reedinstruments
www com
7
Temp (type-K)
-58 to
1832°F 1°F ±(1.0%reading+4.5°F)
±(1.0%reading+2.5°C)
(probe accuracy not included)
-50 to
1000°C 1°C
4-20mA% -25 to
125% 0.01% ±50 digits
0mA=-25%,4mA=0%,20mA=100%,24mA=125%
Note:Accuracyspecicationsconsistoftwoelements:
•(%reading)–Thisistheaccuracyofthemeasurementcircuit
•(+digits)–Thisistheaccuracyoftheanalogtodigitalconverter
Enclosure: Double molded, waterproof
Shock (Drop Test): 6.5 feet (2 meters)
Diode Test: Test current of 0.9mA maximum, open circuit
voltage 2.8V DC typical
Continuity Check: Audible signal will sound if the resistance is less
than35Ω(approx.),testcurrent<0.35mA
Peak: Capturespeaks>1ms
Temperature Sensor: Requires type K thermocouple
InputImpedance: >10MΩVDC&>3MΩVAC
AC Response: True RMS
AC True RMS: The term stands for “Root-Mean-Square”,
which represents the method of calculation of
the voltage or current value.
Average responding multimeters are calibrated
to read correctly only on sine waves and they
will read inaccurately on non-sine wave or
distorted signals. True rms meters read
accurately on either type of signal
ACV Bandwidth: 50Hz to 1000Hz
CrestFactor: ≤3atfullscaleupto500V,decreasinglinearly
to≤1.5at1000V
Display: 40,000 count backlit liquid crystal with bargraph
Overrange indication: “OL” is displayed
Auto Power Off: 15 minutes (approximately) with disable feature
Polarity: Automatic (no indication for positive); Minus (-)
sign for negative
Measurement Rate: 2 times per second, nominal continued ...