Reinhardt MWS 5MV-10 User manual

Manual
MWS 5MV-10 / MWS55MV
Microprocessor Weather Station
with datalogger (SD-card)
REINHARDT System- und
Messelectronic GmbH
Bergstr. 33, D-86911 Dießen-Obermühlhausen
Tel. 0049 - (0) 8196/934100 Fax 0049 - (0)8196/7005 or 1414
E-Mail: [email protected]
http://www.reinhardt-wetterstationen.de

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 2
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
Table of contents
1 Caution .......................................................................................................................................................5
1.1 Intended Use..................................................................................................................................5
1.2 Safety Regulations .........................................................................................................................5
1.3 Mounting.........................................................................................................................................5
2 Setting Up the Weather Station................................................................................................................6
2.1 Installation of Hardware..................................................................................................................6
2.1.1 Overvoltage protection ...........................................................................................................7
2.2 Software Installation.......................................................................................................................8
2.3 Starting the Software......................................................................................................................9
3 Technical details........................................................................................................................................10
3.1 The data logger ..............................................................................................................................10
3.2 In case of power failure ..................................................................................................................10
3.3 Maintenance...................................................................................................................................10
3.4 The sensors....................................................................................................................................11
3.4.1 The Temperature Sensor .......................................................................................................11
3.4.2 The Humidity Sensor..............................................................................................................11
3.4.3 The Pressure Sensor .............................................................................................................11
3.4.4 The Wind Speed Sensor ........................................................................................................12
3.4.5 The Wind Direction Sensor ....................................................................................................12
3.4.6 Rain / Precipitation Sensor (Option).......................................................................................12
3.4.7 Light Intensity Sensor (Lux-Sensor) (Option).........................................................................12
3.4.8 Ultraviolet-radiationsensor (UV-Sensor) (Option) ..................................................................12
3.4.9 Additional Sensor ...................................................................................................................13
3.4.10 Connecting an Additional Sensor.........................................................................................13
3.5 Sensor Accuracy ............................................................................................................................14
3.5.1 Measuring Ranges .................................................................................................................14
3.6 Adjustment of Additional Sensors ..................................................................................................15
3.7 Firmware Update............................................................................................................................19
3.8 Power Supply .................................................................................................................................20
3.8.1 Power consumption................................................................................................................20
3.9 Data Format ...................................................................................................................................21
3.9.1 Data access onto the SD-card ...............................................................................................22
3.9.1.1 Available commands: ...............................................................................................23
3.9.1.2 Description of the commands ..................................................................................23
3.9.1.3 Saving settings onto the SD-card .............................................................................25
3.9.1.4 Load settings from the SD-card ................................................................................25
3.10 System requirements ...................................................................................................................25

