when the unit is cold
and dry (before any other cycle has been run).
Set the autoclave in “Standby Mode” by pressing ready
button (A). The LED next to button (A) will illuminate.
Press and hold button (G) for 6 seconds. The display
(L1) will begin to count down. Hold button (G) until the
countdown reaches zero. The test will then start.
During the test, the segments of the cycle status graphic
(S6) will flash. The test will take been 15 to 30 minutes. On
completion the print out will advise if the unit has passed or
failed the test.
If a fail occurs clean the gasket and the chamber rim before
repeating the test. If the unit repeatedly fails contact Prestige
Medical or your dealer.
Auto-cycle start.
The unit can be programmed to enable any cycle to be
started at any time of the day when the unit is unattended.
This option is recommended particularly when the
operator performs “Air leak detection test”.
•Set the autoclave in “Standby Mode” by pressing ready
button (A). The LED next to button (A) will illuminate.
•Press and hold button (D) for 6 seconds and release.
The time display (L1) will flash with the default time of
06.30 being displayed.
•Press (F) to accept. If another time is required for the
auto-cycle start, change minutes then
hours; by pressing button (D) up and (E) down.
Accept by pressing button (F).
•The unit returns to “Ready Mode” and the LED to the
left of the clock symbol continues to flash.
•Select the required sterilizing cycle using buttons (G),
(H).
•Press the start button (B). The door LED (C) and the
start LED (B) both flash. The clock symbol LED is still
flashing.
•The selected cycle will start when the programmed
start time is reached. The clock symbol LED flashes
until the cycle is completed.
•To abort a timed cycle (before the cycle starts) press
the “Standby button” (A).
Note:
If the unit is used in a very cold environment, water
vapour may be seen coming from the cooling fan.
This is normal and will only last a few minutes.
7

304278B
To ensure the autoclave continues to operate correctly
it is important to follow the following points and to carry
out the necessary care and maintenance procedures as
specified.
Do ensure that..…
…you read and follow these operating instructions.
…the load is suitable for sterilizing and the cycle selected.
…the load can be sterilized at the selected temperature.
…the load has been cleaned.
…the load has been rinsed thoroughly in clean water prior
to sterilization to avoid any chemical residues left after
cleaning contaminating the autoclave.
This product is not a washing / cleaning machine.
…when placing instruments on trays, ensure that they are
placed on the ribs of the tray (to help drainage), they
must not touch each other and must not interfere with
other trays or the chamber above.
…only distilled, de-ionised or sterile water is used (as
recommended).
…the autoclave is in a draught free area.
…the autoclave is not installed in an enclosed cupboard
space.
…all other exterior product panels are 50mm. clear of
adjoining surfaces to allow air circulation.
…the door is left ajar when not in use.
…you quote model / serial number (which are located
inside the door) and date of purchase in all
correspondence.
…only qualified personnel service the autoclave.
It is recommended that a Chemical Indicator strip
be used every cycle to verify that the sterilizing
cycle is effective. If the Chemical Indicator strip fails
to change colour repeat the cycle. If it still fails to
change colour then arrange for a service.
Do not….
…lose this handbook.
…lose this handbook.
…add any chemicals whatsoever to the water.
…attempt to sterilize volatile substances, toxic materials
or other unsuitable loads. Refer to your “Responsible
Person” for advice.
…place the autoclave in direct sunlight.
…place the autoclave on heat sensitive surfaces.
…use inappropriate cleaning materials.
…drop or abuse the autoclave.
…use in areas of risk associated with flammable materials
or gases.
…remove the casing or attempt to service or repair the
autoclave.
Section 4: Essential Information
8

