Quidel Lyme FIA Sofia 2 User manual

DO NOT open the foil pouch containing the Test Cassette until ready to test the sample. Place the Test Cassette on a clean and level surface.
All specimens must be at room temperature before testing. All specimens must be mixed prior to testing.
Expiration Date: Check expiration on each individual test package or outer box before using. Do not use any test past the expiration date on the label.
Precautions: Users should adhere to standard Universal Precautions when performing this Test Procedure. Precautions may include the use of nitrile gloves, safety glasses and the
covering of any cuts or abrasions on the hands. Wash hands thoroughly after performing this Test Procedure.
Verify that Sofia 2 is set to the
desired mode: WALK AWAY or
READ NOW. See the "Using Sofia
2" section for more information.
Follow Package Insert for
Specimen Collection.
Collect 25 μL (first Graduation) of
Specimen using the Graduated
Pipette provided in the kit.
Note: For Specimens in Transport
Media (preserved), collect 100 μL
(2nd Graduation) using the
Graduated Pipette provided in the kit
Transfer the Specimen to the
Specimen Diluent Tube and mix
the solution by squeezing and
releasing the top bulb of the
Graduated Pipette 3 times.
Remove the Graduated Pipette
from the Specimen Diluent Tube.
Screw the green Dropper
Tip to Specimen Diluent
Tube and mix well.
Remove the small clear cap,
hold the Specimen Diluent
Tube vertical and dispense 5
drops into the Test Cassette
sample well.
7
Proceed to the next section,
"Using Sofia 2," to complete
the test.
Study the Package Insert and User Manual
thoroughly before using Quick Reference
Instructions. This is not a complete Package Insert.
IMPORTANT! Read instructions carefully before beginning. The
test procedure below is unique to the Sofia 2 Campylobacter FIA
and may differ from other Sofia and Sofia 2 FIA procedures.
QUICK REFERENCE
INSTRUCTIONS
For use with Sofia 2.
Test Procedure
1
2
4
5
3
6
x5

WALK AWAY/READ NOW Modes
Refer to the Sofia 2 User Manual.
Sofia 2 may be set to two different modes (WALK AWAY and READ NOW). The procedures for each mode are described below. Note: The Supervisor has
the option to set the Sofia 2 to Locked WALK AWAY mode, which will prevent an operator from selecting any other mode than WALK AWAY for running a
test.
In WALK AWAY Mode, the user immediately inserts the Test Cassette into Sofia 2.
Sofia 2 will automatically time the test development, and the results will be
displayed in 15 minutes.
The user places the Test Cassette onto the counter or bench top for 15 minutes (outside of
Sofia 2) and manually times this development step. Then, the user inserts the Test Cassette
into Sofia 2. In READ NOW Mode, Sofia 2 will scan and display the test result within 1 minute.
Warning: Results must not be interpreted past 30 minutes after inoculation. Using the Sofia
2 past this time may result in false results.
Allow the test to develop for the FULL 15 minutes BEFORE placing it into Sofia 2.
WALK AWAY Mode
READ NOW Mode
Using Sofia 2

RUN TEST
1. Input the User ID with the integrated barcode
scanner or enter the data using the on-screen
keypad.
NOTE: If you scan the incorrect barcode, select the
field again to re-highlight it. Then simply rescan
using the correct barcode, and the previous one
will be overwritten with the correct barcode.
4. Insert the Test Cassette into
the drawer. Then gently close
the drawer.
2. Input the Patient ID and/or Order # (if applicable)
using the integrated barcode scanner or enter the
data using the on-screen keypad.
3. Verify that the correct mode (WALK AWAY or
READ NOW) has been selected. Press to open
the Sofia 2 drawer.
NOTE: If the instrument is set in Locked Walk
Away mode, users will only be able to use the
“walk away” mode option for running tests.
Read Now Mode will be grayed out when in
Locked Walk Away Mode.
5. Sofia 2 will start and display the progress. In WALK AWAY Mode, the test
results will be displayed after 15 minutes. In READ NOW Mode, the test results
will be displayed on the screen within 1 minute. See Interpretation of Results
section.
When the test is complete, the results will be displayed on the Sofia 2 screen. Test
Lines will not be visible to the naked eye.
Results: The Sofia 2 screen will display results for the procedural control as being
or . If the control is , retest starting with Step 1 and a new Test Cassette.
Reader Display
Interpretation
Campy
Control
Positive Test for Campylobacter
Campy
Control
Negative Test for Campylobacter
Campy
Control
Result Invalid
Interpretation of Results

From the main menu, select
Run QC.
Follow the prompt on the screen. Scan the QC Card (located on the assay kit box).
Sofia 2 prompts the user to select the desired mode (WALK AWAY or READ NOW).
Follow the External Quality Control Test Procedure in the Package Insert to test each
Control, first the Positive Control followed by the Negative Control.
After both the Positive and Negative Controls have been run, the results will be
displayed as or .
Quidel recommends that Positive and Negative External Controls be run:
▪
Once for each untrained operator.
▪
Once for each new shipment of kits –provided that each different lot received in the shipment is tested.
▪
As deemed additionally necessary by your internal quality control procedures, and in accordance with Local, State and Federal regulations or accreditation
requirements.
Intended Use
Sofia 2 Campylobacter FIA employs immunofluorescence for the rapid qualitative detection of a Campylobacter-specific antigen in human fecal specimens. Sofia 2
Campylobacter FIA is designed to detect C. jejuni, C. coli, C. lari and C. upsaliensis from patients with signs and symptoms of gastroenteritis. The test is intended for use
with preserved fecal specimens in transport media and unpreserved fecal specimens. Test results should be considered in conjunction with clinical findings and patient
history.
ASSISTANCE
If you have any questions regarding the use of this product or if you want to report a test system problem, please call Quidel Technical Support 800.874.1517 (in the
Reference the Package Insert for Warnings and Precautions, Specimen Collection and Handling, and Quality Control.
1394301EN00 (01/22)
Study the Package Insert and User Manual thoroughly
before using Quick Reference Instructions. This is not a
complete Package Insert.
Quidel Corporation
10165 McKellar Court
San Diego, CA 92121 USA
quidel.com
External Quality Control
1
2
3
5
4
Other manuals for Lyme FIA Sofia 2
6
Other Quidel Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Fritsch
Fritsch PULVERISETTE 0 operating instructions

BIO RAD
BIO RAD S3 instruction manual

Lifescan
Lifescan OneTouch Ultra user guide

Labnet
Labnet TC9600-G Operation manual

enviromental express
enviromental express HotBlock Pro Operation & instruction manual

Bright Instruments
Bright Instruments 2212 Starlet manual