manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Reichert
  6. •
  7. Measuring Instrument
  8. •
  9. Reichert Tono-Pen AVIA Vet User manual

Reichert Tono-Pen AVIA Vet User manual

Tono-Pen AVIA Vet®Applanation Tonometer
Quick Reference Guide
230750-104 Rev. F, 02-05-2020
1. Hold the Tono-Pen®with the transducer end pointed straight down toward the floor.
2. Press and hold the button for 5 seconds. The instrument will “beep” 5 times and
the LCD will display [dn].
3. Hold the Tono-Pen steady for 15 seconds until the device “beeps” and the LCD
displays [UP].
4. Quickly and smoothly flip the Tono-Pen up, so the transducer is pointing straight up.
5. A properly functioning Tono-Pen will display [Pass] and the device will “beep”.
Press the operating button again to ready the pen for a measurement.
6. If [Fail] is displayed, repeat this verification procedure.
Ocu-Film®+ Instructions
CORRECT
Proper set up of the Ocu-Film® + tip cover is essential for the precision and
accuracy of the tonometry measurement. A new Ocu-Film + tip cover must be
used for each new patient. Be certain that the film is flat across the tip, but not
too taut or too loose.
Position the Ocu-Film +
applicator on the transducer.
Roll the Ocu-Film + tip
cover onto the tip.
Remove the
applicator.
Taking an IOP Measurement
To take a measurement, press the button once. [88888]will briey be dis-
played, followed by a double row of dashes [= = = =], indicating the Tono-Pen is
ready to measure.
The corneal surface needs only to be momentarily contacted. Indentation or
excessive pressure is not required and may lead to injury to the eye.
CORRECT INCORRECT
Transducer
Button LCD Display
Verication Procedure
Verication must be successfully performed
whenever the measurements are in question.
Verication can be performed with or with-
out an Ocu-Film + tip cover installed.
Note: The rst thing to be displayed on the
LCD screen during verication mode
is [dn] for down.
Keep the Tono-Pen
AVIA Vet covered
with an Ocu-Film +
tip cover at all times,
even in storage.
Tono-Pen AVIA Vet®Applanation Tonometer
Quick Reference Guide
230750-104 Rev. F, 02-05-2020
Reichert, Inc. • 3362 Walden Ave, Suite 100, Depew, New York USA 14043
Toll Free: 888-849-8955 • Tel: (716) 686-4500 • Fax: (716) 686-4545
www.reichert.com • Tono-Pen, Tono-Pen AVIA Vet and Ocu-Film are registered trademarks of Reichert, Inc.
Tono-Pen AVIA Vet Does Not Taking A Reading
• Tip may be dirty. Perform the cleaning procedure. If problem persists, send the
pen in for service.
• Ocu-Film + may be installed incorrectly. Refer to Ocu-Film + Instructions on
reverse.
• Use blue Reichert brand Ocu-Film + ONLY. Use of other brand tip cover may a󰀨ect
readings or may void your warranty.
• If the above does not resolve the issue, STOP and send the unit in for repair.
Tono-Pen AVIA Vet Displays [FAIL]
• Unit may be in verication mode. Perform the verication procedure.
• Verication procedure incorrectly performed. Pen is ipped up when [dn] is dis-
played. Perform the verication procedure.
• Tip may be dirty. Perform the cleaning procedure. If problem persists, send the
pen in for service.
Tono-Pen AVIA Vet Turns O󰀨
• Unit is going into power saving mode. If no measurements are taken within 15 seconds,
the Tono-Pen will go into battery saving mode. Turn it on by pressing the button once.
• Battery may be low. Replace the battery.
Tono-Pen AVIA Vet [FAIL]s Verication
• Tip may be dirty. Perform the cleaning procedure. If problem persists, send the
pen in for service.
• Pen is being ipped up too slowly. The pen should be turned up quickly and smoothly
when [UP] appears.
Troubleshooting Cleaning Procedure
1. Remove the Ocu-Film + tip cover.
2. Using canned air, place the outlet of the canned air firmly against the Trans-
ducer Tip as shown in the below image.
Note: Try to get a good seal between the canned air and the tip to blow any con-
taminants stuck in the tip safely into the internal cavity.
3. Blow air into the tip for approximately 3 seconds. Do not perform more than 3
times consecutively.
4. The tip will be cold; allow the tip to warm to room temperature.
5. Wipe the outside of the Tono-Pen with a clean, dust-free, cotton cloth dampened
with either a 70% isopropyl alcohol solution, 7.5% Hydrogen Peroxide solution, or
5.25% Bleach solution, and distilled water, or a Clorox¹ Bleach Germicidal Wipes
(0.55% sodium hypochlorite) or a Sani-Cloth² Prime Germicidal wipes (Didecyl
Dimethyl Ammonium Chloride, isopropyl alcohol and ethyl alcohol).
6. Perform the verification procedure after cleaning.
7. Install a new Ocu-Film tip cover as indicated in the User Guide.
Number of
Patients per Week
Number of Days
Between Cleaning
10 30
100 15
300 7
600 1
Cleaning ScheduleCleaning Tip
Tono-Pen Service
Go to www.reichert.com/rma.Fill out the RMA request form. An RMA will be sent to you via
email. Send in your instrument, inside the protective carrying case, with a copy of the RMA.
¹ Clorox is a registerd trademark of the Clorox Company, Oakland, CA.
² Sani-Cloth is a registered trademark of PDI, Inc., Woodcliff Lake, NJ.

