Settings menu
The following table lists key parameter settings. Refer to the Clinical Guide for further information.
Therapy
Parameter Description S ST T PAC iVAPS CPAP Range
Mode Sets the therapy mode available on the
device.
IPAP Sets the pressure to be delivered to the
patient when the device is triggered into
inspiration.
4–25 cm H2O (4–25 hPa),
0.2 cm H2O (0.2 hPa) increments.
EPAP Sets the pressure to be delivered to the
patient when the device is cycled into
expiration.
2–[IPAP] cm H2O (2–[IPAP] hPa),
0.2 cm H2O (0.2 hPa) increments.
Min PS Sets the minimum pressure support
delivered by the device.
0–20 cm H2O (0–20 hPa),
0.2 cm H2O (0.2 hPa) increments.
Max PS Sets the maximum pressure support in
iVAPS mode.
0–23 cm H2O (0–23 hPa), 0.2 cm H2O
(0.2 hPa) increments.
Max EPAP Sets the maximum EPAP delivery by the
device.
2–25 cm H2O (2–25 hPa),
0.2 cm H2O (0.2 hPa) increments.
Min EPAP Sets the minimum EPAP delivered by
the device.
2–25 cm H2O (2–25 hPa),
0.2 cm H2O (0.2 hPa) increments.
Ti Max Sets the maximum limit on the time the
device spends in IPAP.
0.3–4.0 sec, 0.1 sec increments.
Ti Min Sets the minimum limit on the time the
device spends in IPAP.
0.1–[Ti Max] sec, 0.1 sec increments.
Rise Time Sets the time taken for pressure to
increase from EPAP to IPAP.
Min / 150–900 ms, 50 ms increments.
Mask Select the type of mask used by the
patient.
Full Face / Nasal / Pillows / Pediatric
Comfort
Parameter Description S ST T PAC iVAPS CPAP Range
Ramp Down Enable / Disable the Ramp Down
feature.
On / Off
Climate Ctrl Available when HumidAir humidifier
is used and ClimateLineAir heated air
tubing is connected.
Manual / Auto
Tube Temp. Set the minimum temperature of air
delivered by heated air tubing such as
ClimateLineAir.
Off / 16–30°C,
1° increments
Humidity
Level
Set the humidity level. Off / 1–8
Read your entire Clinical Guide before use.