Index Introduction
Congratulations on choosing the SYM-
PRO high-performance cleaning device.
This appliance is to be used exclusively
for the professional cleaning of removable
dentures as well as orthodontic appliances
and splints.
This appliance sets new standards of
functionality, performance, and ergono-
mics.
To guarantee long and trouble-
free use please read the follow-
ing instruction manual carefully
and always follow the included
safety notices.
Symbols
This manual contains symbols also found
on the appliance itself with the following
meaning:
Danger
Immediate risk of injury
High Voltage
Danger – high voltage
Caution:
Ignoring this notice may cause
damage to both appliance and
object.
Notice
A general notice that improves
and eases use.
For indoor use only.
Disconnect from mains supply
before opening the appliance.
The appliance complies with
current EU directives.
Introduction ...................................................................... 17
Symbols ............................................................................. 17
User Notices.................................................................... 18
Instruction manual
1. Unpacking and installing..................................... 18
2. Appliance Description .......................................... 18
3. Connecting and switching on ........................... 18
4. Operation.................................................................... 19
4.1 Preparing the object ..................................... 19
4.2 Cleaning fluid ................................................... 19
4.3 Filling the cleaning vessel.......................... 19
4.4 Choose cleaning time and RPM............. 20
4.5 Starting, pausing, and
stopping cleaning........................................... 20
4.6 Remove the cleaning object..................... 20
4.7 Rinsing and refilling cleaning needles. 20
4.7.1 Rinsing .......................................................... 20
4.7.2 Adding cleaning needles............................... 21
4.8 Rinsing the vessel cover ............................ 21
5. Cleaning and Maintenance................................ 21
5.1 Replacing cleaning needles...................... 21
5.2 Cleaning vessel.............................................. 22
5.3 Changing a fuse ............................................. 22
6. Spare Parts ............................................................... 22
7. Presentation.............................................................. 22
8. Available delivery quantities: ............................ 22
9. Accessories............................................................... 22
10.Troubleshooting....................................................... 23
Information for the purchaser
A. Application ................................................................. 24
A.1 Correct use of the unit................................. 24
A.2 Unapproved use............................................. 24
A.3 Ambient conditions
(according to DIN EN 61010-1) .............. 24
B. Hazards and warnings......................................... 24
C. Approved personnel.............................................. 25
D. Prior to installation ................................................. 25
E. Repair........................................................................... 25
F. Disposal instructions............................................. 25
F.1 Disposal instructions for
countries in the EU........................................ 25
G. Technical data.......................................................... 26
H. Disclaimer .................................................................. 26
I. Guarantee.................................................................. 26

EN - 18 -
Within the EU, this device is
subject to the provisions of the
WEEE directive.
Other symbols are explained when displa-
yed.
(E) Retaining magnet
(F) Cleaning time indicator
(G) RPM indicator
(H) RPM selection key
(I) +/- buttons, change cleaning time
(K) Start/ pause button
(L) Stop button
(M) Mains switch
(N) Mains connection
(O) Fuse
3. Connecting and switching
on
Ensure the mains supply and
voltage marked on the name-
plate is identical.
Choose a power cable with the proper•
plug for your mains supply.
Connect the supplied power cable to the•
mains connection socket (N, Fig 3).
Connect the power cable to the mains•
supply.
Turn on the appliance at the mains•
switch (M, Fig. 2).
Instruction manual
1. Unpacking and installing
Remove the appliance and all accessori-•
es from the packaging.
The cleaning needles are found
in the white cleaning vessel.
Place the appliance on a bench of com-•
fortable height.
Place the tweezers supplied with the unit•
and the retaining magnets in the holder
on the back of the appliance (Fig. 4).
Danger of injury through falling
equipment!
Remove the retaining magnets
from the holder when transpor-
ting the appliance.
2. Appliance Description
(See fig. 1, 2, and 3)
(A) High-performance cleaning device
SYMPRO
(B) Cleaning vessel (white) for needle
cleaning using SYMPROfluid Univer-
sal
(C) Cleaning vessel (gray) for SYMPRO-
fluid Nicoclean
(D) Tweezers
User Notices
Instruct the user according to
the information in this manual
on application, possible dan-
gers, and operation of the SYM-
PRO high-performance cleaning
device.
This user information should be made
available to the operator at all times.
Further information can be found in the
‘User notices’
section at the end of this manual.

