1. DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digits
CUB5VR00: Reflective LCD with full viewing angle
CUB5VB00: Transmissive LCD with selectable red or green LED backlight,
viewing angle optimized. Display color change capability with output state
when using an output module.
2. POWER: Input voltage range is +9 to +28 VDC with short circuit and input
polarity protection.
3. INPUT RANGES: Jumper Selectable
D.C. Voltages: 200 mV, 2 V, 20 V, 200 V
4. SIGNAL INPUTS:
5. OVERRANGE RATINGS, PROTECTION & INDICATION:
9 to 28 VDC power circuit is not isolated from the signal circuit.
Input Overrange Indication: “OLOL”.
Input Underrange Indication: “ULUL”.
Display Overrange/Underrange Indication: “.....”/“-.....”
6. A/D CONVERTER: 16 bit resolution
7. RESPONSE TIME:
Display: 500 msec min.
Output: 800 msec max (with input filter setting of 0)
8. NORMAL MODE REJECTION: 60 dB 50/60 Hz
9. USER INPUT (USR): Programmable input. Connect terminal to common
(USR COMM) to activate function. Internal 10Kpull-up resistor to +9 to
28 VDC.
Threshold Levels: VIL = 1.0 V max; VIH = 2.4 V min; VMAX = 28 VDC
Response Time: 5 msec typ.; 50 msec debounce (activation and release)
10. CONNECTIONS: Wire clamping screw terminals
Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 5 inch-lbs (0.565 N-m) max.
11. MEMORY: Nonvolatile E2PROM memory retains all programming
parameters and max/min values when power is removed.
12. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 requirements for
outdoor use. Installation Category I, Pollution Degree 2. High impact plastic
case with clear viewing window. Panel gasket and mounting clip included.
13. ENVIRONMENTAL CONDITIONS:
Operating Temperature Range for CUB5VR00: -35 to 75°C
Operating Temperature Range for CUB5VB00 depends on display color
and intensity level as per below:
INTENSITY LEVEL TEMPERATURE
Red Display 1 & 2 -35 to 75°C
3 -35 to 70°C
4 -35 to 60°C
5 -35 to 50°C
INTENSITY LEVEL TEMPERATURE
Green Display 1 & 2 -35 to 75°C
3 -35 to 65°C
4 -35 to 50°C
5 -35 to 35°C
Storage Temperature: -35 to 85°C
Operating and Storage Humidity: 0 to 85% max. relative humidity (non-
condensing)
Vibration According to IEC 68-2-6: Operational 5 to 500 Hz, in X, Y, Z
direction for 1.5 hours, 5 g’s.
Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directions.
Altitude: Up to 2000 meters
14. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL Recognized Component, File #E179259, UL61010-1, CSA C22.2 No.
61010-1
Recognized to U.S. and Canadian requirements under the Component
Recognition Program of Underwriters Laboratories, Inc.
UL Listed, File # E137808, UL508, CSA C22.2 No. 14-M95
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
Type 4X Outdoor Enclosure rating (Face only), UL50
IECEE CB Scheme Test Report #E179259-V01-S02
Issued by Underwriters Laboratories, Inc.
IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment
for measurement, control, and laboratory use, Part 1.
IP65 Enclosure rating (Face only), IEC 529
ELECTROMAGNETIC COMPATIBILITY
EmissionsandImmunitytoEN61326:ElectricalEquipmentforMeasurement,
Control and Laboratory use.
Notes:
1. Criterion A: Normal operation within specified limits.
15. WEIGHT: 3.2 oz (100 g)
2
GENERAL METER SPECIFICATIONS
ORDERING INFORMATION
MODEL
NO. DISPLAY COLOR INPUT CURRENT
@ 9 VDC WITH
CUB5RLY0
CUB5VR00 --- 10 mA 40 mA
CUB5VB00 Red (max intensity) 85 mA 115 mA
CUB5VB00 Green (max intensity) 95 mA 125 mA
Class AEN 55011Emissions
Emissions: 30 A/m
3 V/rms
Criterion A
Criterion A
EN 61000-4-8
EN 61000-4-6RF conducted interference
1 kV L-L,
Criterion AEN 61000-4-5Surge 1 kV signal
2 kV power
Criterion AEN 61000-4-4Fast transients (burst)
2 kV L&N-E power
10 V/m
Criterion AEN 61000-4-3Electromagnetic RF fields 8 kV air discharge
4 kV contact discharge
Criterion AEN 61000-4-2Electrostatic discharge
Immunity to Industrial Locations:
Power frequency magnetic fields
INPUT CURRENT
@ 9 VDC WITHOUT
CUB5RLY0
INPUT
RANGE
ACCURACY
@23 °C, less
than 85% RH
INPUT
IMPEDANCE
MAX
INPUT
SIGNAL RESOLUTION TEMP.
COEFFICIENT
0.1% of span 1.027 M75 VDC 10 μV70 ppm / °C
2 VDC 0.1% of span 1.027 M75 VDC .1 mV 70 ppm / °C
20 VDC 0.1% of span 1.027 M250 VDC 1 mV 70 ppm / °C
200 VDC 0.1% of span 1.027 M250 VDC 10 mV 70 ppm / °C
200 mVDC
TYPE MODEL NO. DESCRIPTION PART NUMBER
CUB5 CUB5V DC Volt Meter with reflective display CUB5VR00
DC Volt Meter with backlight display CUB5VB00
Optional Plug-in Cards
CUB5RLY Single Relay Option Card CUB5RLY0
CUB5SNK Dual Sinking Open Collector Output card CUB5SNK0
CUB5COM RS485 Serial Communications Card CUB5COM1
RS232 Serial Communications Card CUB5COM2
CUB5USB USB Programming Card CUB5USB0
Accessories
MLPS +12 VDC Micro-Line Power Supply, 85 to 250 VAC source, 400 mA max out MLPS1000
+24 VDC Micro-Line Power Supply, 85 to 250 VAC source, 200 mA max out MLPS2000
CBLPRO Programming Cable RS232 (RJ11-DB9) CBLPROG0
CBPRO Programming Cable RS485 (RJ11-DB9) CBPRO007
SFCRD Crimson 2 PC Configuration Software for Windows 98, ME, 2000, XP 1SFCRD200
CBLUSB USB Programming Cable CBLUSB00
1 Crimson software is a free download from http://www.redlion.net/