MWS5M_8M_e.p65 Page 3
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
4 Connections and Pin assignments..........................................................................................................26
4.1 Cables ............................................................................................................................................26
4.1.1 Data Cable - Allocation of the Connection Cable for 5M(V)-10 / MWS 55M(V) .....................26
4.1.1.1 Connection Diagram Standard data cable ...............................................................26
4.1.2 Allocation of the Connection Cable for MWS Heating............................................................27
4.1.2.1 Connection Diagram Cable for heating device .........................................................27
4.1.3 Allocation of the Connection Cable for GPS-Receiver (Garmin GPS 18x LVC) : ..................27
4.1.4 Allocation of the Connection Cable for Heating and GPS-Receiver (Garmin GPS18x LVC): 28
4.1.5 Allocation of the Connection Cable for Sensors with RS422-port..........................................28
4.2 Pin assignments.............................................................................................................................29
4.2.1 Allocation of the MWS 5M(V)-10 / MWS 55M(V)-Connection Sockets ..................................29
5 Excerpts from the Directions of DWD for Automatic Weather Stations...............................................30
6 Trouble Shooting.......................................................................................................................................31
6.1 Transfer problems ..........................................................................................................................31
6.1.1 Allowed cable lengths ............................................................................................................31
6.2 No data / wrong baud rate..............................................................................................................31
6.3 RS422/RS485 interface..................................................................................................................32
6.4 Protocol-files (weather32 software)................................................................................................32
6.4.1 Log-file in case of error (ErrLog.txt) .......................................................................................32
6.4.2 Logfile when starting up (log.dat) ...........................................................................................32
6.4.3 Show output (internal) ............................................................................................................32
7 Options.......................................................................................................................................................33
7.1 Available displays...........................................................................................................................33
7.1.1 Meteograph ............................................................................................................................33
7.1.2 DKA1......................................................................................................................................33
7.1.3 DMMK ....................................................................................................................................33
8 Technical Appendix...................................................................................................................................34
8.1 Control parameters for calibrating an additional sensor.................................................................34
8.2 Controlling the Microcontroller........................................................................................................35
8.2.1 Input parameters of MWS 5M(V)-10 / MWS 55M(V) -Microcontroller....................................35
(Excerpt from the instruction set - The most important commands) ...............................................35
8.2.2 SECURE-Mode ......................................................................................................................38
8.2.3 Requesting the microcontroller ..............................................................................................38
8.2.3 Meaning of the listing with !?0................................................................................................39
8.2.4 Order of the MWS 5M(V)-10 / MWS 55M(V) Sensors ...........................................................41
9 Instructions for Mounting the MWS5M(V)-10 / MWS 55M(V) Weather Station.....................................46
10 Packing MWS 5M(V)-10 / MWS 55M(V) properly...................................................................................47
11 Exchange Connectors.............................................................................................................................47

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 5
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
1 Caution
1.1 Intended Use
REINHARDT-Weather stations and sensors are exclusively built for stationary operation on a
fixed 1" pipe for automatically collecting climatic parameters outside.
Any use other than described above may cause damage of the product or lead to other dangers.
Do not mount the weather station in reach of children and pets.
Carefully read the complete operating manual. It contains important information about the in-
stallation and operation.
If the MWS 5M(V)-10 /MWS 55M(V) is not used for a longer period of time you should store it
lying to prevent escaping the lubricant from the ball bearings!
1.2 Safety Regulations
The instruments are manufactured according to modern technical standards and can be operated
without danger when used as directed.
Damage caused by non-observance of this operating manual can lead to forfeiture of warranty.
We shall not assume any liability for subsequent damage.
We shall not assume any liability for damage of items or persons caused by improper handling
or non-observance of the safety instructions! In such cases any guarantee claims shall become
null and void.
Dear customer, the following safety and hazard notices not only serve the protection of your health
but also the protection of the appliance. Please read the following points carefully.
The supply voltage is converted by isolated transformers into voltages of maximum 24VDC.
(The stations can be operated at voltages up to 28VDC).
Please do only use the supplied power supply units.
The weather station includes pointed and sharp-edged parts (i.e. windvane and edges of the
sensor's housing), which may cause injuries when handled without care.
Do not leave the packaging material lying around. These parts are dangerous toys in the hands
of children.
Handle the product with care. Blows or impact, or dropping it even from a small heigth will da-
mage it.
1.3 Mounting
The weather station is mounted on a 1" water-pipe. This pipe must be fixed very well to stand
high wind speeds above 150 km/h when the weather station is mounted.
The weather station is to be fixed well onto the 1" pipe with the screws. The stability of the weather
station on the pipe must be established. Really check this after mounting the weather station!
Please mount the weather station on a place you easy can reach for maintenance pur-
poses for the station can not be guaranteed to be absolutely maintenance free!!
You'll also find advice for mounting the weather station on your weatherCD.