EN- 19 -
4. Operation
4.1 Preparing the object
Inspect the object for unstable
areas or damage before cleaning
it in the appliance.
These areas may be affected by
chemical and mechanical stress
during cleaning.
Delicate areas and function ele-
ments (such as thin crown mar-
gin, electroplated parts, friction
surfaces of external telescopes)
may be damaged.
As a rule, these will have to be
covered with an appropriate
wax/silicone layer.
High-gloss areas may become
dulled through cleaning.
If necessary, cover the high-
gloss areas with an appropriate
wax/silicone layer.
4.2 Cleaning fluid
SYMPROfluid• Universal:
Special cleaner for use in the SYMPRO
high-performance cleaning device.
Ready-to-use solution for the removal
of stubborn soils, bleak, calculus, stains
(tea, coffee or red wine etc.), as well as
adhesive cream.
SYMPROfluid• “Nicoclean”:
Special cleaner for post-cleaning stub-
born nicotine deposits.
Ready-to-use solution.
Please observe the processing order:
Without denture adhesive deposits,
continue directly with 2.:
1. When soiled with denture adhesive
deposits, roughly clean the denture
using a dry disposable cloth first.
2. Clean using 200 ml SYMPROfluid
Universal in the cleaning device for
approx. 10-30min. (white vessel).
With nicotine stains (otherwise continue
with 4.)
3. Place into the grey vessel and im-
merse entirely with SYMPROfluid
Nicoclean.
Immersion time approx. 5 – 10 min
(without the cleaning device).
Remove any residual stains by simple-
mechanical cleaning with a toothbrush
or denture brush under warm running
water.
4. Rinse and disinfect the cleaned object
thoroughly before transferring to the
patient.
The submerged object may
cause bacterial contamination of
the cleaning fluid.
For this reason: Always use
fresh cleaning liquid for every
new cleaning sequence!
For information on how to dispose of the
cleaning solutions, please consult the
appropriate product data sheets.
4.3 Filling the cleaning vessel
For use in the high-performance cleaning
device.
Fill the white cleaning vessel (B, Fig. 1)•
with:
75 g cleaning needles (corresponding-
to one packaging unit),
200 ml- SYMPROfluid Universal,
equivalent to filling the vessel to the
MIN mark.
the object to be cleaned.-
Make sure the liquid level stays clear of
the MAX mark.
Close the vessel.•
Place the vessel in the appliance.•
Do not overfill the vessel.
Never use the appliance with the
cleaning vessel open (Fig. 5).

EN - 20 -
before the cleaning time has elapsed
by pressing the Stop button.
After the cleaning time has elapsed the•
indicator shows „00“.
Stop cleaning altogether by pressing the•
Stop button (L, Fig. 2).
The display shows the previously set
cleaning time.
To improve the cleaning perfor-
mance, the sense of rotation will
change every 30 seconds.
With especially stubborn stains
it may be necessary to alter to a
longer running time (+ max.
10min) and then short, manual
cleaning, under running water,
using a hard toothbrush.
4.6 Remove the cleaning object
After cleaning:
Remove the cleaning vessel and open•
the lid while holding over a washbasin or
sink unit.
Never open the vessel while still
inside the appliance (Fig. 5).
Place the retaining magnet on the outer•
wall of the vessel such that the cleaning
needles accumulate on the inner wall
(Fig. 6).
You can now remove the object with the•
enclosed pliers (Fig. 7).
Before delivering the object to the pa-•
tient, rinse it with a copious amount of
water and disinfect.
Carefully check the object for
any adhering cleaning needles
before delivering.
4.7 Rinsing and refilling
cleaning needles
The cleaning needles can be
used more than once.
4.7.1 RINSING
After each cleaning cycle, rinse the clea-
ning needles and prepare fresh cleaning
fluid.
4.4 Choose cleaning time and
RPM
Adjust the cleaning time with the +-/ but-•
tons (I, Fig. 2) (range 0 to 30 minutes).
Adjust RPM with the selector (H, Fig. 2).•
The indicator now shows the chosen
RPM.
The cleaning time in the appli-
ance will depend on the nature
of the cleaning solution, the
selected RPM and the degree of
soil and staining.
It is advisable to check the
results several times during
cleaning if you have no previous
experience with this appliance.
Never open the vessel while still
inside the appliance (Fig. 5).
Recommended RPM:
1200 RPM:•
Cleaning objects with delicate areas and
function elements.
1600 RPM:•
Cleaning objects partly made of precious
metal alloys, orthodontic appliances.
2000 RPM:•
Cleaning objects made of non-precious
alloys, complete dentures, dentures with
cast-metal frameworks and splints.
4.5 Starting, pausing, and
stopping cleaning
Start cleaning with the Start/ Pause•
button (K, Fig. 2).
The appliance stops automatical--
ly when the set cleaning time has
elapsed.
The remaining time is displayed in the-
indicator during cleaning.
Cleaning can be stopped and restarted-
by pressing the Start/ Pause button.
Once cleaning is stopped, the remai--
ning cleaning time can be adjusted.
You can change the set RPM at any-
time during cleaning.
Cleaning can be stopped at any time-

