Reichert Model 30 User manual

Model 30™
Pneumatonometer
User’s Guide

©2017 AMETEK, Inc.
Reichert and Reichert Technologies are registered trademarks of Reichert, Inc.
Model 30 is a trademark of Reichert, Inc.
AMETEK is a registered trademark of AMETEK, Inc.
All other trademarks are property of their respective owners.
The information contained in this document was accurate at time of publication. Specifications subject to change without notice.
Reichert, Inc. reserves the right to make changes in the product described in this manual without notice and without incorporating those
changes in any products already sold.
ISO 9001/13485 Certified – Reichert products are designed and manufactured under quality processes meeting
ISO 9001/13485 requirements.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic,
mechanical, recording, or otherwise, without the prior written permission of Reichert, Inc.
Caution: Federal law restricts this device to sale by or on the order of a licensed physician. Rx only.

16030-101 Rev. J 3
Contents
Warnings and Cautions ................................................................................... 4
Symbol Information.......................................................................................... 6
Introduction...................................................................................................... 7
Indications for Use..................................................................................... 7
Contraindications....................................................................................... 7
Instrument Setup ............................................................................................... 8
Unpacking Instructions .............................................................................. 8
Parts Identication ..................................................................................... 9
Models and Options............................................................................ 10
Accessories......................................................................................... 10
Icon Denition.......................................................................................... 11
Default Settings ....................................................................................... 13
General Setup ......................................................................................... 13
Printout Setup.......................................................................................... 14
Service Screen ........................................................................................ 14
Installation ............................................................................................... 15
Pneumatonometer Installation................................................................. 15
Application of Input Power....................................................................... 17
Disconnection of Input Power.................................................................. 18
Quick Verication Check.......................................................................... 19
Operating Principle .................................................................................. 20
Instructions for Use ..........................................................................................21
Patient Preparation.................................................................................. 21
Measurement Protocol ............................................................................ 22
Measurements......................................................................................... 23
Manual Tonometry .............................................................................. 23
Pulse Tonometry................................................................................. 24
Tonography......................................................................................... 25
Seated IOP .................................................................................... 26
Supine IOP .................................................................................... 26
Cleaning & Maintenance ................................................................................. 28
Tip and Membrane Assembly .................................................................. 28
Cleaning the Tip and Membrane Assembly ........................................ 28
Disinfection of the Tip and Membrane Assembly................................ 28
Inspection of the Tip and Membrane Assembly.................................. 28
Fuses....................................................................................................... 29
Calibration Verier Cornea Replacement ................................................ 29
External Cleaning .................................................................................... 29
Printer Paper ........................................................................................... 30
Filter Replacement .................................................................................. 30
Compressor Filter Replacement......................................................... 30
Inline Filter Replacement.................................................................... 30
Troubleshooting................................................................................................31
Specications .................................................................................................. 32
Classications ................................................................................................. 33
Guidance & Manufacturer’s Declarations...................................................... 34
Warranty ........................................................................................................ 39