MWS5M_8M_e.p65 Page 6
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
2 Setting Up the Weather Station
2.1 Installation of Hardware
MWS 5MV and MWS8 Weather Station has been designed for measuring the most important
weather parameters temperature, humidity, barometric pressure, wind direction and wind speed.
All parameters can be displayed graphically, digitally or as an "multiple display".
There is also a statistics and a history function.
Mount the weather station on a 1 " pipe. Orient the North-marker to North and clamp MWS
5M(V)-10 / MWS 55M(V) in place with the two pre-mounted high quality steel screws. Take care
that the pipe is mounted as perpendicularly as possible, otherwise the windvane will not work
properly and will preferably rotate in one direction.
N.B.
The wind vanes of the 5M(V)-10 / MWS 55M(V) weather station are very sensitive sensory equip-
ment which can easily be destroyed by mechanical influence.
Therefore do keep the packing of your weather station. Whenever you want to ship your weather
station later, this packing guarantees that the weather station is not damaged in transport !! Time
and again we have noticed that the units are damaged by wrong packing. This is very annoying
for the customer because he has to pay for all costs that may arise.
Please ask for a quotation of an original packing if it is no longer available to you (see last page).
The weather station must be set up at a place which is exposed to wind, because otherwise wind
direction and wind speed cannot be measured correctly. (Please see "Directions of Deutscher
Wetterdienst").
Connect the enclosed cable as follows:
Connect the 9pole connector to a free serial interface of your PC or to an USB port with a suitable
COM to USB converter (with driver software which creates a virtual COM port).
Plug the power supply into a power outlet 100-240VAC / 50-60Hz.
Note on Security
Please note that you must in any case use the power supplies which we have provided with your
MWS 5M(V)-10 / MWS 55M(V weather station or which are technically identical. Nominal value
must be between 4 V and 28 V DC voltage; it must be possible to apply at least 100 mA (with op-
tional heating element another 18 or 24VDC 1 A). Please note also that all the power supplies we
provide are only made for use in dry rooms.
The 7 pole plug is connected to the Power / data socket of the weather station.
There are other sockets for additional sensors, for the power supply of a heating device and for
an optional GPS-receiver. Power supply of the additional sensors can be provided by the weather
station or by their own power supply. The output voltage of the additional sensors must not exceed
+4.095 V, sensors with pulse output must be TTL-compatible.
IMPORTANT !
When removing any additional cables from the weather station, please don't forget to protect the
socket with the cap and don't leave the socket open. Otherways the contacts of the socket will start
corroding and be damaged within a short period of time!
Also don't forget to protect the plug of the cable from moisture by covering it when removing the
weather station for calibration i.e. In this case also remove the wall adaptor of the power supply to
ensure that no voltage is applied to the contacts of the weather station's plug when it is not used!
Otherways the plug will start corroding due to moisture and the cable has to be replaced within
a short period of time.

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 7
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
2.1.1 Overvoltage protection
The MWS 5M(V)-10 / MWS 55M(V) contains an integrated overvoltage protection (Suppressordi-
ods) at it's supply- and data wires.
The power linie is protected by a SMCJ26CA-diode. This diode breaks through at voltages above
31VDC and is able to carry up to 1500W for a few milli seconds.
CAUTION: If you connect voltages above 30VDC permanently to the power jack, this
diode will be destroyed within a few seconds!
The data lines are protected by SMBJ15CA-diodes able to carry transients up to 600W.
Also the housing made of high quality steel is connected to system ground.
This protection is suitable for protecting the station from damage due to overvoltage at close
lightning strikes.
For protection of the connected peripherals (computer, camera, i.e.) the customer has to take
further precautions (line protection, opto couplers, i.e.).
But please note that this protection is ineffective in case of direct lightning strike
due to extremely high energy of lightning!
Currents of up to 200.000A can cause voltages up to 20.000V on the housing which
leads to flashovers onto the internal electronic components and it's destruction.

MWS5M_8M_e.p65 Page 8
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
After installing the hardware you install the software.
2.2 Software Installation
Insert the WeatherCD into your CD-drive.
You need an HTML-capable browser (Netscape, Internet-Explorer or else).
If Autostart is activated, the CD starts on its own, if not, you execute STARTER.EXE in the CD root
directory. Now you follow the instructions of the WeatherCD.
Note
The weather 32 software creates the HKEY_CURRENT_USER\Software\ReinhardtGmbH\Wet-
ter\... key in the registry which is only created when running and is not deleted when the software
is uninstalled.
After the software has been installed you connect the 7-pole plug to the POWER connector of the
MWS 5M(V)-10 / MWS 55M(V). At the MWS 5MV you need to remove the lower 3 elements of the
radiation shield by loosen the upper 3 pagoda screws to perform.
See also the mounting hints on the weather CD!