EN- 21 -
To do this:
Hold the retainer magnet against the•
underside of the cleaning vessel (Fig. 8).
Carefully pour out the cleaning fluid (for•
information on how to dispose of the
cleaning solutions, please consult the
appropriate product data sheets)..
The magnet holds the cleaning
needles in the vessel.
Rinse the vessel and cleaning needles•
with clean water.
Hold the retainer magnet against the un-•
derside of the cleaning vessel and pour
out the water.
Once rinsed, the needles can be kept for
longer periods of time in an open, well
dried cleaning vessel.
4.7.2 ADDING CLEANING NEEDLES
Some cleaning needles will invariably be
lost when cleaning the needles and the
denture.
This will reduce the cleaning performance.
To bring the amount of cleaning needles
backup to the required amount of 75 g,
proceed as follows:
Place the cleaning vessel (with cleaning•
needles, but without liquid) on a scales.
A dry vessel and cleaning needles:•
Add cleaning needles until the scales-
indicates a weight of between 144 g
and 145 g.
A wet vessel and cleaning needles:•
Add cleaning needles until the scales-
indicates a weight of between 154 g
and 155 g.
4.8 Rinsing the vessel cover
The cover of the vessel, especially the
membrane (Fig. 9) and seals must be
rinsed under clean water after each
cleaning cycle for hygienic and functional
reasons.
5. Cleaning and
Maintenance
The owner and user of the appli-
ance are responsible for proper
disinfection of the cleaning
needles, cleaning beaker and lid,
as well as the compliance with
national hygiene regulations.
Clean the appliance using only a damp
cloth.
Never clean the appliance under
running water or immerse in
water.
Do not steam clean the appli-
ance.
Do not use solvent-based, ag-
gressive or scouring cleaners.
5.1 Replacing cleaning needles
Replace the cleaning needles when the
cleaning and polishing effect starts to
subside.
To do this:
Hold the retainer magnet against the•
underside of the cleaning vessel (Fig. 8).
Carefully pour out the cleaning liquid.•
(For disposal please see the product
data sheet for the cleaning liquid).
Rinse the cleaning beaker and the•
needles which are to be disposed of with
water.
Remove the retaining magnet and tip out•
the old needles.
Rinse out the cleaning vessel.•
Fill with new cleaning needles (75 g,•
corresponding to one packaging unit).
Cleaned and disinfected cleaning needles
can be disposed of with the regular recy-
cleable waste.

EN - 22 -
5.2 Cleaning vessel.
Cleaning vessels are not heat-proof.
Do not place cleaning vessels in
a dish washer or sterilise in an
autoclave.
5.3 Changing a fuse
Before replacing a fuse, dis-
connect the appliance from the
mains.
Never use a fuse of larger impe-
dance.
The fuse is found next to the mains con-
nection (O, Fig. 3).
Disconnect from mains supply.•
Unlatch the fuse holder on both sides•
and remove.
Remove faulty fuse and replace with a•
new one.
Replace the fuse holder by sliding back•
into the socket until both sides lock into
place.
6. Spare Parts
Refer to the spares list at the end of this
manual for numbers of wearing and re-
placement parts.
7. Presentation
1 SYMPRO, high-performance cleaning
device
1 Cleaning vessel, white, with cover (for
use in the cleaning device with
SYMPROfluid Universal)
1 Cleaning vessel, grey, with cover (for
use with SYMPROfluid Nicoclean)
1 Cleaning needles (container, 75 g)
1 SYMPROfluid Universal (400ml)
1 Tweezers
1 Retaining Magnet
1 Mains Cable (shockproof plug, 230 V)
1 Mains Cable (NEMA plug, 120 V)
1 Instruction manual
8. Available delivery
quantities:
6500-0000 SYMPRO, 100-240 V /
50/60 Hz
9. Accessories
6500-0501 Cleaning vessel, white
1850-0007 Cleaning vessel, grey
6500-0550 Cleaning needles (75 g)
6500-0600 SYMPROfluid Universal
(2x 2l)
6500-0610 SYMPROfluid Nicoclean
(2x 2l)