304278B
WARNING!
Disconnect the autoclave from the mains power
supply before cleaning.
Daily Maintenance
Gasket
THE GASKET MUST BE CLEANED ON A DAILY BASIS,
BEFORE USING THE AUTOCLAVE.
Wipe exposed surface of the gasket and
the surface of the vessel with warm soapy
water using a lint free damp cloth.
Wipe both the gasket and the vessel again
with water using a lint free damp cloth to
remove any residual soap.
EVERY 250 CYCLES.
Clean the autoclave using Autoclave cleaning kit
(Part No. 289138A).
(Part No. 289138).
WARNING
Failure to perform these procedures may result in the unit
showing UOD1 on the display (UOD1 indicates that the unit may
have been contaminated and could fail to operate correctly).
Change the air filter situated at the
front of the autoclave.
Gently unscrew the filter, and fit the
replacement.
Exterior surfaces (as required).
Exterior surfaces should be cleaned with warm soapy
water using a damp cloth.
For persistent marks, use a gentle cream cleaner.
Gasket Replacement
Fresh water tank / Waste water container.
On a monthly basis, fully drain the water tank and leave
overnight (use Autoclave cleaning kit Part No. 289138).
Drain, then refill with fresh water. Repeat this operation
twice more to remove any residue.
Always use de-ionised, distilled or sterile water as
recommended. NEVER USE TAP WATER.
Routine Maintenance.
Annually replace the water filters.
Should the gasket develop a persistent leak it should be
removed, cleaned thoroughly in warm soapy water and
shaken dry. Wiping with a lint-free cloth is acceptable
(Other materials may lead to contamination of the gasket
with fibres).
The door gasket plate must also be cleaned.
If the leak persists you should obtain and fit a new gasket.
To remove the gasket, undo the two thumb-nuts (309070)
in the centre of the gasket plate.
Remove the plate/gasket assembly and then remove the
gasket from the plate.
CAUTION: Ensure the gasket is located the right way
around on the plate, the lip should face outwards.
During re-assembly, place new black sealing
washers (supplied with the gasket) in the machined
recesses in the door casting.
Ensure that the entry port plug aligns with the hole in the
cast lid.
Prior to tightening the two thumb-nuts ensure that the
gasket is correctly seated on the gasket plate and in the
door recess.
DO NOT over-tighten the thumb-nuts as this may damage
the thread.
Section 5: Routine Care and Maintenance
Thumb-Nuts
Sealing Lip - ensure
the gasket is
orientated correctly
Entry port plug
9

304278B
User Message Cause Remedy.
UOD1 and UOD2 Contamination. Conduct routine maintenance procedures.
Replace gasket.
If fault persists Engineer call-out required.
Low water LED (E2) Insufficient water to run a cycle. Press button (A) twice. Top up with water to mark (S3).
“door” illuminates
on display (L2)
Cycle start button pressed whilst door
is open or door micro-switch requires
adjustment.
If door closed but door LED (C) not illuminated - push on door to see if
LED can be illuminated.
If door open, close door and try again.
If fault occurs during a cycle – follow Recovery sequence “i”
01 Power failure during a cycle. Recovery sequence “i”
Check power supply – repeat the cycle.
b02 Sterilizing Temperature out of range. Recovery sequence “i”.
Clean gasket and chamber face then repeat the cycle.
If fault persists Engineer call-out required.
d02 Probe difference out of range.
Recovery sequence “i”.
Conduct routine maintenance procedures.
Clean gasket and chamber face then repeat the cycle.
If fault persists Engineer call-out required.
t02 Clock error. Recovery sequence “i”.
Check & rest the clock – repeat the cycle.
If fault persists Engineer call-out required.
p02 Sterilizing Pressure out of range. Recovery sequence “i”.
Clean gasket and chamber face then repeat the cycle.
If fault persists Engineer call-out required.
03 Air bleed has not been successful. Recovery sequence “i”.
Clean gasket and chamber face then repeat the cycle.
04 Vacuum failure. Recovery sequence “i”.
Clean gasket and chamber face then repeat the cycle.
07 Sensor fault – Boiler thermistor. Recovery Sequence “ii” – Repeat cycle.
If fault persists Engineer call-out required.
08 Sensor fault – Pressure transducer. Recovery Sequence “ii” – Repeat cycle.
If fault persists Engineer call-out required.
10 Water in the boiler. Recovery Sequence “iii” – Repeat cycle.
UOD3 Vacuum failure in drying stage. Recovery sequence “i”.
Clean or change gasket.
If fault persists Engineer call-out required.
13 Water fill time out. Recovery sequence “i”.
Drain water out of the fresh water tank and refill with de-ionised, distilled
or sterile water. Repeat the cycle.
14 Sensor fault – Chamber probe.
Recovery Sequence “ii” – Repeat cycle.
If fault persists Engineer call-out required.
15 System Leak. Recovery Sequence “ii” – Clean gasket and chamber face then
repeat the cycle.
If fault persists Engineer call-out required.
Should a fault occur a visual and audible indication would be given. The nature of the fault can be determined by reference
to the fault guide below. The recovery sequence allows access to any instruments within the autoclave and is the first step
when rectifying a fault mode.
As non-warranty related calls can be expensive it is advisable to ensure that all consumable items have been replaced or
cleaned as appropriate, and that the water quality is as described in Section 13 before contacting Prestige Medical Ltd.
Section 6: Trouble Shooting
10