Other Reichert Measuring Instrument manuals

Reichert AR200 User manual

Reichert

Reichert AR200 User manual

Reichert AR700 User manual

Reichert

Reichert AR700 User manual

Reichert OE200 Operating instructions

Reichert

Reichert OE200 Operating instructions

Reichert VET 360 User manual

Reichert

Reichert VET 360 User manual

Reichert DENSITY4 User manual

Reichert

Reichert DENSITY4 User manual

Reichert AT550 User manual

Reichert

Reichert AT550 User manual

Reichert 13940000 User manual

Reichert

Reichert 13940000 User manual

Reichert AT555 User manual

Reichert

Reichert AT555 User manual

Reichert Model 30 User manual

Reichert

Reichert Model 30 User manual

Reichert AR6 Series User manual

Reichert

Reichert AR6 Series User manual

Reichert AT550 User manual

Reichert

Reichert AT550 User manual

Reichert ABBE MARK III User manual

Reichert

Reichert ABBE MARK III User manual

Reichert Tono-Pen AVIA User manual

Reichert

Reichert Tono-Pen AVIA User manual

Popular Measuring Instrument manuals by other brands

Opt Lasers PLH3D 6W quick start guide

Opt Lasers

Opt Lasers PLH3D 6W quick start guide

Mahr Pocket Surf IV instructions

Mahr

Mahr Pocket Surf IV instructions

Gima BC401 user manual

Gima

Gima BC401 user manual

Fluidwell F112-P manual

Fluidwell

Fluidwell F112-P manual

Termoprodukt DT-1 user manual

Termoprodukt

Termoprodukt DT-1 user manual

ABB PointMaster 200 Series operating manual

ABB

ABB PointMaster 200 Series operating manual

DeWalt DW089K-XJ manual

DeWalt

DeWalt DW089K-XJ manual

Anritsu MS2830A Operation manual

Anritsu

Anritsu MS2830A Operation manual

Hanna Instruments HI 4211 instruction manual

Hanna Instruments

Hanna Instruments HI 4211 instruction manual

Sensitech TempTale 4 Instructions for use

Sensitech

Sensitech TempTale 4 Instructions for use

Bosch Rexroth VDP 92.1 manual

Bosch

Bosch Rexroth VDP 92.1 manual

Major tech MT775 instruction manual

Major tech

Major tech MT775 instruction manual

Atomos Samurai Blade quick start guide

Atomos

Atomos Samurai Blade quick start guide

Fluke 80PK-1 instruction sheet

Fluke

Fluke 80PK-1 instruction sheet

EDAN SONOTRAX Lite user manual

EDAN

EDAN SONOTRAX Lite user manual

Ludlum Measurements 2401-ECA manual

Ludlum Measurements

Ludlum Measurements 2401-ECA manual

wtw pH/ION 340i operating manual

wtw

wtw pH/ION 340i operating manual

Konica Minolta CHROMA METER CS-200 instruction manual

Konica Minolta

Konica Minolta CHROMA METER CS-200 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.