EN- 23 -
10. Troubleshooting
Error Cause Solution
No display• Appliance not connected•
to mains.
Appliance not turned on.•
No power from the•
socket.
Faulty fuse.•
Check correct fit of power cable•
in the appliance socket and wall
socket.
Turn on appliance at the mains•
switch (M, Fig. 3).
Check power socket, for example,•
with another appliance.
Replace the fuse (see section 5.3).•
If the fuse blows repeatedly, have
the appliance inspected by a ser-
vice technician.
Cannot change•
cleaning time.
Appliance still running.• Interrupt the cleaning process with•
the Start/ Pause button, and then
adjust cleaning time.
Cannot save•
cleaning time.
Cleaning time was•
entered during an inter-
rupted cleaning process.
Stop the cleaning process with the•
Stop button, and then enter new
cleaning time.
Insufficient•
cleaning perfor-
mance
cleaning time too short•
Active ingredients of the•
cleaning fluid spent.
Insufficient number of•
cleaning needles.
Cleaning needles spent.•
see chapt. 4.5.•
Prepare fresh cleaning fluid after•
each application.
Fill up the cleaning needles (see•
section 4.7.2).
Replace the cleaning needles (see•
section 5.1).
Cleaning need-•
les have corro-
ded.
Cleaning needles have•
been insufficiently
rinsed.
Cleaning needles have•
been stored incorrectly.
Rinsing and storage see chap-•
ter 4.7.1.
Removing the corrosion:•
Clean the needles with 100 ml
SYMPROfluid Universal, wi-
thout contents, for 5 minutes at
2000 rpm.
Dispose of cleaning liquid.
Cleaning needles in use•
for longer period of time.
The needles are so-called con-•
sumable goods. If it is not possible
to regenerate the needles, or the
needles‘ performance deteriorates,
see chapter 5.1

EN - 24 -
The following information is designed
to help you use the SYMPRO high-per-
formance cleaning device safely in your
laboratory or your dental office.
Instruct the user according to
the information in this manual on
application, possible dangers,
and use of the SYMPRO high-
performance cleaning device.
Ensure that this user information is availa-
ble to operators.
A. Application
A.1 Correct use of the unit
This appliance is to be used exclusively
for the professional cleaning of removable
dentures as well as orthodontic appliances
and splints.
The appliance is designed for non-sterile
use only.
A.2 Unapproved use
Use only Renfert original accessories and
materials (such as cleaning fluid) with this
appliance.
The use of other accessories is against
intended use stipulations.
A.3 Ambient conditions
(according to DIN EN
61010-1)
The appliance should only be operated:
indoors•
up to 2,000 metres above sea level•
at an ambient temperature of 5ºC - 40ºC•
[41ºF - 104ºF] *)
at a maximum relative humidity of 80% at•
31ºC [87.8ºF], linear reduction up to 50%
relative humidity at 40ºC [104ºF] *)
with mains electricity supply provided•
that the voltage fluctuation is within 10%
of the rated value
with Pollution Degree 2•
with Overvoltage Category II•
*) The unit can be used at a temperature of 5ºC - 30ºC
[41ºF - 86ºF] and at a humidity of up to 80%. At
temperatures of 31ºC - 40ºC [87.8ºF - 104ºF] the humidity
must reduce proportionately to ensure that the unit can
be operated (e.g. at 35ºC [95ºF] = 65% humidity, at 40ºC
[104ºF] = 50% humidity). The unit should not be operated at
temperatures above 40ºC [104ºF].
B. Hazards and warnings
Granted protection is not given
any longer in case the unit is
not operated according to the
instruction manual on hand.
Only to be used indoors. The
appliance is only intended for
dry operation and should not be
used or stored outdoors or in
wet conditions.
The appliance should only be
operated using a mains cable
with a country-specific plug sy-
stem. Any modification required
should only be carried out by an
electrician.
The appliance should only be
operated if the data on the rating
plate corresponds with the data
of regional mains voltage.
The appliance should only be
plugged into earthed sockets.
Regularly check connecting
cables and hoses (e.g. mains
cable) for damage
(e.g. kinks, tears, porosity) or
deterioration. Do not operate
appliance with damaged mains
wires, hoses or other defects.
Risk of injury!
Risk of injury if the components
used are not approved.
Always use original Renfert
components.
Information for the purchaser