16030-101 Rev. J
4
Reichert Technologies is not responsible for the safety and reliability of this instrument when:
• Assembly, disassembly, repair, or modication is made by unauthorized dealers or persons.
• Instrument is not used in accordance with this User’s Guide.
WARNING: AN INSTRUCTION THAT DRAWS ATTENTION TO RISK OF INJURY OR DEATH.
WARNING: UNITED STATES FEDERAL LAW AND EUROPEAN REGULATIONS REQUIRE THAT THIS DE-
VICE BE PURCHASED ONLY BY A PHYSICIAN OR A PERSON ACTING ON BEHALF OF A PHYSICIAN.
WARNING: THIS INSTRUMENT SHOULD BE USED IN STRICT ACCORDANCE WITH THE INSTRUC-
TIONS OUTLINED IN THIS USER’S GUIDE. THE SAFETY OF THE OPERATOR AND THE PERFOR-
MANCE OF THE INSTRUMENT CANNOT BE GUARANTEED IF USED IN A MANNER NOT SPECIFIED BY
REICHERT TECHNOLOGIES.
WARNING: DO NOT REPAIR OR SERVICE THIS INSTRUMENT WITHOUT AUTHORIZATION FROM
THE MANUFACTURER. ANY REPAIR OR SERVICE TO THIS INSTRUMENT MUST BE PERFORMED BY
EXPERIENCED PERSONNEL OR DEALERS WHO ARE TRAINED BY REICHERT SO THAT CORRECT
OPERATION OF THIS INSTRUMENT IS MAINTAINED OR SERIOUS INJURY TO THE OPERATOR OR
PATIENT MAY OCCUR.
WARNING: MODIFICATIONS TO THIS INSTRUMENT IS NOT ALLOWED. ANY MODIFICATION TO THIS
UNIT MUST BE AUTHORIZED BY REICHERT SO THAT CORRECT OPERATION IS MAINTAINED OR
SERIOUS INJURY TO THE OPERATOR OR PATIENT MAY OCCUR.
WARNING: IF THIS INSTRUMENT IS MODIFIED, APPROPRIATE INSPECTION AND TESTING MUST BE
CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT.
WARNING: TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE CONNECTED TO
A SUPPLY MAINS WITH PROTECTIVE EARTH OR DAMAGE TO THE INSTRUMENT AND/OR INJURY TO
THE OPERATOR OR PATIENT MAY OCCUR.
WARNING: ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS THE VOLTAGE
THAT IS INDICATED ON THE DATA PLATE OR DAMAGE TO THE INSTRUMENT AND/OR INJURY TO
THE OPERATOR OR PATIENT MAY OCCUR.
WARNING: THIS INSTRUMENT MUST BE PLUGGED INTO AN OUTLET WITH AN EARTH GROUND. DO
NOT REMOVE OR DEFEAT THE EARTH GROUND CONNECTION ON POWER INPUT CONNECTOR
OR THE UNIT’S POWER CORD OF THIS INSTRUMENT OR DAMAGE TO IT AND/OR INJURY TO THE
OPERATOR OR PATIENT MAY OCCUR.
WARNING: THE EQUIPMENT OR SYSTEM SHOULD NOT BE USED ADJACENT TO OR STACKED WITH
OTHER EQUIPMENT AND THAT IF ADJACENT OR STACKED USE IS NECESSARY, THE EQUIPMENT
OR SYSTEM SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE CONFIGURATION IN
WHICH IT WILL BE USED.
WARNING: THIS INSTRUMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF FLAMMABLE AN-
ESTHETIC MIXTURES, SUCH AS OXYGEN OR NITROUS OXIDE.
WARNING: DO NOT USE ANY OTHER PROBE WITH THIS INSTRUMENT THAN THOSE SUPPLIED BY
REICHERT EXPRESSLY FOR USE WITH THIS INSTRUMENT. OTHER PROBES MAY CAUSE DAMAGE
OR INJURY TO THE EYE.
WARNING: DO NOT USE THIS INSTRUMENT IF THE MEASUREMENT TIP IS CRACKED, CHIPPED OR
SHOWS ANY IRREGULARITY OF THE SURFACE, TO PREVENT PATIENT INJURY AND OR INACCU-
RATE READINGS.
WARNING: IN ORDER TO PREVENT PATIENT-TO-PATIENT TRANSFER OF INFECTION, AFTER EACH
USE DISINFECT THE MEASUREMENT TIP FOLLOWING ACCEPTED LOCAL CLINICAL PROCEDURES
REGARDING THE USE OF DISINFECTANTS. ANY CLINICALLY APPROVED CHEMICAL DISINFECTANT
CAN BE USED.
Warnings & Cautions

