Declassification Guide
Digital Oscilloscope
May 2012
RIGOL Technologies, Inc.
RIGOL
Declassification Guide for Digital Oscilloscope
1
DS2000 Series
DS2000 series digital oscilloscope consists of DS2202, DS2102 and DS2072.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents Data
input
method
Location in
instrument
and
remarks
Sanitization
procedure
SDRAM
32MB Yes No Program Boot Main board
in CPU area Cycle power
NOR
FLASH
16MB
No Yes Firmware,
Bin,
Internal
saving
Firmware
update,
Internal
saving,
ref/setup
/pf
Main board
in CPU area Security clear or
save files to C:
FRAM
2kB Yes Yes Settings Program Main board
in CPU area Default
DDR2
32MB*16B
*3
Yes No Acquire
memory Acquire
system Main board
in acquire
area
Cycle power
SRAM
512kB*36B
*2
Yes No Display
memory Display
system Main board
in display
area
Cycle power
RIGOL
Declassification Guide for Digital Oscilloscope
2
Position of Instrument Memory on Main Board
Security Clear Procedure
To clear the memories, press Storage Disk.Manage Security Clear OK and the
system will be restored to the default settings.
RIGOL
Declassification Guide for Digital Oscilloscope
3
DS4000 Series
DS4000 series digital oscilloscope consists of DS4054, DS4052, DS4034, DS4032, DS4024, DS4022,
DS4014 and DS4012.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents Data
input
method
Location in
instrument
and
remarks
Sanitization
procedure
SDRAM
32MB Yes No Program Boot Main board
in CPU area Cycle power
NOR
FLASH
16MB
Yes Yes Firmware,
Bin,
Internal
saving
Firmware
update,
Internal
saving
ref/setup/
pf
Main board
in CPU area Security clear
or Save files to
C:
FRAM
2kB Yes Yes Settings Program Main board
in CPU area Default
DDR2
32MB*16B
*17
Yes No Acquire
memory Acquire
system Main board
in acquire
area
Cycle power
SRAM
512kB*36
B*8
Yes No Display
memory Display
system Main board
in display
area
Cycle power
RIGOL
Declassification Guide for Digital Oscilloscope
4
Position of Instrument Memory on Main Board
Security Clear Procedure
To clear the memories, press Storage Disk.Manage Security Clear OK and the
system will be restored to the default settings.
RIGOL
Declassification Guide for Digital Oscilloscope
5
DS6000 Series
DS6000 series digital oscilloscope consists of DS6104, DS6102, DS6064 and DS6062.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents Data
input
method
Location in
instrument
and
remarks
Sanitization
procedure
SDRAM
32MB Yes No Program Boot Main board
in CPU area Cycle power
NAND
FLASH
1GB
Yes Yes Internal
savings,
setup/wfm/ref
/pf/trc
Internal
saving Main board
in CPU area Save files to
internal C:
NOR
FLASH
16MB
No Yes Firmware, Bin Firmware
update Main board
in CPU area No user data
FRAM
2kB Yes Yes Setting Program Main board
in CPU area Default
DDR2
32MB*16
B*17
Yes No Acquire
memory Acquire
system Main board
in acquire
area
Cycle power
QDR2
22MB*18
B*8
Yes No Display
memory Display
System Main board
in draw area Cycle power
RIGOL
Declassification Guide for Digital Oscilloscope
6
Position of Instrument Memory on Main Board
Security Clear Procedure
To clear the memories, press SAVE Disk.Manage Security Clear OK and the
system will be restored to the default settings.
RIGOL
Declassification Guide for Digital Oscilloscope
7
DS1000B Series
DS1000B series digital oscilloscope consists of DS1204B, DS1104B and DS1074B.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents Data
input
method
Location in
instrument
and
remarks
Sanitization
procedure
SDRAM
16MB Yes No Program Boot Main board
in CPU area Cycle power
NOR
FLASH
8MB
No Yes Firmware,
Bin,
Internal
saving
Firmware
update,
Internal
saving,
ref/setup
/pf
/record
/wfm
Main board
in CPU area Save files to
inter
FRAM
1kB Yes Yes Settings Program Main board
in CPU area Default
SDRAM
16MB Yes No Drawing Draw Main board
in display
area
Cycle power
RIGOL
Declassification Guide for Digital Oscilloscope
8
Position of Instrument Memory on Main Board
RIGOL
Declassification Guide for Digital Oscilloscope
9
DS1000CA Series
DS1000CA digital oscilloscope consists of DS1302CA, DS1202CA, DS1102CA and DS 1072CA.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents Data
input
method
Location in
instrument
and
remarks
Sanitization
procedure
SDRAM
16MB Yes No Program Boot Main board
in CPU area Cycle power
NOR
FLASH
8MB
No Yes Firmware,
Bin,
Internal
saving
Firmware
update,
Internal
saving,
ref/setup/
pf/record
/wfm
Main board
in CPU area Save files to
inter
FRAM
1kB Yes Yes Settings Program Main board
in CPU area Default
SDRAM
16MB Yes No Drawing Draw Main board
in display
area
Cycle power
RIGOL
Declassification Guide for Digital Oscilloscope
10
Position of Instrument Memory on Main Board
RIGOL
Declassification Guide for Digital Oscilloscope
11
DS1000E Series
DS1000E series digital oscilloscope consists of DS1102E and DS1052E.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents Data
input
method
Location in
instrument
and
remarks
Sanitization
procedure
SDRAM
16MB Yes No Program Boot Main board
in CPU area Cycle power
NOR
FLASH
8MB
No Yes Firmware,
Bin,
Internal
saving
Firmware
update,
Internal
saving,
ref/setup/
pf/record
/wfm
Main board
in CPU area Save files to
inter
FRAM
1kB Yes Yes Settings Program Main board
in CPU area Default
SRAM
512kB Yes No wfm points Acquire Main board
in acquire
area
Cycle power
RIGOL
Declassification Guide for Digital Oscilloscope
12
Position of Instrument Memory on Main Board
RIGOL
Declassification Guide for Digital Oscilloscope
13
DS1000D Series
DS1000D series digital oscilloscope consists of DS1102D and DS1052D.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents Data
input
method
Location in
instrument
and remarks
Sanitization
procedure
SDRAM
16MB Yes No Program Boot Main board in
CPU area Cycle power
NOR
FLASH
8MB
No Yes Firmware,
Bin,
Internal
saving
Firmware
update,
Internal
saving,
ref/setup
/pf/
record
/wfm
Main board in
CPU area Save files to
inter
FRAM
1kB Yes Yes Settings Program Main board in
CPU area Default
SRAM
512kB Yes No wfm points Acquire Main board in
acquire area Cycle power
SRAM
512kB Yes No LA points Acquire Main board in
acquire area Cycle power
RIGOL
Declassification Guide for Digital Oscilloscope
14
Position of Instrument Memory on Main Board