EN- 25 -
Cleaning needles contain a
slight amount of nickel
(see product data sheet).
Rinse the object under running
water after cleaning and disin-
fect.
Check the object after cleaning
and remove any adhering
cleaning needles.
Contact with cleaning fluids or
prolonged inhalation of their
vapours may cause irritation.
Use appropriate personal
protective gear (protective
goggles, gloves).
Please note the information
contained within the Material
Safety Data Sheet.
Damage or discolouration may
occur should the cleaning liquid
come into contact with other
objects or laboratory furniture.
This is particularly true of PVC,
linoleum, marble, quarry stone,
stainless steel and wood sur-
faces.
In case of contamination,
immediate team with a copious
amount of water.
Danger of scalding!
Prolonged use of the appliance
may cause the liquid to heat up.
Always use of pliers to remove
the object.
The appliance contains strong
magnets!
People with cardiac pacemakers
should keep 50 cm from the
appliance.
Do not place data storage
mediums, credit cards or wat-
ches in the vicinity.
C. Approved personnel
The SYMPRO high-performance cleaning
device should be operated and maintained
by trained personnel only.
D. Prior to installation
Compare the data on the rating
plate with the data of the regio-
nal mains voltage before instal-
lation.
The appliance should only be
plugged into earthed sockets.
E. Repair
Repairs should only be carried out by
qualified personnel.
Repairs to the electrical equipment,
which are not specified in these operating
instructions, should only be carried out by
an electrician.
F. Disposal instructions
The appliance should be disposed of by a
specialist firm. The specialist firm should
be informed of any hazardous residue in
the appliance.
F.1 Disposal instructions for
countries in the EU
To conserve and protect the environment,
prevent environmental pollution and
improve the recycling of raw materials, the
European Commission adopted a directive
that requires the manufacturer to accept
the return of electrical and electronic units
for proper disposal or recycling.
Within the European Union appliances
with this symbol should not therefore be
disposed of in unsorted domestic waste:
For more information regarding proper
disposal please apply at your local authori-
ties.

EN - 26 -
Subject to alteration
G. Technical data
Rated voltage: 100 - 240 V / 50/60 Hz
Power consumption: 90 VA
Mains power fuse T1,6AL / 250V
Cleaning RPM: 1200 / 1600 / 2000 rpm
Dimensions (W x H x D):
150 x 240 x 280 mm
Weight (without vessel) approx.: ca. 3 kg
Noise level: < 70 dB(A)
H. Disclaimer
Renfert GmbH is not liable for claims for
compensation or claims under guarantee
if:
the product is used for purposes•
other than those stated in the opera-
ting instructions
the product is modified in any way –•
apart from modifications described in
the operating instructions
the product has not been repaired by•
a specialist firm or original Renfert
replacement parts have not been used
there is continued use of the product•
despite obvious safety defects or
damage
the product has been subjected to•
mechanical knocks or has been drop-
ped.
I. Guarantee
Used according to manufacturer’s instruc-
tions, all parts of the SYMPRO carry a
3-year guarantee.
The original sales invoice of the specialist
depot is required for a claim under guaran-
tee.
Parts subjected to normal wear and
consumables (for example, the cleaning
needles) and tear are not covered by the
guarantee.
The guarantee is rendered void in the
case of incorrect use, non-adherence
to the operating, cleaning, servicing or
connection instructions, repairs carried out
by the owner or repairs that are not carried
out by a specialist firm, use of replace-
ment parts from another manufacturer or
unusual or unapproved uses not specified
in the user guidelines.
Successful claims under guarantee do not
extend the guarantee period.
Full operational safety can only be
guaranteed if RENFERT system com-
ponents (cleaning appliance, needles,
vessels, fluids) are used exclusively.
Other manuals for SYMPRO
4
Table of contents
Other Renfert Medical Equipment manuals
Popular Medical Equipment manuals by other brands

SSCOR Duet
SSCOR Duet 2314 series Operating Instructions & Maintenance Manual

Cholestech
Cholestech LDX Quick reference guide

Cardinal Health
Cardinal Health Kangaroo Joey instructions

Defibtech
Defibtech DDU-2300 operating guide

Otto Bock
Otto Bock 7000P Instructions for use

EarlySense
EarlySense noomi installation guide