304278B
Recovery Sequence (allows instruments to be removed from the unit).
Recovery Sequence “i” Recovery Sequence “ii” Recovery Sequence “iii”
Press button (A)
Stabilize (no flashing or bleeping)
Press button (A) “Recover”
Press button (A)
Stabilize (no flashing or bleeping)
NB: Cannot proceed. Switch off at mains supply.
Service required.
Press button (A)
Stabilize (no flashing or bleeping)
Press button (A) to enter “Ready Mode”
The Recovery Sequence (depending on where the fault occurs) will flush the water from the boiler and eventually
complete the cycle with a continuous bleeping to indicate recovery is completed.
Important.
Before restarting a cycle, check that the mains plug is fully inserted into the mains outlet socket and the outlet is of the
earthed/grounded type.
Should all power be lost the door cannot be opened until power is restored and pressure returned to ambient.
Should an internal power failure occur, the door cannot be opened until: -
1. The unit has cooled down to atmospheric pressure. Access can then be made using the supplied tool.
2. A service has been carried out.
In the event of failure of an indication device a service will be required to correct the condition.
Should a safety feature operate, unplug the unit and call for a service – do not attempt to correct the fault.
Primary safety features:
Two primary features have been fitted – a pressure release valve and a boiler over temperature safety cut out.
IMPORTANT NOTE ON EMERGENCY DOOR OPENING PROCEDURES.
If the door interlock lever does not fully release during door opening and the door led has gone out, light pressure on the
door will re-illuminate the door open led. It will then be possible to open the door in the normal manner using the door
open button.
Door opening – Power loss.
If the machine has suffered a loss of power and it is essential to open the machine to remove instruments, the following
procedure should be followed. This should only be performed by the nominated responsible person: -
1. Disconnect the unit from the mains supply.
2. Allow the machine to cool down to room temperature.
3. Remove the protective white protective cap on the top of the machine.
4. Using the hook on the special tool, release any pressure/vacuum inside the vessel by
hooking the ring on the pressure relief valve and pulling this towards you. As shown
opposite.
5. Insert the straight end of the special tool into the interlock lever release guide hole and
push inwards. If the tool starts to buckle at the preformed strain relief repeat from 3
above.
See Photograph opposite.
DO NOT ATTEMPT TO OPEN THE VESSEL IF PRESSURISED.
ONLY USE THE TOOL PROVIDED AND FOLLOW THE INSTRUCTIONS ABOVE
TO GAIN ACCESS TO THE MACHINE.
Section 7: Recovery Sequence
(allows instruments to be removed from the unit).
Door
release
tool hole
11