16030-101 Rev. J 5
WARNING: THE USE OF ACCESSORIES OR CABLES OTHER THAN THOSE SPECIFIED, WITH THE
EXCEPTION OF THOSE SOLD BY THE MANUFACTURER AS REPLACEMENT PARTS FOR INTER-
NAL COMPONENTS, MAY RESULT IN INCREASED EMISSIONS OR DECREASED IMMUNITY OF THE
EQUIPMENT OR SYSTEM.
CAUTION: AN INSTRUCTION THAT DRAWS ATTENTION TO THE RISK OF DAMAGE TO THE PRODUCT.
CAUTION: THE INTERNAL CIRCUITRY OF THE INSTRUMENT CONTAINS ELECTROSTATIC DIS-
CHARGE SENSITIVE DEVICES (ESDS) THAT MAY BE SENSITIVE TO STATIC CHARGES PRODUCED
BY THE HUMAN BODY. DO NOT REMOVE THE COVERS WITHOUT TAKING PROPER PRECAUTIONS.
CAUTION: DO NOT USE SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANY PART OF THIS
INSTRUMENT AS DAMAGE TO THE UNIT MAY OCCUR. SEE MAINTENANCE SECTION FOR DETAILED
CLEANING INSTRUCTION.
CAUTION: USE OF AMMONIA BASED CLEANERS ON THE LIQUID CRYSTAL DISPLAY (LCD) MAY
CAUSE DAMAGE TO DISPLAY. SEE MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUC-
TION.
CAUTION: DO NOT AUTOCLAVE OR DISINFECT USING HIGH TEMPERATURES EXCEEDING THE REC-
OMMENDED TEMPERATURES INDICATED IN THE SPECIFICATIONS SECTION OF THIS MANUAL OR
DAMAGE TO THE UNIT MAY OCCUR.
CAUTION: MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING EMC
AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO THE EMC INFORMATION
PROVIDED IN THE ACCOMPANYING DOCUMENTS.
CAUTION: PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL
ELECTRICAL EQUIPMENT.
CAUTION: THIS INSTRUMENT IS NOT TO BE USED NEAR HIGH-FREQUENCY EMITTING SURGICAL
EQUIPMENT.
CAUTION: THIS INSTRUMENT IS NOT INTENDED TO BE CONNECTED TO EQUIPMENT OUTSIDE THE
CONTROL OF REICHERT INC. OR MUST BE TESTED TO AN APPLICABLE IEC OR ISO STANDARDS.
CAUTION: DO NOT INSTALL ANY ADDITIONAL SOFTWARE OTHER THAN WHAT WAS SUPPLIED WITH
THIS INSTRUMENT. INSTALLATION OF ADDITIONAL SOFTWARE MAY CAUSE UNEXPECTED OPERA-
TION RESULTING IN MALFUNCTION OF THIS INSTRUMENT.
Warnings & Cautions (continued)

16030-101 Rev. J
6
Symbol Information
Symbol Information
The following symbols appear on the instrument:
Caution
Type B Applied Part
Alternating Current Power
Protective Earth Connection
ON / OFF
Manufacturer
2017
Date of Manufacture
REF Catalog Number
S/N Serial Number
Waste of Electrical and Electronic Equipment
Compliance to Medical Device Directive 93/42/EEC
Authorized to mark given by Intertek ETL Semko for conformance
with electrical standards
Accompanying Documents must be consulted
Authorized Representative in European Community
Fragile Contents in Shipping Container - handle with care
Keep Dry - Package shall be kept away from rain
This Way Up - Indicates correct upright position of package
Important Instruction - remove Shipping Bracket

16030-101 Rev. J 7
Congratulations on your purchase of the Reichert®Model 30™Pneumatonometer.
The Model 30 Pneumatonometer is a highly accurate instrument used to measure intraocular pressure
(IOP) non-invasively through applanation tonometry and represents a signicant advance in tonometry and
tonography technology over its predecessors. The pneumatonometer system consists of the base unit,
a probe, and the necessary accessories to accurately measure tonometry/tonography in one easy-to-use
unit. This manual describes its operation and details. Most individuals will nd this instrument easy to use
following the software and audible prompts.
This User’s Guide is designed as a training and reference manual for operation, maintenance, and
troubleshooting. We recommend that you read it carefully prior to use and follow the instructions in the
guide to ensure optimum performance of your new instrument. Properly trained eyecare professionals
such as ophthalmologists, optometrists, opticians and eye care technicians should operate this instrument.
Please retain this guide for future reference and to share with other users. Additional copies can be
obtained from your authorized Reichert Technologies dealer or contact our Customer Service department
directly at:
Tel: 716-686-4500
Fax: 716-686-4555
Email: [email protected]
Please indicate the following when contacting Reichert to ensure that you receive the correct information:
• Model Number
• Serial Number
• Contact Phone Number or Email Address
Indications for use
The Model 30 Classic Pneumatonometer is intended for the measurement of intraocular pressure. The
Model 30 Classic Pneumatonometer is indicated for use as a screening / monitoring tool for glaucoma or
when increased intraocular pressure is suspected.
Contraindications
None.
Introduction

