211846B
To ensure your autoclave gives you the years of service
for which it was designed, it is important to remember a
few “do’s” and “don’t’s” with regards to the operation of
the unit and to carry out simple care and maintenance
procedures on a weekly basis.
Do ensure that....
1…you read these instructions and always follow the
operating sequence.
2…the media will withstand the sterilizing temperature
selected, bottles are not sealed and screw tops are
loose - do not overload the autoclave.
3…the water level is maintained regularly with clean
distilled or de-ionised water only.
4…the unit is in a “draft free” environment and is
positioned not less than 250mm from adjacent walls.
5…you only use green sealing gasket (219500) and that it
is changed at the end of its life, if visibly damaged, or
when shrinkage has occurred.
6…the lid is securely closed when the unit is not in use,
to avoid the risk of accidental damage. Never leave in
position (F1).
7…you quote your model details, serial number and date
of purchase when contacting Prestige Medical or your
supplier.
Do not....
1…lose this operating instruction manual.
2…add any chemicals whatsoever to the water.
3…attempt to sterilize volatile substances, toxic materials
or inappropriate loads.
4…place the unit on heat sensitive surfaces ie polished
wood or glass.
5…open the Depressurization Valve (A) during the
sterilization cycle.
6…leave the Depressurization Valve (A) in the “open”
position when placing the lid upside down on a work
surface.
7…immerse the unit or electrical cord in water when
cleaning.
8…use abrasive materials or lubricants when cleaning.
9…drop or abuse the unit.
10...use in areas of risk associated with flammable
materials or gases.
11...attempt to change any fuses until the unit has been
unplugged from the mains*.
12...reach over the unit when removing cover, to do so
may cause burns from rising heat and steam.
13...cover lid or obstruct steam venting area when
manually depressurizing.
* These operations should only be undertaken by a
qualified person.
Care and maintenance
Green sealing gasket
1. Remove from inside the lid and clean with warm,
soapy water.
2. Rinse thoroughly, shake dry, do not wipe.
3. Replace in the lid by tucking evenly under all lugs.
It may appear slightly wrinkled until used.
4. Replace when it begins to show signs of leakage.
Autoclave
1. If a new gasket leaks, or a persistent leak develops,
gently clean the sealing surface of both the lid and
body of the unit with a “Scotchbrite” scrubbing pad
making sure you do not remove any metal. Rinse both
surfaces but do not dry.
2. Clean both interior and exterior with warm, soapy water
ensuring the electrical parts are kept dry.
3. Monitor the first cycle of the day to check the Air Bleed
Device, which is located inside the lid, audibly “clicks”
shut.
4. Prestige Medical recommends that your unit is
calibrated at six monthly intervals.
5. Lubricate underside of body lugs with “vaseline” if the
lid becomes stiff.
DO NOT LUBRICATE GASKET
Section 4: Do’s and Dont’s
5