Richard Wolf 2303 User manual

GA-A 202 /en /2012-04 V6.0 / PDG 11-5360
POWER CONTROL
2303
POWER STICK M4 8564.121
POWER STICK M3 8563.111/.311
Instruction Manual
2303
POWER CONTROL

0
GA-A 202
Important general notes and instructions for use
Make sure that this product is used only as intended and described in this instruction manual, by adequately
trained and qualified medical personnel, and that maintenance and repair are only carried out by authorized
experts.
Use the product only in the combinations and with the accessories and spare parts specified in this instruction
manual. Use other combinations, accessories and replacement parts only if they are expressly intended for
the planned application and if the performance characteristics and safety requirements are not impaired. Do
not alter the product.
Reprocess the products before each use and before return shipment as required by the instruction manual
to protect the patient, user and others.
Immediately upon receipt, check the product and its accessories for completeness and possible damage.
Should the shipment give right to complaints, please inform the manufacturer or supplier immediately.
Subject to technical changes!
Due to the continuous development of our products, the illustrations and technical data may deviate slightly.
Safety instructions and levels of danger
Symbol Level of danger
WARNING!
Failure to observe can result in death or serious injury.
CAUTION!
Failure to observe can result in slight injury or damage to the product.
.
IMPORTANT!
Failure to observe can result in damage to the product or surroundings.
.
NOTE!
Tips for optimum use and other useful information.
GERMANY
RICHARD WOLF GmbH
75438 Knittlingen
Pforzheimerstr. 32
Telephone: +49 70 43 35-0
Telefax: +49 70 43 35-300
MANUFACTURER
www.richard-wolf.com
USA
RICHARD WOLF
Medical Instruments Corporation
353 Corporate Woods Parkway
Vernon Hills, Illinois 60061
Toll Free: 001 (800) 323 - 9653
Phone: 001 (847) 913 - 1113
Fax: 001 (847) 913 - 1488
www.richardwolfusa.com
UK
RICHARD WOLF UK Ltd.
Waterside Way
Wimbledon
SW17 0HB
Telephone: + 44 20 89 44 74 47
Telefax: +442089441311
www.richardwolf.uk.com
BELGIUM / NETHERLANDS
N.V. Endoscopie
RICHARD WOLF Belgium S.A.
Industriezone Drongen
Landegemstraat 6
9031 Gent Drongen
Telephone: +32 92 80 81 00
Telefax: +3292829216
www.richard-wolf.be
FRANCE
RICHARD WOLF France S.A.R.L.
Rue Daniel Berger
Z.A.C. La Neuvillette
F-51100 Reims
Telephone: +33 3 26 87 02 89
Telefax: +33 3 26 87 60 33
AUSTRIA
RICHARD WOLF Austria
Ges.m.b.H.
Wilhelminenstraße 93 a
A-1160 Vienna
Telephone: +43 14 05 51 51
Telefax: +431405515145
www.richard-wolf.com
Marketing Office
U.A.E
RICHARD WOLF Middle East
P.O. Box 500283
AL Thuraya Tower 1
9th Floor,
Room 904, Dubai
Telephone: + 9 71 43 68 19 20
Telefax: + 9 71 43 68 61 12
www.richard-wolf.com
INDIA
RICHARD WOLF India Private Ltd.
JMD Pacific Square
No. 211 A, Second Floor
Behind 32nd Milestone
Gurgaon - 122 001
National Capitol Region
Telephone: + 91 12 44 31 57 00
Telefax: +911244315705
www.richard-wolf.com