16030-101 Rev. J
8
Instrument Setup
Great care has been taken to deliver this instrument to you safely. The container and packaging was
specially designed to transport this unit. Please retain the packaging if future transportation is required.
Please remove the packaging material from the outer box and then remove the pneumatonometer and its
accessories from the box. Refer to Figures 1-1 through 1-3.
Unpacking Instructions
Please remove the packaging material from the instrument in the following
manner. Refer to Figures 1-1 through 1-3.
The instrument is packaged in a shipping container to protect the
instrument from damage during shipment. Please read the User’s Guide
before operating the unit. The pneumatonometer includes the following
accessories:
1. User’s Guide (not shown)
(P/N 16030-101)
2. Spare Tip and Membrane Assembly
(P/N 230676)
3. Pneumatonometer Probe
(P/N 232349)
4. Power Cord
(P/N WCBL10018 for 16030 & 16032
or P/N WCBL10027 for 16031)
5. Footswitch
(P/N 232345 - only available with P/N 16033 kit)
6. Thermal Paper
(P/N 232348)
7. Calibration Verier
(P/N 232373)
8. Yellow Shipping Bracket (not shown)
(P/N 16030-009)
(attached to the unit when shipped).
Figure 1-1, Unpacking Unit
Figure 1-2, Unit with Foam
2
Figure 1-3, Accessories
6
7
34
5

16030-101 Rev. J 9
Instrument Setup (continued)
Parts Identication
The Model 30 Pneumatonometer is housed in a single compact metal case that fits easily on most coun-
ters or an appropriate stand.
The front panel of the Model 30 Pneumatonometer includes: (Refer to the Figure 1-4.)
1. ON/ OFF switch
2. Built-in chart
3. LCD readout
4. Connector for the pneumatic probe.
The rear panel of the Model 30 Pneumatonometer (Figure 1-5) includes:
1. Access door for changing filters and removing the yellow shipping bracket.
2. Receptacle for the optional footswitch, which allows the operator to select system functions while using
his/her hands to restrain the patient’s eyelid and position the probe.
3. Three-prong plug for the AC power cord that connects the Model 30 Pneumatonometer to an electrical
outlet with an integral fuse holder.
4. USB connection for factory data communication.
Note: The USB connection is only for use by the manufacturer, for service repairs. This port cannot be
used to transfer data or connect to an external printer.
Figure 1-5, Pneumatonometer Parts (Rear)
1
2
3
4
Figure 1-4, Pneumatonometer Parts (Front)
1
23
4

16030-101 Rev. J
10
Instrument Setup (continued)
Parts Identication
Models and Options
The specifications for the different models of the pneumatonometer are found in the General Specifi-
cations section of this manual.
The pneumatonometer is a tonometer that provides the following tests: manual tonometry, pulse
tonometry or tonography.
In the manual mode, the pneumatonometer provides a single pressure value and ends the test when
the probe is removed from the surface of the patient’s eye. The pulse mode provides pressure
output with pulse fluctuations and ends the test after it has read ten ocular pulses or the probe is
removed from the eye surface. The tonography test records changes in IOP when a weighted probe
is held on the cornea for a specified time. The tonography test enables the operator to conduct a
two-minute or four-minute test for each eye.
The Model 30 Pneumatonometer handpiece and probe assembly (Figure 1-6) includes:
1. Tip and membrane assembly (Applied Part)
2. Auto sensor assembly and probe handle
3. Double lumen tubing
4. Quick-flow connector
Accessories (Optional)
Part Number Description
16033 Tonography Kit (includes Foot Pedal & 10 Gram Weight)
230676 Tip and Membrane Assembly, 1 pk
230677 Tip and Membrane Assembly, 3 pk
230678 Tip and Membrane Assembly, 10 pk
232346 Filter Kit
232348 Printer Paper
232373 Calibration Verifier
(To order any of these accessories, contact your local authorized Reichert dealer.)
Figure 1-6, Pneumatonometer Probe
2
1
3
4

16030-101 Rev. J 11
Instrument Setup (continued)
Icon Denition
The Pneumatonometer incorporates a user-friendly icon/menu-based operating system that
will increase the speed of measurements, training and use. Below are the Icons that are
used during the operation of this instrument.
Icon Icon Description
SETUP Access the setup menu for changing default pa-
rameters.
MEASURE Initiates the measurement process.
MANUAL MODE Initiates the manual measurement mode process.
PULSE MODE Initiates the pulse measurement mode process.
TONOGRAPHY Initiates the tonography mode.
LEFT EYE Left eye measurement.
RIGHT EYE Right eye measurement.
PAPERFEED Advances the printer paper.
2 MINUTE Initiates the two minute measurement mode.
4 MINUTE Initiates the four minute measurement mode.
ENTER DATA Allows data to be entered manually.
CALCULATE Provides calculation of data.

