CALIBRATION CHECK
◗ Use the up/down keys to highlight CALIBRATION from the menu.
◗ Press the soft key under the lower right corner of the display screen under SELECT.
◗ Remove Calibration Cassette from the pouch.
Insert the Calibration Cassette into the
Analyzer and close the drawer. The Analyzer
will perform the calibration automatically.
The Analyzer display will indicate when the
calibration is completed.
Reader
FLU
A+B
NOTE: If calibration cannot be completed
successfully, notify the on-site
Supervisor or contact Quidel Service for
assistance.
After successful calibration, select OK to
return to the Main Menu.
Press eject to remove and return the
Calibration Cassette to the storage pouch.
Important: Store the Calibration
Cassette in the provided storage pouch
between uses.
Reader
FLU
A+B
Run QC according to assay specific package insert.