Contents
I
GA--A 202
1 General information 1.......................................................
1.1 Symbols 1..................................................................
1.2 Intended use 3..............................................................
1.3 Indications and field of use 3..................................................
1.4 Contraindications 3...........................................................
1.5 General notes and instructions for use 3........................................
1.6 Combinations 4..............................................................
1.6.1 Potential equalization 4.......................................................
1.6.2 Requirements for the products / components of a combination 5....................
1.7 Electromagnetic compatibility (EMC) 6..........................................
1.7.1 Combination with POWER STICK M4 and, alternatively, POWER STICK M3 8.......
1.7.2 Legend 8...................................................................
2 Illustration 9................................................................
2.1 Figure: POWER CONTROL front panel 9.......................................
2.1.1 Legend 9...................................................................
2.2 Figure: LCD display 9.........................................................
2.2.1 Legend 9...................................................................
2.3 Figure: POWER CONTROL rear panel 10........................................
2.3.1 Legend 10...................................................................
2.4 Figure: Footswitch 10..........................................................
2.4.1 Legend 10...................................................................
2.5 Figure: POWER STICK M3 (8563.111 / .311) 11...................................
2.5.1 Legend 11...................................................................
2.6 Figure: POWER STICK M4 (8564.121) 11........................................
2.6.1 Legend 11...................................................................
3Setup 12....................................................................
3.1 Preparation of POWER CONTROL 12...........................................
3.2 Preparation of POWER STICK M4 13............................................
3.3 Preparation of POWER STICK M3 (8563.111 / .311) 13............................
3.3.1 Attaching the irrigation adaptor (only 8563.311) 14.................................
4 Checks 15...................................................................
4.1 Visual check 15...............................................................
4.2 Function check 15.............................................................
4.2.1 Function control of footswitch 16................................................
4.2.2 Function check -- POWER STICK M4 (8564.121) 16...............................
4.2.3 Function check -- POWER STICK M3 (8563.111 / .311) 17..........................
5Use 18......................................................................
5.1 Operating principle 18.........................................................
5.2 Control elements / modes 18...................................................
5.2.1 General notes and instructions for use 18........................................
5.2.2 Operation of POWER CONTROL via the sensor surface 19........................
5.2.3 Language selection 20.........................................................
5.2.4 Setting the parameters with the footswitch connected 20...........................
5.2.5 Setting the parameters at the POWER CONTROL, with no footswitch connected 22...
5.2.6 Changing the factory settings 23................................................

II GA--A 202
5.2.7 POWER STICK M3 24.........................................................
5.2.8 POWER STICK M4 24.........................................................
5.3 Operation of device 25.........................................................
5.3.1 Speed adjustment 26..........................................................
5.3.2 Instructions and notes on morcellation in urology 27...............................
5.4 Alarms 28....................................................................
5.5 Emergency shutdown of POWER STICK 28......................................
5.5.1 Taking out of service 28........................................................
6 Operation in the RIWO NET SYSTEM 29.......................................
6.1 Combination with RIWO NET SYSTEM 29.......................................
6.2 Operation 29.................................................................
6.3 Setting of “ID” rotary coding switch 29...........................................
6.4 Connection to the RIWO NET SYSTEM 30.......................................
6.5 Controlling the devices using the RIWO--NET menu 31.............................
6.5.1 Controlling the devices via the different input media 31.............................
6.5.2 Illustration of menu 31.........................................................
6.5.3 Main menu 32................................................................
6.5.4 Speed range function 32.......................................................
6.5.5 Speed function 32.............................................................
6.5.6 Oscillation function 32.........................................................
6.5.7 Positioning function 32.........................................................
6.6 System messages 33..........................................................
6.6.1 Status messages 33...........................................................
6.6.2 Warnings 33..................................................................
6.6.3 Errors 33.....................................................................
7 Reprocessing and maintenance 34............................................
7.1 Reprocessing of device 34.....................................................
7.2 Maintenance Device 34........................................................
7.2.1 Maintenance intervals 34.......................................................
7.3 Reprocessing of POWER STICK M4 (8564.121) 35................................
7.3.1 Manual reprocessing 35.......................................................
7.3.2 Machine reprocessing 36......................................................
7.3.3 Assembly 36.................................................................
7.3.4 Function check 36.............................................................
7.3.5 Sterilization 36................................................................
7.4 Reprocessing of POWER STICK M3 (8563.111 / .311) 37..........................
7.4.1 Manual reprocessing 37.......................................................
7.4.2 Machine reprocessing 38.......................................................
7.4.3 Assembly 39.................................................................
7.4.4 Function check 39.............................................................
7.4.5 Sterilization 40................................................................
7.5 Reprocessing of re--usable irrigation adapter 40...................................
7.5.1 Manual reprocessing 40.......................................................
7.5.2 Machine reprocessing 41.......................................................
7.5.3 Visual check 41...............................................................
7.5.4 Sterilization 41................................................................
7.6 Reprocessing of the re--usable rotary blades, abraders and morcellators 42...........
7.6.1 Manual reprocessing 42.......................................................