16030-101 Rev. J
12
Instrument Setup (continued)
Icon Denition (continued)
Icon Icon Description
RETURN Returns to preceding screen.
RIGHT ARROW Used in the setup menus to move right horizontally.
LEFT ARROW Used in the setup menus to move left horizontally.
UP ARROW Used in the setup menus to move up vertically.
DOWN ARROW Used in the setup menus to move down vertically.
PLUS Increases a numerical value displayed on the screen.
MINUS Decreases a numerical value displayed on the
screen.
SELECT Used in the setup menus to activate the new param-
eter or setting.
HOME Changes the display to the main menu screen.
PRINT Provides a printout of the current data.

16030-101 Rev. J 13
Instrument Setup (continued)
Default Settings
This instrument is sent from the factory with measurement, chart, and other general parameters assigned
to default settings. These settings can be changed to the preferences of the operator/clinician. A
summary of these settings is given below. The optional settings follow on subsequent pages.
General Setup:
Chart: On
Footswitch:Off
Volume: Middle
Brightness: Middle
Printout Setup:
Date: (current date)
Date Fmt: MDY
Time: (current time)
Time Fmt: AM/PM
Service:
Service Screen with:
Software Version
Calibration Date
Customer Service Telephone Number
General Setup
Parameters: Settings:
Chart: Options are either On or Off.
Footswitch: Options are either On or Off.
Volume: This option is a left or right arrow to either decrease or increase
the volume.
Brightness: This option is a left or right arrow to either decrease or increase
the brightness.

16030-101 Rev. J
14
Printout Setup
Parameters: Settings:
Date: When the unit is received, the date and time should be changed to
the local time zone.
Date Fmt: The options are: MDY, DMY, YMD (D=day, M=month, Y=year)
Time: The correct time should be changed to agree with the local time
zone.
Time Fmt: The options are: AM/PM, 24HR
Service Screen
This sample screen provides the details of the Software Version, Calibration Date, and the customer
service telephone number. If assistance is needed with this unit, consult the Troubleshooting section
of this manual. If further assistance is needed, please contact the customer service department at
Reichert, Inc.
Instrument Setup (continued)

16030-101 Rev. J 15
Instrument Setup (continued)
Installation
Consider the following factors as you find a location for the Model 30 Pneumatonometer:
• When taking tonometry measurements with the pneumatonometer, the patient can be either seated
or supine. When conducting tonography tests, the patient must be supine.
• The console should be close enough to an electrical outlet so that the operator can connect the sys-
tem with the power cord.
• The operator should be in a position to look directly at the contact between the probe and the sur-
face of the eye.
• The operator should be able to see the instrument’s LCD readout and chart recorder easily during a
test.
• This medical device complies with IEC/EN 60601-1-2 Safety Standard for Electromagnetic Compat-
ibility, Requirements and Test. However, if the equipment is operated in the presence of high levels
of EMI or highly sensitive equipment, interference may be encountered and the user should take
whatever steps are necessary to eliminate or reduce the interference.
Pneumatonometer Installation
WARNING: CARE MUST BE TAKEN TO ARRANGE THE CABLES FOR THE ACCESSORIES SUCH THAT THEY DO
NOT PRESENT A TRIPPING HAZARD TO THE EXAMINER OR A DANGER TO THE PATIENT.
WARNING: POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION
DEVICE (PLUG).
1. Set the ON/OFF switch on the front panel to the OFF position (the O is
pushed in). Refer to Figure 1-7.
2. Remove the yellow shipping bracket according to the following steps. Refer
to Figure 1-8.
A. Remove the screw at the top of the air vent and remove the air vent.
B. Loosen the screw that secures the yellow shipping bracket onto the
back of the compressor.
C. Remove the bracket and then tighten the screw.
D. Replace the air vent and secure it using the screw that was removed.
E. Leave the yellow shipping bracket off and store it in a safe place so that
if it is needed for future shipping, it is available.
3. Plug the power cord into the back of the unit and the other end into an electrical outlet that has the
correct input power parameters as indicated in the General Specifications section of this manual.
Note: Leave the On/Off switch to the OFF position and do not apply power to the unit at this time.
Figure 1-8, Shipping Bracket Removed
Shipping
Bracket
Air Vent
Compressor
Figure 1-7, ON/OFF
ON
OFF