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 9
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
2.3 Starting the Software
Start the software by double-clicking the program icon
After every start, the software checks the clock of MWS 5M(V)-10 / MWS 55M(V) and compares it
to your computer clock. If it differs more than set in the file WEATHER32STATION.INI the soft-
ware sets the MWS 5M(V)-10 / MWS 55M(V) clock automatically. Needless to say, the clock of the
connected PC must give the correct time, as MWS 5M(V)-10 / MWS 55M(V) is set to the PC-clock.
If your MWS 5M(V)-10 / MWS 55M(V) holds a GPS-receiver, it provides the correct time for MWS
5M(V)-10 / MWS 55M(V) (UTC time).
NOTE
In this case the weather station's clock CAN'T be set by the software! With the GPS receiver you
start the software with the parameter -GPS (See sotware manual) for otherways there may be
problems showing and storing data due to wrong time synchronisation!
Data recording to hard disk is started in the selected time interval (5 minutes by default). The
software should run for at least one memory interval so that at least one data file is stored on the
hard disk. This is important for reading out the datalogger later. (If there is no data file on the
hard disk, the datalogger cannot be read out as the software does not find a start date.)
NOTE
MWS 5M(V)-10 / MWS 55M(V) only starts recording data to its datalogger after the time has been
set. This guarantees that the datalogger holds only data with the correct time. The clock of the
weather station is preset to CET when delivered. If your local time is different please ensure to set
the correct timezone when using a GPS-receiver at the first start!
(See commands under 8.2 - Controlling the micro controller. (!ZZ)).
The MWS 5M(V)-10 / MWS 55M(V) has got an internal Lithium cell supplying the clock so that
storing of data is performed immediately after connecting a the power supply without the need to
set the clock before logging.
When using a GPS-receiver for setting the clock there may occur different problems, i.e. when the
software tries to set the clock! (See FAQ on the weather CD)
If the software is not started, there might be a faulty data file on the hard disk or not enough free
memory. This causes an abort.
The manual for the latest software you'll find here.

MWS5M_8M_e.p65 Page 10
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
3 Technical details
3.1 The data logger
The data logger of the MWS 5M(V)-10 / MWS 55M(V) has got an internal SD-Card with 4GB.
When storing data with an interval of 10 seconds the device stores 267840 data sets a month (31
days). With 10 active sensors (clock and date also counts as a sensor) the monthly data file has
got a size of app. 21 MB.
So there are app. 250MB of data a year which allows storing of data for nearly 15 years without
the need of deleting any data.
The data files are stored on the SD-card in the folder "/log/data".
Each month a new data file is being created which contains the month and the year in it's name.
The file suffix is "mws".
The name of the data file of march 2015 therefore is 03_2015.mws.
The data file are ASCII files (text files).
The SD-card contains other file important for operating the weather station. These files must not
be deleted or edited.
3.2 In case of power failure
The datalogger of MWS 5M(V)-10 / MWS 55M(V) is kept up in case of power failure (EEPROMS),
but no new data are stored.
In case of power failure, the clock of MWS 5M(V)-10 / MWS 55M(V) is kept running and doesn't
need to be set when power is applied again!
Due to power failure there are data missing in the data files (time gaps). These gaps are filled
with dummy values (-99999) by the weather32 software when reading the logger.
So missing data are indicated by measure values of -99999 in the datafile. The software then re-
cognizes in the values of -99999 that data are missing and creates a gap in the graphical displays.
3.3 Maintenance
Because of its elaborated sensors, MWS 5M(V)-10 / MWS 55M(V) needs no maintenance.
There are wear parts like the ball bearings and the fan which have to be replaced from time to time.
These parts have got a life time of app. 5-6 years under normal environment conditions.
Extreme conditions (operation directly at the sea, high mountin regions, extreme pollution, i.e.)
may cause very reduced life time of these parts.
MWS 5M(V)-10 / MWS 55M(V) weather station has been developed for stationary use under nor-
mal climatic conditions (temperate zone). Use under extreme conditions such as e.g. on board of a
ship, mobile use on a measuring vehicle etc. has not been tested. It is therefore not recommended
to set up the weather station where it is exposed to salt or salt water ( e.g. right at the coast etc.).
It can be used on a measuring car under certain conditions although the measured values of the
wind sensors cannot be reproduced.
Our warranty ends if there is any intervention into hardware or software from your side.