Contents
III
GA--A 202
7.6.2 Machine reprocessing 42......................................................
7.6.3 Checks 42...................................................................
7.6.4 Sterilization 42................................................................
8 Technical description 43.....................................................
8.1 Trouble shooting 43...........................................................
8.1.1 POWER CONTROL 43........................................................
8.1.2 POWER STICK 43............................................................
8.2 Technical data of POWER CONTROL 44.........................................
8.2.1 Interfaces 44.................................................................
8.3 Technical data of POWER STICK 45............................................
8.4 Technical data of footswitch 45..................................................
8.5 Operating, storage, transport and shipping conditions 45...........................
8.6 Spare parts and accessories 46.................................................
8.6.1 POWER CONTROL 46........................................................
8.6.2 POWER STICK M4 46.........................................................
8.6.3 POWER STICK M3 46.........................................................
8.6.4 Reusable rotary blades and abraders 47.........................................
8.6.5 Disposable rotary blades and abraders 47........................................
8.6.6 Reusable rotary blades and abraders for spinal surgery (POWER STICK M4) 48......
8.6.7 Reusable rotary morcellator for urology 48........................................
8.6.8 Disposable Rotary Morcellator for urology. 48.....................................
8.7 Replacing parts on POWER STICK M4 (8564.121) 49.............................
8.7.1 POWER STICK M4 connection cable 49.........................................
8.8 Replacing parts on POWER STICK M3 (8563.111 / .311) 49........................
8.8.1 O--ring for twist--lock mechanism 49.............................................
8.8.2 O--ring for suction/irrigation plunger 49...........................................
8.8.3 Sealing cap of suction/irrigation system (connecting hole to handle) 49...............
8.8.4 POWER STICK M3 connection cable 49.........................................
8.9 Device fuses 50...............................................................
8.9.1 Disposal of the product, packing material and accessories 50.......................

1GA--A 202
1 General information
1.1 Symbols
Symbols Meaning
Attention, Caution
Follow the instruction manual
Off (power: disconnection from power/mains)
On (power: connection to the power/mains)
Equipotentality
Fuse
μAlternating current (AC)
TYPE BF APPLIED PART
Socket for footswitch
Alarm
Ready (Triggering of function check)
Signal input (CAN-BUS)
Signal output (CAN-BUS)
Identification of sensor field
Display blinks: “No footswitch"
Display blinks: “No POWER STICK"
Indication “CCW rotation"
Indication “CW rotation"
Indication “Oscillation"
Selected speed range display
Counter-clockwise rotation
Clockwise rotation
Oscillation
Back to previous menu
Increase value by one
Decrease value by one

2
GA--A 202
Meaning
Symbols
Selection
Confirm value
Single step
“Adjust speed" selector
“Select speed range" selector
“Device settings“ selector
Oscillation rate adjustment
max
min
12
34 Speed range adjustment
Speed adjustment
Adjust stop position (works only without footswitch)
Reset to factory setting
Language selection
Control level “B" activated
Footswitch
Change between control levels “A" and “B"
A: Speed range 1 through 4 B: Single step
A: “CCW“ rotation B: Reduce speed
A: “CW“ rotation B: Increase speed
A: “Oscillating“ rotation B: Speed control
REF Order number
Lot identification number
Manufacturing date
To be used before (date):
Do not reuse!
Sterilized with ethylene oxide
Recycle the product separately. Do not discard together with other waste.
A Registered Trademark of ETL, a Recognized Testing Laboratory, listing compliance as
Medical Electrical Equipment to standard CAN/CSA C 22.2 No. 601.1 (c) and UL 60601--1 (us)