16030-101 Rev. J
16
Pneumatonometer Installation (continued)
4. Remove the probe from its container and place the clean tip and
membrane assembly that was sent with the unit onto the probe, if
one is not already installed. If a tip is not installed, perform the fol-
lowing:
A. Perform the steps of the Disinfection of the Tip and Membrane
Assembly as indicated in the Reprocessing Guidelines section
of this manual
B. Install the tip and membrane onto the probe tip so that the tub-
ing is positioned approximately half way down the metal tube
as shown in Figure 1-9.
Note: The correct installation of the tip and membrane assembly is
that the tubing is installed approximately half way down the metal tube.
5. Plug the other end of the probe tubing into the front panel connector as follows: (Refer to Figure 1-10)
A. Push in the probe connector into its mating connector.
B. Ensure the connection is tight and that there is no sound of leaking air from around the connector.
6. Loosely coil the probe tubing around the probe tray using the cutouts at either end of the tray and place
the probe in the tray so that the tip is not in contact with any other item.
7. If the unit includes the optional footswitch, connect the footswitch cord into the receptacle on the back
panel. Refer to Figure 1-5 for the location of the footswitch receptacle.
8. Make sure that the printer has the correct thermal paper installed properly. If the unit needs thermal
paper installed, refer to the Maintenance section of this manual for the installation instructions.
Figure 1-10, Probe Connection
Instrument Setup (continued)
Figure 1-8, Tip

16030-101 Rev. J 17
Instrument Setup (continued)
Application of Input Power
This section describes the Model 30 Pneumatonometer initialization.
1. Turn the instrument ON by pressing down the “I”
on the front panel ON /OFF switch. At start-up,
the Model 30 Pneumatonometer LCD readout will
display the initial screen for several seconds as the
instrument conducts a self-check test. After initial-
ization, the unit will display the Main Menu Screen.
Refer to Figure 2-1.
2. Touch the SETUP icon to enter the setup mode.
3. Touch the OK icon to set the options in the General
Setup mode. Refer to Figures 2-2 and 2-3.
A. Chart – this option sets the Chart function ON or
OFF. Default is ON.
B. Footswitch - this option sets the footswitch ON
or OFF. Default is OFF.
C. Volume – sets the volume level for the audible
prompts.
D. Brightness – sets the brightness intensity level
for the LCD Screen.
4. Touch the RETURN icon to return to the Main Menu
screen.
Figure 2-1, Main Menu Screen
Figure 2-2, Setup Menu Screen
Figure 2-3, General Menu Option Screen

16030-101 Rev. J
18
Instrument Setup (continued)
Application of Input Power (continued)
5. Touch the DOWN ARROW icon to highlight the Printout
Setup option.
6. Touch the OK icon to select the Printout Setup menu.
Refer to Figure 2-4.
7. Touch the UP or DOWN icons to highlight the option that
needs to be changed and touch the OK icon to select
the option. Refer to Figure 2-5.
8. Touch the LEFT, RIGHT, “+” or “–“ icons to change the
option to the appropriate setting. Refer to Figure 2-6.
9. Touch the OK icon to activate the Printout Setup menu
screen.
10. If all the options are set to the desired settings, touch the
RETURN icon to display to the Setup Menu screen.
11. Touch the RETURN icon to return to the Main Menu
screen.
Disconnection of Input Power
1. At any time, the power switch can be set to OFF. The
unit does not have a power down sequence. To termi-
nate operation of this instrument, press the ON / OFF
switch to the OFF position (O).
2. If this instrument is intended to be OFF for an extended
period of time, it can be disconnected from power by
detaching the power cord from the its receptacle.
Figure 2-4, Printout Setup Screen
Figure 2-5, Setup Menu Option Screen
Figure 2-6, Changing Setup Options

