ResMed lumis series User manual

Clinical quick start guide
English
Air outlet for connecting the air tubing
Dial for navigating and changing options
Home button for returning to
the Home screen at any time
Start/Stop button for starting therapy
LCD screen which displays
therapy parameters and data
SD card cover to access
the SD card
Accessing the Clinical Menu
+
To access the Clinical Menu, press and hold the
dial and the Home button for three seconds.
The Home screen is displayed with an
unlockicon in the top right corner of the
screen.
The Home screen of the Clinical Menu consists
of two areas:
• Settings - to change therapy and device settings
according to the patient’s needs.
• Sleep Report - provides a summary of the
patient’s therapy session.
Adjusting Settings
3. Turn the dial to adjust the
setting and press the dial
to save the change.
1. Access the Clinical Menu,
highlight Settings and press
the dial. The Settings menu
is displayed.
2. Turn the dial to highlight
the setting you want to
adjust and then press
the dial.
StartingTherapy
To start therapy, press the Start/Stop button. Once therapy begins, the
Monitoring screen is displayed. This screen may differ according to the
configuration of the device and displays:
• Current treatment pressures - The pressure bar shows the inspiratory
and expiratory pressures in green. The green bar will expand and
contract as the patient breathes in and out.
• Pressure settings - (Displays Mode, IPAP, EPAP and pressure support).
• Humidifier heating icon - If a humidifier is connected.
• Humidifier settings - If a humidifier is connected.
• Ramp setting - If enabled.
QuickNav
QuickNav is a feature that provides quick navigation between the Monitoring screens and the
Settings menu. Changes can be made to the patient’s settings whilst the patient is receiving
treatment. To use QuickNav:
Double click the home button to toggle
between the Settings screen (where you
can make adjustments to your patient’s
therapy) and the Monitoring screen
(where you can check the impact of those
adjustments on the patient’s therapy).
Viewing the Sleep Report
Once therapy is stopped, the Sleep Report screen is displayed.
The Sleep Report screen
shows sleep quality and
mask seal status for the
most recent therapy
session.
Turn the dial to scroll down
to view more detailed
usage data.
The parameters displayed will depend on the therapy mode.
Refer to the Clinical Guide for further information on the Sleep Report screen parameters
ResMed.com
ResMed Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153 Australia
DISTRIBUTED BY
ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA
ResMed (UK) Ltd 96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK
See www.resmed.com for other ResMed locations worldwide. For patent information, see www.resmed.com/ip. Lumis, ClimateLine and
HumidAir are trademarks of ResMed Ltd. ClimateLine is registered in U.S. Patent and Trademark Office.
© 2015 ResMed Ltd. 288341/1 2015-04
ResMed.com
Settings menu
The following table lists key parameter settings. Refer to the Clinical Guide for further information.
Therapy
Parameter Description S ST T PAC iVAPS CPAP Range
Mode Sets the therapy mode available on the
device.
IPAP Sets the pressure to be delivered to the
patient when the device is triggered into
inspiration.
4–25 cm H2O (4–25 hPa),
0.2 cm H2O (0.2 hPa) increments.
EPAP Sets the pressure to be delivered to the
patient when the device is cycled into
expiration.
2–[IPAP] cm H2O (2–[IPAP] hPa),
0.2 cm H2O (0.2 hPa) increments.
Min PS Sets the minimum pressure support
delivered by the device.
0–20 cm H2O (0–20 hPa),
0.2 cm H2O (0.2 hPa) increments.
Max PS Sets the maximum pressure support in
iVAPS mode.
0–23 cm H2O (0–23 hPa), 0.2 cm H2O
(0.2 hPa) increments.
Max EPAP Sets the maximum EPAP delivery by the
device.
2–25 cm H2O (2–25 hPa),
0.2 cm H2O (0.2 hPa) increments.
Min EPAP Sets the minimum EPAP delivered by
the device.
2–25 cm H2O (2–25 hPa),
0.2 cm H2O (0.2 hPa) increments.
Ti Max Sets the maximum limit on the time the
device spends in IPAP.
0.3–4.0 sec, 0.1 sec increments.
Ti Min Sets the minimum limit on the time the
device spends in IPAP.
0.1–[Ti Max] sec, 0.1 sec increments.
Rise Time Sets the time taken for pressure to
increase from EPAP to IPAP.
Min / 150–900 ms, 50 ms increments.
Mask Select the type of mask used by the
patient.
Full Face / Nasal / Pillows / Pediatric
Comfort
Parameter Description S ST T PAC iVAPS CPAP Range
Ramp Down Enable / Disable the Ramp Down
feature.
On / Off
Climate Ctrl Available when HumidAir humidifier
is used and ClimateLineAir heated air
tubing is connected.
Manual / Auto
Tube Temp. Set the minimum temperature of air
delivered by heated air tubing such as
ClimateLineAir.
Off / 16–30°C,
1° increments
Humidity
Level
Set the humidity level. Off / 1–8
Read your entire Clinical Guide before use.

Other manuals for lumis series

4

Other ResMed Fan manuals

ResMed Stellar 100 User manual

ResMed

ResMed Stellar 100 User manual

ResMed Stellar 150 User manual

ResMed

ResMed Stellar 150 User manual

ResMed Stellar 150 Guide

ResMed

ResMed Stellar 150 Guide

Popular Fan manuals by other brands

Ebmpapst A4D500-AM03-01 operating instructions

Ebmpapst

Ebmpapst A4D500-AM03-01 operating instructions

Ebmpapst W2E200-HH86-90 operating instructions

Ebmpapst

Ebmpapst W2E200-HH86-90 operating instructions

Ebmpapst M4Q045-EF01-75 operating instructions

Ebmpapst

Ebmpapst M4Q045-EF01-75 operating instructions

Ebmpapst A3G990-AZ02-01 operating instructions

Ebmpapst

Ebmpapst A3G990-AZ02-01 operating instructions

TriStar VE - 5930 instruction manual

TriStar

TriStar VE - 5930 instruction manual

Concept2 conclima VT6930 manual

Concept2

Concept2 conclima VT6930 manual

System air AR 200E2 sileo operating instructions

System air

System air AR 200E2 sileo operating instructions

Ebmpapst A2E170-AF23-34 operating instructions

Ebmpapst

Ebmpapst A2E170-AF23-34 operating instructions

HOME NETWERKS 7130-08-BT user manual

HOME NETWERKS

HOME NETWERKS 7130-08-BT user manual

Ebmpapst R3G140-AW05-12 operating instructions

Ebmpapst

Ebmpapst R3G140-AW05-12 operating instructions

Ebmpapst R2E133-BH66-07 operating instructions

Ebmpapst

Ebmpapst R2E133-BH66-07 operating instructions

Ebmpapst R3G250-RE07-05 operating instructions

Ebmpapst

Ebmpapst R3G250-RE07-05 operating instructions

Ebmpapst W3G990-GY32-11 operating instructions

Ebmpapst

Ebmpapst W3G990-GY32-11 operating instructions

Ebmpapst R2E133-BH66-14 operating instructions

Ebmpapst

Ebmpapst R2E133-BH66-14 operating instructions

Ebmpapst G2D140-AC30-18 operating instructions

Ebmpapst

Ebmpapst G2D140-AC30-18 operating instructions

Ebmpapst S2E200-AH38-01 operating instructions

Ebmpapst

Ebmpapst S2E200-AH38-01 operating instructions

Ebmpapst R2E220-AB06-84 operating instructions

Ebmpapst

Ebmpapst R2E220-AB06-84 operating instructions

Super Vac 724BCS user guide

Super Vac

Super Vac 724BCS user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.