3GA--A 202
Meaning
Symbols
A Registered Trademark of a Recognized Testing Laboratory, confirm the compliance to the standard
of Medical Electrical Equipment CAN/CSA C22.2 No.60601-1 (c) and ANSI/AAMI ES60601-1 (us)
Identification in conformity with Medical Devices Directive 93/42/EEC only valid if the product and/
or packaging is marked with this symbol. Products of category IIa and above, as well as
sterile products or products with measuring function of category I, are additionally marked with the
code number of the notified body (0124)
1.2 Intended use
The POWER CONTROL 2303 in conjunction with POWER STICK M4
(8564.121) or, alternatively, POWER STICK M3 (8563.111/.311) serves
to drive WOLF rotary blades/abraders and morcellators (tissue punches)
for the removal of tissue in endoscopic operations.
At the same time aspiration allows continuous removal of ablated tissue.
This product is designed exclusively for use by specialized medical per
sonnel and may only be applied by medically qualified and adequately
trained persons.
1.3 Indications and field of use
For use with endoscopic accessories:
- In arthroscopy e.g. for meniscal resection, for the removal of soft tissue
as well as for intra-articular transsection or abrasion of osseous tissue
(e.g. in ACL or shoulder applications).
- In thorax surgery e.g. to remove hematomas.
- In sinus surgery e.g. to remove polyps.
- In spinal surgery (arthroscopic microdiscectomy (AMD)) e.g. for
removing degenerated tissue.
- In urology for fragmentation and removing of detached prostatic
adenomas after Laser-TURP.
1.4 Contraindications
Relative or absolute contraindications may result from the general condi-
tion of the patient, i.e. the general findings, or in special cases, where the
risk for the patient is significantly higher if motorized tools are used. Make
sure you observe the cases described in the relevant medical literature.
1.5 General notes and instructions for use
CAUTION!
All R.Wolf disposable rotary blades and abraders are sterily packed
in double wrapping and are designed for single use only.
Any products which are not used should be kept in the outer pack-
aging (cardboard box). Do not use products with damaged outer
sterile wrapping/packaging (plastic bag).
CAUTION!
The rotary blades and abraders are packed individually in pairs
(inner and outer blade / abrader together).
These components must not be combined with inner or outer
blades/abraders from other packages.

4
GA--A 202
CAUTION!
Reduced cutting performance and wear on the rotary blade/abrader !
If the rotary blade/abrader is pushed against the tissue at high
pressure this will not improve the cutting performance but may ins-
tead cause damage, increased wear and tear on the inner blade/ab-
rader.
.IMPORTANT!
Quick cooling of a hot motorized handle can cause stresses in the mater-
ial and may lead to failure of the device or a reduced service life.
1.6 Combinations
.IMPORTANT!
In addition to this instruction manual be sure to follow the instruction manuals of
the products used in combination with this product.
If other products are used in combination with this product, make sure that the
intended uses and relevant technical data (working length, diameter etc.) are the
same.
When using rotary blades/abraders for spinal surgery, ensure that the working
channel of discoscopes and the internal diameter of working sleeves is sufficiently
large. Use instruments with 2.5 mm, 3.5 mm and 4.0 mm diameter through disco-
scopes and instruments with 4.5 mm diameter through suitable working sleeves.
Only suction/aspiration devices and irrigation fluid supply systems that do not
reduce the protection rating ”BF” of the motorized handle in accordance with IEC /
EN 60601 may be used. The user is responsible for selecting a suitable irrigation
fluid.
.IMPORTANT!
As a result of differing anatomy, it is not possible to quantify irrigation and suction
rates. It is the user’s responsibility to adapt the settings to the requirements.
1.6.1 Potential equalization
The potential equalization cable establishes a direct connection between
a medical electrical device and an equipotential bonding rail.
It serves to equalize differences in potential between enclosures of elec
trical equipment and firmly installed conductive parts in the patient envi
ronment.