16030-101 Rev. J 19
Quick Verication Check
Verify the operation of the probe and base unit using the calibration
verifier in the Manual mode. Refer to Figure 2-7.
1. Fill the tube of the Calibration Verier up to the line marked 15
mmHg with ltered water.
2. From the Main Menu, touch the Manual IOP icon. A menu screen
will be displayed describing a summary of the Manual IOP mea-
surement process. Touch the OK icon to continue to the Manual
IOP screen.
3. The Model 30 Pneumatonometer will display the OD and OS
menu. Touch the OD or the OS icon and touch the probe of the
unit to the membrane and align the tip of the probe so that white
part of the probe handle is between the red and black lines on the
metal tube. When the probe is between the red and black lines,
the Model 30 Pneumatonometer will be ready to display the aver-
age IOP readings and its standard deviation. Refer to Figure 2-8.
Note: If the unit makes a loud vibration, remove the shipping bracket located on the back of the unit.
Note: The tone will settle to a lower tone when the probe is aligned properly. During an actual test,
the tone will change to a noticeably lower pitch when the standard deviation is below 1.0
mmHg for at least three seconds.
A lower tone change signies that the instrument has acquired usable data.
After the reading is acquired, the probe can be removed from the Calibration Verifier.
Note: If the IOP is not 15 mmHg (± 2.0 mmHg), go to the Troubleshooting section of this manual for
assistance. If the Troubleshooting section of the manual did not help, contact Reichert at the
address or phone number in the Introduction section of this manual.
4. Touch the HOME icon to display the Main Menu.
Instrument Setup (continued)
Figure 2-7, Calibration Verifier
15 mmHg Line
Membrane
Figure 2-8, Probe Red and Black Lines
Red
Red
Black
Alignment for
Measurement
Red and
Black Lines

16030-101 Rev. J
20
Operating principle
The Model 30 Pneumatonometer measures intraocular
pressure (IOP) non-invasively through applanation tonome-
try. The Pneumatonometer probe contains a gentle, floating
pneumatic sensor that touches the surface of the anesthe-
tized cornea with the exact amount of applanating force
required to take a tonometry or tonography measurement.
Refer to Figure 3-1.
The sensing element is a lightweight plastic tip covered with
a thin, highly elastic silicone membrane. The tip is mounted
on a floating piston supported by a porous bearing.
A precisely regulated flow of filtered air enters the piston from the Model 30 Pneumatonometer and travels
through the end of the sensor tip until it is blocked by the membrane. When nothing is touching the mem-
brane, air flows to the periphery of the tip, where it escapes through venting ports. However, when the tip
touches the eye, the pressure against the membrane causes it to seal the vents, blocking the escape of
air and building up pressure in the
system. The pressure increases until
it matches the IOP and stops when
the eye is applanated. At this point,
the membrane can no longer maintain
the seal. Any increased back pres-
sure in the system is released through
the venting ports. Applanation is
then automatically maintained by the
pneumatic feedback system. Refer to
Figure 3-2.
The Model 30 Pneumatonometer monitors the back pressure in the sensor and displays a real-time mov-
ing average of the IOP as well as its standard deviation. The Model 30 Pneumatonometer generates an
audible tone to tell you when the most accurate reading is obtained. The tone will change to a noticeably
lower pitch when the standard deviation among IOP readings remains below 1.0 mm Hg for three seconds.
The reading remains on the LCD readout when the probe is removed from the eye.
The Model 30 Pneumatonometer in the tonography mode calculates and displays the C value (aqueous
outflow coefficient) using either a two-minute or a four-minute tonography examination.
Figure 3-1, Pneumatonometer Probe
Figure 3-2, Probe Tip Open
Instrument Setup (continued)
Table of contents
Other Reichert Measuring Instrument manuals

Reichert
Reichert Tono-Pen AVIA User manual

Reichert
Reichert Tono-Pen AVIA Vet User manual

Reichert
Reichert AR700 User manual

Reichert
Reichert AR200 User manual

Reichert
Reichert OE200 Operating instructions

Reichert
Reichert AT550 User manual

Reichert
Reichert AT555 User manual

Reichert
Reichert AT550 User manual

Reichert
Reichert VET 360 User manual

Reichert
Reichert AR6 Series User manual