
Instructions for Use –NPS3
Manufacturer: RAUMEDIC AG, Hermann-Staudinger-Str. 2, 95233 Helmbrechts, Germany
zwo-188 EN Rev. 11 / 2022-05-04
Page 3 of 10
Device corresponds the essential requirements of Directive 93/42/EEC
6 Initial operation of the NPS3
The device (housing, foil keyboard, cable insulation, plug, label) must be visually inspected for intactness
every time before being started up. The device must not be used if defects are detected. It must be
returned to the manufacturer immediately.
Visual inspection
•Front film undamaged (watertight, button OK)?
•Name plate fitted and legible?
•Housing undamaged (no cracks or sharp edges)?
•Cable connection and Redel coupling visually free from damage?
•Housing screwed in three places?
•Are there any loose parts within the device (shake the device gently)?
•Is the device or accessories dirty?
NPS3 has no separate circuit closer. It is activated by plugging the ICP-TEMP-Cable (cf. chapter 14 –
accessories) into the NPS3.
Connection of measuring chain
•Connect the ICP-TEMP-Cable to the NPS3
The red “No sensor present” LED display has to be lit. If no LED is lit or the “Battery low” LED display
is on only, there is either no battery in the appliance or it has run too low. In either case, a new
battery has to be inserted (cf. chapter 7 on “Change of battery”).
•Connect the pressure catheter to the ICP-TEMP-Cable
The red “No sensor present” LED display will go out. At the same time, the numeric LCD will be
activated and will display the pressure value of the pressure catheter.
If the “No sensor present” LED display continues to be lit, the pressure catheter has not been
plugged in correctly or it is defective (due to broken cable, for example). In this case, the pressure
catheter has to be replaced. A defective ICP-TEMP cable may also be the reason for the fault, which
has to be replaced as well in this case.
Caution:
•For self-test of the zero point of the NPS3, a test shall have to be carried out prior to every use:
With the pressure catheter connected, press the ”Test 0 mmHg” button →0 mmHg must appear in
the display. Deviation of ±1 mmHg is acceptible. If a larger deviation is displayed, the NPS3 may not
be used anymore! Return the NPS3 to the manufacturer immediately!
Calibration of the zero point is not required for RAUMEDIC precision pressure catheters.
This document is for information only