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 11
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
3.4 The sensors
3.4.1 The Temperature Sensor
Temperature measurement is based on a precision precalibrated sensor SHT25 from Sensirion.
The values are read internally via a I²C bus. The resulting measured value is additionally linea-
rised by the weather station.
By standard, the temperature sensor is mounted on the lower side of the weather station. A white
laquered pagoda protects against radiation and prevents a buildup of heat.
Range: from -40 °C to + 60 °C, measuring accuracy ± 0.3 °C, (display also possible in Kelvin or
°Fahrenheit)
CAUTION: Compared to temperature measurements in big shielded cabins the measured values
can be higher when the sun is shining. If the temperature measurements must correlate with
the measurements in big shielded cabins you should measure temperature in the shaodow or
measure with an additional temperature sensor placed in the shadow or in a big shielded cabin!
Unit: [°C]
3.4.2 The Humidity Sensor
is also based on a precision precalibrated sensor SHT25 from Sensirion. The values are read in-
ternally via a I²C bus. The resulting measured value is additionally linearised and compensated
by the weather station. The sensor is also mounted on the lower side of the weather station. A
Gore-Tex cover protects it from pollution or destruction by dust or insects.
The humidity sensor can be used in a temperature range between -40 °C to + 60 °C. It is linearized
to an accuracy of 2 %.
Range: from 10 to 100 %, measuring accuracy ±2 %, display also as dewpoint measurement in
°C or °F
NOTE
This sensor is very responsive to static charge and air pollution (dust, aggressive gases, but also salt).
Please note that under unfavourable conditions (i.e. microbic stress caused by moulds, bacteria)
this sensor ages faster than under normal conditions.
Unit: [%]
3.4.3 The Pressure Sensor
is a precalibrated digital sensor from Bosch (BMP085) which is read internally via a I²C bus. This
sensor already is linearized to 4 hPa for the whole temperature range, the barometer is tempe-
rature compensated. Another temperature linearisation reduces the deviation to less than 2 hPa
over the whole temperature range.
The sensor can be used in the temperature range of -40 °C to + 60 °C.
Measuring range: from 300 hPa to 1100 hPa with ± 0.8 hPa accuracy. Display can be reduced to
0 m above sea level (input of the local altitude in [m], display also in mm mercury column or Inch
mercury column).
An additional output of barametric pressure (reduced to sea level) is possible. 3 different formulas
are predefined within the weather station (simple formula, international barometric formula and
barometric formula of DWD considering temperature and humidity).
Further formulas you may use within the weather32 software.
This sensor can be transported by air cargo!
Unit: [hPa]