Other manuals for DS4000 Series

2

This manual suits for next models

32

Other Rigol Test Equipment manuals

Rigol DS1000Z-E Series User manual

Rigol

Rigol DS1000Z-E Series User manual

Rigol DS70000 Series User manual

Rigol

Rigol DS70000 Series User manual

Rigol MSO8000A Series User manual

Rigol

Rigol MSO8000A Series User manual

Rigol MSO5000-E Series User manual

Rigol

Rigol MSO5000-E Series User manual

Rigol DS4024E User manual

Rigol

Rigol DS4024E User manual

Rigol MSO4000 Series User manual

Rigol

Rigol MSO4000 Series User manual

Rigol DHO800 Series User manual

Rigol

Rigol DHO800 Series User manual

Rigol DG2000 Series Installation and operation manual

Rigol

Rigol DG2000 Series Installation and operation manual

Rigol DS1000B Series User manual

Rigol

Rigol DS1000B Series User manual

Rigol DS1102E User manual

Rigol

Rigol DS1102E User manual

Rigol DS70000 Series User manual

Rigol

Rigol DS70000 Series User manual

Rigol DS1000 Series User manual

Rigol

Rigol DS1000 Series User manual

Rigol MSO2000A Series User manual

Rigol

Rigol MSO2000A Series User manual

Rigol RP7080S User manual

Rigol

Rigol RP7080S User manual

Rigol MSO4000 Series User manual

Rigol

Rigol MSO4000 Series User manual

Rigol MS05000-4CH User manual

Rigol

Rigol MS05000-4CH User manual

Rigol DS1302CA User manual

Rigol

Rigol DS1302CA User manual

Rigol MSO5000-E Series User manual

Rigol

Rigol MSO5000-E Series User manual

Rigol DS2000 Series User manual

Rigol

Rigol DS2000 Series User manual

Rigol DS1000Z Series Operating instructions

Rigol

Rigol DS1000Z Series Operating instructions

Rigol DS1000Z Series User manual

Rigol

Rigol DS1000Z Series User manual

Rigol DS70000 Series User manual

Rigol

Rigol DS70000 Series User manual

Rigol DS1000Z Series User manual

Rigol

Rigol DS1000Z Series User manual

Rigol HDO4000 Series User manual

Rigol

Rigol HDO4000 Series User manual

Popular Test Equipment manuals by other brands

Viavi SmartClass OLP-85 Series operating manual

Viavi

Viavi SmartClass OLP-85 Series operating manual

Testifire 25 quick start guide

Testifire

Testifire 25 quick start guide

Hanna FC 300 instruction manual

Hanna

Hanna FC 300 instruction manual

Chroma 63200A Series Operation & programming manual

Chroma

Chroma 63200A Series Operation & programming manual

Sealey VSE2020.V3 instructions

Sealey

Sealey VSE2020.V3 instructions

RS PRO RS-SI 7005 manual

RS PRO

RS PRO RS-SI 7005 manual

Knightsbridge XT468 Installation & maintenance manual

Knightsbridge

Knightsbridge XT468 Installation & maintenance manual

NIEUWKOOP PH2010 user manual

NIEUWKOOP

NIEUWKOOP PH2010 user manual

MBE REMOTE EXAMINER user manual

MBE

MBE REMOTE EXAMINER user manual

BI COMMUNICATIONS Lan Caster ProST user manual

BI COMMUNICATIONS

BI COMMUNICATIONS Lan Caster ProST user manual

LINSHANG LS252C user manual

LINSHANG

LINSHANG LS252C user manual

Emos M0014A manual

Emos

Emos M0014A manual

Teledyne Lecroy WaveSurfer 4000HD quick start guide

Teledyne Lecroy

Teledyne Lecroy WaveSurfer 4000HD quick start guide

Kewtech KT45 instruction manual

Kewtech

Kewtech KT45 instruction manual

FLO-DYNAMICS Test-Quick LPVLT-100 Operation manual

FLO-DYNAMICS

FLO-DYNAMICS Test-Quick LPVLT-100 Operation manual

Hioki FT6031-03 instruction manual

Hioki

Hioki FT6031-03 instruction manual

Micro-Pulse B5 user manual

Micro-Pulse

Micro-Pulse B5 user manual

Tektronix TBS2000B Series Installation and safety instructions

Tektronix

Tektronix TBS2000B Series Installation and safety instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.