9
4.7 DC Voltage or AC Voltage Measurement
To avoid any electric shock and/or damage to the meter, do not attempt a
voltage measurement if the voltage (valid value) is over 600V DC or AC
current.
To avoid any electric shock and/or damage to the meter, don't attempt to
impose between any public terminals and ground any voltage whose valid
value is over 600V for DC or AC current.
The meter provides DC voltage measuring ranges as follows: 600mV, 6V, 60V and
600V. AC voltage measuring ranges: 6V, 60V and 600V.
Measurement DC of voltage or AC voltage
1)Turn the rotary switch to the position “V”and press “FUNC” to choose DCV or
ACV.
2)Connect the test leads to COM input socket and V/input socket respectively.
3)Use another two ends of the test leads to measure the voltage of the circuit to be
measured. (In parallel connection with the circuit to be measured.)
4)Read the measured voltage value on LCD screen. When DC voltage measurement
is attempted, the display unit will show the voltage polarity of the circuit connected to
the lead-shaped meter in red.
Notes:
Within the measuring range of DC or AC voltage, even if there is no input or no
connection to the test lead, the meter will display some information. In this situation,
press short circuit “V” and “COM” terminal to reset the meter to zero.
The value of the AC voltage measured with this meter is True RMS (root mean
square). These measurements are accurate for sine wave and other waves (without
DC offset), square wave, triangular wave and step wave.
If you measure the AC voltage from the VFC (Variable-frequency converter) power,
press the “FUNC /VFC”key for about 2s to turn on the VFC voltage measurement.
4.8 LoZ Volts (DC or AC voltage) Measurement
Low impedance voltage measurement function to eliminate false voltage. This meter
automatic identified of DC voltage and AC voltage.
The meter provides DC or AC voltage measuring ranges 600V.
1)Turn the rotary switch to the LoZ V.
2)Connect the test leads to COM input socket and V/input socket respectively.
3)Use another two ends of the test lead to measure the voltage of the circuit to be
measured. (In parallel connection with the circuit to be measured.)
4)Read the measured voltage value on LCD screen. When DC voltage measurement
is attempted, the display unit will show the voltage polarity of the circuit connected to
the lead-shaped meter in red.