MWS5M_8M_e.p65 Page 12
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
3.4.4 The Wind Speed Sensor
is made up of an anemometer with magnetic scanning. Wind speed is measured without touch
using a Hall-sensor and a magnetic wheel with 12 magnets. A peak detector finds every wind
peak and hands them on the measuring software. An average value is determined within the
respective memory intervals.
Range: in km/h from 0 to 150 km/h with ± 2.5km/h measuring accuracy, (display also in m/s,
miles/h, Knot or Beaufort), starting speed < 0.5 m/s.
As we have a very comfortable, 3-fold way of measuring wind speed with current wind speed
(WG), average wind speed (WD) and wind peaks (WS), you can conform your wind measurement
to your very needs.
Please note that dependent on the current winds, the 3 different methods of measuring wind
speed can result in very differing graphs: When measuring WG, only a current value is written
in the selected measuring interval, when measuring WD and WS, there is continuous evaluation
and the whole measuring period is monitored.
Unit: [km/h]
3.4.5 The Wind Direction Sensor
There is a weather vane with a precision magnetical encoder and a rotation angle of 360 ° for
measuring wind direction. Wind direction is given in °, with 90° being East, 180° being South,
270° being West and 0 ° being North.
Range: in 360 °, measuring accuracy 5 °, starting speed, < 0.6 m/s, hysteresis < 8°.
Output is performed as WR (current winddirection) and WV (main wind direction within a storage
interval).
Unit: [°]
3.4.6 Rain / Precipitation Sensor (Option)
A self-emptying bucket is tilted by the collected rain. All the water that has been collected on
the normed area of 200 cm2is led through a funnel to the bucket. The bucket tilts whenever a
certain quantity of water has been collected. The tilting creates pulses which are counted. Out of
the pulses, the software calculates the rain that has fallen per m2. The current intensity of rain
is also found and displayed. Unit: [mm] [l/m²]
Range: from 0 to 5000 ltr/m2,
Measuring accuracy ± 0.2 ltr/m2
3.4.7 Light Intensity Sensor (Lux-Sensor) (Option)
measures light intensity in Lux. Measuring range is within the human eye response.
Range: from 0 to 220000 lx
Measuring resolution: 4 lux
3.4.8 Ultraviolet-radiationsensor (UV-Sensor) (Option)
measures ultraviolet radiation (UV-A) in mW/m². The spectral range is 320nm..395nm.
Range: 0 to 50000 mW/m²
Measuring accuracy: +/-10%

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 13
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
3.4.9 Additional Sensor
Any kind of sensors can be integrated as additional sensors. The information signal must be ap-
plied as voltage in the range between 0 V and + 4.095 V or it has to be adapted to MWS 5M(V)-10
/ MWS 55M(V) by a special amplifier. Negative or higher voltages are blocked, but out of security
reasons, should be avoided in any case!
The signals are linearized with up to 8 programmable inpterpolation value pairs by the MWS
5M(V)-10 / MWS 55M(V).
When can deliver a temperature sensor for measuring the soil or water temperature and a sensor
for measuring the temperature of the road surface.
The MWS 5M(V)-10 / MWS 55M(V) can be equipped with an input for current signals (0..20mA).
The signal of the current input is applied to pin6 and pin7 of the additional sensors socket. The
sensors output has got the sensor identifier ZC in that case while sensor ZB can't be used in par-
alled to the current measurment!
Additional sensors can be voltage supplied by the weather station as long as their current con-
sumption is below 25 mA.
NOTE
If current consumption exceeds this value, we cannot grant that MWS 5M weather station works
properly !!!
If they need higher currents, the sensors must have an external power supply and its ground has
to be connected to the MWS 5M(V)-10 / MWS 55M(V)-ground (Pin 1 of the 7-pole connector socket).
You can call for adjustment instructions of the additional inputs at any time.
3.4.10 Connecting an Additional Sensor
Additional sensors are connected to MWS 5M at the 8-pole socket.
The socket is allocated as follows:
Pin1 : GND
Pin2 : Input for precipitation sensor (TTL-pulse)
Pin3 : Input for analog additional sensor 4 (0..4.095V)
Pin4 : Input for analog additional sensor 5 (0..4.095V)
Pin5 : Input for analog additional sensor 1 (0..4.095V)
Pin6 : Input for analog additional sensor 2 (0..4.095V)
(option: I+ for current input 0..20mA)
Pin7 : Input for analog additional sensor 3 (0..4.095V)
(option: I- for current input 0..20mA)
Pin8 : Output operating voltage (9..20VDC) maximum 25mA