5GA--A 202
1.6.2 Requirements for the products / components of a combination
1.5m
IEC 60601-1, 3. Ed.
1.5m
1.5m
The general requirements depend on whether the products /
components are inside or outside the patient environment.
Medically used room Non-medically
used room
Requirements / measures
inside the
patient environment
outside the
patient environment
Leakage currents to section 16.6
IEC 60601-1:2005 / EN 60601-1:2006 *
MP MP
~~- - Verification of the total patient leakage current
MP MP
~**
-
-
Verification of leakage currents
a) additional protective earth connection
(consult the corresponding manufacturer),
or
b) additional isolating transformer for medical applications **
MP
~
NMP
~**
MP NMP
~**
NMP
~
MP
MP NMP
~
MP MP / NMP
~**
MP
~
NMP
~
-
Verification of leakage currents
a) no plugs with metal housing, or
b) additional isolation device (to avoid voltage differentials)
MP
~
MP / NMP
~
Verification of leakage currents
a) common protective earth connection, or
b) additional protective earth connection at MP (clarify with the corre
ponding manufacturer), or
c) additional isolation device (to avoid voltage differentials), or
d) no plugs with metal housing in the patient environment
additional "isolating transformer" to IEC/ EN60601-1
**
additional isolating device to IEC/
EN 60601-1 Multiple socket strip
Functional connection ~Power supply grid
MP =medical electrical device to IEC/ EN 60601-1, ANSI/AAMI es60601-1, CSA C22.2 No. 60601-1 08
NMP =non-medical electrical device in accordance with product-specific IEC/EN/UL standards
*When connected via a power bar under standard conditions the earth leakage current of the power bar must not exceed 5 mA.
** e.g. Richard Wolf video cart with "isolating transformer"

6
GA--A 202
.
IMPORTANT!
The persons combining products to form a system are responsible for not impairing the system's compliance with
performance and safety requirements, and that the technical data and the intended use are adequately fulfilled.
Possible electromagnetic or other interference that may occur between the product and other products can cause
faults or malfunctions.
When selecting the system components, make sure that they meet the necessary requirements of the medical envi
ronment they are used in, in particular IEC/ EN 60601-1 (3. Edition IEC/EN 60601-1, section 16). In case of doubt
contact the manufacturer(s) of the system components.
Do not touch connectors for electrical connections between various components (such as signal input connectors
and signal output connectors for video signals, data exchange, controls etc.) and the patient at the same time.
1.7 Electromagnetic compatibility (EMC)
NOTE: The device or system in the following called product always relates to the POWER CONTROL 2303.
The product does not have any performance features classified as essential performance features in accordance with IEC/EN 60601-1.
Guidelines and manufacturer's declaration - Electromagnetic emissions
The product is intended for use in the environment specified below. The user should assure that the product is used in such an environment.
Emissions measurement / test Compliance Electromagnetic environment - Guidelines
HF emissions to CISPR 11 Group 1
The product uses HF energy for its internal function.
The HF emission level is extremely low and it is not likely to cause any
interference in nearby electronic equipment.
HF emissions to CISPR 11 Class B
The product is suitable for use in all establishments, including domestic
establishments. This also includes establishments directly connected to
the public low voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
to IEC 61000-3-2 Class A
In conformity with IEC 61000-3-3 “Emission of voltage fluctuations
/flicker"
Guidelines and manufacturer's declaration - Electromagnetic immunity
The product is intended for use in the environment specified below. The user should assure that the product is used in such an environment.
Immunity tests IEC 60601 test level Compliance Electromagnetic environment - Guidelines
Electrostatic discharge (ESD)
to IEC 61000-4-2
6 KV contact discharge
8 KV air discharge Yes
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the re
lative humidity should be at least 30%.
Electrical fast transience, bursts
to IEC 61000-4-4
2 KV for power supply lines
1 KV for input and output
lines
Yes Mains/line power quality should be that of a typical
commercial or hospital environment.
Surge voltage (surges)
to IEC 61000-4-5
1KVlinetoline
voltage
2 KV line to ground
voltage
Yes Mains/line power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions and sup
ply voltage variations
to IEC 61000-4-11
Voltage dip for 0.5 cycle
> 95% UT*
Voltage dip for 5 cycles
60% UT*
Voltage dip for 25 cycles
30% UT*
Voltage dip for 5 sec
> 95% UT*
Yes
Mains/line power quality should be that of a typical
commercial or hospital environment. If the user of
the product requires continued operation during
power mains/line interruptions it is recommended
that the product be powered from an uninterrupti
ble power supply or battery.
Power frequency (50/60 Hz) magnetic
field,
to IEC 61000-4-8
3A/m Yes
Power frequency magnetic fields should be at le
vels characteristic of a typical location in a com
mercial or hospital environment.
*NOTE: UTis the line / mains voltage prior to application of the test level.

