TECHNICAL DATA (According to ANSI standard S3.22 2003)
Maximum-OSPL90 125 dB (1000 Hz)
HFA-OSPL90 122 dB
HFA-full-on Acoustic Gain 55 dB
Reference Test Gain 45 dB
Frequency Range 200 Hz to 5400 Hz
Equivalent Input Noise Level
21 dB
Total Harmonic Distortion
500 Hz: 2% 800 Hz: 2% 1600 Hz: 1%
AGC CRC (Input AGC)/OPC (Output AGC)
Operating Switch O-M2-M
Output Limiting Control OPC (range: 18 dB)
Tone Control TONE H, TONE L
Gain Control Main VR (range: 30 dB)
GAIN (range: 34 dB)
Battery Type/Supply Voltage
13/1.3 V
Battery Current 0.94 mA
Battery Life Approx. 280 hours
Dimensions/Weight
3.75 × 1.27 × 0.88 cm/4.0 g
(excludingbattery)
(Typical value with E1 hook)
TECHNICAL DATA ()
Reference Test Frequency 1600 Hz
OSPL90 125 dB
500 Hz 122 dB
Peak 132 dB
Full-on Acoustic Gain 58 dB
Equivalent Input Noise Level
22 dB
Total Harmonic Distortion
500 Hz: 3% 800 Hz: 2% 1600 Hz: 2%
AGC CRC (Input AGC)/OPC (Output AGC)
Operating Switch O-M2-M
Output Limiting Control OPC (range: 18 dB)
Tone Control TONE H, TONE L
Gain Control Main VR (range: 30 dB)
GAIN (range: 34 dB)
Battery Type/Supply Voltage
PR48/1.3 V
Battery Current 1.0 mA
Battery Life Approx. 260 hours
Dimensions/Weight
3.75 × 1.27 × 0.88 cm/4.0 g
(excludingbattery)
(Typical value with E1 hook)
zMemory selection (using selector switch)
If sounds are hard to hear
depending on the environment,
such as in a noisy or quiet
environment, change the setting of
the memory selector to find the
position that yields best results.
* The memory selector switch (M2/
M) is programmed to match your
hearing pattern. Operate the switch
to find the position that yields best
results in a given situation.
zSwitch operation sound
Memory 1 Memory 2
M2ÆM: one beep M ÆM2: two beeps
Removing Hearing Aid
1. Set the switch to "O" (power off) before
removing the hearing aid. If the switch is left
No. 40240 04-10
in the "M" or "M2" position, acoustic feedback (howling) may
occur.
2. Be sure to set the switch to "O" and take out the battery,
when not using the hearing aid.
Turn-on precaution
If you switch the hearing aid off and then immediately on again,
there may be no sound, but this is not a defect. In such a case,
return the switch to "O" (Off), wait for at least 10 seconds, and
then set the switch to "M" or "M2" again.
Adjusting the Controls
This hearing aid is a digital type. To adjust it, a computer (with
NOAH and the Rionet Selector installed) and a HI-PRO unit are
required. Have the adjustment done at the dealer.
Factory Default Setting
zO - M2 - M switch : "O"
zVolume control (1 - 4) : "1"
According to IEC standard Pub. 60118-0-1983
Amendment 1-1994
Full-on gain curve
Frequency response curve and
effect of tone control OSPL90 curve and effect of
output limiting control
Full-on acoustic gain
frequency response curve
Basic frequency response curve
and effect of tone control OSPL90 curve and effect of
output limiting control