10
You will also notice that the green Selected LED light on Ch.
1 went out but its green Active LED light is still illuminated.
This means that there is a treatment currently on Ch. 1 but its
parameters are not selected for display. If you pressed the
waveform or Profile button the treatment parameters on Ch.2
are now displayed. If you pressed a channel button you now
need to select the waveform or Profile you want to use and its
parameters will be displayed. Any adjustments can now be
made to the parameters for this second treatment. Then, set
Intensity and press the Start/Accept button to begin the
treatment time countdown. NOTE: If the Quadpolar IFC
therapy waveform is selected, it requires both channels one
and two or channels three and four and will subsequently set
the same intensity level on both channels. Once the intensity
levels are set in Quadpolar IFC and the Start/Accept button
has been pressed, you have the option of Balancing the output
by pressing the Balance button. When pressed, the green LED
light on the Balance button will illuminate and the green
Adjust LED light will illuminate under the intensity
dial. Use the dial to balance the patient’s sensation. You will
see only the selected channel’s intensity level increase or
decrease, but understand that as it increases, the other Quad
IFC channel’s intensity decreases proportionately. Once the
balance has been completed, press the Start/Accept button.
To View Stimulation Treatment
If you are running two or more concurrent stim treatments,
you will only see one treatment’s parameters at a time. Press
the desired channel button to illuminate its Selected LED
light. You will then see all the parameters for that particular
channel’s treatment. Whichever channel button you press will
activate the Selected LED and display that treatment’s
parameters.
To Stop a Stimulation Treatment
If you only want to stop a single treatment without affecting
the other treatments, press and hold the desired channel button
and the stimulation Stop button at the same time.
NOTE: If you just press the stimulation’s Stop button it WILL
STOP ALL STIM TREATMENTS.
Waveform Parameter Options
Waveform choices include Quadpolar IFC, Pre-Mod IFC,
Russian, Hi-Volt (Monophasic), and Microcurrent.
Quadpolar IFC will only output classic Interferential current
in a Normal mode and it requires both Ch.1 and Ch. 2 or Ch. 3
and Ch. 4 (if applicable). The Balance button allows for better
patient sensation balance if that button is selected - as you
increase or decrease intensity on the selected channel it will do
the opposite on the other channel in the pair. Quadpolar IFC
also allows you to turn an amplitude modulating Vector on or
off. The Vector provides a soothing massage effect and also
helps to lessen patient accommodation. The Pulse Rate
available is either the High, Low or a Hi-Lo combination.
Pre-Mod IFC will output pre-modulated, two-pad
Interferential current in any of the desired Modes (Normal,
Surge, Co-Cont and Alternating). The Vector can be turned
on or off in the Normal mode. The Pulse Rates available in
the Normal mode are High, Low, and Hi-Lo scan. If the
Surge, Co-Contraction, or Alternating mode is selected, only
the Fixed Pulse Rate is available. Any of the six On/Off
timing options are available in the Surge or CoCont modes. In
the Alternating mode the 10/10 or the 5/5 timing cycles are
available.
Russian and Hi-Volt- Both waveforms are available in any of
the Modes (Normal, Surge, Co-Cont, and Alternating). The
Fixed Pulse Rate is only available for Russian.
* The Hi-Volt Pulse Rates available in the Normal mode are
High, Low, and Hi-Lo scan. If the Surge, Co-Contraction, or
Alternating mode is selected, only the Fixed Pulse Rate is
available. Any of the six On/Off timing options are available
in the Surge or Co-Cont modes. In the Alternating mode, the
10/10 or the 5/5 timing cycles are available.
NOTE: The Hi-Volt in the Winner is a non-dispersive style.
This means that for each channel, one of the electrodes is
positive and the other is negative. When using the Hi-Volt
waveform, the red pin is positive and the white pin is negative.
Microcurrent will only output in a Normal mode and its Pulse
Rate is only available as a Fixed output.
Stimulation Modes
Depending upon the waveform selected, the user can choose
from Normal, Surge, Co-Contract and Alternating modes.
Normal Mode is basic continuous, un-interrupted output of the
selected waveform.
Surge Mode is a single channel of output that is interrupted in
an “on/off” method of a certain time, in seconds, of output
followed by a certain time, in seconds, of no output. For
example, if Surge mode is selected and the “10/20” option is
selected, this means that for 10 seconds there will be output to
the set intensity level followed by 20 seconds of zero output.
The cycle continues until treatment time expires.
* NOTE: There is a built-in ramp up time of three seconds in
the “on” time and a ramp down of one second in the “off”
time.
Co-Contraction Mode is the same as the Surge mode except
that it will concurrently run two channels, either 1 and 2 or 3
and 4, to the same “on” and “off” time. Independent intensity
levels can be set.
Alternating Channel Mode will alternately turn one channel
“on” and then a second channel “on” as soon as the first
channel goes into the “off” time. For example, in an Alt mode
both Ch. 1 and Ch. 2 will be used. If the “10/10” option is
selected in the Time On/Off then, after intensity levels have
been set for each channel and treatment started, Ch. 1 will be
“on” for 10 seconds. Once Ch.1 goes “off”, Ch. 2 will be “on”
for 10 seconds. This alternating will continue until treatment
time expires.