7GA--A 202
Guidelines and manufacturer's declaration - Electromagnetic immunity for products that are not life-supporting
The product is intended for use in the environment specified below. The user should assure that the product is used in such an environment.
Immunity tests IEC 60601 test level Compliance Electromagnetic environment - Guidelines
Conducted HF interference
to IEC 61000-4-6
Radiated HF interference
to IEC 61000-4-3
3V
rms
150kHzto80MHz
3V/m
80MHzto2.5GHz
Yes
Portable and mobile RF communications equipment should
be used no closer to any part of the product, including ca
bles, than the recommended separation distance calculated
from equation applicable to the frequency of the transmitter.
Recommended separation distance:
d=1.2pP
d=1.2pP for 80 MHz to 800 MHz
d=2.3pP for 800 MHz to 2.5 GHz
P = Nominal power output rating of the transmitter in watts
(W)
(according to the transmitter manufacturer)
d = recommended separation distance in meters (m)
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey1, should be less than the
compliance level in each frequency range2.
Interference may occur in the vicinity of devices with the
following symbol:
REMARKS:
At 80 MHz and 800 MHz the higher frequency range applies. These guidelines may not apply in all situations. The propagation of elec
tromagnetic waves is affected by absorption and reflexion from buildings, objects and people.
1= The field strength of stationary transmitters (e.g. base station for radio phone, earth to earth radio stations,, amateur radio stations, radio
and television transmitters) cannot be exactly predicted in theory. To assess the EMC environment due to fixed transmitters an electromagnetic
site survey should be conducted. If the measured field strength in the location in which the product is used exceeds the applicable compliance
level above, the product should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be required, such as reorienting or relocating the product.
2= Over the frequency range between 150 kHz and 80 MHz the field strength should be below 3 V/m.
Recommended separation distances between portable and mobile HF telecommunication devices and devices which
are not life-supporting
The product is intended for use in an electromagnetic environment where HF disturbances are controlled.
The user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile HF telecommunica
tions equipment and the product.
Rated nominal output power of the
transmitter (Watts)
Separation distance as a function of transmitter frequency (m)
150kHzto80MHz
d=1.2pP
80MHzto800MHz
d=1.2pP
800 MHz to 2.5 GHz
d=2.3pP
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a nominal output power not listed in the table above, the recommended separation distance (d) in meters (m) can be
determined using the applicable equation (observe frequency). P = nominal power of the transmitter in Watts (W).
REMARKS:
At 80 MHz and 800 MHz the higher frequency range applies. These guidelines may not apply in all situations. The propagation of elec
tromagnetic waves is affected by absorption and reflexion from buildings, objects and people.

8
GA--A 202
1.7.1 Combination with POWER STICK M4 and, alternatively, POWER STICK M3
1.0
3.0
2.0
4.0
4.0
4.1
4.1
5.0
6.0
5.1
6.1
1.7.2 Legend
1.0 POWER CONTROL 2303 5.0 POWER STICK M3 with irrigation connector
(8563.311)
2.0 POWER STICK M4 5.1 Adapter for M3
3.0 Footswitch 6.0 Irrigation fluid container (only for 8563.311)
4.0 Suction pump 6.1 Irrigation tube kit (only for 8563.311)
4.1 Suction tube
.NOTE!
Instead of POWER STICK M4, POWER STICK M3 with irrigation connector (8563.311) can be
connected to POWER CONTROL 2303 using adapter 2303.911.

9GA--A 202
2 Illustration
2.1 Figure: POWER CONTROL front panel
2303
POWER CONTROL
2
4
6
7
1
3
5
2.1.1 Legend
1Power switch 5Sensor field (sensor surface)
2”Alarm” indicator 6Socket for motorized handle (Type BF applied part)
3“Ready” button 7Socket for footswitch
4LCD
2.2 Figure: LCD display
5.5
4.3
5.1
4.44.1
5.2
5.3
4.24.5
4.6
4.8
4.7
5.4
2.2.1 Legend
4.1 “Minimum speed” 4.8 “No footswitch” indicator
4.2 “Preselected speed” 5.1 “Rotate CCW” sensor
4.3 “Maximum speed” 5.2 “Rotate CW” sensor
4.4 “Actual direction of rotation” indication 5.3 “Oscillation” sensor
4.5 “Actual speed” 5.4 “Device settings” sensor
4.6 “Preselected speed range” 5.5 “Speed / stop position adjustment” sensor field
4.7 Speed indication