304278B
Operator:
The person assigned to use the autoclave.
Responsible Person:
The person who is responsible for the management of the
equipment, load assignment, care and maintenance. This
person is also responsible for ensuring that all applicable
Health & Safety Regulations are applied including those
relating to the pressure vessel.
This person must verify that only suitably qualified persons
undertake repair and maintenance work other than that
described under “Routine Care and Maintenance” within
this handbook.
Qualified Person:
A person who is qualified by training or experience to a
recognised level in respect of the work to be undertaken.
Service:
Calibration and maintenance as required.
Manual Handling:
Due to the weight of the unit two people are required when
unpacking or moving the product.
UNPACKING.
When lifting the unit out of the box ensure there is one
person on either side of the unit. Lift out of the box and
place on the work surface.
POSITIONING.
Start lifting by holding the unit below the front bezel (large
plastic moulding). As clearance is gained, lift at the other
corners. Place in position, and release in the reverse
order to lifting. NOTE Always drain the water tank before
moving. Before moving always allow 30 minutes after use
for the unit to cool down.
Cleaning Materials:
Mild washing up liquid. Non-abrasive cream cleaner.
Disinfectant diluted in water. Autoclave Cleaning Kit
289138.
Product decontamination.
Should the unit require repair, it must be decontaminated
in accordance with a recognised procedure prior to return
or on-site repair. A statement of equipment contamination
status must be available with the product.
Details of a suitable procedure are available on request
from Prestige Medical Ltd.
Approvals:
Approvals are all model specific. However, the following
standards apply in whole or part:
•Medical Devices Directive (MDD 93/42/EEC)
•Electro Magnetic Compatibility Directive.
•European Sterilizer Standard (EN13060)
•Pressure vessel to ASME Section 8.
•BS EN 61010 Parts 1 and 2
Spares:
Only those spare parts supplied or specified by Prestige
Medical should be used in the maintenance of the
autoclave. Use of unauthorised parts will invalidate any
warranty and may adversely affect the performance or
safety of the unit.
Accessories / Consumables
Printer roll (Part No.279505): Ten replacement rolls.
Furniture:
Tray / Rack lifter (Part No.279007):
Optional Furniture
5 Clip tray set (Part No. 309093)
Door tool (Part No.309068):
Door plate thumb screw (Part No.309070)
Sealing gasket x 4 (Part No 279571 )
Gasket for a 250mm. diameter chamber.
“O” Ring door (Part No.279100):
Waste Water Container (Part No.279503
Container for single use water.
Test Pack (Part No.309031)
Prestige Medical Ltd approved TST Helix available from:
Albert Browne Ltd.
Leicester
LE5 1QZ
UK
Order Code 3781.
Autoclave cleaner (Part No.289138):
For cleaning the autoclave to ensure continued operation.
Air filter (Part No.273096A):
Replacement air filter.
Section 8: Additional Information
12

304278B
Prestige Medical Ltd will, in the first twelve months from
the date of purchase (or 18 months from the date of
factory despatch – whichever is first), repair or replace free
of charge any parts* inclusive of labour which prove to be
defective in workmanship and / or materials.
In the following twelve months Prestige Medical Ltd
will replace free of charge any parts which prove to be
defective in workmanship and / or materials exclusive of
labour.
This warranty cover is dependent on authorised servicing
of the unit at least every 12 months and that Prestige
Medical Ltd received the warranty card within 3 months of
installation.
Prestige Medical Ltd will not be liable in the event
that the purchaser has failed to adhere to the
instructions contained herein or if the autoclave has
been abused, interfered with, altered, repaired or
serviced by any unauthorised party. This may result in
the protection provided by the equipment being impaired.
* This warranty excludes the door gasket, all internal
furniture and consumables.
Consumer’s statutory rights are not affected.
The Prestige Medical Ltd policy is one of continuous
development and as such reserves the right to change
the specification of the models and items illustrated and
described herein at any time.
Disposal
WEEE Statement
(Waste, Electrical and Electronic Equipment)
The WEEE directive places an obligation
on all EU-based manufacturers and
importers to take-back electronic
products at the end of their useful life.
Prestige Medical Limited accepts
its responsibility to finance the cost
of treatment of redundant WEEE in
accordance with the specific recycling
requirements.
The symbol (shown below) is present on all Prestige
Medical products, which indicates that the product must
NOT be disposed of with other waste. Instead it is the
user’s responsibility to dispose of their waste electrical and
electronic equipment by handing it over to an approved
reprocessor, or by returning it to Prestige Medical for
reprocessing. For more information about where you can
send your waste equipment for recycling, please contact
your local city office or Prestige Medical.
Power on procedure
Connect the printer to the power supply and switch on
at the mains. The LED light will illuminate on the printer.
Connect the printer to the autoclave with the supplied
cable. The printer connection at the autoclave is connector
K - see page 4
Replacing Paper Roll – Part Number 279505.
If the paper roll needs replacing, open the paper cup lid
and remove the remaining paper using the button, do not
pull paper through the printer mechanism. Reel off
a few centimetres from a new roll and check that the end
has a clean straight edge. Slide the leading edge of the
paper through the paper entry slot, with the leading edge
feeding forwards from the bottom of the roll, until you feel
resistance. Press the button and feed paper through the
paper exit slot. Sit the new paper in the paper cup and
close the lid.
Should the paper become creased or out of line when
feeding in a new roll, cut the end off the paper roll, feed
out the creased paper using the button, and reload
ensuring the paper has a clean straight edge.
Status LED.
The printer incorporates an LED indicator to report its
condition. If there is a fault the LED will flash in sequence.
The fault can be identified by counting the number of
flashes.
LED
Indication.
Condition Solution.
On Printer On -
Off Printer Off or Asleep -
* * * Paper Out Fit new paper
** ** ** Thermal head too hot Allow head to
cool
Place the printer on the worktop at the side of the
autoclave.
Do not place the printer on the top of the autoclave.
Make sure the paper is kept away from hot surfaces
such as the dump pipe etc.
Store the printouts in folders away from direct or
indirect sunlight.
Data Logger Instructions (Optional)
Refer to the instructions supplied with the data logger for
setting up and operating the data logger.
Section 9: Warranty
Section 10: Printer Instruction
(Optional)
13

