Instruction Manual
6.2 FUNCTIONS
Ambient concentration reset
When entering a contaminated area, it is important to have a function which
allows you to reset the leak detector. Upon initial power on, the TRITECTOR
automatically resets itself to ambient concentration, to ignore the level of
refrigerant present at the tip. Only a level or concentration greater than the
ambient concentration will cause an alarm.
Automatic reset
After 30 seconds, the leak detector resets itself. This function takes into account
the changes in the environment and resets itself just if there are no significant
changes. By default, this function activated, but you have the possibility to turn
this function off. By pressing & simultaneously for three seconds
this function is disabled.
By disabling this function, the symbol appears in the display. When starting
the TRITECTOR, it will reset itself for the first time (Ambient concentration
reset).
Manual reset
The manual reset allows you to reset the TRITECTOR whenever you want and
you feel that it is necessary.
To manually reset the TRITECTOR, press briefly . A double tone sounds
and all rings on the display light up briefly, to indicate that the manual reset has
been performed.
Peak
The Peak function facilitates searching for a leak in a hard to reach area without
being able to see the LCD display.
The Peak function shows the highest concentration detected by the
TRITECTOR. To enable the peak function, press and hold the peak button