10
GA--A 202
2.3 Figure: POWER CONTROL rear panel
10
8
9
11.2 11.3
11.1
FEDERAL LAW RESTRICTS THIS UNIT TO BE
USED OR SOLD, EXCEPT UNDER THE SUPER--
VISION OF A MEDICAL DOCTOR.
2.3.1 Legend
8Power connector with fuse holder 11.1 CAN--BUS address selector
9Equipotential connector 11.2 CAN--BUS input
10 Indentification plate 11.3 CAN--BUS output
2.4 Figure: Footswitch
1234
12.1
12
12.2
12.3 12.4
12.5
2.4.1 Legend
12 Footswitch
12.1 Switchover between functions “A” and “B”
12.2 A: Speed range 1 through 4 B: Single step (Blade position)
12.3 A: “CCW“ rotation B: Reduce speed
12.4 A:“CW“rotation B: Increase speed
12.5 A: “Oscillation“ B: Speed control

11 GA--A 202
2.5 Figure: POWER STICK M3 (8563.111 / .311)
13.3
13.1
13.2
13.4
13.5
13.6
13.7
13.8
13.9
13.10
2.5.1 Legend
13.1 Clamping chuck 13.6 Irrigation control valve (only 8563.311)
13.2 Suction connector 13.7 Rubber sleeve
13.3 Connection cable with connectors 13.8 Suction/suction and irrigation system
13.4 Irrigation connector (only 8563.311) 13.9 Drive unit
13.5 Suction control valve 13.10 Cover
2.6 Figure: POWER STICK M4 (8564.121)
14.1
14.2
14.3
14.4 14.5
2.6.1 Legend
14.1 Clamping chuck 14.4 Suction connector
14.2 Suction control valve 14.5 Connection cable with connectors
14.3 Grip sleeve

12
GA--A 202
3Setup
WARNING!
The device is not protected against explosions.
Explosion hazard.
Do not operate this device in areas where there is the danger of
explosion.
WARNING!
Danger if a power supply without protective earth is used.
Danger of electric shock!
Connect the device only to a power supply with protective earth.
.NOTE!
Check that the mains/line voltage is the same as the voltage specified on
the identification plate. Connect the device only with the supplied power
cable or a power cable meeting the same specifications.
Make sure that the vent slots are not blocked.
CAUTION!
Danger of faults and malfunctions.
To guarantee the safety of the user, the patient and others use only
accessories and spare parts specified by the manufacturer of this
product.
Other accessories or spare parts can cause the emission of
increased electromagnetic radiation or reduced immunity against
interference.
.IMPORTANT!
Medical devices are subject to special precautions with regard to electro-
magnetic compatibility (EMC).
Make sure you observe the notes on EMC for installation and operation.
Medical electrical devices can be influenced by mobile HF communication
devices.
If it is necessary to stack the devices or place them next to each other
and HF interference is observed, make sure you observe the intended
use of the devices.
3.1 Preparation of POWER CONTROL
ZConnect device to mains/power.
ZConnect footswitch cable to device.
ZSwitch on the power switch.
.NOTE:
To disconnect the cable from the device, pull back locking sleeve (B) on
connector (front portion with red marking).
B