304278B
Chamber component materials.
Vessel: Stainless Steel – 304 –S15
Boiler: Aluminium – LM25
Lid: Aluminium – ASME SB26 356.0 T6
Door Plate: Stainless Steel – 304 –S15
Chamber Capacities 22litre
Overall product width 480mm.
Overall product height 410mm.
Overall product length 610mm.
Unpacked weight 43Kg. (Max)
Chamber Diameter 250mm.
Chamber lengths 430mm.
Max. Instrument length 420mm.
Max. load non-vacuum 6 kg
Max. load vacuum
*Pack size - 3 x standard vet
pack @ 500g each
6 kg (un-pouched) 2 kg (pouched)
1.5 kg (porous)*
Sterilizing temp/time. 134ºC/3½mins.
121ºC/15mins.
Operating pressure
(Minimum)
2.05bar.(gauge)
Voltage/Wattage 230v. /2200W
Frequency 50 – 60 Hz.
Safety Shutdown:
Refer to “Temperature cut out”. Only a qualified person
should reset this.
Environmental conditions:
Indoor use at an altitude of up to 2,000m.
Ambient temperature range +10ºC to +40ºC
Maximum relative humidity 80% for temperatures up to
30ºC, decreasing linearly to 50% at 40ºC.
Mains supply voltage range 207 to 254volts.
Drying performance maybe affected by local environmental
conditions.
Maximum Water usage:
Vacuum cycles = 600mls per cycle (typical).
Non-Vacuum cycles = 450mls per cycle (typical).
WARNING
Disconnect the autoclave from the mains power supply
before changing fuses.
Only qualified personnel should replace fuses.
Mains plug top fuse (user replaceable)
F13A to BS1362 UK only.
Maximum single fault temperature:
143ºC determined by the pressure release valve.
Pressure Release Valve:
Operates at 2.9bar. Accumulation is <10%.
Storage:
When leaving the unit standing idle for any length of time,
switch mains off & drain the water tank especially if there
is a possibility of the room temperature dropping below
freezing point.
Packaging:
Packing materials used have been selected for ease of
recycling. Please ensure you use the correct disposal
system for disposal of packing materials.
Maximum power used:
1.18kwh
LEQ Sound level = 60db.
NB. The overall cycle time will increase as the mains
supply voltage decreases.
Rating: All products are rated for intermittent use,
continuously.
Heaters: Cast into the boiler.
Internal heater.
Temperature cut out:
Bi-metallic type rated at 180ºC. with automatic reset.
Over Voltage category: Group II
Pollution Degree: Group II
Insulation: Class I.
The machine is fitted with an automatic thermal reset
mechanism. In the event of the boiler overheating
(for example due to accidental overload), the unit will
power down. After approx. 10 minutes, power will
automatically be restored and the display will show
Error 01. The load will need to be re-processed, ie
start the cycle again.
Section 11: Technical Specification
14