MWS5M_8M_e.p65 Page 14
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
3.5 Sensor Accuracy
Temperature: ± 0.3 °C
Humidity: ± 2.0 % (at 10°C..35°C)
Pressure: ± 0.8 hPa (at 0°C..50°C), ±2 hPa below 0°C
Rain/precipitation: ± 0.2 mm
Wind direction: ± 5° (at -10°C..50°C), hysteresis < 8°
Start speed: < 0.5m/s (at -10°C..50°C)
Wind speed: ± 2.5 km/h (at -10°C..50°C)
Start speed: < 0.6 m/s (at -10°C..50°C)
Global radiation: +/- 50W (at 10°C..40°C)
UV-radiation: +/- 10%
Light intensity: +/- 6%
Additional sensor: ± 1 % of final value (at 0°C..50°C)
3.5.1 Measuring Ranges
Temperature: from -40 ° to + 60 °, resolution 0.01 °
Relative humidity:f rom 0 to 100 % resolution 0.01 %
Dewpoint: from -40 ° to + 60 °, resolution 0.01 °
Barometric pressure: from 950 hPa to 1050 hPa in 0.02 hPa resolution
Absolute pressure: from 300 hPa to 1100hPa in 0.02 hPa resolution
Precipitation: from 0 to 5000 mm with 0.1mm resolution
Wind direction: 0 to 360 °, resolution 0.02 °
Wind speed: in km/h from 0 to 150 km/h with 0.05 km/h resolution
Global radiaion from 0 to 1300 W/m² (Spectral range 0.3..2.8µm)
Light intensity: from 0 to 220000 Lux (Spectral range: human eye response)
UV-radiation: from 0 to 50000 mW/m² (Spectral range 320..395nm)
Additional inputs: 0- 4.095 V
Supply 4-28 V, 150 mA at 18 VDC, standard equipment
Dimensions
Size: Outer diameter 90 mm, height 240 mm (MWS 5MV-10)
Weight: app. 1.05 kg (MWS 5MV-10)
Weight: app. 800 g (10m ready made cable with PSU)
The housing must be mounted centrally on a 1" pipe.
Standard cabling is 10 m - longer cabling on request.

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 15
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
3.6 Adjustment of Additional Sensors
You will find the adjustment value pairs of your weather station in the MWS5M-10.XXX or
MWS55M.XXX, with XXX a three-digit number.
These files are not in the scope of supply but you may get these files on request.
The additional sensor 1 ... 5 have got the identifiers ZA ... ZE.
By default the additional inputs are precalibrated in mV.
0mV cause an output of 0, 4095mV are a raw value of 65535.
Now you connect the additional sensor to the 8-pole socket.
Start a terminal program (e.g. Windows-terminal) and configure as follows: 9600 Baud, 8bit, no
parity, 1 Stopbit.
You now change into the adjustment mode of the weather station with the command !W.
All sensors will be output with the internal measured raw values.
Here you can find out which raw value is created by your additional sensor at a respective analog
value.
Example:
Expose a temperature sensor to 0°C aus, wait until the sensor has adapted to this temperature
and then you see the respective raw value (i.e. 12500). You proceed with e.g. 30°C etc.
Note these pairs of values and replace the resprective lines in the adjustment file. It is most im-
portant that the first pair of values (e.g. !L5,1,F...) must contain the lowest voltage and the last
pair of values must contain the highest voltage.
At least 2 pairs of values must be determined per sensor (e.g. !L5,1F,.... to L5,2,W..).
All pairs of values which are not needed must contain a raw value of 999999!!
After you have determined the necessary adjustment values and entered them in the adjustment
file, you can transmit the changed file to the weather station as follows:
Start program MWS_TranCfg.exe and then chose the name of your adjustment file. The adjust-
ment file contains many commands which are sent to the weather station.
Here you can see the steps of data transfer to your weather station with the tool MWS_Trancfg.
exe. Transferring the data will need app. 5 minutes at 9600 baud.
Either click into the window for automatic detection of your weather station or selcet the COM
port directly by right click as shown on the following pages:

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 17
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
Chose the file for transfer to the wea-
ther station.
After chosing the file click into the
window to start the transfer.
Sending the file to the weather station
now is running. When errors occur the
error codes will be highlighted in red
text color.