13 GA--A 202
3.2 Preparation of POWER STICK M4
ZConnecting the cable to POWER STICK M4.
'Align connector (the red marking (C) must be positioned exactly be-
low the suction connector) and push home.
ZConnect suction tube between suction pump and suction connector on
POWER STICK M4. Set suction valve to “OFF”.
ZInserting the rotary blades.
'Retract ring (14.8) on clamping chuck.
'Insertrotarybladeinsuchawaythatthecamisalignedwithgroove
(14.10) on the clamping chuck.
'Let go of ring, the rotary blade is locked in place.
ZConnectthecableofPOWERSTICKM4tothePOWERCONTROL
unit.
14.10
14.8
.IMPORTANT!
The inner and outer blades are supplied in matched pairs. Do not mix
with other blades.
3.3 Preparation of POWER STICK M3 (8563.111 / .311)
ZConnecting the cable to POWER STICK M3:
'Turn the plug with the rubber sleeve under slight pressure until it en-
gages in the socket connector. Then push both the plug and the pro-
tection sleeve home (for this the plug must be in home position).
ZConnecting the suction tube between the suction pump and the suction
connector of the motorized handle.
'Set the suction control valve (13.5) to ”closed”.
ZIn the case of motorized handle 8563.311 also connect the irrigation
connector.
'Set irrigation control valve (13.6) to ”closed”.
.NOTE!
Theirrigationpressurecanbeadjustedbyvaryingtheheightdifference
(hydrostatic pressure) between the working height and the height of the
irrigation fluid container or suction pump.
ZInserting the rotary blades.
'Retract ring (13.1.1) of clamping chuck.
'Insert rotary blades in such a way that cam (13.12) is aligned with
groove (13.11) on clamping chuck (13.1).
'Let go of ring, the rotary blade is locked in place.
C
13.5
13.6

14
GA--A 202
.IMPORTANT!
The inner and outer blades are supplied in matched pairs. Do not mix
with other blades.
ZUse the adapter to connect the POWER STICK M3 connection cable
to the POWER CONTROL unit.
3.3.1 Attaching the irrigation adaptor (only 8563.311)
ZCheck that the sealing element (13.14.2) is installed on irrigation tube
(13.14.1).
ZPush irrigation adapter (13.14) onto the rotary blade.
ZInsert irrigation tube (13.14.1) in opening (13.13) on the clamping
chuck, then push the irrigation adapter onto the rotary blade as far as it
will go.
13.1.1
13.12
13.11
13.14 13.14.1
13.13
13.14.2
13.1

15 GA--A 202
4Checks
.
IMPORTANT!
Run through the checks before and after each use.
Do not use the products if they are damaged or incomplete or have loose
parts.
Return damaged products together with any loose parts for repair.
Do not attempt to do any repairs yourself.
4.1 Visual check
ZCarry out the following checks:
'Check devices, instruments and accessories for damage, hygienic
condition, completeness and corrosion.
'Check all connection cables and tubes for damage.
'Check lettering/labeling for completeness and good legibility.
'Check that the cable connector on the motorized handle is in home
position, i.e. fully inserted.
'Check that the valve is plugged in.
'Check that the cutting edges and windows of the blades/abraders
areinperfectworkingorder.
'Check footswitch and connection cables for damage and cracks in
the insulation of the connection cable.
'Check the sensor surfaces on the POWER CONTROL unit.
4.2 Function check
.NOTE!
Make sure that the tools are immersed in liquid (cooling of the heat gen-
erated by friction). Do not operate in dry condition!
The POWER CONTROL M4 self--test can be activated at any time.
ZSwitch on the power switch.
ZPress the blinking “Ready” button.
'The device carries out a self--test. In the case of an error the result is
displayedontheLCD.
'If no footswitch is connected, the “No footswitch” indicator blinks.
'If no POWER STICK is connected, the “No POWER STICK” indicator
blinks.
This manual suits for next models
2
Table of contents
Other Richard Wolf Medical Equipment manuals
Popular Medical Equipment manuals by other brands

ADC
ADC Adscope Sprague Use, care & maintenance

DJO
DJO Donjoy Drytex Hinged Air quick start guide

Azoi
Azoi Kito user manual

Boston Scientific
Boston Scientific VERCISE DBS Handbook

MacroMedics
MacroMedics CouchStrip CS53FCF-IB manual

TASKA Prosthetics
TASKA Prosthetics Low Profile Wrist Maintenance and Care Guide

BioCare Medical
BioCare Medical VALENT manual

Shukla Medical
Shukla Medical Xtract-All Nail Winquist III MSK1017 Surgical Technique Guide

Thomashilfen
Thomashilfen Plate Rim Raiser Instructions for use

Nellcor
Nellcor N-180 Service manual

EKOM
EKOM DK50 2V/M MOBILE user manual

CognitiveTPG
CognitiveTPG Thrive installation guide