304278B
Must be suitable to produce steam in accordance with international standards.
Example specification for clean steam
Determinant Value Recommended test for compliance
Based on Sterilized Water for Injections BP:
Acid or Alkalinity. NQ BP test. Tests for pH are not an acceptable
substitute.
Ammonium 0.2mg/litre BP test or other suitable method
Oxidisable substances. NQ BP test.
Calcium and Magnesium. NQ BP test. Tests for hardness are not an
acceptable substitute.
Heavy Metals. 0.1mg/litre BP test. Tests for individual elements are not
an acceptable substitute.
Chloride 0.5mg/litre BP test or other suitable method.
Nitrate 0.2mg/litre BP test or other suitable method.
Sulphate NQ BP test.
Residue on evaporation 30mg/litre BP test. Conductivity measurement is not an
acceptable substitute.
Pyrogens 0.25EU/ml BP test.
Based on EN 285:
Phosphate 0.1mg/litre Any suitable method.
Silicate. 0.1mg/litre Any suitable method.
Routine monitoring only:
Electrical Conductivity at 25°C 35μS/cm
NQ = Not Quantified; BP = British Pharmacopeia; EU = Endotoxin unit. (Consideration should be given to single shot type sterilizers,
where water is not returned to the reservoir at the end of the cycle).
USE OF WATER OUTSIDE THE ABOVE SPECIFICATION MAY INVALIDATE THE MACHINE WARRANTY.
Section 12: Water Quality
Section 13: Loading the Autoclave
Loading has a significant impact on how the autoclave
performs.
The maximum permissible loads are as follows: -
6 kg for Non Vacuum Cycles
6 kg for Vacuum cycles (un-pouched)
2 kg for Vacuum cycles (pouched)
1.5 kg for Vacuum cycles (porous load)
Always use the instrument trays or racks, which are
supplied.
Load instruments so that they do not touch other
instruments or the chamber and are resting on the ribs of
the tray.
Only one item should be placed in a pouch.
Pouches may be placed on the pouch trays provided.
Pouches should be arranged so that they do not touch or
overlap neighbouring pouches.
If using pouch trays, these should be spaced so that there
is a good airflow between the trays.
When loading the trays into the rack use the top rungs of
the rack first and work down.
Pouches should be placed on pouch trays with the paper
side down as shown below.
NB The use of indicator tape for sealing pouches is
not recommended in this autoclave.
15

304278B
L OAD TYPE.
(See KEY below)
CYCLE
SELECTED.
DATE CYCLE NUMBER PASS OR FAIL * OPERATOR
KEY FOR LOAD
TYPES: -
B= Hollow
Instruments
P= Pouched
or bagged
Instruments
U= Unwrapped
Instruments
M= Mixed
*If a cycle fails record the Error Code in the Pass/Fail column
Section 14: Daily Record Sheet
16
DLC Australia Pty Ltd
17-19 Horne St,
Hoppers Crossing, VIC, 3029
Tel:03 9360 9700
Fax:03 9360 9994
For more information
VIC:1300 785 405
QLD: 1300 785 401
www.dlc.com.au
DLC Australia Pty Ltd
Unit 6/1 Lear-Jet Drive,
Caboolture, QLD, 4510
Tel:07 5428 3666
Fax:07 5428 3777
Table of contents
Other Prestige medical Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Brookfield
Brookfield TC Series Operator's manual

KLA Tencor
KLA Tencor Alpha-StepIQ operating guide

Lotus cooker
Lotus cooker DRC 700 Instruction manual for installation, maintenance and use

IKA
IKA ULTRA-TURRAZ UTL 1000/10 operating instructions

Diesse
Diesse cube 30 touch user manual

Selecta
Selecta AGIMATIC-HS quick start guide