MWS5M_8M_e.p65 Page 18
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
Do not in any case change the pairs of values of the other sensors as this would lead to wrong
measurements of your weather station.
Example for a value calculation:
Additional temperature sensor at ZA with linear output of 10mV / °C, 0V at -40°C and a maximum
value of +200°C.
The calibration value pairs for the first basic value are (raw value 0 and temperature -40°C):
!L5,1,F0 und !L5,1,W-40
The calibration value pairs for the second basic value must be in between the min and max values.
(i.e. raw value 500mV and temperature +10°C):
!L5,2,F8002 und !L5,2,W10
The calibration value pairs for the last basic value are (raw value 2400mV and temperature +200°C):
!L5,3,F38409 und !L5,3,W200
Between these basic values the real measured values are calculated by simple rule of the three.
I.e. if the raw value is 19500 you'll need the calibration value pairs 2 and 3.
Calculation:
(38409-8002)/(200-10) results in 160/°C.
Then (19500-8002)/160 leads to 71.9.
At last you add the value of basic value 2 to 71.9 (71.9 + 10).
The temperature at a raw value 19500 therefore is 81.9°C.

REINHARDT System- und Messelectronic GmbH
Bergstr. 33 86911 Diessen-Obermühlhausen Postfach 12 41 D-86908 Diessen Tel. 08196 - 934100
MWS5M_8M_e Page 19
geändert am 08.04.2015 von DO
Manual MWS 5M(V)-10 / MWS 55M(V)
3.7 Firmware Update
The MWS 5M(V)-10 / MWS 55M(V) allows updating the firmware via the serial port.
Start the tool "MWS_WUpda.exe" to perform.
(On the WeatherCD under SOFTWARE/MWS_WUpda)
In this folder also the current firmware version is stored.
Current firmware 25.03.2015: MWS_UPV3_0115.hex (Version 3.0115)
Updating the firmware is quite the same as transferring calibration files decribed above.
Updating the firmware takes app. 2 minutes at 115200 baud, app. 25 minutes at 9600 baud.

MWS5M_8M_e.p65 Page 20
Manual MWS 5M(V)-10 / MWS 55M(V)
geändert am 08.04.2015 von DO
REINHARDT System- und Messelectronic GmbH
3.8 Power Supply
4-28 VDC, app. 70mA at 18VDC. 18VDC PSU with 10m ready made cable is standard equipment.
When connecting any additional sensors, please note, that the supply voltage for the additional
sensors is lead via a reverse polarity protection diode from the weather station's PSU.
Caution: Some additional sensors (LUX-sensor, Rain detector;..) need at least 9 VDC for proper
operation!
3.8.1 Power consumption
Typical power consumption of MWS 5M(V)-10 / MWS 55M(V) with fan but without additional
sensors at different voltages:
Current at 4VDC : 370mA, power consumption: 1.48W
Current at 5VDC : 255mA, power consumption: 1.28W
Current at 8VDC : 145mA, power consumption: 1.16W
Current at 10VDC : 110mA, power consumption: 1.10W
Current at 12VDC : 100mA, power consumption: 1.20W
Current at 15VDC : 83mA, power consumption: 1.25W
Current at 18VDC : 70mA, power consumption: 1.26W
Current at 24VDC : 56mA, power consumption: 1.34W
This manual suits for next models
4
Table of contents
Other Reinhardt Weather Station manuals
Popular Weather Station manuals by other brands

Honeywell
Honeywell TE653ELW - Portable Barometric Weather... user manual

TFA
TFA 35.1130 instruction manual

Bresser
Bresser ClimaTemp XXL + Multisensor-Set instruction manual

BALDR
BALDR B0359WST4H4PR Quick setup guide

BEARWARE
BEARWARE 303717/20200528HK016 manual

Levenhuk
Levenhuk Discovery Report W10 manual